2D multislice MP2RAGE sequence for fast T1 mapping at 7 T: Application to mouse imaging and MR thermometry
Purpose To develop a 2D radial multislice MP2RAGE sequence for fast and reliable T1 mapping at 7 T in mice and for MR thermometry. Methods The 2D‐MP2RAGE sequence was performed with the following parameters: TI1‐TI2‐MP2RAGETR = 1000‐3000‐9000 ms. The multiple dead times within the sequence were used...
Gespeichert in:
Veröffentlicht in: | Magnetic resonance in medicine 2020-09, Vol.84 (3), p.1430-1440 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 1440 |
---|---|
container_issue | 3 |
container_start_page | 1430 |
container_title | Magnetic resonance in medicine |
container_volume | 84 |
creator | Faller, Thibaut L. Trotier, Aurélien J. Rousseau, Alice F. Franconi, Jean‐Michel Miraux, Sylvain Ribot, Emeline J. |
description | Purpose
To develop a 2D radial multislice MP2RAGE sequence for fast and reliable T1 mapping at 7 T in mice and for MR thermometry.
Methods
The 2D‐MP2RAGE sequence was performed with the following parameters: TI1‐TI2‐MP2RAGETR = 1000‐3000‐9000 ms. The multiple dead times within the sequence were used for interleaved multislice acquisition, enabling one to acquire six slices in 9 seconds. The excitation pulse shape, inversion selectivity, and interslice gap were optimized. In vitro comparison with the inversion‐recovery sequence was performed. The T1 variations with temperature were measured on tubes with T1 ranging from 800 ms to 2000 ms. The sequence was used to acquire T1 maps continuously during 30 minutes on the brain and abdomen of healthy mice.
Results
A three‐lobe cardinal sine excitation pulse, combined with an inversion slice thickness and an interslice gap of respectively 150% and 50% of the imaging slice thickness, led to a SD and bias of the T1 measurements below 1% and 2%, respectively. A linear dependence of T1 with temperature was measured between 10°C and 60°C. In vivo, less than 1% variation was measured between successive T1 maps in the mouse brain. In the abdomen, no obvious in‐plane motion artifacts were observed but respiratory motion in the slice dimension led to 6% T1 underestimation.
Conclusion
The multislice MP2RAGE sequence could be used for fast whole‐body T1 mapping and MR thermometry. Its reconstruction method would enable on‐the‐fly reconstruction. |
doi_str_mv | 10.1002/mrm.28220 |
format | Article |
fullrecord | <record><control><sourceid>proquest_wiley</sourceid><recordid>TN_cdi_proquest_miscellaneous_2408539705</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2408539705</sourcerecordid><originalsourceid>FETCH-LOGICAL-p2910-2cf29acc64b9ee25a06bf72716126487614fd6f62b844bdcb11fc501c7ad21f43</originalsourceid><addsrcrecordid>eNpdkMFOwzAQRC0EEqVw4A8sceGS1t44ccytKqUgNQJV5Ww5qV1SxUmIHaH-PW7LidOuVm9WM4PQPSUTSghMbW8nkAGQCzSiCUAEiWCXaEQ4I1FMBbtGN87tCSFCcDZCe3jGdqh95eqq1Dj_gPVsucBOfw-6CQfT9tgo5_GGYqu6rmp2WHnM8eYJz7ouiJSv2gb7Ftt2cBpXVu1OULPF-Rr7L93b1mrfH27RlVG103d_c4w-Xxab-Wu0el--zWerqANBSQSlAaHKMmWF0BoSRdLCcOA0pZCyjKeUmW1qUigyxoptWVBqyoTQkqstUMPiMXo8_-36NqRwXtrKlbquVaODRQmMZEksOEkC-vAP3bdD3wR3R4pnjNI0C9T0TP1UtT7Irg8Z-4OkRB4rl6Fyeapc5uv8tMS_t4JzzA</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2407841168</pqid></control><display><type>article</type><title>2D multislice MP2RAGE sequence for fast T1 mapping at 7 T: Application to mouse imaging and MR thermometry</title><source>Wiley Online Library Journals Frontfile Complete</source><creator>Faller, Thibaut L. ; Trotier, Aurélien J. ; Rousseau, Alice F. ; Franconi, Jean‐Michel ; Miraux, Sylvain ; Ribot, Emeline J.</creator><creatorcontrib>Faller, Thibaut L. ; Trotier, Aurélien J. ; Rousseau, Alice F. ; Franconi, Jean‐Michel ; Miraux, Sylvain ; Ribot, Emeline J.</creatorcontrib><description>Purpose
To develop a 2D radial multislice MP2RAGE sequence for fast and reliable T1 mapping at 7 T in mice and for MR thermometry.
Methods
The 2D‐MP2RAGE sequence was performed with the following parameters: TI1‐TI2‐MP2RAGETR = 1000‐3000‐9000 ms. The multiple dead times within the sequence were used for interleaved multislice acquisition, enabling one to acquire six slices in 9 seconds. The excitation pulse shape, inversion selectivity, and interslice gap were optimized. In vitro comparison with the inversion‐recovery sequence was performed. The T1 variations with temperature were measured on tubes with T1 ranging from 800 ms to 2000 ms. The sequence was used to acquire T1 maps continuously during 30 minutes on the brain and abdomen of healthy mice.
Results
A three‐lobe cardinal sine excitation pulse, combined with an inversion slice thickness and an interslice gap of respectively 150% and 50% of the imaging slice thickness, led to a SD and bias of the T1 measurements below 1% and 2%, respectively. A linear dependence of T1 with temperature was measured between 10°C and 60°C. In vivo, less than 1% variation was measured between successive T1 maps in the mouse brain. In the abdomen, no obvious in‐plane motion artifacts were observed but respiratory motion in the slice dimension led to 6% T1 underestimation.
Conclusion
The multislice MP2RAGE sequence could be used for fast whole‐body T1 mapping and MR thermometry. Its reconstruction method would enable on‐the‐fly reconstruction.</description><identifier>ISSN: 0740-3194</identifier><identifier>EISSN: 1522-2594</identifier><identifier>DOI: 10.1002/mrm.28220</identifier><language>eng</language><publisher>Hoboken: Wiley Subscription Services, Inc</publisher><subject>2D multislice ; 7 T ; Abdomen ; Brain ; Brain mapping ; Excitation ; Image reconstruction ; Inversion ; Magnetic resonance imaging ; Mapping ; mouse ; MP2RAGE ; Neuroimaging ; Pulse shape ; Selectivity ; T1 mapping ; Temperature dependence ; Thermal imaging ; Thermometry ; Thickness ; Tubes</subject><ispartof>Magnetic resonance in medicine, 2020-09, Vol.84 (3), p.1430-1440</ispartof><rights>2020 International Society for Magnetic Resonance in Medicine</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><orcidid>0000-0002-3063-1263</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://onlinelibrary.wiley.com/doi/pdf/10.1002%2Fmrm.28220$$EPDF$$P50$$Gwiley$$H</linktopdf><linktohtml>$$Uhttps://onlinelibrary.wiley.com/doi/full/10.1002%2Fmrm.28220$$EHTML$$P50$$Gwiley$$H</linktohtml><link.rule.ids>314,776,780,1411,27903,27904,45553,45554</link.rule.ids></links><search><creatorcontrib>Faller, Thibaut L.</creatorcontrib><creatorcontrib>Trotier, Aurélien J.</creatorcontrib><creatorcontrib>Rousseau, Alice F.</creatorcontrib><creatorcontrib>Franconi, Jean‐Michel</creatorcontrib><creatorcontrib>Miraux, Sylvain</creatorcontrib><creatorcontrib>Ribot, Emeline J.</creatorcontrib><title>2D multislice MP2RAGE sequence for fast T1 mapping at 7 T: Application to mouse imaging and MR thermometry</title><title>Magnetic resonance in medicine</title><description>Purpose
To develop a 2D radial multislice MP2RAGE sequence for fast and reliable T1 mapping at 7 T in mice and for MR thermometry.
Methods
The 2D‐MP2RAGE sequence was performed with the following parameters: TI1‐TI2‐MP2RAGETR = 1000‐3000‐9000 ms. The multiple dead times within the sequence were used for interleaved multislice acquisition, enabling one to acquire six slices in 9 seconds. The excitation pulse shape, inversion selectivity, and interslice gap were optimized. In vitro comparison with the inversion‐recovery sequence was performed. The T1 variations with temperature were measured on tubes with T1 ranging from 800 ms to 2000 ms. The sequence was used to acquire T1 maps continuously during 30 minutes on the brain and abdomen of healthy mice.
Results
A three‐lobe cardinal sine excitation pulse, combined with an inversion slice thickness and an interslice gap of respectively 150% and 50% of the imaging slice thickness, led to a SD and bias of the T1 measurements below 1% and 2%, respectively. A linear dependence of T1 with temperature was measured between 10°C and 60°C. In vivo, less than 1% variation was measured between successive T1 maps in the mouse brain. In the abdomen, no obvious in‐plane motion artifacts were observed but respiratory motion in the slice dimension led to 6% T1 underestimation.
Conclusion
The multislice MP2RAGE sequence could be used for fast whole‐body T1 mapping and MR thermometry. Its reconstruction method would enable on‐the‐fly reconstruction.</description><subject>2D multislice</subject><subject>7 T</subject><subject>Abdomen</subject><subject>Brain</subject><subject>Brain mapping</subject><subject>Excitation</subject><subject>Image reconstruction</subject><subject>Inversion</subject><subject>Magnetic resonance imaging</subject><subject>Mapping</subject><subject>mouse</subject><subject>MP2RAGE</subject><subject>Neuroimaging</subject><subject>Pulse shape</subject><subject>Selectivity</subject><subject>T1 mapping</subject><subject>Temperature dependence</subject><subject>Thermal imaging</subject><subject>Thermometry</subject><subject>Thickness</subject><subject>Tubes</subject><issn>0740-3194</issn><issn>1522-2594</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2020</creationdate><recordtype>article</recordtype><recordid>eNpdkMFOwzAQRC0EEqVw4A8sceGS1t44ccytKqUgNQJV5Ww5qV1SxUmIHaH-PW7LidOuVm9WM4PQPSUTSghMbW8nkAGQCzSiCUAEiWCXaEQ4I1FMBbtGN87tCSFCcDZCe3jGdqh95eqq1Dj_gPVsucBOfw-6CQfT9tgo5_GGYqu6rmp2WHnM8eYJz7ouiJSv2gb7Ftt2cBpXVu1OULPF-Rr7L93b1mrfH27RlVG103d_c4w-Xxab-Wu0el--zWerqANBSQSlAaHKMmWF0BoSRdLCcOA0pZCyjKeUmW1qUigyxoptWVBqyoTQkqstUMPiMXo8_-36NqRwXtrKlbquVaODRQmMZEksOEkC-vAP3bdD3wR3R4pnjNI0C9T0TP1UtT7Irg8Z-4OkRB4rl6Fyeapc5uv8tMS_t4JzzA</recordid><startdate>202009</startdate><enddate>202009</enddate><creator>Faller, Thibaut L.</creator><creator>Trotier, Aurélien J.</creator><creator>Rousseau, Alice F.</creator><creator>Franconi, Jean‐Michel</creator><creator>Miraux, Sylvain</creator><creator>Ribot, Emeline J.</creator><general>Wiley Subscription Services, Inc</general><scope>8FD</scope><scope>FR3</scope><scope>K9.</scope><scope>M7Z</scope><scope>P64</scope><scope>7X8</scope><orcidid>https://orcid.org/0000-0002-3063-1263</orcidid></search><sort><creationdate>202009</creationdate><title>2D multislice MP2RAGE sequence for fast T1 mapping at 7 T: Application to mouse imaging and MR thermometry</title><author>Faller, Thibaut L. ; Trotier, Aurélien J. ; Rousseau, Alice F. ; Franconi, Jean‐Michel ; Miraux, Sylvain ; Ribot, Emeline J.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-p2910-2cf29acc64b9ee25a06bf72716126487614fd6f62b844bdcb11fc501c7ad21f43</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2020</creationdate><topic>2D multislice</topic><topic>7 T</topic><topic>Abdomen</topic><topic>Brain</topic><topic>Brain mapping</topic><topic>Excitation</topic><topic>Image reconstruction</topic><topic>Inversion</topic><topic>Magnetic resonance imaging</topic><topic>Mapping</topic><topic>mouse</topic><topic>MP2RAGE</topic><topic>Neuroimaging</topic><topic>Pulse shape</topic><topic>Selectivity</topic><topic>T1 mapping</topic><topic>Temperature dependence</topic><topic>Thermal imaging</topic><topic>Thermometry</topic><topic>Thickness</topic><topic>Tubes</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Faller, Thibaut L.</creatorcontrib><creatorcontrib>Trotier, Aurélien J.</creatorcontrib><creatorcontrib>Rousseau, Alice F.</creatorcontrib><creatorcontrib>Franconi, Jean‐Michel</creatorcontrib><creatorcontrib>Miraux, Sylvain</creatorcontrib><creatorcontrib>Ribot, Emeline J.</creatorcontrib><collection>Technology Research Database</collection><collection>Engineering Research Database</collection><collection>ProQuest Health & Medical Complete (Alumni)</collection><collection>Biochemistry Abstracts 1</collection><collection>Biotechnology and BioEngineering Abstracts</collection><collection>MEDLINE - Academic</collection><jtitle>Magnetic resonance in medicine</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Faller, Thibaut L.</au><au>Trotier, Aurélien J.</au><au>Rousseau, Alice F.</au><au>Franconi, Jean‐Michel</au><au>Miraux, Sylvain</au><au>Ribot, Emeline J.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>2D multislice MP2RAGE sequence for fast T1 mapping at 7 T: Application to mouse imaging and MR thermometry</atitle><jtitle>Magnetic resonance in medicine</jtitle><date>2020-09</date><risdate>2020</risdate><volume>84</volume><issue>3</issue><spage>1430</spage><epage>1440</epage><pages>1430-1440</pages><issn>0740-3194</issn><eissn>1522-2594</eissn><abstract>Purpose
To develop a 2D radial multislice MP2RAGE sequence for fast and reliable T1 mapping at 7 T in mice and for MR thermometry.
Methods
The 2D‐MP2RAGE sequence was performed with the following parameters: TI1‐TI2‐MP2RAGETR = 1000‐3000‐9000 ms. The multiple dead times within the sequence were used for interleaved multislice acquisition, enabling one to acquire six slices in 9 seconds. The excitation pulse shape, inversion selectivity, and interslice gap were optimized. In vitro comparison with the inversion‐recovery sequence was performed. The T1 variations with temperature were measured on tubes with T1 ranging from 800 ms to 2000 ms. The sequence was used to acquire T1 maps continuously during 30 minutes on the brain and abdomen of healthy mice.
Results
A three‐lobe cardinal sine excitation pulse, combined with an inversion slice thickness and an interslice gap of respectively 150% and 50% of the imaging slice thickness, led to a SD and bias of the T1 measurements below 1% and 2%, respectively. A linear dependence of T1 with temperature was measured between 10°C and 60°C. In vivo, less than 1% variation was measured between successive T1 maps in the mouse brain. In the abdomen, no obvious in‐plane motion artifacts were observed but respiratory motion in the slice dimension led to 6% T1 underestimation.
Conclusion
The multislice MP2RAGE sequence could be used for fast whole‐body T1 mapping and MR thermometry. Its reconstruction method would enable on‐the‐fly reconstruction.</abstract><cop>Hoboken</cop><pub>Wiley Subscription Services, Inc</pub><doi>10.1002/mrm.28220</doi><tpages>11</tpages><orcidid>https://orcid.org/0000-0002-3063-1263</orcidid><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0740-3194 |
ispartof | Magnetic resonance in medicine, 2020-09, Vol.84 (3), p.1430-1440 |
issn | 0740-3194 1522-2594 |
language | eng |
recordid | cdi_proquest_miscellaneous_2408539705 |
source | Wiley Online Library Journals Frontfile Complete |
subjects | 2D multislice 7 T Abdomen Brain Brain mapping Excitation Image reconstruction Inversion Magnetic resonance imaging Mapping mouse MP2RAGE Neuroimaging Pulse shape Selectivity T1 mapping Temperature dependence Thermal imaging Thermometry Thickness Tubes |
title | 2D multislice MP2RAGE sequence for fast T1 mapping at 7 T: Application to mouse imaging and MR thermometry |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-27T02%3A19%3A51IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_wiley&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=2D%20multislice%20MP2RAGE%20sequence%20for%20fast%20T1%20mapping%20at%207%20T:%20Application%20to%20mouse%20imaging%20and%20MR%20thermometry&rft.jtitle=Magnetic%20resonance%20in%20medicine&rft.au=Faller,%20Thibaut%20L.&rft.date=2020-09&rft.volume=84&rft.issue=3&rft.spage=1430&rft.epage=1440&rft.pages=1430-1440&rft.issn=0740-3194&rft.eissn=1522-2594&rft_id=info:doi/10.1002/mrm.28220&rft_dat=%3Cproquest_wiley%3E2408539705%3C/proquest_wiley%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2407841168&rft_id=info:pmid/&rfr_iscdi=true |