Development of a dynamic deformable thorax phantom for the quality management of deformable image registration
•We developed a dynamic deformable thorax phantom for DIR commissioning and QA.•Our phantom can create various deformation patterns with various marker patterns.•DIR results with different marker settings and DIR parameters were obtained using this phantom. The purpose of this study was to develop a...
Gespeichert in:
Veröffentlicht in: | Physica medica 2020-09, Vol.77, p.100-107 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 107 |
---|---|
container_issue | |
container_start_page | 100 |
container_title | Physica medica |
container_volume | 77 |
creator | Sugawara, Yasuharu Kadoya, Noriyuki Kotabe, Kazuki Nakajima, Yujiro Ikeda, Ryutaro Tanabe, Shunpei Ohashi, Haruna Jingu, Keiichi |
description | •We developed a dynamic deformable thorax phantom for DIR commissioning and QA.•Our phantom can create various deformation patterns with various marker patterns.•DIR results with different marker settings and DIR parameters were obtained using this phantom.
The purpose of this study was to develop a novel dynamic deformable thorax phantom for deformable image registration (DIR) quality assurance (QA) and to verify as a tool for commissioning and DIR QA.
The phantom consists of a base phantom, an inner phantom, and a motor-derived piston. The base phantom is an acrylic cylinder phantom with a diameter of 180 mm. The inner phantom consists of deformable, 20 mm thick disk-shaped sponges. To evaluate the physical characteristics of the phantom, we evaluated its image quality and deformation. DIR accuracies were evaluated using the three types of commercially DIR software (MIM, RayStation, and Velocity AI) to test the feasibility of this phantom. We used different DIR parameters to test the impact of parameters on DIR accuracy in various phantom settings. To evaluate DIR accuracy, a target registration error (TRE) was calculated using the anatomical landmark points.
The three locations (i.e., distal, middle, and proximal positions) had different displacement amounts. This result indicated that the inner phantom was not moved but deformed. In cases with different phantom settings and marker settings, the ranges of the average TRE were 0.63–15.60 mm (MIM). In cases with different DIR parameters settings, the ranges of the average TRE were as follows: 0.73–7.10 mm (MIM), 8.25–8.66 mm (RayStation), and 8.26–8.43 mm (Velocity). These results suggest that our phantom could evaluate the detailed DIR behaviors with TRE. Therefore, this is indicative of the potential usefulness of our phantom in DIR commissioning and QA. |
doi_str_mv | 10.1016/j.ejmp.2020.08.007 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2436394305</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S112017972030199X</els_id><sourcerecordid>2436394305</sourcerecordid><originalsourceid>FETCH-LOGICAL-c399t-40c5a6687b1bef1941ab0a74da572488f42fdf5add534e31a0eaf928a200b6b83</originalsourceid><addsrcrecordid>eNp9kM1LxDAQxYso-PkPeMrRS-sk_QYvsn6C4EXBW5i2k92UtukmWXH_e7OsgidPMzzeb5j3ouiSQ8KBF9d9Qv04JwIEJFAlAOVBdMJLkcW85h-HYecCYl7W5XF06lwPkAqR5yfRdEefNJh5pMkzoxiybjvhqFvWkTJ2xGYg5lfG4hebVzh5M7KgB4nYeoOD9ls24oRL-r3wh9Nj0JmlpXbeotdmOo-OFA6OLn7mWfT-cP-2eIpfXh-fF7cvcZvWtY8zaHMsiqpseEOK1xnHBrDMOsxDpqpSmVCdyrHr8jSjlCMQqlpUKACaoqnSs-hqf3e2Zr0h5-WoXUvDgBOZjZMiS4u0zlLIg1Xsra01zllScrbhcbuVHOSuXNnLXblyV66ESoZyA3SzhyiE-NRkpWs1TS112lLrZWf0f_g3vluFbQ</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2436394305</pqid></control><display><type>article</type><title>Development of a dynamic deformable thorax phantom for the quality management of deformable image registration</title><source>Elsevier ScienceDirect Journals</source><creator>Sugawara, Yasuharu ; Kadoya, Noriyuki ; Kotabe, Kazuki ; Nakajima, Yujiro ; Ikeda, Ryutaro ; Tanabe, Shunpei ; Ohashi, Haruna ; Jingu, Keiichi</creator><creatorcontrib>Sugawara, Yasuharu ; Kadoya, Noriyuki ; Kotabe, Kazuki ; Nakajima, Yujiro ; Ikeda, Ryutaro ; Tanabe, Shunpei ; Ohashi, Haruna ; Jingu, Keiichi</creatorcontrib><description>•We developed a dynamic deformable thorax phantom for DIR commissioning and QA.•Our phantom can create various deformation patterns with various marker patterns.•DIR results with different marker settings and DIR parameters were obtained using this phantom.
The purpose of this study was to develop a novel dynamic deformable thorax phantom for deformable image registration (DIR) quality assurance (QA) and to verify as a tool for commissioning and DIR QA.
The phantom consists of a base phantom, an inner phantom, and a motor-derived piston. The base phantom is an acrylic cylinder phantom with a diameter of 180 mm. The inner phantom consists of deformable, 20 mm thick disk-shaped sponges. To evaluate the physical characteristics of the phantom, we evaluated its image quality and deformation. DIR accuracies were evaluated using the three types of commercially DIR software (MIM, RayStation, and Velocity AI) to test the feasibility of this phantom. We used different DIR parameters to test the impact of parameters on DIR accuracy in various phantom settings. To evaluate DIR accuracy, a target registration error (TRE) was calculated using the anatomical landmark points.
The three locations (i.e., distal, middle, and proximal positions) had different displacement amounts. This result indicated that the inner phantom was not moved but deformed. In cases with different phantom settings and marker settings, the ranges of the average TRE were 0.63–15.60 mm (MIM). In cases with different DIR parameters settings, the ranges of the average TRE were as follows: 0.73–7.10 mm (MIM), 8.25–8.66 mm (RayStation), and 8.26–8.43 mm (Velocity). These results suggest that our phantom could evaluate the detailed DIR behaviors with TRE. Therefore, this is indicative of the potential usefulness of our phantom in DIR commissioning and QA.</description><identifier>ISSN: 1120-1797</identifier><identifier>EISSN: 1724-191X</identifier><identifier>DOI: 10.1016/j.ejmp.2020.08.007</identifier><language>eng</language><publisher>Elsevier Ltd</publisher><subject>Deformable image registration ; Physical phantom ; Quality assurance ; Radiotherapy ; Target registration error</subject><ispartof>Physica medica, 2020-09, Vol.77, p.100-107</ispartof><rights>2020 Associazione Italiana di Fisica Medica</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c399t-40c5a6687b1bef1941ab0a74da572488f42fdf5add534e31a0eaf928a200b6b83</citedby><cites>FETCH-LOGICAL-c399t-40c5a6687b1bef1941ab0a74da572488f42fdf5add534e31a0eaf928a200b6b83</cites><orcidid>0000-0001-5018-3800</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://www.sciencedirect.com/science/article/pii/S112017972030199X$$EHTML$$P50$$Gelsevier$$H</linktohtml><link.rule.ids>314,776,780,3537,27901,27902,65306</link.rule.ids></links><search><creatorcontrib>Sugawara, Yasuharu</creatorcontrib><creatorcontrib>Kadoya, Noriyuki</creatorcontrib><creatorcontrib>Kotabe, Kazuki</creatorcontrib><creatorcontrib>Nakajima, Yujiro</creatorcontrib><creatorcontrib>Ikeda, Ryutaro</creatorcontrib><creatorcontrib>Tanabe, Shunpei</creatorcontrib><creatorcontrib>Ohashi, Haruna</creatorcontrib><creatorcontrib>Jingu, Keiichi</creatorcontrib><title>Development of a dynamic deformable thorax phantom for the quality management of deformable image registration</title><title>Physica medica</title><description>•We developed a dynamic deformable thorax phantom for DIR commissioning and QA.•Our phantom can create various deformation patterns with various marker patterns.•DIR results with different marker settings and DIR parameters were obtained using this phantom.
The purpose of this study was to develop a novel dynamic deformable thorax phantom for deformable image registration (DIR) quality assurance (QA) and to verify as a tool for commissioning and DIR QA.
The phantom consists of a base phantom, an inner phantom, and a motor-derived piston. The base phantom is an acrylic cylinder phantom with a diameter of 180 mm. The inner phantom consists of deformable, 20 mm thick disk-shaped sponges. To evaluate the physical characteristics of the phantom, we evaluated its image quality and deformation. DIR accuracies were evaluated using the three types of commercially DIR software (MIM, RayStation, and Velocity AI) to test the feasibility of this phantom. We used different DIR parameters to test the impact of parameters on DIR accuracy in various phantom settings. To evaluate DIR accuracy, a target registration error (TRE) was calculated using the anatomical landmark points.
The three locations (i.e., distal, middle, and proximal positions) had different displacement amounts. This result indicated that the inner phantom was not moved but deformed. In cases with different phantom settings and marker settings, the ranges of the average TRE were 0.63–15.60 mm (MIM). In cases with different DIR parameters settings, the ranges of the average TRE were as follows: 0.73–7.10 mm (MIM), 8.25–8.66 mm (RayStation), and 8.26–8.43 mm (Velocity). These results suggest that our phantom could evaluate the detailed DIR behaviors with TRE. Therefore, this is indicative of the potential usefulness of our phantom in DIR commissioning and QA.</description><subject>Deformable image registration</subject><subject>Physical phantom</subject><subject>Quality assurance</subject><subject>Radiotherapy</subject><subject>Target registration error</subject><issn>1120-1797</issn><issn>1724-191X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2020</creationdate><recordtype>article</recordtype><recordid>eNp9kM1LxDAQxYso-PkPeMrRS-sk_QYvsn6C4EXBW5i2k92UtukmWXH_e7OsgidPMzzeb5j3ouiSQ8KBF9d9Qv04JwIEJFAlAOVBdMJLkcW85h-HYecCYl7W5XF06lwPkAqR5yfRdEefNJh5pMkzoxiybjvhqFvWkTJ2xGYg5lfG4hebVzh5M7KgB4nYeoOD9ls24oRL-r3wh9Nj0JmlpXbeotdmOo-OFA6OLn7mWfT-cP-2eIpfXh-fF7cvcZvWtY8zaHMsiqpseEOK1xnHBrDMOsxDpqpSmVCdyrHr8jSjlCMQqlpUKACaoqnSs-hqf3e2Zr0h5-WoXUvDgBOZjZMiS4u0zlLIg1Xsra01zllScrbhcbuVHOSuXNnLXblyV66ESoZyA3SzhyiE-NRkpWs1TS112lLrZWf0f_g3vluFbQ</recordid><startdate>202009</startdate><enddate>202009</enddate><creator>Sugawara, Yasuharu</creator><creator>Kadoya, Noriyuki</creator><creator>Kotabe, Kazuki</creator><creator>Nakajima, Yujiro</creator><creator>Ikeda, Ryutaro</creator><creator>Tanabe, Shunpei</creator><creator>Ohashi, Haruna</creator><creator>Jingu, Keiichi</creator><general>Elsevier Ltd</general><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><orcidid>https://orcid.org/0000-0001-5018-3800</orcidid></search><sort><creationdate>202009</creationdate><title>Development of a dynamic deformable thorax phantom for the quality management of deformable image registration</title><author>Sugawara, Yasuharu ; Kadoya, Noriyuki ; Kotabe, Kazuki ; Nakajima, Yujiro ; Ikeda, Ryutaro ; Tanabe, Shunpei ; Ohashi, Haruna ; Jingu, Keiichi</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c399t-40c5a6687b1bef1941ab0a74da572488f42fdf5add534e31a0eaf928a200b6b83</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2020</creationdate><topic>Deformable image registration</topic><topic>Physical phantom</topic><topic>Quality assurance</topic><topic>Radiotherapy</topic><topic>Target registration error</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Sugawara, Yasuharu</creatorcontrib><creatorcontrib>Kadoya, Noriyuki</creatorcontrib><creatorcontrib>Kotabe, Kazuki</creatorcontrib><creatorcontrib>Nakajima, Yujiro</creatorcontrib><creatorcontrib>Ikeda, Ryutaro</creatorcontrib><creatorcontrib>Tanabe, Shunpei</creatorcontrib><creatorcontrib>Ohashi, Haruna</creatorcontrib><creatorcontrib>Jingu, Keiichi</creatorcontrib><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Physica medica</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Sugawara, Yasuharu</au><au>Kadoya, Noriyuki</au><au>Kotabe, Kazuki</au><au>Nakajima, Yujiro</au><au>Ikeda, Ryutaro</au><au>Tanabe, Shunpei</au><au>Ohashi, Haruna</au><au>Jingu, Keiichi</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Development of a dynamic deformable thorax phantom for the quality management of deformable image registration</atitle><jtitle>Physica medica</jtitle><date>2020-09</date><risdate>2020</risdate><volume>77</volume><spage>100</spage><epage>107</epage><pages>100-107</pages><issn>1120-1797</issn><eissn>1724-191X</eissn><abstract>•We developed a dynamic deformable thorax phantom for DIR commissioning and QA.•Our phantom can create various deformation patterns with various marker patterns.•DIR results with different marker settings and DIR parameters were obtained using this phantom.
The purpose of this study was to develop a novel dynamic deformable thorax phantom for deformable image registration (DIR) quality assurance (QA) and to verify as a tool for commissioning and DIR QA.
The phantom consists of a base phantom, an inner phantom, and a motor-derived piston. The base phantom is an acrylic cylinder phantom with a diameter of 180 mm. The inner phantom consists of deformable, 20 mm thick disk-shaped sponges. To evaluate the physical characteristics of the phantom, we evaluated its image quality and deformation. DIR accuracies were evaluated using the three types of commercially DIR software (MIM, RayStation, and Velocity AI) to test the feasibility of this phantom. We used different DIR parameters to test the impact of parameters on DIR accuracy in various phantom settings. To evaluate DIR accuracy, a target registration error (TRE) was calculated using the anatomical landmark points.
The three locations (i.e., distal, middle, and proximal positions) had different displacement amounts. This result indicated that the inner phantom was not moved but deformed. In cases with different phantom settings and marker settings, the ranges of the average TRE were 0.63–15.60 mm (MIM). In cases with different DIR parameters settings, the ranges of the average TRE were as follows: 0.73–7.10 mm (MIM), 8.25–8.66 mm (RayStation), and 8.26–8.43 mm (Velocity). These results suggest that our phantom could evaluate the detailed DIR behaviors with TRE. Therefore, this is indicative of the potential usefulness of our phantom in DIR commissioning and QA.</abstract><pub>Elsevier Ltd</pub><doi>10.1016/j.ejmp.2020.08.007</doi><tpages>8</tpages><orcidid>https://orcid.org/0000-0001-5018-3800</orcidid></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1120-1797 |
ispartof | Physica medica, 2020-09, Vol.77, p.100-107 |
issn | 1120-1797 1724-191X |
language | eng |
recordid | cdi_proquest_miscellaneous_2436394305 |
source | Elsevier ScienceDirect Journals |
subjects | Deformable image registration Physical phantom Quality assurance Radiotherapy Target registration error |
title | Development of a dynamic deformable thorax phantom for the quality management of deformable image registration |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-08T17%3A38%3A22IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Development%20of%20a%20dynamic%20deformable%20thorax%20phantom%20for%20the%20quality%20management%20of%20deformable%20image%20registration&rft.jtitle=Physica%20medica&rft.au=Sugawara,%20Yasuharu&rft.date=2020-09&rft.volume=77&rft.spage=100&rft.epage=107&rft.pages=100-107&rft.issn=1120-1797&rft.eissn=1724-191X&rft_id=info:doi/10.1016/j.ejmp.2020.08.007&rft_dat=%3Cproquest_cross%3E2436394305%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2436394305&rft_id=info:pmid/&rft_els_id=S112017972030199X&rfr_iscdi=true |