Investigation of the relation between administered dose and image quality for pediatric 99mTc-DMSA renal scintigraphy: clinical study applying the JSNM (Japanese Society of Nuclear Medicine) pediatric dosage card: The Japanese Society of Nuclear Medicine Technology (JSNMT), the Optimization of Imaging Technique Committee for Pediatric Nuclear Medicine Studies
Purpose In 2013, the Japanese Society of Nuclear Medicine (JSNM) announced consensus guidelines for pediatric nuclear medicine. These JSNM guidelines proposed use of lower administered doses compared with traditionally determined doses, which were estimated from age, weight or body surface area (BSA...
Gespeichert in:
Veröffentlicht in: | Annals of nuclear medicine 2019-03, Vol.33 (3), p.153-159 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 159 |
---|---|
container_issue | 3 |
container_start_page | 153 |
container_title | Annals of nuclear medicine |
container_volume | 33 |
creator | Fujiwara, Takahiro Hidaka, Kuniyuki Sugibayashi, Keiichi Matsumoto, Makoto Kida, Tetsuo Shiina, Katsuya |
description | Purpose
In 2013, the Japanese Society of Nuclear Medicine (JSNM) announced consensus guidelines for pediatric nuclear medicine. These JSNM guidelines proposed use of lower administered doses compared with traditionally determined doses, which were estimated from age, weight or body surface area (BSA) based on the administered dose for adults in Japan. When the JSNM guidelines are used, the relationship between this recommended administered dose and image quality remains unclear. In this study, we clarified the relationship between administered dose and image quality for pediatric
99m
Tc-DMSA renal scan retrospectively, and verified the diagnosable image quality with the recommended administered dose of the JSNM guidelines.
Materials and methods
Data from 7 pediatric patients who underwent
99m
Tc-DMSA dynamic renal scans according to the guidelines’ recommended doses were collected. Scan frame rate was 1 frame/min, and scan time was up to 8 min. Eight images, which had different acquired time periods from 1 min to 8 min were prepared by adding each frame. Nine nuclear medicine specialists determined 8 images with different acquired times as diagnosable or undiagnosable. A region of interest (ROI) with 50% thresholds was placed on each kidney of every image. Coefficient of variation (CV) was calculated by dividing the standard deviation (
σ
) by the mean counts (
µ
) of each ROI (CV =
σ
/
µ
× 100).
99m
Tc-DMSA renal scans (total of 2821 cases) that were performed previously in collaboration with 6 hospitals were collected, and CVs of these images were calculated in all cases. These 2821 cases were separated into 5 groups for every 10 kg weight; i.e., (1) less than 10 kg, (2) 10–19.9 kg, (3) 20–29.9 kg, (4) 30–39.9 kg, and (5) above 40 kg. Regression line of each group was analyzed in relation to the CV and administered dose. The CV at the point of intersection with the recommended dose range from the guideline was determined for each group. This CV value was considered as the estimated CV of the image obtained when the recommended dose of the guideline was used. Thus, if the CV was equal to or less than the estimated CV value, then the diagnostic image quality was deemed satisfactory.
Results
Average CV of the lower limit of diagnosable images in 7 cases as determined by 9 nuclear medicine specialists was 19.9%. Estimated CV was 21.2–24.2% in the group weighing 10 kg and |
doi_str_mv | 10.1007/s12149-018-1320-6 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2138649430</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2138649430</sourcerecordid><originalsourceid>FETCH-LOGICAL-c1186-53b368d73d7e77e8b0e635d6b20d322a3eb3ab1384097fd44daedee485d8bc603</originalsourceid><addsrcrecordid>eNp9Uctu1DAUtRCVGAofwM7LsnCxY4_jsKtKC6067WLK2nLsm6mrjJ3aCSj_yQfhEBasWFnXOi-dg9AHRs8ZpfWnzComGkKZIoxXlMhXaMOUFEQKzl-jDW2YIDVT9Rv0NudnSiu1VdUG_boJPyCP_mBGHwOOHR6fACfo17uF8SdAwMYdffB5hAQOu5gBm-CwP5oD4JfJ9H6ccRcTHsB5MyZvcdMcHy35sttfFLVgepytD8UnmeFp_oxtX_Ts8j1ObsZmGPrZh8Mf99v9_Q6f3ZrBBChO-2g9FP2S7X6yPZiEd8WmyMHHfwxLqiWNNcm9Qyed6TO8__ueou_XV4-X38jdw9eby4s7Ylnphmx5y6VyNXc11DWoloLkWyfbijpeVYZDy03LuBK0qTsnhDPgAITaOtVaSfkpOlt1hxRfplKjPvpsoe9L8DhlXRWuFI3gC5StUJtizgk6PaRSX5o1o3pZUK8L6rKgXhbUsnCqlZMLNhwg6ec4pdJl_g_pNyy5ong</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2138649430</pqid></control><display><type>article</type><title>Investigation of the relation between administered dose and image quality for pediatric 99mTc-DMSA renal scintigraphy: clinical study applying the JSNM (Japanese Society of Nuclear Medicine) pediatric dosage card: The Japanese Society of Nuclear Medicine Technology (JSNMT), the Optimization of Imaging Technique Committee for Pediatric Nuclear Medicine Studies</title><source>SpringerLink Journals - AutoHoldings</source><creator>Fujiwara, Takahiro ; Hidaka, Kuniyuki ; Sugibayashi, Keiichi ; Matsumoto, Makoto ; Kida, Tetsuo ; Shiina, Katsuya</creator><creatorcontrib>Fujiwara, Takahiro ; Hidaka, Kuniyuki ; Sugibayashi, Keiichi ; Matsumoto, Makoto ; Kida, Tetsuo ; Shiina, Katsuya</creatorcontrib><description><![CDATA[Purpose
In 2013, the Japanese Society of Nuclear Medicine (JSNM) announced consensus guidelines for pediatric nuclear medicine. These JSNM guidelines proposed use of lower administered doses compared with traditionally determined doses, which were estimated from age, weight or body surface area (BSA) based on the administered dose for adults in Japan. When the JSNM guidelines are used, the relationship between this recommended administered dose and image quality remains unclear. In this study, we clarified the relationship between administered dose and image quality for pediatric
99m
Tc-DMSA renal scan retrospectively, and verified the diagnosable image quality with the recommended administered dose of the JSNM guidelines.
Materials and methods
Data from 7 pediatric patients who underwent
99m
Tc-DMSA dynamic renal scans according to the guidelines’ recommended doses were collected. Scan frame rate was 1 frame/min, and scan time was up to 8 min. Eight images, which had different acquired time periods from 1 min to 8 min were prepared by adding each frame. Nine nuclear medicine specialists determined 8 images with different acquired times as diagnosable or undiagnosable. A region of interest (ROI) with 50% thresholds was placed on each kidney of every image. Coefficient of variation (CV) was calculated by dividing the standard deviation (
σ
) by the mean counts (
µ
) of each ROI (CV =
σ
/
µ
× 100).
99m
Tc-DMSA renal scans (total of 2821 cases) that were performed previously in collaboration with 6 hospitals were collected, and CVs of these images were calculated in all cases. These 2821 cases were separated into 5 groups for every 10 kg weight; i.e., (1) less than 10 kg, (2) 10–19.9 kg, (3) 20–29.9 kg, (4) 30–39.9 kg, and (5) above 40 kg. Regression line of each group was analyzed in relation to the CV and administered dose. The CV at the point of intersection with the recommended dose range from the guideline was determined for each group. This CV value was considered as the estimated CV of the image obtained when the recommended dose of the guideline was used. Thus, if the CV was equal to or less than the estimated CV value, then the diagnostic image quality was deemed satisfactory.
Results
Average CV of the lower limit of diagnosable images in 7 cases as determined by 9 nuclear medicine specialists was 19.9%. Estimated CV was 21.2–24.2% in the group weighing < 10 kg (group 1), 19.9–20.6% in the group weighing > 10 kg and < 20 kg (group 2), 19.6% in group weighing > 20 kg and < 30 kg (group 3), 19.4–19.5% in the group weighing > 30 kg and < 40 kg (group 4), and 19.8% in the group weighing > 40 kg (group 5). The estimated CVs from groups 1 and 2 with weight < 20 kg exceeded 19.9%.
Conclusions
Although
99m
Tc-DMSA renal scan can be carried out using the guidelines’ recommended dose with conventional image acquisition time in patients weighing 20 kg or more, those < 20 kg need consideration for a longer image acquisition time to obtain diagnosable images.]]></description><identifier>ISSN: 0914-7187</identifier><identifier>EISSN: 1864-6433</identifier><identifier>DOI: 10.1007/s12149-018-1320-6</identifier><language>eng</language><publisher>Singapore: Springer Singapore</publisher><subject>Imaging ; Medicine ; Medicine & Public Health ; Nuclear Medicine ; Original Article ; Radiology</subject><ispartof>Annals of nuclear medicine, 2019-03, Vol.33 (3), p.153-159</ispartof><rights>The Japanese Society of Nuclear Medicine 2018</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><cites>FETCH-LOGICAL-c1186-53b368d73d7e77e8b0e635d6b20d322a3eb3ab1384097fd44daedee485d8bc603</cites><orcidid>0000-0002-9675-7803</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://link.springer.com/content/pdf/10.1007/s12149-018-1320-6$$EPDF$$P50$$Gspringer$$H</linktopdf><linktohtml>$$Uhttps://link.springer.com/10.1007/s12149-018-1320-6$$EHTML$$P50$$Gspringer$$H</linktohtml><link.rule.ids>314,780,784,27924,27925,41488,42557,51319</link.rule.ids></links><search><creatorcontrib>Fujiwara, Takahiro</creatorcontrib><creatorcontrib>Hidaka, Kuniyuki</creatorcontrib><creatorcontrib>Sugibayashi, Keiichi</creatorcontrib><creatorcontrib>Matsumoto, Makoto</creatorcontrib><creatorcontrib>Kida, Tetsuo</creatorcontrib><creatorcontrib>Shiina, Katsuya</creatorcontrib><title>Investigation of the relation between administered dose and image quality for pediatric 99mTc-DMSA renal scintigraphy: clinical study applying the JSNM (Japanese Society of Nuclear Medicine) pediatric dosage card: The Japanese Society of Nuclear Medicine Technology (JSNMT), the Optimization of Imaging Technique Committee for Pediatric Nuclear Medicine Studies</title><title>Annals of nuclear medicine</title><addtitle>Ann Nucl Med</addtitle><description><![CDATA[Purpose
In 2013, the Japanese Society of Nuclear Medicine (JSNM) announced consensus guidelines for pediatric nuclear medicine. These JSNM guidelines proposed use of lower administered doses compared with traditionally determined doses, which were estimated from age, weight or body surface area (BSA) based on the administered dose for adults in Japan. When the JSNM guidelines are used, the relationship between this recommended administered dose and image quality remains unclear. In this study, we clarified the relationship between administered dose and image quality for pediatric
99m
Tc-DMSA renal scan retrospectively, and verified the diagnosable image quality with the recommended administered dose of the JSNM guidelines.
Materials and methods
Data from 7 pediatric patients who underwent
99m
Tc-DMSA dynamic renal scans according to the guidelines’ recommended doses were collected. Scan frame rate was 1 frame/min, and scan time was up to 8 min. Eight images, which had different acquired time periods from 1 min to 8 min were prepared by adding each frame. Nine nuclear medicine specialists determined 8 images with different acquired times as diagnosable or undiagnosable. A region of interest (ROI) with 50% thresholds was placed on each kidney of every image. Coefficient of variation (CV) was calculated by dividing the standard deviation (
σ
) by the mean counts (
µ
) of each ROI (CV =
σ
/
µ
× 100).
99m
Tc-DMSA renal scans (total of 2821 cases) that were performed previously in collaboration with 6 hospitals were collected, and CVs of these images were calculated in all cases. These 2821 cases were separated into 5 groups for every 10 kg weight; i.e., (1) less than 10 kg, (2) 10–19.9 kg, (3) 20–29.9 kg, (4) 30–39.9 kg, and (5) above 40 kg. Regression line of each group was analyzed in relation to the CV and administered dose. The CV at the point of intersection with the recommended dose range from the guideline was determined for each group. This CV value was considered as the estimated CV of the image obtained when the recommended dose of the guideline was used. Thus, if the CV was equal to or less than the estimated CV value, then the diagnostic image quality was deemed satisfactory.
Results
Average CV of the lower limit of diagnosable images in 7 cases as determined by 9 nuclear medicine specialists was 19.9%. Estimated CV was 21.2–24.2% in the group weighing < 10 kg (group 1), 19.9–20.6% in the group weighing > 10 kg and < 20 kg (group 2), 19.6% in group weighing > 20 kg and < 30 kg (group 3), 19.4–19.5% in the group weighing > 30 kg and < 40 kg (group 4), and 19.8% in the group weighing > 40 kg (group 5). The estimated CVs from groups 1 and 2 with weight < 20 kg exceeded 19.9%.
Conclusions
Although
99m
Tc-DMSA renal scan can be carried out using the guidelines’ recommended dose with conventional image acquisition time in patients weighing 20 kg or more, those < 20 kg need consideration for a longer image acquisition time to obtain diagnosable images.]]></description><subject>Imaging</subject><subject>Medicine</subject><subject>Medicine & Public Health</subject><subject>Nuclear Medicine</subject><subject>Original Article</subject><subject>Radiology</subject><issn>0914-7187</issn><issn>1864-6433</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2019</creationdate><recordtype>article</recordtype><recordid>eNp9Uctu1DAUtRCVGAofwM7LsnCxY4_jsKtKC6067WLK2nLsm6mrjJ3aCSj_yQfhEBasWFnXOi-dg9AHRs8ZpfWnzComGkKZIoxXlMhXaMOUFEQKzl-jDW2YIDVT9Rv0NudnSiu1VdUG_boJPyCP_mBGHwOOHR6fACfo17uF8SdAwMYdffB5hAQOu5gBm-CwP5oD4JfJ9H6ccRcTHsB5MyZvcdMcHy35sttfFLVgepytD8UnmeFp_oxtX_Ts8j1ObsZmGPrZh8Mf99v9_Q6f3ZrBBChO-2g9FP2S7X6yPZiEd8WmyMHHfwxLqiWNNcm9Qyed6TO8__ueou_XV4-X38jdw9eby4s7Ylnphmx5y6VyNXc11DWoloLkWyfbijpeVYZDy03LuBK0qTsnhDPgAITaOtVaSfkpOlt1hxRfplKjPvpsoe9L8DhlXRWuFI3gC5StUJtizgk6PaRSX5o1o3pZUK8L6rKgXhbUsnCqlZMLNhwg6ec4pdJl_g_pNyy5ong</recordid><startdate>20190301</startdate><enddate>20190301</enddate><creator>Fujiwara, Takahiro</creator><creator>Hidaka, Kuniyuki</creator><creator>Sugibayashi, Keiichi</creator><creator>Matsumoto, Makoto</creator><creator>Kida, Tetsuo</creator><creator>Shiina, Katsuya</creator><general>Springer Singapore</general><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><orcidid>https://orcid.org/0000-0002-9675-7803</orcidid></search><sort><creationdate>20190301</creationdate><title>Investigation of the relation between administered dose and image quality for pediatric 99mTc-DMSA renal scintigraphy: clinical study applying the JSNM (Japanese Society of Nuclear Medicine) pediatric dosage card</title><author>Fujiwara, Takahiro ; Hidaka, Kuniyuki ; Sugibayashi, Keiichi ; Matsumoto, Makoto ; Kida, Tetsuo ; Shiina, Katsuya</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c1186-53b368d73d7e77e8b0e635d6b20d322a3eb3ab1384097fd44daedee485d8bc603</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2019</creationdate><topic>Imaging</topic><topic>Medicine</topic><topic>Medicine & Public Health</topic><topic>Nuclear Medicine</topic><topic>Original Article</topic><topic>Radiology</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Fujiwara, Takahiro</creatorcontrib><creatorcontrib>Hidaka, Kuniyuki</creatorcontrib><creatorcontrib>Sugibayashi, Keiichi</creatorcontrib><creatorcontrib>Matsumoto, Makoto</creatorcontrib><creatorcontrib>Kida, Tetsuo</creatorcontrib><creatorcontrib>Shiina, Katsuya</creatorcontrib><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Annals of nuclear medicine</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Fujiwara, Takahiro</au><au>Hidaka, Kuniyuki</au><au>Sugibayashi, Keiichi</au><au>Matsumoto, Makoto</au><au>Kida, Tetsuo</au><au>Shiina, Katsuya</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Investigation of the relation between administered dose and image quality for pediatric 99mTc-DMSA renal scintigraphy: clinical study applying the JSNM (Japanese Society of Nuclear Medicine) pediatric dosage card: The Japanese Society of Nuclear Medicine Technology (JSNMT), the Optimization of Imaging Technique Committee for Pediatric Nuclear Medicine Studies</atitle><jtitle>Annals of nuclear medicine</jtitle><stitle>Ann Nucl Med</stitle><date>2019-03-01</date><risdate>2019</risdate><volume>33</volume><issue>3</issue><spage>153</spage><epage>159</epage><pages>153-159</pages><issn>0914-7187</issn><eissn>1864-6433</eissn><abstract><![CDATA[Purpose
In 2013, the Japanese Society of Nuclear Medicine (JSNM) announced consensus guidelines for pediatric nuclear medicine. These JSNM guidelines proposed use of lower administered doses compared with traditionally determined doses, which were estimated from age, weight or body surface area (BSA) based on the administered dose for adults in Japan. When the JSNM guidelines are used, the relationship between this recommended administered dose and image quality remains unclear. In this study, we clarified the relationship between administered dose and image quality for pediatric
99m
Tc-DMSA renal scan retrospectively, and verified the diagnosable image quality with the recommended administered dose of the JSNM guidelines.
Materials and methods
Data from 7 pediatric patients who underwent
99m
Tc-DMSA dynamic renal scans according to the guidelines’ recommended doses were collected. Scan frame rate was 1 frame/min, and scan time was up to 8 min. Eight images, which had different acquired time periods from 1 min to 8 min were prepared by adding each frame. Nine nuclear medicine specialists determined 8 images with different acquired times as diagnosable or undiagnosable. A region of interest (ROI) with 50% thresholds was placed on each kidney of every image. Coefficient of variation (CV) was calculated by dividing the standard deviation (
σ
) by the mean counts (
µ
) of each ROI (CV =
σ
/
µ
× 100).
99m
Tc-DMSA renal scans (total of 2821 cases) that were performed previously in collaboration with 6 hospitals were collected, and CVs of these images were calculated in all cases. These 2821 cases were separated into 5 groups for every 10 kg weight; i.e., (1) less than 10 kg, (2) 10–19.9 kg, (3) 20–29.9 kg, (4) 30–39.9 kg, and (5) above 40 kg. Regression line of each group was analyzed in relation to the CV and administered dose. The CV at the point of intersection with the recommended dose range from the guideline was determined for each group. This CV value was considered as the estimated CV of the image obtained when the recommended dose of the guideline was used. Thus, if the CV was equal to or less than the estimated CV value, then the diagnostic image quality was deemed satisfactory.
Results
Average CV of the lower limit of diagnosable images in 7 cases as determined by 9 nuclear medicine specialists was 19.9%. Estimated CV was 21.2–24.2% in the group weighing < 10 kg (group 1), 19.9–20.6% in the group weighing > 10 kg and < 20 kg (group 2), 19.6% in group weighing > 20 kg and < 30 kg (group 3), 19.4–19.5% in the group weighing > 30 kg and < 40 kg (group 4), and 19.8% in the group weighing > 40 kg (group 5). The estimated CVs from groups 1 and 2 with weight < 20 kg exceeded 19.9%.
Conclusions
Although
99m
Tc-DMSA renal scan can be carried out using the guidelines’ recommended dose with conventional image acquisition time in patients weighing 20 kg or more, those < 20 kg need consideration for a longer image acquisition time to obtain diagnosable images.]]></abstract><cop>Singapore</cop><pub>Springer Singapore</pub><doi>10.1007/s12149-018-1320-6</doi><tpages>7</tpages><orcidid>https://orcid.org/0000-0002-9675-7803</orcidid></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0914-7187 |
ispartof | Annals of nuclear medicine, 2019-03, Vol.33 (3), p.153-159 |
issn | 0914-7187 1864-6433 |
language | eng |
recordid | cdi_proquest_miscellaneous_2138649430 |
source | SpringerLink Journals - AutoHoldings |
subjects | Imaging Medicine Medicine & Public Health Nuclear Medicine Original Article Radiology |
title | Investigation of the relation between administered dose and image quality for pediatric 99mTc-DMSA renal scintigraphy: clinical study applying the JSNM (Japanese Society of Nuclear Medicine) pediatric dosage card: The Japanese Society of Nuclear Medicine Technology (JSNMT), the Optimization of Imaging Technique Committee for Pediatric Nuclear Medicine Studies |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-05T23%3A18%3A00IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Investigation%20of%20the%20relation%20between%20administered%20dose%20and%20image%20quality%20for%20pediatric%2099mTc-DMSA%20renal%20scintigraphy:%20clinical%20study%20applying%20the%20JSNM%20(Japanese%20Society%20of%20Nuclear%20Medicine)%20pediatric%20dosage%20card:%20The%20Japanese%20Society%20of%20Nuclear%20Medicine%20Technology%20(JSNMT),%20the%20Optimization%20of%20Imaging%20Technique%20Committee%20for%20Pediatric%20Nuclear%20Medicine%20Studies&rft.jtitle=Annals%20of%20nuclear%20medicine&rft.au=Fujiwara,%20Takahiro&rft.date=2019-03-01&rft.volume=33&rft.issue=3&rft.spage=153&rft.epage=159&rft.pages=153-159&rft.issn=0914-7187&rft.eissn=1864-6433&rft_id=info:doi/10.1007/s12149-018-1320-6&rft_dat=%3Cproquest_cross%3E2138649430%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2138649430&rft_id=info:pmid/&rfr_iscdi=true |