Accuracy of fluid delivery devices for the neonate: Are the measures assured?

Introduction: Delivery of accurate volumes of fluid in surgical neonates and children is crucial for the good outcome of treatment. But how accurate are the calibrations on the fluid delivery devices? Aims: This study seeks to verify the accuracy of these devices in common use in our practice. Mater...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Nigerian journal of clinical practice 2020-08, Vol.23 (8), p.1044-1047
Hauptverfasser: Okoro, P, Gbobo, I, Igwe, P, Umeh, D, Okoro, C, Nwiwu, P
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 1047
container_issue 8
container_start_page 1044
container_title Nigerian journal of clinical practice
container_volume 23
creator Okoro, P
Gbobo, I
Igwe, P
Umeh, D
Okoro, C
Nwiwu, P
description Introduction: Delivery of accurate volumes of fluid in surgical neonates and children is crucial for the good outcome of treatment. But how accurate are the calibrations on the fluid delivery devices? Aims: This study seeks to verify the accuracy of these devices in common use in our practice. Materials and Methods: This is a cross-sectional experimental study carried out in our center; a tertiary health facility in Southern Nigeria in May 2019. Fluid delivery devices (FDDs) used in the course of treatment of our pediatric patients were randomly included in the study. The number of drops per ml of each device was obtained by counting while the fluid dropped until a 1 ml volume was delivered. The data was then collated and analyzed. Results: A total of 215 FDDs were included in this study. They comprised infusion giving set, Soluset (Burette) giving set, and blood giving set. The rate of delivery was 20 drops/ml (infusion giving sets), 60 drops/min (Burette/Soluset), and 15 drops/ml (Blood giving set). They were all in keeping with the labeled/assumed calibration in each of the types of FDDs P < 0.05. Therefore, the mean, median, and mode were the same. Conclusion: This study has demonstrated that the FDDs used our center are accurately calibrated and safe as they deliver volumes of fluid as labeled. The findings in this study reassure us of the dependability and accuracy of delivery of the FDDs we use in children in our center.
doi_str_mv 10.4103/njcp.njcp_502_19
format Article
fullrecord <record><control><sourceid>gale_proqu</sourceid><recordid>TN_cdi_proquest_miscellaneous_2434053814</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><galeid>A632841619</galeid><sourcerecordid>A632841619</sourcerecordid><originalsourceid>FETCH-LOGICAL-c457y-b995885475478b9ee65a11c456290aa8b771f5cfaa2e468943efcdd8efebba033</originalsourceid><addsrcrecordid>eNp1kc1L9DAQxnNQ_L57koIgXnZNmrRN3svLsvgFihc9hzSduNG0WZPWZf97s66KgpKQCTO_Z5jhQeiQ4DEjmJ51T3o-Xj2ywLkkYgPtEELEiOKq2ka7MT5hXArKyRbapnnFOavEDrqdaD0EpZeZN5lxg22yBpx9hbBMn1erIWbGh6yfQdaB71QP_7JJgPdECyoOIREqrmLzfx9tGuUiHHzEPfRwcX4_vRrd3F1eTyc3I82KajmqhSg4L1iVLq8FQFkoQlKtzAVWitdVRUyhjVI5sJILRsHopuFgoK4VpnQPna77zoN_GSD2srVRg3MqzThEmTPKcJF2ZQk9XqOPyoG0nfF9WneFy0lJc85ISUSixr9Q6TTQWu07MDblfwhOvglmoFw_i94NvfVd_AniNaiDjzGAkfNgWxWWkmC5Mk6-e_bNuCQ5-thuqFtovgSfriVgugYW3vUQ4rMbFhBkYp87v_izcSowJj8Np29MA63Z</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2434053814</pqid></control><display><type>article</type><title>Accuracy of fluid delivery devices for the neonate: Are the measures assured?</title><source>MEDLINE</source><source>Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals</source><source>African Journals Online (Open Access)</source><creator>Okoro, P ; Gbobo, I ; Igwe, P ; Umeh, D ; Okoro, C ; Nwiwu, P</creator><creatorcontrib>Okoro, P ; Gbobo, I ; Igwe, P ; Umeh, D ; Okoro, C ; Nwiwu, P</creatorcontrib><description>Introduction: Delivery of accurate volumes of fluid in surgical neonates and children is crucial for the good outcome of treatment. But how accurate are the calibrations on the fluid delivery devices? Aims: This study seeks to verify the accuracy of these devices in common use in our practice. Materials and Methods: This is a cross-sectional experimental study carried out in our center; a tertiary health facility in Southern Nigeria in May 2019. Fluid delivery devices (FDDs) used in the course of treatment of our pediatric patients were randomly included in the study. The number of drops per ml of each device was obtained by counting while the fluid dropped until a 1 ml volume was delivered. The data was then collated and analyzed. Results: A total of 215 FDDs were included in this study. They comprised infusion giving set, Soluset (Burette) giving set, and blood giving set. The rate of delivery was 20 drops/ml (infusion giving sets), 60 drops/min (Burette/Soluset), and 15 drops/ml (Blood giving set). They were all in keeping with the labeled/assumed calibration in each of the types of FDDs P &lt; 0.05. Therefore, the mean, median, and mode were the same. Conclusion: This study has demonstrated that the FDDs used our center are accurately calibrated and safe as they deliver volumes of fluid as labeled. The findings in this study reassure us of the dependability and accuracy of delivery of the FDDs we use in children in our center.</description><identifier>ISSN: 1119-3077</identifier><identifier>DOI: 10.4103/njcp.njcp_502_19</identifier><identifier>PMID: 32788479</identifier><language>eng</language><publisher>India: Wolters Kluwer India Pvt. Ltd</publisher><subject>Administration, Oral ; Care and treatment ; Child ; Child health ; Cross-Sectional Studies ; Equipment Safety - standards ; Female ; Fluid Therapy - instrumentation ; Fluid Therapy - methods ; Humans ; Infant, Newborn ; Infusion Pumps - standards ; Newborn infants ; Nigeria</subject><ispartof>Nigerian journal of clinical practice, 2020-08, Vol.23 (8), p.1044-1047</ispartof><rights>COPYRIGHT 2020 Medknow Publications and Media Pvt. Ltd.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c457y-b995885475478b9ee65a11c456290aa8b771f5cfaa2e468943efcdd8efebba033</citedby><cites>FETCH-LOGICAL-c457y-b995885475478b9ee65a11c456290aa8b771f5cfaa2e468943efcdd8efebba033</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27924,27925</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/32788479$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Okoro, P</creatorcontrib><creatorcontrib>Gbobo, I</creatorcontrib><creatorcontrib>Igwe, P</creatorcontrib><creatorcontrib>Umeh, D</creatorcontrib><creatorcontrib>Okoro, C</creatorcontrib><creatorcontrib>Nwiwu, P</creatorcontrib><title>Accuracy of fluid delivery devices for the neonate: Are the measures assured?</title><title>Nigerian journal of clinical practice</title><addtitle>Niger J Clin Pract</addtitle><description>Introduction: Delivery of accurate volumes of fluid in surgical neonates and children is crucial for the good outcome of treatment. But how accurate are the calibrations on the fluid delivery devices? Aims: This study seeks to verify the accuracy of these devices in common use in our practice. Materials and Methods: This is a cross-sectional experimental study carried out in our center; a tertiary health facility in Southern Nigeria in May 2019. Fluid delivery devices (FDDs) used in the course of treatment of our pediatric patients were randomly included in the study. The number of drops per ml of each device was obtained by counting while the fluid dropped until a 1 ml volume was delivered. The data was then collated and analyzed. Results: A total of 215 FDDs were included in this study. They comprised infusion giving set, Soluset (Burette) giving set, and blood giving set. The rate of delivery was 20 drops/ml (infusion giving sets), 60 drops/min (Burette/Soluset), and 15 drops/ml (Blood giving set). They were all in keeping with the labeled/assumed calibration in each of the types of FDDs P &lt; 0.05. Therefore, the mean, median, and mode were the same. Conclusion: This study has demonstrated that the FDDs used our center are accurately calibrated and safe as they deliver volumes of fluid as labeled. The findings in this study reassure us of the dependability and accuracy of delivery of the FDDs we use in children in our center.</description><subject>Administration, Oral</subject><subject>Care and treatment</subject><subject>Child</subject><subject>Child health</subject><subject>Cross-Sectional Studies</subject><subject>Equipment Safety - standards</subject><subject>Female</subject><subject>Fluid Therapy - instrumentation</subject><subject>Fluid Therapy - methods</subject><subject>Humans</subject><subject>Infant, Newborn</subject><subject>Infusion Pumps - standards</subject><subject>Newborn infants</subject><subject>Nigeria</subject><issn>1119-3077</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2020</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp1kc1L9DAQxnNQ_L57koIgXnZNmrRN3svLsvgFihc9hzSduNG0WZPWZf97s66KgpKQCTO_Z5jhQeiQ4DEjmJ51T3o-Xj2ywLkkYgPtEELEiOKq2ka7MT5hXArKyRbapnnFOavEDrqdaD0EpZeZN5lxg22yBpx9hbBMn1erIWbGh6yfQdaB71QP_7JJgPdECyoOIREqrmLzfx9tGuUiHHzEPfRwcX4_vRrd3F1eTyc3I82KajmqhSg4L1iVLq8FQFkoQlKtzAVWitdVRUyhjVI5sJILRsHopuFgoK4VpnQPna77zoN_GSD2srVRg3MqzThEmTPKcJF2ZQk9XqOPyoG0nfF9WneFy0lJc85ISUSixr9Q6TTQWu07MDblfwhOvglmoFw_i94NvfVd_AniNaiDjzGAkfNgWxWWkmC5Mk6-e_bNuCQ5-thuqFtovgSfriVgugYW3vUQ4rMbFhBkYp87v_izcSowJj8Np29MA63Z</recordid><startdate>20200801</startdate><enddate>20200801</enddate><creator>Okoro, P</creator><creator>Gbobo, I</creator><creator>Igwe, P</creator><creator>Umeh, D</creator><creator>Okoro, C</creator><creator>Nwiwu, P</creator><general>Wolters Kluwer India Pvt. Ltd</general><general>Medknow Publications and Media Pvt. Ltd</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>20200801</creationdate><title>Accuracy of fluid delivery devices for the neonate: Are the measures assured?</title><author>Okoro, P ; Gbobo, I ; Igwe, P ; Umeh, D ; Okoro, C ; Nwiwu, P</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c457y-b995885475478b9ee65a11c456290aa8b771f5cfaa2e468943efcdd8efebba033</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2020</creationdate><topic>Administration, Oral</topic><topic>Care and treatment</topic><topic>Child</topic><topic>Child health</topic><topic>Cross-Sectional Studies</topic><topic>Equipment Safety - standards</topic><topic>Female</topic><topic>Fluid Therapy - instrumentation</topic><topic>Fluid Therapy - methods</topic><topic>Humans</topic><topic>Infant, Newborn</topic><topic>Infusion Pumps - standards</topic><topic>Newborn infants</topic><topic>Nigeria</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Okoro, P</creatorcontrib><creatorcontrib>Gbobo, I</creatorcontrib><creatorcontrib>Igwe, P</creatorcontrib><creatorcontrib>Umeh, D</creatorcontrib><creatorcontrib>Okoro, C</creatorcontrib><creatorcontrib>Nwiwu, P</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Nigerian journal of clinical practice</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Okoro, P</au><au>Gbobo, I</au><au>Igwe, P</au><au>Umeh, D</au><au>Okoro, C</au><au>Nwiwu, P</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Accuracy of fluid delivery devices for the neonate: Are the measures assured?</atitle><jtitle>Nigerian journal of clinical practice</jtitle><addtitle>Niger J Clin Pract</addtitle><date>2020-08-01</date><risdate>2020</risdate><volume>23</volume><issue>8</issue><spage>1044</spage><epage>1047</epage><pages>1044-1047</pages><issn>1119-3077</issn><abstract>Introduction: Delivery of accurate volumes of fluid in surgical neonates and children is crucial for the good outcome of treatment. But how accurate are the calibrations on the fluid delivery devices? Aims: This study seeks to verify the accuracy of these devices in common use in our practice. Materials and Methods: This is a cross-sectional experimental study carried out in our center; a tertiary health facility in Southern Nigeria in May 2019. Fluid delivery devices (FDDs) used in the course of treatment of our pediatric patients were randomly included in the study. The number of drops per ml of each device was obtained by counting while the fluid dropped until a 1 ml volume was delivered. The data was then collated and analyzed. Results: A total of 215 FDDs were included in this study. They comprised infusion giving set, Soluset (Burette) giving set, and blood giving set. The rate of delivery was 20 drops/ml (infusion giving sets), 60 drops/min (Burette/Soluset), and 15 drops/ml (Blood giving set). They were all in keeping with the labeled/assumed calibration in each of the types of FDDs P &lt; 0.05. Therefore, the mean, median, and mode were the same. Conclusion: This study has demonstrated that the FDDs used our center are accurately calibrated and safe as they deliver volumes of fluid as labeled. The findings in this study reassure us of the dependability and accuracy of delivery of the FDDs we use in children in our center.</abstract><cop>India</cop><pub>Wolters Kluwer India Pvt. Ltd</pub><pmid>32788479</pmid><doi>10.4103/njcp.njcp_502_19</doi><tpages>4</tpages><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 1119-3077
ispartof Nigerian journal of clinical practice, 2020-08, Vol.23 (8), p.1044-1047
issn 1119-3077
language eng
recordid cdi_proquest_miscellaneous_2434053814
source MEDLINE; Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals; African Journals Online (Open Access)
subjects Administration, Oral
Care and treatment
Child
Child health
Cross-Sectional Studies
Equipment Safety - standards
Female
Fluid Therapy - instrumentation
Fluid Therapy - methods
Humans
Infant, Newborn
Infusion Pumps - standards
Newborn infants
Nigeria
title Accuracy of fluid delivery devices for the neonate: Are the measures assured?
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-03T03%3A34%3A48IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-gale_proqu&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Accuracy%20of%20fluid%20delivery%20devices%20for%20the%20neonate:%20Are%20the%20measures%20assured?&rft.jtitle=Nigerian%20journal%20of%20clinical%20practice&rft.au=Okoro,%20P&rft.date=2020-08-01&rft.volume=23&rft.issue=8&rft.spage=1044&rft.epage=1047&rft.pages=1044-1047&rft.issn=1119-3077&rft_id=info:doi/10.4103/njcp.njcp_502_19&rft_dat=%3Cgale_proqu%3EA632841619%3C/gale_proqu%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2434053814&rft_id=info:pmid/32788479&rft_galeid=A632841619&rfr_iscdi=true