The Natural History of Radiolucencies Following Uncemented Total Knee Arthroplasty at 9 Years
Aseptic loosening remains a common cause of failure in total knee arthroplasty (TKA). There is an increased interest in using uncemented TKA to reduce this complication. Radiolucencies (RLs) following uncemented TKA can be concerning. We report on the 9-year history of RLs in patients with uncemente...
Gespeichert in:
Veröffentlicht in: | The Journal of arthroplasty 2020-01, Vol.35 (1), p.127-131 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Aseptic loosening remains a common cause of failure in total knee arthroplasty (TKA). There is an increased interest in using uncemented TKA to reduce this complication. Radiolucencies (RLs) following uncemented TKA can be concerning. We report on the 9-year history of RLs in patients with uncemented TKA.
Twenty-one patients (26 knees) were treated with a cruciate-retaining fully porous coated femur/tibia and cemented patella. At final follow-up, 17 patients (22 knees) were available for review. Average follow-up was 9.6 years, average age was 59.1, and average body mass index was 34.1. X-rays were taken at 6 weeks, 1 year, and at final follow-up. RLs were measured using the Knee Society scoring system and read by two separate surgeons.
At 6 weeks, we identified RL in all patients on both the tibia and femur. The majority were beneath the tibial tray and femoral chamfer. At 1 year, 4 femurs and 4 tibias showed new RLs ( |
---|---|