Geometric and Dosimetric Evaluation of a Commercially Available Auto-segmentation Tool for Gross Tumour Volume Delineation in Locally Advanced Non-small Cell Lung Cancer: a Feasibility Study
To quantify the reliability of a commercially available auto-segmentation tool in locally advanced non-small cell lung cancer using serial four-dimensional computed tomography (4DCT) scans during conventionally fractionated radiotherapy. Eight patients with serial 4DCT scans (n = 44) acquired over t...
Gespeichert in:
Veröffentlicht in: | Clinical oncology (Royal College of Radiologists (Great Britain)) 2021-03, Vol.33 (3), p.155-162 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 162 |
---|---|
container_issue | 3 |
container_start_page | 155 |
container_title | Clinical oncology (Royal College of Radiologists (Great Britain)) |
container_volume | 33 |
creator | Barrett, S. Simpkin, A.J. Walls, G.M. Leech, M. Marignol, L. |
description | To quantify the reliability of a commercially available auto-segmentation tool in locally advanced non-small cell lung cancer using serial four-dimensional computed tomography (4DCT) scans during conventionally fractionated radiotherapy.
Eight patients with serial 4DCT scans (n = 44) acquired over the course of radiotherapy were assessed. Each 4DCT had a physician-defined primary tumour manual contour (MC). An auto-contour (AC) and a user-adjusted auto-contour (UA-AC) were created for each scan. Geometric agreement of the AC and the UA-AC to the MC was assessed using the dice similarity coefficient (DSC), the centre of mass (COM) shift from the MC and the structure volume difference from the MC. Bland Altman analysis was carried out to assess agreement between contouring methods. Dosimetric reliability was assessed by comparison of planning target volume dose coverage on the MC and UA-AC. The time trend analysis of the geometric accuracy measures from the initial planning scan through to the final scan for each patient was evaluated using a Wilcoxon signed ranks test to assess the reliability of the UA-AC over the duration of radiotherapy.
User adjustment significantly improved all geometric comparison metrics over the AC alone. Improved agreement was observed in smaller tumours not abutting normal soft tissue and median values for geometric comparisons to the MC for DSC, tumour volume difference and COM offset were 0.80 (range 0.49–0.89), 0.8 cm3 (range 0.0–5.9 cm3) and 0.16 cm (range 0.09–0.69 cm), respectively. There were no significant differences in dose metrics measured from the MC and the UA-AC after Bonferroni correction. Variation in geometric agreement between the MC and the UA-AC were observed over the course of radiotherapy with both DSC (P = 0.035) and COM shift from the MC (ns) worsening. The median tumour volume difference from the MC improved at the later time point.
These findings suggest that the UA-AC can produce geometrically and dosimetrically acceptable contours for appropriately selected patients with non-small cell lung cancer. Larger studies are required to confirm the findings.
•User-adjustment of auto-contours improves geometric agreement towards the gold standard.•Appropriate case selection improves the accuracy of contours generated.•No significant dosimetric variations measured between PTVs generated from manual contours and user-adjusted auto-contours.•The unique dataset within the study seems to confirm the reliability of us |
doi_str_mv | 10.1016/j.clon.2020.07.019 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2434478293</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S0936655520303095</els_id><sourcerecordid>2434478293</sourcerecordid><originalsourceid>FETCH-LOGICAL-c400t-833c83c1a0b1b4cbc6852a8258c44141e78bc924fba7032c5669d3f77dac29ec3</originalsourceid><addsrcrecordid>eNp9kcuO0zAUhi0EYsrAC7BAXrJJ8S2Jg9hUmZmCVMGCwtZynJORK18GO6nUl-PZSJTCko0t29_5dY4_hN5SsqWEVh9OW-Ni2DLCyJbUW0KbZ2hDBecFayR9jjak4VVRlWV5g17lfCKEMCmbl-iGs3omSrlBv_cQPYzJGqxDj-9ittfj_Vm7SY82BhwHrHEbvYdkrHbugndnbZ3uHODdNMYiw6OHMK70MUaHh5jwPsWc8XHycUr4Z3STB3wHzgZYQRvwIZo1rz_rYKDHX2Mosp_vcAvzcpjCI26Xp_Rx7uEBdLaddXa84O_j1F9eoxeDdhneXPdb9OPh_th-Lg7f9l_a3aEwgpCxkJwbyQ3VpKOdMJ2pZMm0ZKU0QlBBoZadaZgYOl0TzkxZVU3Ph7rutWENGH6L3q-5Tyn-miCPytts5g51gDhlxQQXopas4TPKVtQs4ycY1FOyXqeLokQt3tRJLd7U4k2RWs3e5qJ31_yp89D_K_kragY-rQDMU54tJJWNheXLbAIzqj7a_-X_AfJMrMk</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2434478293</pqid></control><display><type>article</type><title>Geometric and Dosimetric Evaluation of a Commercially Available Auto-segmentation Tool for Gross Tumour Volume Delineation in Locally Advanced Non-small Cell Lung Cancer: a Feasibility Study</title><source>Elsevier ScienceDirect Journals Complete</source><creator>Barrett, S. ; Simpkin, A.J. ; Walls, G.M. ; Leech, M. ; Marignol, L.</creator><creatorcontrib>Barrett, S. ; Simpkin, A.J. ; Walls, G.M. ; Leech, M. ; Marignol, L.</creatorcontrib><description>To quantify the reliability of a commercially available auto-segmentation tool in locally advanced non-small cell lung cancer using serial four-dimensional computed tomography (4DCT) scans during conventionally fractionated radiotherapy.
Eight patients with serial 4DCT scans (n = 44) acquired over the course of radiotherapy were assessed. Each 4DCT had a physician-defined primary tumour manual contour (MC). An auto-contour (AC) and a user-adjusted auto-contour (UA-AC) were created for each scan. Geometric agreement of the AC and the UA-AC to the MC was assessed using the dice similarity coefficient (DSC), the centre of mass (COM) shift from the MC and the structure volume difference from the MC. Bland Altman analysis was carried out to assess agreement between contouring methods. Dosimetric reliability was assessed by comparison of planning target volume dose coverage on the MC and UA-AC. The time trend analysis of the geometric accuracy measures from the initial planning scan through to the final scan for each patient was evaluated using a Wilcoxon signed ranks test to assess the reliability of the UA-AC over the duration of radiotherapy.
User adjustment significantly improved all geometric comparison metrics over the AC alone. Improved agreement was observed in smaller tumours not abutting normal soft tissue and median values for geometric comparisons to the MC for DSC, tumour volume difference and COM offset were 0.80 (range 0.49–0.89), 0.8 cm3 (range 0.0–5.9 cm3) and 0.16 cm (range 0.09–0.69 cm), respectively. There were no significant differences in dose metrics measured from the MC and the UA-AC after Bonferroni correction. Variation in geometric agreement between the MC and the UA-AC were observed over the course of radiotherapy with both DSC (P = 0.035) and COM shift from the MC (ns) worsening. The median tumour volume difference from the MC improved at the later time point.
These findings suggest that the UA-AC can produce geometrically and dosimetrically acceptable contours for appropriately selected patients with non-small cell lung cancer. Larger studies are required to confirm the findings.
•User-adjustment of auto-contours improves geometric agreement towards the gold standard.•Appropriate case selection improves the accuracy of contours generated.•No significant dosimetric variations measured between PTVs generated from manual contours and user-adjusted auto-contours.•The unique dataset within the study seems to confirm the reliability of user-adjusted contours over time.•Longitudinal auto-contouring may have a role in biologically guided radiotherapy for non-small cell lung cancer.</description><identifier>ISSN: 0936-6555</identifier><identifier>EISSN: 1433-2981</identifier><identifier>DOI: 10.1016/j.clon.2020.07.019</identifier><identifier>PMID: 32798158</identifier><language>eng</language><publisher>England: Elsevier Ltd</publisher><subject>auto-contouring ; delineation ; non-small cell lung cancer ; target volume definition</subject><ispartof>Clinical oncology (Royal College of Radiologists (Great Britain)), 2021-03, Vol.33 (3), p.155-162</ispartof><rights>2021 The Authors</rights><rights>Copyright © 2021 The Royal College of Radiologists. Published by Elsevier Ltd. All rights reserved.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c400t-833c83c1a0b1b4cbc6852a8258c44141e78bc924fba7032c5669d3f77dac29ec3</citedby><cites>FETCH-LOGICAL-c400t-833c83c1a0b1b4cbc6852a8258c44141e78bc924fba7032c5669d3f77dac29ec3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://dx.doi.org/10.1016/j.clon.2020.07.019$$EHTML$$P50$$Gelsevier$$Hfree_for_read</linktohtml><link.rule.ids>314,780,784,3550,27924,27925,45995</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/32798158$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Barrett, S.</creatorcontrib><creatorcontrib>Simpkin, A.J.</creatorcontrib><creatorcontrib>Walls, G.M.</creatorcontrib><creatorcontrib>Leech, M.</creatorcontrib><creatorcontrib>Marignol, L.</creatorcontrib><title>Geometric and Dosimetric Evaluation of a Commercially Available Auto-segmentation Tool for Gross Tumour Volume Delineation in Locally Advanced Non-small Cell Lung Cancer: a Feasibility Study</title><title>Clinical oncology (Royal College of Radiologists (Great Britain))</title><addtitle>Clin Oncol (R Coll Radiol)</addtitle><description>To quantify the reliability of a commercially available auto-segmentation tool in locally advanced non-small cell lung cancer using serial four-dimensional computed tomography (4DCT) scans during conventionally fractionated radiotherapy.
Eight patients with serial 4DCT scans (n = 44) acquired over the course of radiotherapy were assessed. Each 4DCT had a physician-defined primary tumour manual contour (MC). An auto-contour (AC) and a user-adjusted auto-contour (UA-AC) were created for each scan. Geometric agreement of the AC and the UA-AC to the MC was assessed using the dice similarity coefficient (DSC), the centre of mass (COM) shift from the MC and the structure volume difference from the MC. Bland Altman analysis was carried out to assess agreement between contouring methods. Dosimetric reliability was assessed by comparison of planning target volume dose coverage on the MC and UA-AC. The time trend analysis of the geometric accuracy measures from the initial planning scan through to the final scan for each patient was evaluated using a Wilcoxon signed ranks test to assess the reliability of the UA-AC over the duration of radiotherapy.
User adjustment significantly improved all geometric comparison metrics over the AC alone. Improved agreement was observed in smaller tumours not abutting normal soft tissue and median values for geometric comparisons to the MC for DSC, tumour volume difference and COM offset were 0.80 (range 0.49–0.89), 0.8 cm3 (range 0.0–5.9 cm3) and 0.16 cm (range 0.09–0.69 cm), respectively. There were no significant differences in dose metrics measured from the MC and the UA-AC after Bonferroni correction. Variation in geometric agreement between the MC and the UA-AC were observed over the course of radiotherapy with both DSC (P = 0.035) and COM shift from the MC (ns) worsening. The median tumour volume difference from the MC improved at the later time point.
These findings suggest that the UA-AC can produce geometrically and dosimetrically acceptable contours for appropriately selected patients with non-small cell lung cancer. Larger studies are required to confirm the findings.
•User-adjustment of auto-contours improves geometric agreement towards the gold standard.•Appropriate case selection improves the accuracy of contours generated.•No significant dosimetric variations measured between PTVs generated from manual contours and user-adjusted auto-contours.•The unique dataset within the study seems to confirm the reliability of user-adjusted contours over time.•Longitudinal auto-contouring may have a role in biologically guided radiotherapy for non-small cell lung cancer.</description><subject>auto-contouring</subject><subject>delineation</subject><subject>non-small cell lung cancer</subject><subject>target volume definition</subject><issn>0936-6555</issn><issn>1433-2981</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2021</creationdate><recordtype>article</recordtype><recordid>eNp9kcuO0zAUhi0EYsrAC7BAXrJJ8S2Jg9hUmZmCVMGCwtZynJORK18GO6nUl-PZSJTCko0t29_5dY4_hN5SsqWEVh9OW-Ni2DLCyJbUW0KbZ2hDBecFayR9jjak4VVRlWV5g17lfCKEMCmbl-iGs3omSrlBv_cQPYzJGqxDj-9ittfj_Vm7SY82BhwHrHEbvYdkrHbugndnbZ3uHODdNMYiw6OHMK70MUaHh5jwPsWc8XHycUr4Z3STB3wHzgZYQRvwIZo1rz_rYKDHX2Mosp_vcAvzcpjCI26Xp_Rx7uEBdLaddXa84O_j1F9eoxeDdhneXPdb9OPh_th-Lg7f9l_a3aEwgpCxkJwbyQ3VpKOdMJ2pZMm0ZKU0QlBBoZadaZgYOl0TzkxZVU3Ph7rutWENGH6L3q-5Tyn-miCPytts5g51gDhlxQQXopas4TPKVtQs4ycY1FOyXqeLokQt3tRJLd7U4k2RWs3e5qJ31_yp89D_K_kragY-rQDMU54tJJWNheXLbAIzqj7a_-X_AfJMrMk</recordid><startdate>202103</startdate><enddate>202103</enddate><creator>Barrett, S.</creator><creator>Simpkin, A.J.</creator><creator>Walls, G.M.</creator><creator>Leech, M.</creator><creator>Marignol, L.</creator><general>Elsevier Ltd</general><scope>6I.</scope><scope>AAFTH</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>202103</creationdate><title>Geometric and Dosimetric Evaluation of a Commercially Available Auto-segmentation Tool for Gross Tumour Volume Delineation in Locally Advanced Non-small Cell Lung Cancer: a Feasibility Study</title><author>Barrett, S. ; Simpkin, A.J. ; Walls, G.M. ; Leech, M. ; Marignol, L.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c400t-833c83c1a0b1b4cbc6852a8258c44141e78bc924fba7032c5669d3f77dac29ec3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2021</creationdate><topic>auto-contouring</topic><topic>delineation</topic><topic>non-small cell lung cancer</topic><topic>target volume definition</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Barrett, S.</creatorcontrib><creatorcontrib>Simpkin, A.J.</creatorcontrib><creatorcontrib>Walls, G.M.</creatorcontrib><creatorcontrib>Leech, M.</creatorcontrib><creatorcontrib>Marignol, L.</creatorcontrib><collection>ScienceDirect Open Access Titles</collection><collection>Elsevier:ScienceDirect:Open Access</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Clinical oncology (Royal College of Radiologists (Great Britain))</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Barrett, S.</au><au>Simpkin, A.J.</au><au>Walls, G.M.</au><au>Leech, M.</au><au>Marignol, L.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Geometric and Dosimetric Evaluation of a Commercially Available Auto-segmentation Tool for Gross Tumour Volume Delineation in Locally Advanced Non-small Cell Lung Cancer: a Feasibility Study</atitle><jtitle>Clinical oncology (Royal College of Radiologists (Great Britain))</jtitle><addtitle>Clin Oncol (R Coll Radiol)</addtitle><date>2021-03</date><risdate>2021</risdate><volume>33</volume><issue>3</issue><spage>155</spage><epage>162</epage><pages>155-162</pages><issn>0936-6555</issn><eissn>1433-2981</eissn><abstract>To quantify the reliability of a commercially available auto-segmentation tool in locally advanced non-small cell lung cancer using serial four-dimensional computed tomography (4DCT) scans during conventionally fractionated radiotherapy.
Eight patients with serial 4DCT scans (n = 44) acquired over the course of radiotherapy were assessed. Each 4DCT had a physician-defined primary tumour manual contour (MC). An auto-contour (AC) and a user-adjusted auto-contour (UA-AC) were created for each scan. Geometric agreement of the AC and the UA-AC to the MC was assessed using the dice similarity coefficient (DSC), the centre of mass (COM) shift from the MC and the structure volume difference from the MC. Bland Altman analysis was carried out to assess agreement between contouring methods. Dosimetric reliability was assessed by comparison of planning target volume dose coverage on the MC and UA-AC. The time trend analysis of the geometric accuracy measures from the initial planning scan through to the final scan for each patient was evaluated using a Wilcoxon signed ranks test to assess the reliability of the UA-AC over the duration of radiotherapy.
User adjustment significantly improved all geometric comparison metrics over the AC alone. Improved agreement was observed in smaller tumours not abutting normal soft tissue and median values for geometric comparisons to the MC for DSC, tumour volume difference and COM offset were 0.80 (range 0.49–0.89), 0.8 cm3 (range 0.0–5.9 cm3) and 0.16 cm (range 0.09–0.69 cm), respectively. There were no significant differences in dose metrics measured from the MC and the UA-AC after Bonferroni correction. Variation in geometric agreement between the MC and the UA-AC were observed over the course of radiotherapy with both DSC (P = 0.035) and COM shift from the MC (ns) worsening. The median tumour volume difference from the MC improved at the later time point.
These findings suggest that the UA-AC can produce geometrically and dosimetrically acceptable contours for appropriately selected patients with non-small cell lung cancer. Larger studies are required to confirm the findings.
•User-adjustment of auto-contours improves geometric agreement towards the gold standard.•Appropriate case selection improves the accuracy of contours generated.•No significant dosimetric variations measured between PTVs generated from manual contours and user-adjusted auto-contours.•The unique dataset within the study seems to confirm the reliability of user-adjusted contours over time.•Longitudinal auto-contouring may have a role in biologically guided radiotherapy for non-small cell lung cancer.</abstract><cop>England</cop><pub>Elsevier Ltd</pub><pmid>32798158</pmid><doi>10.1016/j.clon.2020.07.019</doi><tpages>8</tpages><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0936-6555 |
ispartof | Clinical oncology (Royal College of Radiologists (Great Britain)), 2021-03, Vol.33 (3), p.155-162 |
issn | 0936-6555 1433-2981 |
language | eng |
recordid | cdi_proquest_miscellaneous_2434478293 |
source | Elsevier ScienceDirect Journals Complete |
subjects | auto-contouring delineation non-small cell lung cancer target volume definition |
title | Geometric and Dosimetric Evaluation of a Commercially Available Auto-segmentation Tool for Gross Tumour Volume Delineation in Locally Advanced Non-small Cell Lung Cancer: a Feasibility Study |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-05T00%3A52%3A22IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Geometric%20and%20Dosimetric%20Evaluation%20of%20a%20Commercially%20Available%20Auto-segmentation%20Tool%20for%20Gross%20Tumour%20Volume%20Delineation%20in%20Locally%20Advanced%20Non-small%20Cell%20Lung%20Cancer:%20a%20Feasibility%20Study&rft.jtitle=Clinical%20oncology%20(Royal%20College%20of%20Radiologists%20(Great%20Britain))&rft.au=Barrett,%20S.&rft.date=2021-03&rft.volume=33&rft.issue=3&rft.spage=155&rft.epage=162&rft.pages=155-162&rft.issn=0936-6555&rft.eissn=1433-2981&rft_id=info:doi/10.1016/j.clon.2020.07.019&rft_dat=%3Cproquest_cross%3E2434478293%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2434478293&rft_id=info:pmid/32798158&rft_els_id=S0936655520303095&rfr_iscdi=true |