Gas Bubble Anatomy During Laser Lithotripsy: An Experimental In Vitro Study of a Pulsed Solid-State Tm:YAG and Ho:YAG Device

To examine gas bubbles generated by two laser lithotripsy devices, a pulsed thulium solid-state laser and a holmium:yttrium-aluminum-garnet (Ho:YAG) device, and their possible effects in lithotripsy. We investigated two Dornier laser devices, a Medilas H Solvo 35 and a pulsed solid-state thulium las...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of endourology 2021-07, Vol.35 (7), p.1051-1057
Hauptverfasser: Petzold, Ralf, Suarez-Ibarrola, Rodrigo, Miernik, Arkadiusz
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 1057
container_issue 7
container_start_page 1051
container_title Journal of endourology
container_volume 35
creator Petzold, Ralf
Suarez-Ibarrola, Rodrigo
Miernik, Arkadiusz
description To examine gas bubbles generated by two laser lithotripsy devices, a pulsed thulium solid-state laser and a holmium:yttrium-aluminum-garnet (Ho:YAG) device, and their possible effects in lithotripsy. We investigated two Dornier laser devices, a Medilas H Solvo 35 and a pulsed solid-state thulium laser evaluation model (Dornier MedTech Laser GmbH, Wessling, Germany). Our setup consisted of a water-filled glass tank heated to 37/60/70°C. Different laser power/frequency settings and short/long pulses were examined for both laser devices. We analyzed the impact of degraded, cut, and broken fibers on gas bubble anatomy. Furthermore, high-speed recordings of BegoStone ablation were analyzed. For all recordings, we used a Photron Nova S12 camera. These two devices produced differently shaped gas bubbles under different fiber conditions, temperatures, power settings, and short and long pulse settings, which explain the differing repulsive force and pressure values. Inside the gas bubble, a cone was visible whose angle correlates with the protruding jet. We observed turbulences and swirls moving back and forth the fiber tip. During fragmentation, sparks are generated that demonstrate the photothermal effect, and we recorded stone fragments being pulled toward the fiber. Both devices showed comparable results with differences mainly due to pulse lengths. The shapes of the vapor bubbles formed during laser lithotripsy depend on several factors. Excessive transoperative fiber cleavage seems to be unnecessary. Due to the large gas bubbles observed and because of the amount of potential pressure generated, only low energies should be applied in the ureter.
doi_str_mv 10.1089/end.2020.0526
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2462409190</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2462409190</sourcerecordid><originalsourceid>FETCH-LOGICAL-c293t-81e72191957f97c375ab571a88304ca28191ffb2c39771174e447f1f2bb497773</originalsourceid><addsrcrecordid>eNo9kM9PwjAUxxujEUSPXk2PXob9sdGVGwICyRJNQKOnpds6rdlWbDvjEv94i6Cn9_K-n7yX9wHgEqMhRjG_kU0xJIigIYrI6Aj0cRSxgCP0fAz6PicBYxz1wJm17whhOsL0FPQoJYjxCPXB90JYeNtmWSXhpBFO1x2ctUY1rzARVhqYKPemnVFb2409AedfW2lULRsnKrhq4JNyRsO1a4sO6hIK-NBWVhZwrStVBGsnnISbevwyWUDRFHCpf9uZ_FS5PAcnpfD0xaEOwOPdfDNdBsn9YjWdJEFOOHVBjCUjmGMesZKznLJIZBHDIo4pCnNBYp-VZUZyyhnDmIUyDFmJS5JloZ8wOgDX-71boz9aaV1aK5vLqhKN1K1NSTgiIfIHkEeDPZobba2RZbr13wrTpRilO-GpF57uhKc74Z6_Oqxus1oW__SfYfoDx115hQ</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2462409190</pqid></control><display><type>article</type><title>Gas Bubble Anatomy During Laser Lithotripsy: An Experimental In Vitro Study of a Pulsed Solid-State Tm:YAG and Ho:YAG Device</title><source>Alma/SFX Local Collection</source><creator>Petzold, Ralf ; Suarez-Ibarrola, Rodrigo ; Miernik, Arkadiusz</creator><creatorcontrib>Petzold, Ralf ; Suarez-Ibarrola, Rodrigo ; Miernik, Arkadiusz</creatorcontrib><description>To examine gas bubbles generated by two laser lithotripsy devices, a pulsed thulium solid-state laser and a holmium:yttrium-aluminum-garnet (Ho:YAG) device, and their possible effects in lithotripsy. We investigated two Dornier laser devices, a Medilas H Solvo 35 and a pulsed solid-state thulium laser evaluation model (Dornier MedTech Laser GmbH, Wessling, Germany). Our setup consisted of a water-filled glass tank heated to 37/60/70°C. Different laser power/frequency settings and short/long pulses were examined for both laser devices. We analyzed the impact of degraded, cut, and broken fibers on gas bubble anatomy. Furthermore, high-speed recordings of BegoStone ablation were analyzed. For all recordings, we used a Photron Nova S12 camera. These two devices produced differently shaped gas bubbles under different fiber conditions, temperatures, power settings, and short and long pulse settings, which explain the differing repulsive force and pressure values. Inside the gas bubble, a cone was visible whose angle correlates with the protruding jet. We observed turbulences and swirls moving back and forth the fiber tip. During fragmentation, sparks are generated that demonstrate the photothermal effect, and we recorded stone fragments being pulled toward the fiber. Both devices showed comparable results with differences mainly due to pulse lengths. The shapes of the vapor bubbles formed during laser lithotripsy depend on several factors. Excessive transoperative fiber cleavage seems to be unnecessary. Due to the large gas bubbles observed and because of the amount of potential pressure generated, only low energies should be applied in the ureter.</description><identifier>ISSN: 0892-7790</identifier><identifier>EISSN: 1557-900X</identifier><identifier>DOI: 10.1089/end.2020.0526</identifier><identifier>PMID: 33207950</identifier><language>eng</language><publisher>United States</publisher><ispartof>Journal of endourology, 2021-07, Vol.35 (7), p.1051-1057</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c293t-81e72191957f97c375ab571a88304ca28191ffb2c39771174e447f1f2bb497773</citedby><cites>FETCH-LOGICAL-c293t-81e72191957f97c375ab571a88304ca28191ffb2c39771174e447f1f2bb497773</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,776,780,27901,27902</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/33207950$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Petzold, Ralf</creatorcontrib><creatorcontrib>Suarez-Ibarrola, Rodrigo</creatorcontrib><creatorcontrib>Miernik, Arkadiusz</creatorcontrib><title>Gas Bubble Anatomy During Laser Lithotripsy: An Experimental In Vitro Study of a Pulsed Solid-State Tm:YAG and Ho:YAG Device</title><title>Journal of endourology</title><addtitle>J Endourol</addtitle><description>To examine gas bubbles generated by two laser lithotripsy devices, a pulsed thulium solid-state laser and a holmium:yttrium-aluminum-garnet (Ho:YAG) device, and their possible effects in lithotripsy. We investigated two Dornier laser devices, a Medilas H Solvo 35 and a pulsed solid-state thulium laser evaluation model (Dornier MedTech Laser GmbH, Wessling, Germany). Our setup consisted of a water-filled glass tank heated to 37/60/70°C. Different laser power/frequency settings and short/long pulses were examined for both laser devices. We analyzed the impact of degraded, cut, and broken fibers on gas bubble anatomy. Furthermore, high-speed recordings of BegoStone ablation were analyzed. For all recordings, we used a Photron Nova S12 camera. These two devices produced differently shaped gas bubbles under different fiber conditions, temperatures, power settings, and short and long pulse settings, which explain the differing repulsive force and pressure values. Inside the gas bubble, a cone was visible whose angle correlates with the protruding jet. We observed turbulences and swirls moving back and forth the fiber tip. During fragmentation, sparks are generated that demonstrate the photothermal effect, and we recorded stone fragments being pulled toward the fiber. Both devices showed comparable results with differences mainly due to pulse lengths. The shapes of the vapor bubbles formed during laser lithotripsy depend on several factors. Excessive transoperative fiber cleavage seems to be unnecessary. Due to the large gas bubbles observed and because of the amount of potential pressure generated, only low energies should be applied in the ureter.</description><issn>0892-7790</issn><issn>1557-900X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2021</creationdate><recordtype>article</recordtype><recordid>eNo9kM9PwjAUxxujEUSPXk2PXob9sdGVGwICyRJNQKOnpds6rdlWbDvjEv94i6Cn9_K-n7yX9wHgEqMhRjG_kU0xJIigIYrI6Aj0cRSxgCP0fAz6PicBYxz1wJm17whhOsL0FPQoJYjxCPXB90JYeNtmWSXhpBFO1x2ctUY1rzARVhqYKPemnVFb2409AedfW2lULRsnKrhq4JNyRsO1a4sO6hIK-NBWVhZwrStVBGsnnISbevwyWUDRFHCpf9uZ_FS5PAcnpfD0xaEOwOPdfDNdBsn9YjWdJEFOOHVBjCUjmGMesZKznLJIZBHDIo4pCnNBYp-VZUZyyhnDmIUyDFmJS5JloZ8wOgDX-71boz9aaV1aK5vLqhKN1K1NSTgiIfIHkEeDPZobba2RZbr13wrTpRilO-GpF57uhKc74Z6_Oqxus1oW__SfYfoDx115hQ</recordid><startdate>20210701</startdate><enddate>20210701</enddate><creator>Petzold, Ralf</creator><creator>Suarez-Ibarrola, Rodrigo</creator><creator>Miernik, Arkadiusz</creator><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>20210701</creationdate><title>Gas Bubble Anatomy During Laser Lithotripsy: An Experimental In Vitro Study of a Pulsed Solid-State Tm:YAG and Ho:YAG Device</title><author>Petzold, Ralf ; Suarez-Ibarrola, Rodrigo ; Miernik, Arkadiusz</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c293t-81e72191957f97c375ab571a88304ca28191ffb2c39771174e447f1f2bb497773</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2021</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Petzold, Ralf</creatorcontrib><creatorcontrib>Suarez-Ibarrola, Rodrigo</creatorcontrib><creatorcontrib>Miernik, Arkadiusz</creatorcontrib><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Journal of endourology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Petzold, Ralf</au><au>Suarez-Ibarrola, Rodrigo</au><au>Miernik, Arkadiusz</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Gas Bubble Anatomy During Laser Lithotripsy: An Experimental In Vitro Study of a Pulsed Solid-State Tm:YAG and Ho:YAG Device</atitle><jtitle>Journal of endourology</jtitle><addtitle>J Endourol</addtitle><date>2021-07-01</date><risdate>2021</risdate><volume>35</volume><issue>7</issue><spage>1051</spage><epage>1057</epage><pages>1051-1057</pages><issn>0892-7790</issn><eissn>1557-900X</eissn><abstract>To examine gas bubbles generated by two laser lithotripsy devices, a pulsed thulium solid-state laser and a holmium:yttrium-aluminum-garnet (Ho:YAG) device, and their possible effects in lithotripsy. We investigated two Dornier laser devices, a Medilas H Solvo 35 and a pulsed solid-state thulium laser evaluation model (Dornier MedTech Laser GmbH, Wessling, Germany). Our setup consisted of a water-filled glass tank heated to 37/60/70°C. Different laser power/frequency settings and short/long pulses were examined for both laser devices. We analyzed the impact of degraded, cut, and broken fibers on gas bubble anatomy. Furthermore, high-speed recordings of BegoStone ablation were analyzed. For all recordings, we used a Photron Nova S12 camera. These two devices produced differently shaped gas bubbles under different fiber conditions, temperatures, power settings, and short and long pulse settings, which explain the differing repulsive force and pressure values. Inside the gas bubble, a cone was visible whose angle correlates with the protruding jet. We observed turbulences and swirls moving back and forth the fiber tip. During fragmentation, sparks are generated that demonstrate the photothermal effect, and we recorded stone fragments being pulled toward the fiber. Both devices showed comparable results with differences mainly due to pulse lengths. The shapes of the vapor bubbles formed during laser lithotripsy depend on several factors. Excessive transoperative fiber cleavage seems to be unnecessary. Due to the large gas bubbles observed and because of the amount of potential pressure generated, only low energies should be applied in the ureter.</abstract><cop>United States</cop><pmid>33207950</pmid><doi>10.1089/end.2020.0526</doi><tpages>7</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0892-7790
ispartof Journal of endourology, 2021-07, Vol.35 (7), p.1051-1057
issn 0892-7790
1557-900X
language eng
recordid cdi_proquest_miscellaneous_2462409190
source Alma/SFX Local Collection
title Gas Bubble Anatomy During Laser Lithotripsy: An Experimental In Vitro Study of a Pulsed Solid-State Tm:YAG and Ho:YAG Device
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-07T10%3A29%3A56IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Gas%20Bubble%20Anatomy%20During%20Laser%20Lithotripsy:%20An%20Experimental%20In%20Vitro%20Study%20of%20a%20Pulsed%20Solid-State%20Tm:YAG%20and%20Ho:YAG%20Device&rft.jtitle=Journal%20of%20endourology&rft.au=Petzold,%20Ralf&rft.date=2021-07-01&rft.volume=35&rft.issue=7&rft.spage=1051&rft.epage=1057&rft.pages=1051-1057&rft.issn=0892-7790&rft.eissn=1557-900X&rft_id=info:doi/10.1089/end.2020.0526&rft_dat=%3Cproquest_cross%3E2462409190%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2462409190&rft_id=info:pmid/33207950&rfr_iscdi=true