Decreased RPM reduces von Willebrand factor degradation with the EVAHEART LVAS: implications for device‐specific LVAD management

Background Continuous‐flow left ventricular assist devices (LVADs) produces supraphysiologic shear stress that causes von Willebrand factor (VWF) degradation and a bleeding diathesis. Reduction of revolutions per minute (RPM) with axial‐flow LVADs does not decrease shear stress enough to reduce VWF...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of cardiac surgery 2020-07, Vol.35 (7), p.1477-1483
Hauptverfasser: Bartoli, Carlo R., Kang, Jooeun, Motomura, Tadashi
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 1483
container_issue 7
container_start_page 1477
container_title Journal of cardiac surgery
container_volume 35
creator Bartoli, Carlo R.
Kang, Jooeun
Motomura, Tadashi
description Background Continuous‐flow left ventricular assist devices (LVADs) produces supraphysiologic shear stress that causes von Willebrand factor (VWF) degradation and a bleeding diathesis. Reduction of revolutions per minute (RPM) with axial‐flow LVADs does not decrease shear stress enough to reduce VWF degradation and bleeding. However, it is unknown if RPM reduction with centrifugal flow LVADs may minimize VWF degradation. We tested the hypothesis that RPM reduction preserves VWF multimers in the centrifugal‐flow EVAHEART left ventricular assist system (LVAS), which is designed to minimize shear stress and blood trauma. Methods Whole blood samples were collected from humans (n = 28). Blood was circulated in ex vivo mock circulatory loops for 6 hours with an EVAHEART LVAS at 2300 (n = 12), 2100 (n = 8), or 1800 RPM (n = 8). Immunoblotting was used to resolve and quantify VWF multimers and degradation fragments. Results RPM reduction from 2300 to 2100 to 1800 RPM significantly decreased EVAHEART blood flow from 5.8 ± 0.4 to 4.3 ± 0.6 to 4.1 ± 0.5 L/min (analysis of variance [ANOVA], P = .03). RPM reduction protected VWF from pathologic degradation. At lower RPMs, significantly greater levels of VWF multimers were observed (ANOVA, P = .001). Similarly, at lower RPMs, significantly fewer VWF fragments, a product of VWF degradation, were observed (ANOVA, P = .007). Conclusions RPM reduction significantly reduced VWF degradation with the centrifugal‐flow EVAHEART LVAS, an LVAD specifically designed with low shear stress. Different LVADs have unique hematologic footprints and should be managed with device‐specific protocols. Adjustment of RPM to minimize blood trauma while still maintaining physiologic hemodynamics has the potential to decrease complications related to LVAD‐associated von Willebrand's disease, such as gastrointestinal bleeding and hemorrhagic stroke.
doi_str_mv 10.1111/jocs.14620
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2423069366</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2423069366</sourcerecordid><originalsourceid>FETCH-LOGICAL-c3420-be9f87b287e4d4d39f86ca2b70e50ae88258443cdeb665dc4b375f5bae6220a3</originalsourceid><addsrcrecordid>eNp90M1OwkAQB_CN0URELz7BHo1Jcbvdbou3BlA0GAwQPDbb3Sks6Ze7RcLN-AQ-o09ioZ6dy2Qyv5nDH6Frl_Tcpu42pbQ9l3FKTlDH9RlxQrfvnqIOCUPuEMbIObqwdkMIpcwjHfQ1BGlAWFB49vqCDaitBIs_ygK_6SyDxIhC4VTIujRYwcoIJWrdbHe6XuN6DXi0jMajaLbAk2U0v8c6rzItj8bi9Hj0oSX8fH7bCqROtTzAIc5FIVaQQ1FforNUZBau_noXLR5Gi8HYmUwfnwbRxJEeo8RJoJ-GQULDAJhiymsmLgVNAgI-ERCG1A8Z86SChHNfSZZ4gZ_6iQBOKRFeF920bytTvm_B1nGurYQsEwWUWxtTRj3C-x7nDb1tqTSltQbSuDI6F2YfuyQ-5Bwfco6POTfYbfFOZ7D_R8bP08G8vfkF4oKB0A</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2423069366</pqid></control><display><type>article</type><title>Decreased RPM reduces von Willebrand factor degradation with the EVAHEART LVAS: implications for device‐specific LVAD management</title><source>Access via Wiley Online Library</source><creator>Bartoli, Carlo R. ; Kang, Jooeun ; Motomura, Tadashi</creator><creatorcontrib>Bartoli, Carlo R. ; Kang, Jooeun ; Motomura, Tadashi</creatorcontrib><description>Background Continuous‐flow left ventricular assist devices (LVADs) produces supraphysiologic shear stress that causes von Willebrand factor (VWF) degradation and a bleeding diathesis. Reduction of revolutions per minute (RPM) with axial‐flow LVADs does not decrease shear stress enough to reduce VWF degradation and bleeding. However, it is unknown if RPM reduction with centrifugal flow LVADs may minimize VWF degradation. We tested the hypothesis that RPM reduction preserves VWF multimers in the centrifugal‐flow EVAHEART left ventricular assist system (LVAS), which is designed to minimize shear stress and blood trauma. Methods Whole blood samples were collected from humans (n = 28). Blood was circulated in ex vivo mock circulatory loops for 6 hours with an EVAHEART LVAS at 2300 (n = 12), 2100 (n = 8), or 1800 RPM (n = 8). Immunoblotting was used to resolve and quantify VWF multimers and degradation fragments. Results RPM reduction from 2300 to 2100 to 1800 RPM significantly decreased EVAHEART blood flow from 5.8 ± 0.4 to 4.3 ± 0.6 to 4.1 ± 0.5 L/min (analysis of variance [ANOVA], P = .03). RPM reduction protected VWF from pathologic degradation. At lower RPMs, significantly greater levels of VWF multimers were observed (ANOVA, P = .001). Similarly, at lower RPMs, significantly fewer VWF fragments, a product of VWF degradation, were observed (ANOVA, P = .007). Conclusions RPM reduction significantly reduced VWF degradation with the centrifugal‐flow EVAHEART LVAS, an LVAD specifically designed with low shear stress. Different LVADs have unique hematologic footprints and should be managed with device‐specific protocols. Adjustment of RPM to minimize blood trauma while still maintaining physiologic hemodynamics has the potential to decrease complications related to LVAD‐associated von Willebrand's disease, such as gastrointestinal bleeding and hemorrhagic stroke.</description><identifier>ISSN: 0886-0440</identifier><identifier>EISSN: 1540-8191</identifier><identifier>DOI: 10.1111/jocs.14620</identifier><language>eng</language><subject>bleeding ; left ventricular assist device ; mock circulatory loop ; revolutions per minute ; shear stress ; von Willebrand factor</subject><ispartof>Journal of cardiac surgery, 2020-07, Vol.35 (7), p.1477-1483</ispartof><rights>2020 Wiley Periodicals LLC</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c3420-be9f87b287e4d4d39f86ca2b70e50ae88258443cdeb665dc4b375f5bae6220a3</citedby><cites>FETCH-LOGICAL-c3420-be9f87b287e4d4d39f86ca2b70e50ae88258443cdeb665dc4b375f5bae6220a3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://onlinelibrary.wiley.com/doi/pdf/10.1111%2Fjocs.14620$$EPDF$$P50$$Gwiley$$H</linktopdf><linktohtml>$$Uhttps://onlinelibrary.wiley.com/doi/full/10.1111%2Fjocs.14620$$EHTML$$P50$$Gwiley$$H</linktohtml><link.rule.ids>314,780,784,1417,27924,27925,45574,45575</link.rule.ids></links><search><creatorcontrib>Bartoli, Carlo R.</creatorcontrib><creatorcontrib>Kang, Jooeun</creatorcontrib><creatorcontrib>Motomura, Tadashi</creatorcontrib><title>Decreased RPM reduces von Willebrand factor degradation with the EVAHEART LVAS: implications for device‐specific LVAD management</title><title>Journal of cardiac surgery</title><description>Background Continuous‐flow left ventricular assist devices (LVADs) produces supraphysiologic shear stress that causes von Willebrand factor (VWF) degradation and a bleeding diathesis. Reduction of revolutions per minute (RPM) with axial‐flow LVADs does not decrease shear stress enough to reduce VWF degradation and bleeding. However, it is unknown if RPM reduction with centrifugal flow LVADs may minimize VWF degradation. We tested the hypothesis that RPM reduction preserves VWF multimers in the centrifugal‐flow EVAHEART left ventricular assist system (LVAS), which is designed to minimize shear stress and blood trauma. Methods Whole blood samples were collected from humans (n = 28). Blood was circulated in ex vivo mock circulatory loops for 6 hours with an EVAHEART LVAS at 2300 (n = 12), 2100 (n = 8), or 1800 RPM (n = 8). Immunoblotting was used to resolve and quantify VWF multimers and degradation fragments. Results RPM reduction from 2300 to 2100 to 1800 RPM significantly decreased EVAHEART blood flow from 5.8 ± 0.4 to 4.3 ± 0.6 to 4.1 ± 0.5 L/min (analysis of variance [ANOVA], P = .03). RPM reduction protected VWF from pathologic degradation. At lower RPMs, significantly greater levels of VWF multimers were observed (ANOVA, P = .001). Similarly, at lower RPMs, significantly fewer VWF fragments, a product of VWF degradation, were observed (ANOVA, P = .007). Conclusions RPM reduction significantly reduced VWF degradation with the centrifugal‐flow EVAHEART LVAS, an LVAD specifically designed with low shear stress. Different LVADs have unique hematologic footprints and should be managed with device‐specific protocols. Adjustment of RPM to minimize blood trauma while still maintaining physiologic hemodynamics has the potential to decrease complications related to LVAD‐associated von Willebrand's disease, such as gastrointestinal bleeding and hemorrhagic stroke.</description><subject>bleeding</subject><subject>left ventricular assist device</subject><subject>mock circulatory loop</subject><subject>revolutions per minute</subject><subject>shear stress</subject><subject>von Willebrand factor</subject><issn>0886-0440</issn><issn>1540-8191</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2020</creationdate><recordtype>article</recordtype><recordid>eNp90M1OwkAQB_CN0URELz7BHo1Jcbvdbou3BlA0GAwQPDbb3Sks6Ze7RcLN-AQ-o09ioZ6dy2Qyv5nDH6Frl_Tcpu42pbQ9l3FKTlDH9RlxQrfvnqIOCUPuEMbIObqwdkMIpcwjHfQ1BGlAWFB49vqCDaitBIs_ygK_6SyDxIhC4VTIujRYwcoIJWrdbHe6XuN6DXi0jMajaLbAk2U0v8c6rzItj8bi9Hj0oSX8fH7bCqROtTzAIc5FIVaQQ1FforNUZBau_noXLR5Gi8HYmUwfnwbRxJEeo8RJoJ-GQULDAJhiymsmLgVNAgI-ERCG1A8Z86SChHNfSZZ4gZ_6iQBOKRFeF920bytTvm_B1nGurYQsEwWUWxtTRj3C-x7nDb1tqTSltQbSuDI6F2YfuyQ-5Bwfco6POTfYbfFOZ7D_R8bP08G8vfkF4oKB0A</recordid><startdate>202007</startdate><enddate>202007</enddate><creator>Bartoli, Carlo R.</creator><creator>Kang, Jooeun</creator><creator>Motomura, Tadashi</creator><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>202007</creationdate><title>Decreased RPM reduces von Willebrand factor degradation with the EVAHEART LVAS: implications for device‐specific LVAD management</title><author>Bartoli, Carlo R. ; Kang, Jooeun ; Motomura, Tadashi</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c3420-be9f87b287e4d4d39f86ca2b70e50ae88258443cdeb665dc4b375f5bae6220a3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2020</creationdate><topic>bleeding</topic><topic>left ventricular assist device</topic><topic>mock circulatory loop</topic><topic>revolutions per minute</topic><topic>shear stress</topic><topic>von Willebrand factor</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Bartoli, Carlo R.</creatorcontrib><creatorcontrib>Kang, Jooeun</creatorcontrib><creatorcontrib>Motomura, Tadashi</creatorcontrib><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Journal of cardiac surgery</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Bartoli, Carlo R.</au><au>Kang, Jooeun</au><au>Motomura, Tadashi</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Decreased RPM reduces von Willebrand factor degradation with the EVAHEART LVAS: implications for device‐specific LVAD management</atitle><jtitle>Journal of cardiac surgery</jtitle><date>2020-07</date><risdate>2020</risdate><volume>35</volume><issue>7</issue><spage>1477</spage><epage>1483</epage><pages>1477-1483</pages><issn>0886-0440</issn><eissn>1540-8191</eissn><abstract>Background Continuous‐flow left ventricular assist devices (LVADs) produces supraphysiologic shear stress that causes von Willebrand factor (VWF) degradation and a bleeding diathesis. Reduction of revolutions per minute (RPM) with axial‐flow LVADs does not decrease shear stress enough to reduce VWF degradation and bleeding. However, it is unknown if RPM reduction with centrifugal flow LVADs may minimize VWF degradation. We tested the hypothesis that RPM reduction preserves VWF multimers in the centrifugal‐flow EVAHEART left ventricular assist system (LVAS), which is designed to minimize shear stress and blood trauma. Methods Whole blood samples were collected from humans (n = 28). Blood was circulated in ex vivo mock circulatory loops for 6 hours with an EVAHEART LVAS at 2300 (n = 12), 2100 (n = 8), or 1800 RPM (n = 8). Immunoblotting was used to resolve and quantify VWF multimers and degradation fragments. Results RPM reduction from 2300 to 2100 to 1800 RPM significantly decreased EVAHEART blood flow from 5.8 ± 0.4 to 4.3 ± 0.6 to 4.1 ± 0.5 L/min (analysis of variance [ANOVA], P = .03). RPM reduction protected VWF from pathologic degradation. At lower RPMs, significantly greater levels of VWF multimers were observed (ANOVA, P = .001). Similarly, at lower RPMs, significantly fewer VWF fragments, a product of VWF degradation, were observed (ANOVA, P = .007). Conclusions RPM reduction significantly reduced VWF degradation with the centrifugal‐flow EVAHEART LVAS, an LVAD specifically designed with low shear stress. Different LVADs have unique hematologic footprints and should be managed with device‐specific protocols. Adjustment of RPM to minimize blood trauma while still maintaining physiologic hemodynamics has the potential to decrease complications related to LVAD‐associated von Willebrand's disease, such as gastrointestinal bleeding and hemorrhagic stroke.</abstract><doi>10.1111/jocs.14620</doi><tpages>7</tpages><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 0886-0440
ispartof Journal of cardiac surgery, 2020-07, Vol.35 (7), p.1477-1483
issn 0886-0440
1540-8191
language eng
recordid cdi_proquest_miscellaneous_2423069366
source Access via Wiley Online Library
subjects bleeding
left ventricular assist device
mock circulatory loop
revolutions per minute
shear stress
von Willebrand factor
title Decreased RPM reduces von Willebrand factor degradation with the EVAHEART LVAS: implications for device‐specific LVAD management
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-25T11%3A37%3A38IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Decreased%20RPM%20reduces%20von%20Willebrand%20factor%20degradation%20with%20the%20EVAHEART%20LVAS:%20implications%20for%20device%E2%80%90specific%20LVAD%20management&rft.jtitle=Journal%20of%20cardiac%20surgery&rft.au=Bartoli,%20Carlo%20R.&rft.date=2020-07&rft.volume=35&rft.issue=7&rft.spage=1477&rft.epage=1483&rft.pages=1477-1483&rft.issn=0886-0440&rft.eissn=1540-8191&rft_id=info:doi/10.1111/jocs.14620&rft_dat=%3Cproquest_cross%3E2423069366%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2423069366&rft_id=info:pmid/&rfr_iscdi=true