COMPARISON OF DIFFERENT TECHNIQUES FOR EVALUATION OF DOSE DISTRIBUTIONS IN RADIOTHERAPY USING RADIOCHROMIC EBT3 FILMS
In radiotherapy, radiochromic films can be used for verification of delivery of dose distributions calculated by treatment planning systems. The main objective of this work was to compare three different techniques for evaluation of dose distributions for prostate cancer treatment plans using radioc...
Gespeichert in:
Veröffentlicht in: | Radiation protection dosimetry 2019-12, Vol.186 (2-3), p.357-361 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In radiotherapy, radiochromic films can be used for verification of delivery of dose distributions calculated by treatment planning systems. The main objective of this work was to compare three different techniques for evaluation of dose distributions for prostate cancer treatment plans using radiochromic EBT3 films. These techniques are: red channel evaluation taking into account only a response of irradiated film (R), red channel evaluation taking into account a response of unirradiated and irradiated film (Rcor) and multichannel evaluation in FilmQA software (RGB). Also comparison between film and MatriXX measurement was performed. Comparison showed that gamma analysis passing rates strongly depend on evaluation technique and on a model of scanner for digitizing films. The highest gamma passing rates were obtained with red channel evaluation taking into account a response of unirradiated and irradiated film using Epson V750 scanner (Rcor) and multichannel evaluation in FilmQA using Epson 11000XL scanner. |
---|---|
ISSN: | 0144-8420 1742-3406 |
DOI: | 10.1093/rpd/ncz231 |