Impact of physiological variables and genetic background on myocardial frequency-resistivity relations in the intact beating murine heart

Conductance measurements for generation of an instantaneous left ventricular (LV) volume signal in the mouse are limited, because the volume signal is a combination of blood and LV muscle, and only the blood signal is desired. We have developed a conductance system that operates at two simultaneous...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:American Journal of Physiology: Cell Physiology 2006-10, Vol.291 (4), p.H1659-H1669
Hauptverfasser: Reyes, Maricela, Steinhelper, Mark E, Alvarez, Jorge A, Escobedo, Daniel, Pearce, John, Valvano, Jonathan W, Pollock, Brad H, Wei, Chia-Ling, Kottam, Anil, Altman, David, Bailey, Steven, Thomsen, Sharon, Lee, Shuko, Colston, James T, Oh, Jung Hwan, Freeman, Gregory L, Feldman, Marc D
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page H1669
container_issue 4
container_start_page H1659
container_title American Journal of Physiology: Cell Physiology
container_volume 291
creator Reyes, Maricela
Steinhelper, Mark E
Alvarez, Jorge A
Escobedo, Daniel
Pearce, John
Valvano, Jonathan W
Pollock, Brad H
Wei, Chia-Ling
Kottam, Anil
Altman, David
Bailey, Steven
Thomsen, Sharon
Lee, Shuko
Colston, James T
Oh, Jung Hwan
Freeman, Gregory L
Feldman, Marc D
description Conductance measurements for generation of an instantaneous left ventricular (LV) volume signal in the mouse are limited, because the volume signal is a combination of blood and LV muscle, and only the blood signal is desired. We have developed a conductance system that operates at two simultaneous frequencies to identify and remove the myocardial contribution to the instantaneous volume signal. This system is based on the observation that myocardial resistivity varies with frequency, whereas blood resistivity does not. For calculation of LV blood volume with the dual-frequency conductance system in mice, in vivo murine myocardial resistivity was measured and combined with an analytic approach. The goals of the present study were to identify and minimize the sources of error in the measurement of myocardial resistivity to enhance the accuracy of the dual-frequency conductance system. We extended these findings to a gene-altered mouse model to determine the impact of measured myocardial resistivity on the calculation of LV pressure-volume relations. We examined the impact of temperature, timing of the measurement during the cardiac cycle, breeding strain, anisotropy, and intrameasurement and interanimal variability on the measurement of intact murine myocardial resistivity. Applying this knowledge to diabetic and nondiabetic 11- and 20- to 24-wk-old mice, we demonstrated differences in myocardial resistivity at low frequencies, enhancement of LV systolic function at 11 wk and LV dilation at 20-24 wk, and histological and electron-microscopic studies demonstrating greater glycogen deposition in the diabetic mice. This study demonstrated the accurate technique of measuring myocardial resistivity and its impact on the determination of LV pressure-volume relations in gene-altered mice.
format Article
fullrecord <record><control><sourceid>proquest</sourceid><recordid>TN_cdi_proquest_miscellaneous_21233670</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>21233670</sourcerecordid><originalsourceid>FETCH-proquest_miscellaneous_212336703</originalsourceid><addsrcrecordid>eNqNjcFKxDAURYMoWEf_4a3cFdJkJuJaFN27H14zr-3TNKl56UA_wb82gh_g6sDhcO-FarqDMW13cPZSNdo627pub6_VjciH1npv3GOjvt_mBX2BNMAybcIppJE9BjhjZuwDCWA8wUiRCnvo0X-OOa1VpQjzljzmE9d8yPS1UvRbm0lYCp-5bJApYOEUBThCmaii_L71VHUcYV4zR4KJMJdbdTVgELr7407dvzy_P722S051WspxZvEUAkZKqxxNZ6x1D9r-O_wBylJaFw</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>21233670</pqid></control><display><type>article</type><title>Impact of physiological variables and genetic background on myocardial frequency-resistivity relations in the intact beating murine heart</title><source>American Physiological Society Paid</source><source>EZB-FREE-00999 freely available EZB journals</source><creator>Reyes, Maricela ; Steinhelper, Mark E ; Alvarez, Jorge A ; Escobedo, Daniel ; Pearce, John ; Valvano, Jonathan W ; Pollock, Brad H ; Wei, Chia-Ling ; Kottam, Anil ; Altman, David ; Bailey, Steven ; Thomsen, Sharon ; Lee, Shuko ; Colston, James T ; Oh, Jung Hwan ; Freeman, Gregory L ; Feldman, Marc D</creator><creatorcontrib>Reyes, Maricela ; Steinhelper, Mark E ; Alvarez, Jorge A ; Escobedo, Daniel ; Pearce, John ; Valvano, Jonathan W ; Pollock, Brad H ; Wei, Chia-Ling ; Kottam, Anil ; Altman, David ; Bailey, Steven ; Thomsen, Sharon ; Lee, Shuko ; Colston, James T ; Oh, Jung Hwan ; Freeman, Gregory L ; Feldman, Marc D</creatorcontrib><description>Conductance measurements for generation of an instantaneous left ventricular (LV) volume signal in the mouse are limited, because the volume signal is a combination of blood and LV muscle, and only the blood signal is desired. We have developed a conductance system that operates at two simultaneous frequencies to identify and remove the myocardial contribution to the instantaneous volume signal. This system is based on the observation that myocardial resistivity varies with frequency, whereas blood resistivity does not. For calculation of LV blood volume with the dual-frequency conductance system in mice, in vivo murine myocardial resistivity was measured and combined with an analytic approach. The goals of the present study were to identify and minimize the sources of error in the measurement of myocardial resistivity to enhance the accuracy of the dual-frequency conductance system. We extended these findings to a gene-altered mouse model to determine the impact of measured myocardial resistivity on the calculation of LV pressure-volume relations. We examined the impact of temperature, timing of the measurement during the cardiac cycle, breeding strain, anisotropy, and intrameasurement and interanimal variability on the measurement of intact murine myocardial resistivity. Applying this knowledge to diabetic and nondiabetic 11- and 20- to 24-wk-old mice, we demonstrated differences in myocardial resistivity at low frequencies, enhancement of LV systolic function at 11 wk and LV dilation at 20-24 wk, and histological and electron-microscopic studies demonstrating greater glycogen deposition in the diabetic mice. This study demonstrated the accurate technique of measuring myocardial resistivity and its impact on the determination of LV pressure-volume relations in gene-altered mice.</description><identifier>ISSN: 0363-6143</identifier><identifier>EISSN: 1522-1563</identifier><language>eng</language><ispartof>American Journal of Physiology: Cell Physiology, 2006-10, Vol.291 (4), p.H1659-H1669</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784</link.rule.ids></links><search><creatorcontrib>Reyes, Maricela</creatorcontrib><creatorcontrib>Steinhelper, Mark E</creatorcontrib><creatorcontrib>Alvarez, Jorge A</creatorcontrib><creatorcontrib>Escobedo, Daniel</creatorcontrib><creatorcontrib>Pearce, John</creatorcontrib><creatorcontrib>Valvano, Jonathan W</creatorcontrib><creatorcontrib>Pollock, Brad H</creatorcontrib><creatorcontrib>Wei, Chia-Ling</creatorcontrib><creatorcontrib>Kottam, Anil</creatorcontrib><creatorcontrib>Altman, David</creatorcontrib><creatorcontrib>Bailey, Steven</creatorcontrib><creatorcontrib>Thomsen, Sharon</creatorcontrib><creatorcontrib>Lee, Shuko</creatorcontrib><creatorcontrib>Colston, James T</creatorcontrib><creatorcontrib>Oh, Jung Hwan</creatorcontrib><creatorcontrib>Freeman, Gregory L</creatorcontrib><creatorcontrib>Feldman, Marc D</creatorcontrib><title>Impact of physiological variables and genetic background on myocardial frequency-resistivity relations in the intact beating murine heart</title><title>American Journal of Physiology: Cell Physiology</title><description>Conductance measurements for generation of an instantaneous left ventricular (LV) volume signal in the mouse are limited, because the volume signal is a combination of blood and LV muscle, and only the blood signal is desired. We have developed a conductance system that operates at two simultaneous frequencies to identify and remove the myocardial contribution to the instantaneous volume signal. This system is based on the observation that myocardial resistivity varies with frequency, whereas blood resistivity does not. For calculation of LV blood volume with the dual-frequency conductance system in mice, in vivo murine myocardial resistivity was measured and combined with an analytic approach. The goals of the present study were to identify and minimize the sources of error in the measurement of myocardial resistivity to enhance the accuracy of the dual-frequency conductance system. We extended these findings to a gene-altered mouse model to determine the impact of measured myocardial resistivity on the calculation of LV pressure-volume relations. We examined the impact of temperature, timing of the measurement during the cardiac cycle, breeding strain, anisotropy, and intrameasurement and interanimal variability on the measurement of intact murine myocardial resistivity. Applying this knowledge to diabetic and nondiabetic 11- and 20- to 24-wk-old mice, we demonstrated differences in myocardial resistivity at low frequencies, enhancement of LV systolic function at 11 wk and LV dilation at 20-24 wk, and histological and electron-microscopic studies demonstrating greater glycogen deposition in the diabetic mice. This study demonstrated the accurate technique of measuring myocardial resistivity and its impact on the determination of LV pressure-volume relations in gene-altered mice.</description><issn>0363-6143</issn><issn>1522-1563</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2006</creationdate><recordtype>article</recordtype><recordid>eNqNjcFKxDAURYMoWEf_4a3cFdJkJuJaFN27H14zr-3TNKl56UA_wb82gh_g6sDhcO-FarqDMW13cPZSNdo627pub6_VjciH1npv3GOjvt_mBX2BNMAybcIppJE9BjhjZuwDCWA8wUiRCnvo0X-OOa1VpQjzljzmE9d8yPS1UvRbm0lYCp-5bJApYOEUBThCmaii_L71VHUcYV4zR4KJMJdbdTVgELr7407dvzy_P722S051WspxZvEUAkZKqxxNZ6x1D9r-O_wBylJaFw</recordid><startdate>20061001</startdate><enddate>20061001</enddate><creator>Reyes, Maricela</creator><creator>Steinhelper, Mark E</creator><creator>Alvarez, Jorge A</creator><creator>Escobedo, Daniel</creator><creator>Pearce, John</creator><creator>Valvano, Jonathan W</creator><creator>Pollock, Brad H</creator><creator>Wei, Chia-Ling</creator><creator>Kottam, Anil</creator><creator>Altman, David</creator><creator>Bailey, Steven</creator><creator>Thomsen, Sharon</creator><creator>Lee, Shuko</creator><creator>Colston, James T</creator><creator>Oh, Jung Hwan</creator><creator>Freeman, Gregory L</creator><creator>Feldman, Marc D</creator><scope>8FD</scope><scope>FR3</scope><scope>P64</scope><scope>RC3</scope></search><sort><creationdate>20061001</creationdate><title>Impact of physiological variables and genetic background on myocardial frequency-resistivity relations in the intact beating murine heart</title><author>Reyes, Maricela ; Steinhelper, Mark E ; Alvarez, Jorge A ; Escobedo, Daniel ; Pearce, John ; Valvano, Jonathan W ; Pollock, Brad H ; Wei, Chia-Ling ; Kottam, Anil ; Altman, David ; Bailey, Steven ; Thomsen, Sharon ; Lee, Shuko ; Colston, James T ; Oh, Jung Hwan ; Freeman, Gregory L ; Feldman, Marc D</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-proquest_miscellaneous_212336703</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2006</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Reyes, Maricela</creatorcontrib><creatorcontrib>Steinhelper, Mark E</creatorcontrib><creatorcontrib>Alvarez, Jorge A</creatorcontrib><creatorcontrib>Escobedo, Daniel</creatorcontrib><creatorcontrib>Pearce, John</creatorcontrib><creatorcontrib>Valvano, Jonathan W</creatorcontrib><creatorcontrib>Pollock, Brad H</creatorcontrib><creatorcontrib>Wei, Chia-Ling</creatorcontrib><creatorcontrib>Kottam, Anil</creatorcontrib><creatorcontrib>Altman, David</creatorcontrib><creatorcontrib>Bailey, Steven</creatorcontrib><creatorcontrib>Thomsen, Sharon</creatorcontrib><creatorcontrib>Lee, Shuko</creatorcontrib><creatorcontrib>Colston, James T</creatorcontrib><creatorcontrib>Oh, Jung Hwan</creatorcontrib><creatorcontrib>Freeman, Gregory L</creatorcontrib><creatorcontrib>Feldman, Marc D</creatorcontrib><collection>Technology Research Database</collection><collection>Engineering Research Database</collection><collection>Biotechnology and BioEngineering Abstracts</collection><collection>Genetics Abstracts</collection><jtitle>American Journal of Physiology: Cell Physiology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Reyes, Maricela</au><au>Steinhelper, Mark E</au><au>Alvarez, Jorge A</au><au>Escobedo, Daniel</au><au>Pearce, John</au><au>Valvano, Jonathan W</au><au>Pollock, Brad H</au><au>Wei, Chia-Ling</au><au>Kottam, Anil</au><au>Altman, David</au><au>Bailey, Steven</au><au>Thomsen, Sharon</au><au>Lee, Shuko</au><au>Colston, James T</au><au>Oh, Jung Hwan</au><au>Freeman, Gregory L</au><au>Feldman, Marc D</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Impact of physiological variables and genetic background on myocardial frequency-resistivity relations in the intact beating murine heart</atitle><jtitle>American Journal of Physiology: Cell Physiology</jtitle><date>2006-10-01</date><risdate>2006</risdate><volume>291</volume><issue>4</issue><spage>H1659</spage><epage>H1669</epage><pages>H1659-H1669</pages><issn>0363-6143</issn><eissn>1522-1563</eissn><abstract>Conductance measurements for generation of an instantaneous left ventricular (LV) volume signal in the mouse are limited, because the volume signal is a combination of blood and LV muscle, and only the blood signal is desired. We have developed a conductance system that operates at two simultaneous frequencies to identify and remove the myocardial contribution to the instantaneous volume signal. This system is based on the observation that myocardial resistivity varies with frequency, whereas blood resistivity does not. For calculation of LV blood volume with the dual-frequency conductance system in mice, in vivo murine myocardial resistivity was measured and combined with an analytic approach. The goals of the present study were to identify and minimize the sources of error in the measurement of myocardial resistivity to enhance the accuracy of the dual-frequency conductance system. We extended these findings to a gene-altered mouse model to determine the impact of measured myocardial resistivity on the calculation of LV pressure-volume relations. We examined the impact of temperature, timing of the measurement during the cardiac cycle, breeding strain, anisotropy, and intrameasurement and interanimal variability on the measurement of intact murine myocardial resistivity. Applying this knowledge to diabetic and nondiabetic 11- and 20- to 24-wk-old mice, we demonstrated differences in myocardial resistivity at low frequencies, enhancement of LV systolic function at 11 wk and LV dilation at 20-24 wk, and histological and electron-microscopic studies demonstrating greater glycogen deposition in the diabetic mice. This study demonstrated the accurate technique of measuring myocardial resistivity and its impact on the determination of LV pressure-volume relations in gene-altered mice.</abstract></addata></record>
fulltext fulltext
identifier ISSN: 0363-6143
ispartof American Journal of Physiology: Cell Physiology, 2006-10, Vol.291 (4), p.H1659-H1669
issn 0363-6143
1522-1563
language eng
recordid cdi_proquest_miscellaneous_21233670
source American Physiological Society Paid; EZB-FREE-00999 freely available EZB journals
title Impact of physiological variables and genetic background on myocardial frequency-resistivity relations in the intact beating murine heart
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-25T04%3A41%3A10IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Impact%20of%20physiological%20variables%20and%20genetic%20background%20on%20myocardial%20frequency-resistivity%20relations%20in%20the%20intact%20beating%20murine%20heart&rft.jtitle=American%20Journal%20of%20Physiology:%20Cell%20Physiology&rft.au=Reyes,%20Maricela&rft.date=2006-10-01&rft.volume=291&rft.issue=4&rft.spage=H1659&rft.epage=H1669&rft.pages=H1659-H1669&rft.issn=0363-6143&rft.eissn=1522-1563&rft_id=info:doi/&rft_dat=%3Cproquest%3E21233670%3C/proquest%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=21233670&rft_id=info:pmid/&rfr_iscdi=true