Design, Material, and Seating Load Effects on In Vitro Fretting Corrosion Performance of Modular Head-Neck Tapers

The short-term corrosion and micromechanical behavior of 32 unique head-neck taper design/material/assembly conditions was tested using an incremental cyclic fretting corrosion (ICFC) test method previously developed. Seven materials, design, and simulated surgical parameters were evaluated, each be...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of arthroplasty 2019-05, Vol.34 (5), p.991-1002
Hauptverfasser: Ouellette, Eric S., Mali, Sachin A., Kim, Jua, Grostefon, Justin, Gilbert, Jeremy L.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 1002
container_issue 5
container_start_page 991
container_title The Journal of arthroplasty
container_volume 34
creator Ouellette, Eric S.
Mali, Sachin A.
Kim, Jua
Grostefon, Justin
Gilbert, Jeremy L.
description The short-term corrosion and micromechanical behavior of 32 unique head-neck taper design/material/assembly conditions was tested using an incremental cyclic fretting corrosion (ICFC) test method previously developed. Seven materials, design, and simulated surgical parameters were evaluated, each being assigned 2 conditions for testing, using a 27-2 (7 factor, quarter factorial) design of experiments test matrix. The factors explored were (1) seating load, (2) head-neck offset, (3) material combination, (4) taper diameter, (5) taper roughness, (6) angular mismatch/engagement, and (7) taper length. Each sample underwent assembly, ICFC testing, pull off. Low seating load and high head offset correlated with increased fretting corrosion (P < .05). High head offset also contributed to a lower onset load for fretting current and higher micromotion (P < .05). Head subsidence measured over the ICFC test for samples seated at 100 N was significantly higher than samples seated at 4000 N. Micromotion for 12-mm head offsets was statistically higher than samples with a 1.5-mm head offset. A number of interactive effects were observed. For example, samples seated at 4000 N were less sensitive to head offset than samples seated at 100 N in terms of the resulting fretting current. Taper locking position, material combination, taper engagement length, taper roughness, and taper dimensions all had weak or no correlation with fretting current and taper micromotion. This test method and experimental design is a versatile means of assessing potential new taper designs in the future.
doi_str_mv 10.1016/j.arth.2019.01.043
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2187956250</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S0883540319300889</els_id><sourcerecordid>2187956250</sourcerecordid><originalsourceid>FETCH-LOGICAL-c356t-58f816bf88dea5c38b12c7bf9d740333abd033578c8a4ebf47ed6511c884c8573</originalsourceid><addsrcrecordid>eNp9kMFuEzEURS1U1IbSH2CBvOyiM9jj8diR2KDQ0kppQaKwtTz2c3E6GafPEyT-hm_pl-GQliWrt3jnXukeQt5wVnPGu3er2uL0o24Yn9eM16wVL8iMS9FUumXdAZkxrUUlWyaOyKucV4xxLmV7SI4E041SXM0IfoQc78Yzem0nwGiHM2pHT7-CneJ4R5fJenoeArgp0zTSq_Hx9_c4YaIXCNNfZJEQU47l-QUwJFzb0QFNgV4nvx0s0kuwvroBd09v7QYwvyYvgx0ynDzdY_Lt4vx2cVktP3-6WnxYVk7IbqqkDpp3fdDag5VO6J43TvVh7lVZJITtfTlSaadtC31oFfhOcu60bp2WShyT033vBtPDFvJk1jE7GAY7Qtpm03Ct5rJrJCtos0ddmZIRgtlgXFv8ZTgzO9dmZXauzc61YdwU1yX09ql_26_B_4s8yy3A-z0AZeXPCGiyi1Ds-IhFqPEp_q__D8jhkJI</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2187956250</pqid></control><display><type>article</type><title>Design, Material, and Seating Load Effects on In Vitro Fretting Corrosion Performance of Modular Head-Neck Tapers</title><source>MEDLINE</source><source>Access via ScienceDirect (Elsevier)</source><creator>Ouellette, Eric S. ; Mali, Sachin A. ; Kim, Jua ; Grostefon, Justin ; Gilbert, Jeremy L.</creator><creatorcontrib>Ouellette, Eric S. ; Mali, Sachin A. ; Kim, Jua ; Grostefon, Justin ; Gilbert, Jeremy L.</creatorcontrib><description>The short-term corrosion and micromechanical behavior of 32 unique head-neck taper design/material/assembly conditions was tested using an incremental cyclic fretting corrosion (ICFC) test method previously developed. Seven materials, design, and simulated surgical parameters were evaluated, each being assigned 2 conditions for testing, using a 27-2 (7 factor, quarter factorial) design of experiments test matrix. The factors explored were (1) seating load, (2) head-neck offset, (3) material combination, (4) taper diameter, (5) taper roughness, (6) angular mismatch/engagement, and (7) taper length. Each sample underwent assembly, ICFC testing, pull off. Low seating load and high head offset correlated with increased fretting corrosion (P &lt; .05). High head offset also contributed to a lower onset load for fretting current and higher micromotion (P &lt; .05). Head subsidence measured over the ICFC test for samples seated at 100 N was significantly higher than samples seated at 4000 N. Micromotion for 12-mm head offsets was statistically higher than samples with a 1.5-mm head offset. A number of interactive effects were observed. For example, samples seated at 4000 N were less sensitive to head offset than samples seated at 100 N in terms of the resulting fretting current. Taper locking position, material combination, taper engagement length, taper roughness, and taper dimensions all had weak or no correlation with fretting current and taper micromotion. This test method and experimental design is a versatile means of assessing potential new taper designs in the future.</description><identifier>ISSN: 0883-5403</identifier><identifier>EISSN: 1532-8406</identifier><identifier>DOI: 10.1016/j.arth.2019.01.043</identifier><identifier>PMID: 30827717</identifier><language>eng</language><publisher>United States: Elsevier Inc</publisher><subject>Arthroplasty, Replacement, Hip - adverse effects ; Arthroplasty, Replacement, Hip - instrumentation ; Corrosion ; electrochemistry ; fretting ; Hip Prosthesis - adverse effects ; In Vitro Techniques ; MACC ; micromotion ; Prosthesis Design ; Prosthesis Failure ; Stress, Mechanical ; THA</subject><ispartof>The Journal of arthroplasty, 2019-05, Vol.34 (5), p.991-1002</ispartof><rights>2019 Elsevier Inc.</rights><rights>Copyright © 2019 Elsevier Inc. All rights reserved.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c356t-58f816bf88dea5c38b12c7bf9d740333abd033578c8a4ebf47ed6511c884c8573</citedby><cites>FETCH-LOGICAL-c356t-58f816bf88dea5c38b12c7bf9d740333abd033578c8a4ebf47ed6511c884c8573</cites><orcidid>0000-0002-7428-2851</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://dx.doi.org/10.1016/j.arth.2019.01.043$$EHTML$$P50$$Gelsevier$$H</linktohtml><link.rule.ids>314,780,784,3550,27924,27925,45995</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/30827717$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Ouellette, Eric S.</creatorcontrib><creatorcontrib>Mali, Sachin A.</creatorcontrib><creatorcontrib>Kim, Jua</creatorcontrib><creatorcontrib>Grostefon, Justin</creatorcontrib><creatorcontrib>Gilbert, Jeremy L.</creatorcontrib><title>Design, Material, and Seating Load Effects on In Vitro Fretting Corrosion Performance of Modular Head-Neck Tapers</title><title>The Journal of arthroplasty</title><addtitle>J Arthroplasty</addtitle><description>The short-term corrosion and micromechanical behavior of 32 unique head-neck taper design/material/assembly conditions was tested using an incremental cyclic fretting corrosion (ICFC) test method previously developed. Seven materials, design, and simulated surgical parameters were evaluated, each being assigned 2 conditions for testing, using a 27-2 (7 factor, quarter factorial) design of experiments test matrix. The factors explored were (1) seating load, (2) head-neck offset, (3) material combination, (4) taper diameter, (5) taper roughness, (6) angular mismatch/engagement, and (7) taper length. Each sample underwent assembly, ICFC testing, pull off. Low seating load and high head offset correlated with increased fretting corrosion (P &lt; .05). High head offset also contributed to a lower onset load for fretting current and higher micromotion (P &lt; .05). Head subsidence measured over the ICFC test for samples seated at 100 N was significantly higher than samples seated at 4000 N. Micromotion for 12-mm head offsets was statistically higher than samples with a 1.5-mm head offset. A number of interactive effects were observed. For example, samples seated at 4000 N were less sensitive to head offset than samples seated at 100 N in terms of the resulting fretting current. Taper locking position, material combination, taper engagement length, taper roughness, and taper dimensions all had weak or no correlation with fretting current and taper micromotion. This test method and experimental design is a versatile means of assessing potential new taper designs in the future.</description><subject>Arthroplasty, Replacement, Hip - adverse effects</subject><subject>Arthroplasty, Replacement, Hip - instrumentation</subject><subject>Corrosion</subject><subject>electrochemistry</subject><subject>fretting</subject><subject>Hip Prosthesis - adverse effects</subject><subject>In Vitro Techniques</subject><subject>MACC</subject><subject>micromotion</subject><subject>Prosthesis Design</subject><subject>Prosthesis Failure</subject><subject>Stress, Mechanical</subject><subject>THA</subject><issn>0883-5403</issn><issn>1532-8406</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2019</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp9kMFuEzEURS1U1IbSH2CBvOyiM9jj8diR2KDQ0kppQaKwtTz2c3E6GafPEyT-hm_pl-GQliWrt3jnXukeQt5wVnPGu3er2uL0o24Yn9eM16wVL8iMS9FUumXdAZkxrUUlWyaOyKucV4xxLmV7SI4E041SXM0IfoQc78Yzem0nwGiHM2pHT7-CneJ4R5fJenoeArgp0zTSq_Hx9_c4YaIXCNNfZJEQU47l-QUwJFzb0QFNgV4nvx0s0kuwvroBd09v7QYwvyYvgx0ynDzdY_Lt4vx2cVktP3-6WnxYVk7IbqqkDpp3fdDag5VO6J43TvVh7lVZJITtfTlSaadtC31oFfhOcu60bp2WShyT033vBtPDFvJk1jE7GAY7Qtpm03Ct5rJrJCtos0ddmZIRgtlgXFv8ZTgzO9dmZXauzc61YdwU1yX09ql_26_B_4s8yy3A-z0AZeXPCGiyi1Ds-IhFqPEp_q__D8jhkJI</recordid><startdate>201905</startdate><enddate>201905</enddate><creator>Ouellette, Eric S.</creator><creator>Mali, Sachin A.</creator><creator>Kim, Jua</creator><creator>Grostefon, Justin</creator><creator>Gilbert, Jeremy L.</creator><general>Elsevier Inc</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><orcidid>https://orcid.org/0000-0002-7428-2851</orcidid></search><sort><creationdate>201905</creationdate><title>Design, Material, and Seating Load Effects on In Vitro Fretting Corrosion Performance of Modular Head-Neck Tapers</title><author>Ouellette, Eric S. ; Mali, Sachin A. ; Kim, Jua ; Grostefon, Justin ; Gilbert, Jeremy L.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c356t-58f816bf88dea5c38b12c7bf9d740333abd033578c8a4ebf47ed6511c884c8573</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2019</creationdate><topic>Arthroplasty, Replacement, Hip - adverse effects</topic><topic>Arthroplasty, Replacement, Hip - instrumentation</topic><topic>Corrosion</topic><topic>electrochemistry</topic><topic>fretting</topic><topic>Hip Prosthesis - adverse effects</topic><topic>In Vitro Techniques</topic><topic>MACC</topic><topic>micromotion</topic><topic>Prosthesis Design</topic><topic>Prosthesis Failure</topic><topic>Stress, Mechanical</topic><topic>THA</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Ouellette, Eric S.</creatorcontrib><creatorcontrib>Mali, Sachin A.</creatorcontrib><creatorcontrib>Kim, Jua</creatorcontrib><creatorcontrib>Grostefon, Justin</creatorcontrib><creatorcontrib>Gilbert, Jeremy L.</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>The Journal of arthroplasty</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Ouellette, Eric S.</au><au>Mali, Sachin A.</au><au>Kim, Jua</au><au>Grostefon, Justin</au><au>Gilbert, Jeremy L.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Design, Material, and Seating Load Effects on In Vitro Fretting Corrosion Performance of Modular Head-Neck Tapers</atitle><jtitle>The Journal of arthroplasty</jtitle><addtitle>J Arthroplasty</addtitle><date>2019-05</date><risdate>2019</risdate><volume>34</volume><issue>5</issue><spage>991</spage><epage>1002</epage><pages>991-1002</pages><issn>0883-5403</issn><eissn>1532-8406</eissn><abstract>The short-term corrosion and micromechanical behavior of 32 unique head-neck taper design/material/assembly conditions was tested using an incremental cyclic fretting corrosion (ICFC) test method previously developed. Seven materials, design, and simulated surgical parameters were evaluated, each being assigned 2 conditions for testing, using a 27-2 (7 factor, quarter factorial) design of experiments test matrix. The factors explored were (1) seating load, (2) head-neck offset, (3) material combination, (4) taper diameter, (5) taper roughness, (6) angular mismatch/engagement, and (7) taper length. Each sample underwent assembly, ICFC testing, pull off. Low seating load and high head offset correlated with increased fretting corrosion (P &lt; .05). High head offset also contributed to a lower onset load for fretting current and higher micromotion (P &lt; .05). Head subsidence measured over the ICFC test for samples seated at 100 N was significantly higher than samples seated at 4000 N. Micromotion for 12-mm head offsets was statistically higher than samples with a 1.5-mm head offset. A number of interactive effects were observed. For example, samples seated at 4000 N were less sensitive to head offset than samples seated at 100 N in terms of the resulting fretting current. Taper locking position, material combination, taper engagement length, taper roughness, and taper dimensions all had weak or no correlation with fretting current and taper micromotion. This test method and experimental design is a versatile means of assessing potential new taper designs in the future.</abstract><cop>United States</cop><pub>Elsevier Inc</pub><pmid>30827717</pmid><doi>10.1016/j.arth.2019.01.043</doi><tpages>12</tpages><orcidid>https://orcid.org/0000-0002-7428-2851</orcidid></addata></record>
fulltext fulltext
identifier ISSN: 0883-5403
ispartof The Journal of arthroplasty, 2019-05, Vol.34 (5), p.991-1002
issn 0883-5403
1532-8406
language eng
recordid cdi_proquest_miscellaneous_2187956250
source MEDLINE; Access via ScienceDirect (Elsevier)
subjects Arthroplasty, Replacement, Hip - adverse effects
Arthroplasty, Replacement, Hip - instrumentation
Corrosion
electrochemistry
fretting
Hip Prosthesis - adverse effects
In Vitro Techniques
MACC
micromotion
Prosthesis Design
Prosthesis Failure
Stress, Mechanical
THA
title Design, Material, and Seating Load Effects on In Vitro Fretting Corrosion Performance of Modular Head-Neck Tapers
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-22T13%3A15%3A32IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Design,%20Material,%20and%20Seating%20Load%20Effects%20on%20In%C2%A0Vitro%20Fretting%20Corrosion%20Performance%20of%20Modular%20Head-Neck%20Tapers&rft.jtitle=The%20Journal%20of%20arthroplasty&rft.au=Ouellette,%20Eric%20S.&rft.date=2019-05&rft.volume=34&rft.issue=5&rft.spage=991&rft.epage=1002&rft.pages=991-1002&rft.issn=0883-5403&rft.eissn=1532-8406&rft_id=info:doi/10.1016/j.arth.2019.01.043&rft_dat=%3Cproquest_cross%3E2187956250%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2187956250&rft_id=info:pmid/30827717&rft_els_id=S0883540319300889&rfr_iscdi=true