Ultrasound Capsule Endoscopy With a Mechanically Scanning Micro-ultrasound: A Porcine Study

Wireless capsule endoscopy has been used for the clinical examination of the gastrointestinal (GI) tract for two decades. However, most commercially available devices only utilise optical imaging to examine the GI wall surface. Using this sensing modality, pathology within the GI wall cannot be dete...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Ultrasound in medicine & biology 2020-03, Vol.46 (3), p.796-804
Hauptverfasser: Qiu, Yongqiang, Huang, Yaocai, Zhang, Zhiqiang, Cox, Benjamin F., Liu, Rong, Hong, Jiehan, Mu, Peitian, Lay, Holly S., Cummins, Gerard, Desmulliez, Marc P.Y., Clutton, Eddie, Zheng, Hairong, Qiu, Weibao, Cochran, Sandy
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 804
container_issue 3
container_start_page 796
container_title Ultrasound in medicine & biology
container_volume 46
creator Qiu, Yongqiang
Huang, Yaocai
Zhang, Zhiqiang
Cox, Benjamin F.
Liu, Rong
Hong, Jiehan
Mu, Peitian
Lay, Holly S.
Cummins, Gerard
Desmulliez, Marc P.Y.
Clutton, Eddie
Zheng, Hairong
Qiu, Weibao
Cochran, Sandy
description Wireless capsule endoscopy has been used for the clinical examination of the gastrointestinal (GI) tract for two decades. However, most commercially available devices only utilise optical imaging to examine the GI wall surface. Using this sensing modality, pathology within the GI wall cannot be detected. Micro-ultrasound (μUS) using high-frequency (>20 MHz) ultrasound can provide a means of transmural or cross-sectional image of the GI tract. Depth of imaging is approximately 10 mm with a resolution of between 40–120 μm that is sufficient to differentiate between subsurface histologic layers of the various regions of the GI tract. Ultrasound capsule endoscopy (USCE) uses a capsule equipped with μUS transducers that are capable of imaging below the GI wall surface, offering thereby a complementary sensing technique to optical imaging capsule endoscopy. In this work, a USCE device integrated with a ∼30 MHz ultrasonic transducer was developed to capture a full 360° image of the lumen. The performance of the device was initially evaluated using a wire phantom, indicating an axial resolution of 69.0 μm and lateral resolution of 262.5 μm. Later, in vivo imaging performance was characterised in the oesophagus and small intestine of anaesthetized pigs. The reconstructed images demonstrate clear layer differentiation of the lumen wall. The tissue thicknesses measured from the B-scan images show good agreement with ex vivo images from the literature. The high-resolution ultrasound images in the in vivo porcine model achieved with this device is an encouraging preliminary step in the translation of these devices toward future clinical use.
doi_str_mv 10.1016/j.ultrasmedbio.2019.12.003
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2333926005</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S0301562919316230</els_id><sourcerecordid>2333926005</sourcerecordid><originalsourceid>FETCH-LOGICAL-c432t-2fbca5ad440211a3b563e81ce12c0ebb50df5b4edcb935636066c145fb2be7a43</originalsourceid><addsrcrecordid>eNqNkEtPGzEQx62Kqgm0X6Gyeupll7G962S5ofCUQEUC1Eo9WH7MFkcbO9i7SPn2bJoU9chpDv_HzPwI-cagZMDk8bIcuj7pvEJnfCw5sKZkvAQQH8iUzWdNwRv264BMQQArasmbCTnMeQkAMylmn8hEsAZ4Vckp-f34tyoOwdGFXuehQ3oeXMw2rjf0p--fqKa3aJ908FZ33YbeWx2CD3_orbcpFsNb_oSe0ruYrA9I7_vBbT6Tj63uMn7ZzyPyeHH-sLgqbn5cXi9ObwpbCd4XvDVW19pVFXDGtDC1FDhnFhm3gMbU4NraVOisacSoSZDSsqpuDTc405U4It93vesUnwfMvVr5bLHrdMA4ZMWFEA2XAPVoPdlZx9NzTtiqdfIrnTaKgdrCVUv1P1y1hasYVyPcMfx1v2cwo_wW_UdzNJztDDh---IxqWw9BovOJ7S9ctG_Z88rb_KTZw</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2333926005</pqid></control><display><type>article</type><title>Ultrasound Capsule Endoscopy With a Mechanically Scanning Micro-ultrasound: A Porcine Study</title><source>Elsevier ScienceDirect Journals</source><creator>Qiu, Yongqiang ; Huang, Yaocai ; Zhang, Zhiqiang ; Cox, Benjamin F. ; Liu, Rong ; Hong, Jiehan ; Mu, Peitian ; Lay, Holly S. ; Cummins, Gerard ; Desmulliez, Marc P.Y. ; Clutton, Eddie ; Zheng, Hairong ; Qiu, Weibao ; Cochran, Sandy</creator><creatorcontrib>Qiu, Yongqiang ; Huang, Yaocai ; Zhang, Zhiqiang ; Cox, Benjamin F. ; Liu, Rong ; Hong, Jiehan ; Mu, Peitian ; Lay, Holly S. ; Cummins, Gerard ; Desmulliez, Marc P.Y. ; Clutton, Eddie ; Zheng, Hairong ; Qiu, Weibao ; Cochran, Sandy</creatorcontrib><description>Wireless capsule endoscopy has been used for the clinical examination of the gastrointestinal (GI) tract for two decades. However, most commercially available devices only utilise optical imaging to examine the GI wall surface. Using this sensing modality, pathology within the GI wall cannot be detected. Micro-ultrasound (μUS) using high-frequency (&gt;20 MHz) ultrasound can provide a means of transmural or cross-sectional image of the GI tract. Depth of imaging is approximately 10 mm with a resolution of between 40–120 μm that is sufficient to differentiate between subsurface histologic layers of the various regions of the GI tract. Ultrasound capsule endoscopy (USCE) uses a capsule equipped with μUS transducers that are capable of imaging below the GI wall surface, offering thereby a complementary sensing technique to optical imaging capsule endoscopy. In this work, a USCE device integrated with a ∼30 MHz ultrasonic transducer was developed to capture a full 360° image of the lumen. The performance of the device was initially evaluated using a wire phantom, indicating an axial resolution of 69.0 μm and lateral resolution of 262.5 μm. Later, in vivo imaging performance was characterised in the oesophagus and small intestine of anaesthetized pigs. The reconstructed images demonstrate clear layer differentiation of the lumen wall. The tissue thicknesses measured from the B-scan images show good agreement with ex vivo images from the literature. The high-resolution ultrasound images in the in vivo porcine model achieved with this device is an encouraging preliminary step in the translation of these devices toward future clinical use.</description><identifier>ISSN: 0301-5629</identifier><identifier>EISSN: 1879-291X</identifier><identifier>DOI: 10.1016/j.ultrasmedbio.2019.12.003</identifier><identifier>PMID: 31902446</identifier><language>eng</language><publisher>England: Elsevier Inc</publisher><subject>Capsule endoscopy ; High-frequency ultrasound ; In vivo porcine animal study ; Micro-ultrasound capsule endoscopy</subject><ispartof>Ultrasound in medicine &amp; biology, 2020-03, Vol.46 (3), p.796-804</ispartof><rights>2019 The Authors</rights><rights>Copyright © 2019 World Federation for Ultrasound in Medicine &amp; Biology. Published by Elsevier Inc. All rights reserved.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c432t-2fbca5ad440211a3b563e81ce12c0ebb50df5b4edcb935636066c145fb2be7a43</citedby><cites>FETCH-LOGICAL-c432t-2fbca5ad440211a3b563e81ce12c0ebb50df5b4edcb935636066c145fb2be7a43</cites><orcidid>0000-0003-0266-1294</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://dx.doi.org/10.1016/j.ultrasmedbio.2019.12.003$$EHTML$$P50$$Gelsevier$$Hfree_for_read</linktohtml><link.rule.ids>314,777,781,3537,27905,27906,45976</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/31902446$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Qiu, Yongqiang</creatorcontrib><creatorcontrib>Huang, Yaocai</creatorcontrib><creatorcontrib>Zhang, Zhiqiang</creatorcontrib><creatorcontrib>Cox, Benjamin F.</creatorcontrib><creatorcontrib>Liu, Rong</creatorcontrib><creatorcontrib>Hong, Jiehan</creatorcontrib><creatorcontrib>Mu, Peitian</creatorcontrib><creatorcontrib>Lay, Holly S.</creatorcontrib><creatorcontrib>Cummins, Gerard</creatorcontrib><creatorcontrib>Desmulliez, Marc P.Y.</creatorcontrib><creatorcontrib>Clutton, Eddie</creatorcontrib><creatorcontrib>Zheng, Hairong</creatorcontrib><creatorcontrib>Qiu, Weibao</creatorcontrib><creatorcontrib>Cochran, Sandy</creatorcontrib><title>Ultrasound Capsule Endoscopy With a Mechanically Scanning Micro-ultrasound: A Porcine Study</title><title>Ultrasound in medicine &amp; biology</title><addtitle>Ultrasound Med Biol</addtitle><description>Wireless capsule endoscopy has been used for the clinical examination of the gastrointestinal (GI) tract for two decades. However, most commercially available devices only utilise optical imaging to examine the GI wall surface. Using this sensing modality, pathology within the GI wall cannot be detected. Micro-ultrasound (μUS) using high-frequency (&gt;20 MHz) ultrasound can provide a means of transmural or cross-sectional image of the GI tract. Depth of imaging is approximately 10 mm with a resolution of between 40–120 μm that is sufficient to differentiate between subsurface histologic layers of the various regions of the GI tract. Ultrasound capsule endoscopy (USCE) uses a capsule equipped with μUS transducers that are capable of imaging below the GI wall surface, offering thereby a complementary sensing technique to optical imaging capsule endoscopy. In this work, a USCE device integrated with a ∼30 MHz ultrasonic transducer was developed to capture a full 360° image of the lumen. The performance of the device was initially evaluated using a wire phantom, indicating an axial resolution of 69.0 μm and lateral resolution of 262.5 μm. Later, in vivo imaging performance was characterised in the oesophagus and small intestine of anaesthetized pigs. The reconstructed images demonstrate clear layer differentiation of the lumen wall. The tissue thicknesses measured from the B-scan images show good agreement with ex vivo images from the literature. The high-resolution ultrasound images in the in vivo porcine model achieved with this device is an encouraging preliminary step in the translation of these devices toward future clinical use.</description><subject>Capsule endoscopy</subject><subject>High-frequency ultrasound</subject><subject>In vivo porcine animal study</subject><subject>Micro-ultrasound capsule endoscopy</subject><issn>0301-5629</issn><issn>1879-291X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2020</creationdate><recordtype>article</recordtype><recordid>eNqNkEtPGzEQx62Kqgm0X6Gyeupll7G962S5ofCUQEUC1Eo9WH7MFkcbO9i7SPn2bJoU9chpDv_HzPwI-cagZMDk8bIcuj7pvEJnfCw5sKZkvAQQH8iUzWdNwRv264BMQQArasmbCTnMeQkAMylmn8hEsAZ4Vckp-f34tyoOwdGFXuehQ3oeXMw2rjf0p--fqKa3aJ908FZ33YbeWx2CD3_orbcpFsNb_oSe0ruYrA9I7_vBbT6Tj63uMn7ZzyPyeHH-sLgqbn5cXi9ObwpbCd4XvDVW19pVFXDGtDC1FDhnFhm3gMbU4NraVOisacSoSZDSsqpuDTc405U4It93vesUnwfMvVr5bLHrdMA4ZMWFEA2XAPVoPdlZx9NzTtiqdfIrnTaKgdrCVUv1P1y1hasYVyPcMfx1v2cwo_wW_UdzNJztDDh---IxqWw9BovOJ7S9ctG_Z88rb_KTZw</recordid><startdate>202003</startdate><enddate>202003</enddate><creator>Qiu, Yongqiang</creator><creator>Huang, Yaocai</creator><creator>Zhang, Zhiqiang</creator><creator>Cox, Benjamin F.</creator><creator>Liu, Rong</creator><creator>Hong, Jiehan</creator><creator>Mu, Peitian</creator><creator>Lay, Holly S.</creator><creator>Cummins, Gerard</creator><creator>Desmulliez, Marc P.Y.</creator><creator>Clutton, Eddie</creator><creator>Zheng, Hairong</creator><creator>Qiu, Weibao</creator><creator>Cochran, Sandy</creator><general>Elsevier Inc</general><scope>6I.</scope><scope>AAFTH</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><orcidid>https://orcid.org/0000-0003-0266-1294</orcidid></search><sort><creationdate>202003</creationdate><title>Ultrasound Capsule Endoscopy With a Mechanically Scanning Micro-ultrasound: A Porcine Study</title><author>Qiu, Yongqiang ; Huang, Yaocai ; Zhang, Zhiqiang ; Cox, Benjamin F. ; Liu, Rong ; Hong, Jiehan ; Mu, Peitian ; Lay, Holly S. ; Cummins, Gerard ; Desmulliez, Marc P.Y. ; Clutton, Eddie ; Zheng, Hairong ; Qiu, Weibao ; Cochran, Sandy</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c432t-2fbca5ad440211a3b563e81ce12c0ebb50df5b4edcb935636066c145fb2be7a43</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2020</creationdate><topic>Capsule endoscopy</topic><topic>High-frequency ultrasound</topic><topic>In vivo porcine animal study</topic><topic>Micro-ultrasound capsule endoscopy</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Qiu, Yongqiang</creatorcontrib><creatorcontrib>Huang, Yaocai</creatorcontrib><creatorcontrib>Zhang, Zhiqiang</creatorcontrib><creatorcontrib>Cox, Benjamin F.</creatorcontrib><creatorcontrib>Liu, Rong</creatorcontrib><creatorcontrib>Hong, Jiehan</creatorcontrib><creatorcontrib>Mu, Peitian</creatorcontrib><creatorcontrib>Lay, Holly S.</creatorcontrib><creatorcontrib>Cummins, Gerard</creatorcontrib><creatorcontrib>Desmulliez, Marc P.Y.</creatorcontrib><creatorcontrib>Clutton, Eddie</creatorcontrib><creatorcontrib>Zheng, Hairong</creatorcontrib><creatorcontrib>Qiu, Weibao</creatorcontrib><creatorcontrib>Cochran, Sandy</creatorcontrib><collection>ScienceDirect Open Access Titles</collection><collection>Elsevier:ScienceDirect:Open Access</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Ultrasound in medicine &amp; biology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Qiu, Yongqiang</au><au>Huang, Yaocai</au><au>Zhang, Zhiqiang</au><au>Cox, Benjamin F.</au><au>Liu, Rong</au><au>Hong, Jiehan</au><au>Mu, Peitian</au><au>Lay, Holly S.</au><au>Cummins, Gerard</au><au>Desmulliez, Marc P.Y.</au><au>Clutton, Eddie</au><au>Zheng, Hairong</au><au>Qiu, Weibao</au><au>Cochran, Sandy</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Ultrasound Capsule Endoscopy With a Mechanically Scanning Micro-ultrasound: A Porcine Study</atitle><jtitle>Ultrasound in medicine &amp; biology</jtitle><addtitle>Ultrasound Med Biol</addtitle><date>2020-03</date><risdate>2020</risdate><volume>46</volume><issue>3</issue><spage>796</spage><epage>804</epage><pages>796-804</pages><issn>0301-5629</issn><eissn>1879-291X</eissn><abstract>Wireless capsule endoscopy has been used for the clinical examination of the gastrointestinal (GI) tract for two decades. However, most commercially available devices only utilise optical imaging to examine the GI wall surface. Using this sensing modality, pathology within the GI wall cannot be detected. Micro-ultrasound (μUS) using high-frequency (&gt;20 MHz) ultrasound can provide a means of transmural or cross-sectional image of the GI tract. Depth of imaging is approximately 10 mm with a resolution of between 40–120 μm that is sufficient to differentiate between subsurface histologic layers of the various regions of the GI tract. Ultrasound capsule endoscopy (USCE) uses a capsule equipped with μUS transducers that are capable of imaging below the GI wall surface, offering thereby a complementary sensing technique to optical imaging capsule endoscopy. In this work, a USCE device integrated with a ∼30 MHz ultrasonic transducer was developed to capture a full 360° image of the lumen. The performance of the device was initially evaluated using a wire phantom, indicating an axial resolution of 69.0 μm and lateral resolution of 262.5 μm. Later, in vivo imaging performance was characterised in the oesophagus and small intestine of anaesthetized pigs. The reconstructed images demonstrate clear layer differentiation of the lumen wall. The tissue thicknesses measured from the B-scan images show good agreement with ex vivo images from the literature. The high-resolution ultrasound images in the in vivo porcine model achieved with this device is an encouraging preliminary step in the translation of these devices toward future clinical use.</abstract><cop>England</cop><pub>Elsevier Inc</pub><pmid>31902446</pmid><doi>10.1016/j.ultrasmedbio.2019.12.003</doi><tpages>9</tpages><orcidid>https://orcid.org/0000-0003-0266-1294</orcidid><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 0301-5629
ispartof Ultrasound in medicine & biology, 2020-03, Vol.46 (3), p.796-804
issn 0301-5629
1879-291X
language eng
recordid cdi_proquest_miscellaneous_2333926005
source Elsevier ScienceDirect Journals
subjects Capsule endoscopy
High-frequency ultrasound
In vivo porcine animal study
Micro-ultrasound capsule endoscopy
title Ultrasound Capsule Endoscopy With a Mechanically Scanning Micro-ultrasound: A Porcine Study
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-18T11%3A56%3A26IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Ultrasound%20Capsule%20Endoscopy%20With%20a%20Mechanically%20Scanning%20Micro-ultrasound:%20A%20Porcine%20Study&rft.jtitle=Ultrasound%20in%20medicine%20&%20biology&rft.au=Qiu,%20Yongqiang&rft.date=2020-03&rft.volume=46&rft.issue=3&rft.spage=796&rft.epage=804&rft.pages=796-804&rft.issn=0301-5629&rft.eissn=1879-291X&rft_id=info:doi/10.1016/j.ultrasmedbio.2019.12.003&rft_dat=%3Cproquest_cross%3E2333926005%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2333926005&rft_id=info:pmid/31902446&rft_els_id=S0301562919316230&rfr_iscdi=true