Effect of scan pattern on complete-arch scans with 4 digital scanners
Complete-arch digital scans are becoming popular as digital dentistry is adopted for expanded clinical situations such as complete-arch prostheses, removable prostheses, extensive implant-supported treatment, and orthodontic aligners. Whether the scan pattern technique affects the trueness and preci...
Gespeichert in:
Veröffentlicht in: | The Journal of prosthetic dentistry 2020-01, Vol.123 (1), p.85-95 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 95 |
---|---|
container_issue | 1 |
container_start_page | 85 |
container_title | The Journal of prosthetic dentistry |
container_volume | 123 |
creator | Latham, Jason Ludlow, Mark Mennito, Anthony Kelly, Abigail Evans, Zachary Renne, Walter |
description | Complete-arch digital scans are becoming popular as digital dentistry is adopted for expanded clinical situations such as complete-arch prostheses, removable prostheses, extensive implant-supported treatment, and orthodontic aligners. Whether the scan pattern technique affects the trueness and precision of complete-arch scans and whether differences in accuracy exist among different scanners remain unclear. Furthermore, each manufacturer recommends a different scan pattern, but evidence of the superiority of the manufacturer’s recommended pattern is lacking.
The purpose of this in vitro study was to determine whether the scan pattern affects the trueness, precision, and speed of complete-arch digital scans performed by using 4 different digital scanning systems.
A custom model used as the reference standard was fabricated with teeth having the same refractive index as dentin and enamel to simulate the natural dentition. The scan of the custom typodont was obtained by using an ATOS III Triple Scan 3D optical scanner. This study evaluated the CEREC Omnicam, Planmeca Emerald, Align iTero Element, and 3Shape TRIOS 3. Experimental scans were obtained from each of the 4 different digital scanning systems by using 4 unique scan patterns by experienced clinicians. Four experimental scans were acquired from each of the scanners by using 4 distinct scan patterns for a total of 16 scans for each scanner. Scan patterns 1 to 4 were based on the operator manuals for each different scanner. The scan time was recorded for each scan. All experimental scans were converted to standard tessellation language (STL) format, and a comprehensive metrology program, Geomagic Control X, was used to compare the reference standard scan with the experimental scans.
For trueness, the scanner (P |
doi_str_mv | 10.1016/j.prosdent.2019.02.008 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2210002554</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S0022391319301520</els_id><sourcerecordid>2210002554</sourcerecordid><originalsourceid>FETCH-LOGICAL-c416t-34068bcdccb466e64c7f042e54cae39ba7250a62c7049a6a44a4de782bf7d1b23</originalsourceid><addsrcrecordid>eNqFkMtOwzAQRS0EoqXwC1WWbBLGjuMkO1BVHlIlNrC2HGdCXeWF7YL4e9wXW1YjzZyZ0T2EzCkkFKi42ySjHVyNvU8Y0DIBlgAUZ2RKocxjUXB6TqYAjMVpSdMJuXJuA4HIcnpJJimUBRNUTMly2TSofTQ0kdOqj0blPdo-GvpID93YosdYWb3eT130bfw64lFtPoxX7b7Zo3XX5KJRrcObY52R98fl2-I5Xr0-vSweVrHmVPg45SCKStdaV1wIFFznDXCGGdcK07JSOctACaZz4KUSinPFa8wLVjV5TSuWzsjt4W4I_7lF52VnnMa2VT0OWycZoyEkyzIeUHFAdfDkLDZytKZT9kdSkDuFciNPCuVOoQQmg6CwOD_-2FYd1n9rJ2cBuD8AGJJ-GbTSaYO9xtrYoFLWg_nvxy_MDYWM</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2210002554</pqid></control><display><type>article</type><title>Effect of scan pattern on complete-arch scans with 4 digital scanners</title><source>MEDLINE</source><source>Elsevier ScienceDirect Journals</source><creator>Latham, Jason ; Ludlow, Mark ; Mennito, Anthony ; Kelly, Abigail ; Evans, Zachary ; Renne, Walter</creator><creatorcontrib>Latham, Jason ; Ludlow, Mark ; Mennito, Anthony ; Kelly, Abigail ; Evans, Zachary ; Renne, Walter</creatorcontrib><description>Complete-arch digital scans are becoming popular as digital dentistry is adopted for expanded clinical situations such as complete-arch prostheses, removable prostheses, extensive implant-supported treatment, and orthodontic aligners. Whether the scan pattern technique affects the trueness and precision of complete-arch scans and whether differences in accuracy exist among different scanners remain unclear. Furthermore, each manufacturer recommends a different scan pattern, but evidence of the superiority of the manufacturer’s recommended pattern is lacking.
The purpose of this in vitro study was to determine whether the scan pattern affects the trueness, precision, and speed of complete-arch digital scans performed by using 4 different digital scanning systems.
A custom model used as the reference standard was fabricated with teeth having the same refractive index as dentin and enamel to simulate the natural dentition. The scan of the custom typodont was obtained by using an ATOS III Triple Scan 3D optical scanner. This study evaluated the CEREC Omnicam, Planmeca Emerald, Align iTero Element, and 3Shape TRIOS 3. Experimental scans were obtained from each of the 4 different digital scanning systems by using 4 unique scan patterns by experienced clinicians. Four experimental scans were acquired from each of the scanners by using 4 distinct scan patterns for a total of 16 scans for each scanner. Scan patterns 1 to 4 were based on the operator manuals for each different scanner. The scan time was recorded for each scan. All experimental scans were converted to standard tessellation language (STL) format, and a comprehensive metrology program, Geomagic Control X, was used to compare the reference standard scan with the experimental scans.
For trueness, the scanner (P<.001), scan pattern (P=.001), and their interaction (P<.001) were found to be significant. Overall, scan pattern 2 showed the highest average trueness and precision. Likewise, for overall scan pattern precision, the scanner, scan pattern, and their interaction were found to be significant (P<.001).
Scan pattern affected trueness and precision for some scanners, but not for others. Differences exist in the complete-arch scan speed, trueness, and precision of individual scanners. Scan pattern can play an important role in the success of digital scanning.</description><identifier>ISSN: 0022-3913</identifier><identifier>EISSN: 1097-6841</identifier><identifier>DOI: 10.1016/j.prosdent.2019.02.008</identifier><identifier>PMID: 30982616</identifier><language>eng</language><publisher>United States: Elsevier Inc</publisher><subject>Computer-Aided Design ; Dental Arch ; Dental Impression Technique ; Dental Models ; Dentistry ; Humans ; Imaging, Three-Dimensional ; Mouth, Edentulous</subject><ispartof>The Journal of prosthetic dentistry, 2020-01, Vol.123 (1), p.85-95</ispartof><rights>2019 Editorial Council for the Journal of Prosthetic Dentistry</rights><rights>Copyright © 2019 Editorial Council for the Journal of Prosthetic Dentistry. Published by Elsevier Inc. All rights reserved.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c416t-34068bcdccb466e64c7f042e54cae39ba7250a62c7049a6a44a4de782bf7d1b23</citedby><cites>FETCH-LOGICAL-c416t-34068bcdccb466e64c7f042e54cae39ba7250a62c7049a6a44a4de782bf7d1b23</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://www.sciencedirect.com/science/article/pii/S0022391319301520$$EHTML$$P50$$Gelsevier$$H</linktohtml><link.rule.ids>314,776,780,3537,27901,27902,65306</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/30982616$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Latham, Jason</creatorcontrib><creatorcontrib>Ludlow, Mark</creatorcontrib><creatorcontrib>Mennito, Anthony</creatorcontrib><creatorcontrib>Kelly, Abigail</creatorcontrib><creatorcontrib>Evans, Zachary</creatorcontrib><creatorcontrib>Renne, Walter</creatorcontrib><title>Effect of scan pattern on complete-arch scans with 4 digital scanners</title><title>The Journal of prosthetic dentistry</title><addtitle>J Prosthet Dent</addtitle><description>Complete-arch digital scans are becoming popular as digital dentistry is adopted for expanded clinical situations such as complete-arch prostheses, removable prostheses, extensive implant-supported treatment, and orthodontic aligners. Whether the scan pattern technique affects the trueness and precision of complete-arch scans and whether differences in accuracy exist among different scanners remain unclear. Furthermore, each manufacturer recommends a different scan pattern, but evidence of the superiority of the manufacturer’s recommended pattern is lacking.
The purpose of this in vitro study was to determine whether the scan pattern affects the trueness, precision, and speed of complete-arch digital scans performed by using 4 different digital scanning systems.
A custom model used as the reference standard was fabricated with teeth having the same refractive index as dentin and enamel to simulate the natural dentition. The scan of the custom typodont was obtained by using an ATOS III Triple Scan 3D optical scanner. This study evaluated the CEREC Omnicam, Planmeca Emerald, Align iTero Element, and 3Shape TRIOS 3. Experimental scans were obtained from each of the 4 different digital scanning systems by using 4 unique scan patterns by experienced clinicians. Four experimental scans were acquired from each of the scanners by using 4 distinct scan patterns for a total of 16 scans for each scanner. Scan patterns 1 to 4 were based on the operator manuals for each different scanner. The scan time was recorded for each scan. All experimental scans were converted to standard tessellation language (STL) format, and a comprehensive metrology program, Geomagic Control X, was used to compare the reference standard scan with the experimental scans.
For trueness, the scanner (P<.001), scan pattern (P=.001), and their interaction (P<.001) were found to be significant. Overall, scan pattern 2 showed the highest average trueness and precision. Likewise, for overall scan pattern precision, the scanner, scan pattern, and their interaction were found to be significant (P<.001).
Scan pattern affected trueness and precision for some scanners, but not for others. Differences exist in the complete-arch scan speed, trueness, and precision of individual scanners. Scan pattern can play an important role in the success of digital scanning.</description><subject>Computer-Aided Design</subject><subject>Dental Arch</subject><subject>Dental Impression Technique</subject><subject>Dental Models</subject><subject>Dentistry</subject><subject>Humans</subject><subject>Imaging, Three-Dimensional</subject><subject>Mouth, Edentulous</subject><issn>0022-3913</issn><issn>1097-6841</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2020</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNqFkMtOwzAQRS0EoqXwC1WWbBLGjuMkO1BVHlIlNrC2HGdCXeWF7YL4e9wXW1YjzZyZ0T2EzCkkFKi42ySjHVyNvU8Y0DIBlgAUZ2RKocxjUXB6TqYAjMVpSdMJuXJuA4HIcnpJJimUBRNUTMly2TSofTQ0kdOqj0blPdo-GvpID93YosdYWb3eT130bfw64lFtPoxX7b7Zo3XX5KJRrcObY52R98fl2-I5Xr0-vSweVrHmVPg45SCKStdaV1wIFFznDXCGGdcK07JSOctACaZz4KUSinPFa8wLVjV5TSuWzsjt4W4I_7lF52VnnMa2VT0OWycZoyEkyzIeUHFAdfDkLDZytKZT9kdSkDuFciNPCuVOoQQmg6CwOD_-2FYd1n9rJ2cBuD8AGJJ-GbTSaYO9xtrYoFLWg_nvxy_MDYWM</recordid><startdate>202001</startdate><enddate>202001</enddate><creator>Latham, Jason</creator><creator>Ludlow, Mark</creator><creator>Mennito, Anthony</creator><creator>Kelly, Abigail</creator><creator>Evans, Zachary</creator><creator>Renne, Walter</creator><general>Elsevier Inc</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>202001</creationdate><title>Effect of scan pattern on complete-arch scans with 4 digital scanners</title><author>Latham, Jason ; Ludlow, Mark ; Mennito, Anthony ; Kelly, Abigail ; Evans, Zachary ; Renne, Walter</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c416t-34068bcdccb466e64c7f042e54cae39ba7250a62c7049a6a44a4de782bf7d1b23</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2020</creationdate><topic>Computer-Aided Design</topic><topic>Dental Arch</topic><topic>Dental Impression Technique</topic><topic>Dental Models</topic><topic>Dentistry</topic><topic>Humans</topic><topic>Imaging, Three-Dimensional</topic><topic>Mouth, Edentulous</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Latham, Jason</creatorcontrib><creatorcontrib>Ludlow, Mark</creatorcontrib><creatorcontrib>Mennito, Anthony</creatorcontrib><creatorcontrib>Kelly, Abigail</creatorcontrib><creatorcontrib>Evans, Zachary</creatorcontrib><creatorcontrib>Renne, Walter</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>The Journal of prosthetic dentistry</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Latham, Jason</au><au>Ludlow, Mark</au><au>Mennito, Anthony</au><au>Kelly, Abigail</au><au>Evans, Zachary</au><au>Renne, Walter</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Effect of scan pattern on complete-arch scans with 4 digital scanners</atitle><jtitle>The Journal of prosthetic dentistry</jtitle><addtitle>J Prosthet Dent</addtitle><date>2020-01</date><risdate>2020</risdate><volume>123</volume><issue>1</issue><spage>85</spage><epage>95</epage><pages>85-95</pages><issn>0022-3913</issn><eissn>1097-6841</eissn><abstract>Complete-arch digital scans are becoming popular as digital dentistry is adopted for expanded clinical situations such as complete-arch prostheses, removable prostheses, extensive implant-supported treatment, and orthodontic aligners. Whether the scan pattern technique affects the trueness and precision of complete-arch scans and whether differences in accuracy exist among different scanners remain unclear. Furthermore, each manufacturer recommends a different scan pattern, but evidence of the superiority of the manufacturer’s recommended pattern is lacking.
The purpose of this in vitro study was to determine whether the scan pattern affects the trueness, precision, and speed of complete-arch digital scans performed by using 4 different digital scanning systems.
A custom model used as the reference standard was fabricated with teeth having the same refractive index as dentin and enamel to simulate the natural dentition. The scan of the custom typodont was obtained by using an ATOS III Triple Scan 3D optical scanner. This study evaluated the CEREC Omnicam, Planmeca Emerald, Align iTero Element, and 3Shape TRIOS 3. Experimental scans were obtained from each of the 4 different digital scanning systems by using 4 unique scan patterns by experienced clinicians. Four experimental scans were acquired from each of the scanners by using 4 distinct scan patterns for a total of 16 scans for each scanner. Scan patterns 1 to 4 were based on the operator manuals for each different scanner. The scan time was recorded for each scan. All experimental scans were converted to standard tessellation language (STL) format, and a comprehensive metrology program, Geomagic Control X, was used to compare the reference standard scan with the experimental scans.
For trueness, the scanner (P<.001), scan pattern (P=.001), and their interaction (P<.001) were found to be significant. Overall, scan pattern 2 showed the highest average trueness and precision. Likewise, for overall scan pattern precision, the scanner, scan pattern, and their interaction were found to be significant (P<.001).
Scan pattern affected trueness and precision for some scanners, but not for others. Differences exist in the complete-arch scan speed, trueness, and precision of individual scanners. Scan pattern can play an important role in the success of digital scanning.</abstract><cop>United States</cop><pub>Elsevier Inc</pub><pmid>30982616</pmid><doi>10.1016/j.prosdent.2019.02.008</doi><tpages>11</tpages><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0022-3913 |
ispartof | The Journal of prosthetic dentistry, 2020-01, Vol.123 (1), p.85-95 |
issn | 0022-3913 1097-6841 |
language | eng |
recordid | cdi_proquest_miscellaneous_2210002554 |
source | MEDLINE; Elsevier ScienceDirect Journals |
subjects | Computer-Aided Design Dental Arch Dental Impression Technique Dental Models Dentistry Humans Imaging, Three-Dimensional Mouth, Edentulous |
title | Effect of scan pattern on complete-arch scans with 4 digital scanners |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-30T11%3A43%3A29IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Effect%20of%20scan%20pattern%20on%20complete-arch%20scans%20with%204%20digital%20scanners&rft.jtitle=The%20Journal%20of%20prosthetic%20dentistry&rft.au=Latham,%20Jason&rft.date=2020-01&rft.volume=123&rft.issue=1&rft.spage=85&rft.epage=95&rft.pages=85-95&rft.issn=0022-3913&rft.eissn=1097-6841&rft_id=info:doi/10.1016/j.prosdent.2019.02.008&rft_dat=%3Cproquest_cross%3E2210002554%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2210002554&rft_id=info:pmid/30982616&rft_els_id=S0022391319301520&rfr_iscdi=true |