Flat-panel CT versus 128-slice CT in temporal bone imaging: Assessment of image quality and radiation dose
We compared the image quality and radiation dose of flat-panel CT (FPCT) and multi-slice CT (MSCT) performed respectively with an angiographic unit and a 128-slice CT scanner. We investigated whether the higher spatial resolution of FPCT translated into higher image quality and we sought to eliminat...
Gespeichert in:
Veröffentlicht in: | European journal of radiology 2018-09, Vol.106, p.106-113 |
---|---|
Hauptverfasser: | , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 113 |
---|---|
container_issue | |
container_start_page | 106 |
container_title | European journal of radiology |
container_volume | 106 |
creator | Piergallini, Lorenzo Scola, Elisa Tuscano, Bruno Brambilla, Roberto Campoleoni, Mauro Raimondi, Gabriella Lombardi, Luciano Di Berardino, Federica Zanetti, Diego Sina, Clara Triulzi, Fabio Conte, Giorgio |
description | We compared the image quality and radiation dose of flat-panel CT (FPCT) and multi-slice CT (MSCT) performed respectively with an angiographic unit and a 128-slice CT scanner. We investigated whether the higher spatial resolution of FPCT translated into higher image quality and we sought to eliminate inter-subject variability by scanning temporal bone specimens with both techniques.
Fifteen temporal bone specimens were imaged with FPCT and MSCT. Two neuroradiologists experienced in otoradiology evaluated 30 anatomical structures with a 0–2 score; 18 structures important from a clinical perspective were assigned a twofold value in calculation of the overall score. The radiation dose was calculated through the use of an anthropomorphic phantom.
The image quality was significantly higher for FPCT than MSCT for 10 of the 30 anatomical structures; the overall score was also significantly higher for FPCT (p = 0.001).
The equivalent dose of the two techniques was very similar, but with different effective doses to the organs.
FPCT performed on an angiographic unit provides higher image quality in temporal bone assessment compared to MSCT performed on a 128-slice CT scanner thanks to its higher spatial resolution, with comparable equivalent doses but different effective doses to the organs. |
doi_str_mv | 10.1016/j.ejrad.2018.07.013 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2095532633</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S0720048X18302523</els_id><sourcerecordid>2095532633</sourcerecordid><originalsourceid>FETCH-LOGICAL-c359t-cec6072d5b232542b1921974982111fc045ff0a55f8295804fb0d9efd09f10d3</originalsourceid><addsrcrecordid>eNp9kEFr3DAQhUVoSbZpf0Gh6NiL3ZFkra1CD2FJmkCglz30JmRpFGRsayPZgfz7aLNpjj0NDG_mvfcR8pVBzYBtfww1Dsm4mgPramhrYOKMbFjX8qptefuBbKDlUEHT_b0gn3IeAEA2ip-TCwFMAgi2IcPNaJbqYGYc6W5PnzDlNVPGuyqPweJxF2a64HSIyYy0jzPSMJmHMD_8pFc5Y84TzguN_nWN9HE1Y1ieqZkdLemCWUKcqYsZP5OP3owZv7zNS7K_ud7vbqv7P7_vdlf3lRVSLZVFuy3Bney54LLhPVOcqbZRHWeMeQuN9B6MlL7jSnbQ-B6cQu9AeQZOXJLvp7eHFB9XzIueQrY4jqVjXLPmoKQUfCtEkYqT1KaYc0KvD6m0SM-agT4y1oN-ZayPjDW0ujAuV9_eDNZ-Qvd-8w9qEfw6CbC0fAqYdLYBZ4suJLSLdjH81-AFFg-M1w</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2095532633</pqid></control><display><type>article</type><title>Flat-panel CT versus 128-slice CT in temporal bone imaging: Assessment of image quality and radiation dose</title><source>MEDLINE</source><source>Elsevier ScienceDirect Journals Complete</source><creator>Piergallini, Lorenzo ; Scola, Elisa ; Tuscano, Bruno ; Brambilla, Roberto ; Campoleoni, Mauro ; Raimondi, Gabriella ; Lombardi, Luciano ; Di Berardino, Federica ; Zanetti, Diego ; Sina, Clara ; Triulzi, Fabio ; Conte, Giorgio</creator><creatorcontrib>Piergallini, Lorenzo ; Scola, Elisa ; Tuscano, Bruno ; Brambilla, Roberto ; Campoleoni, Mauro ; Raimondi, Gabriella ; Lombardi, Luciano ; Di Berardino, Federica ; Zanetti, Diego ; Sina, Clara ; Triulzi, Fabio ; Conte, Giorgio</creatorcontrib><description>We compared the image quality and radiation dose of flat-panel CT (FPCT) and multi-slice CT (MSCT) performed respectively with an angiographic unit and a 128-slice CT scanner. We investigated whether the higher spatial resolution of FPCT translated into higher image quality and we sought to eliminate inter-subject variability by scanning temporal bone specimens with both techniques.
Fifteen temporal bone specimens were imaged with FPCT and MSCT. Two neuroradiologists experienced in otoradiology evaluated 30 anatomical structures with a 0–2 score; 18 structures important from a clinical perspective were assigned a twofold value in calculation of the overall score. The radiation dose was calculated through the use of an anthropomorphic phantom.
The image quality was significantly higher for FPCT than MSCT for 10 of the 30 anatomical structures; the overall score was also significantly higher for FPCT (p = 0.001).
The equivalent dose of the two techniques was very similar, but with different effective doses to the organs.
FPCT performed on an angiographic unit provides higher image quality in temporal bone assessment compared to MSCT performed on a 128-slice CT scanner thanks to its higher spatial resolution, with comparable equivalent doses but different effective doses to the organs.</description><identifier>ISSN: 0720-048X</identifier><identifier>EISSN: 1872-7727</identifier><identifier>DOI: 10.1016/j.ejrad.2018.07.013</identifier><identifier>PMID: 30150031</identifier><language>eng</language><publisher>Ireland: Elsevier B.V</publisher><subject>Anatomy ; Cadaver ; Cone-beam computed tomography ; Ear ; Humans ; Multidetector computed tomography ; Radiation Dosage ; Temporal bone ; Temporal Bone - anatomy & histology ; Tomography Scanners, X-Ray Computed ; Tomography, X-Ray Computed - instrumentation ; Tomography, X-Ray Computed - methods</subject><ispartof>European journal of radiology, 2018-09, Vol.106, p.106-113</ispartof><rights>2018 Elsevier B.V.</rights><rights>Copyright © 2018 Elsevier B.V. All rights reserved.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c359t-cec6072d5b232542b1921974982111fc045ff0a55f8295804fb0d9efd09f10d3</citedby><cites>FETCH-LOGICAL-c359t-cec6072d5b232542b1921974982111fc045ff0a55f8295804fb0d9efd09f10d3</cites><orcidid>0000-0002-8116-4108 ; 0000-0001-7514-5624 ; 0000-0001-9056-4122 ; 0000-0002-1562-7389</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://dx.doi.org/10.1016/j.ejrad.2018.07.013$$EHTML$$P50$$Gelsevier$$H</linktohtml><link.rule.ids>314,780,784,3550,27924,27925,45995</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/30150031$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Piergallini, Lorenzo</creatorcontrib><creatorcontrib>Scola, Elisa</creatorcontrib><creatorcontrib>Tuscano, Bruno</creatorcontrib><creatorcontrib>Brambilla, Roberto</creatorcontrib><creatorcontrib>Campoleoni, Mauro</creatorcontrib><creatorcontrib>Raimondi, Gabriella</creatorcontrib><creatorcontrib>Lombardi, Luciano</creatorcontrib><creatorcontrib>Di Berardino, Federica</creatorcontrib><creatorcontrib>Zanetti, Diego</creatorcontrib><creatorcontrib>Sina, Clara</creatorcontrib><creatorcontrib>Triulzi, Fabio</creatorcontrib><creatorcontrib>Conte, Giorgio</creatorcontrib><title>Flat-panel CT versus 128-slice CT in temporal bone imaging: Assessment of image quality and radiation dose</title><title>European journal of radiology</title><addtitle>Eur J Radiol</addtitle><description>We compared the image quality and radiation dose of flat-panel CT (FPCT) and multi-slice CT (MSCT) performed respectively with an angiographic unit and a 128-slice CT scanner. We investigated whether the higher spatial resolution of FPCT translated into higher image quality and we sought to eliminate inter-subject variability by scanning temporal bone specimens with both techniques.
Fifteen temporal bone specimens were imaged with FPCT and MSCT. Two neuroradiologists experienced in otoradiology evaluated 30 anatomical structures with a 0–2 score; 18 structures important from a clinical perspective were assigned a twofold value in calculation of the overall score. The radiation dose was calculated through the use of an anthropomorphic phantom.
The image quality was significantly higher for FPCT than MSCT for 10 of the 30 anatomical structures; the overall score was also significantly higher for FPCT (p = 0.001).
The equivalent dose of the two techniques was very similar, but with different effective doses to the organs.
FPCT performed on an angiographic unit provides higher image quality in temporal bone assessment compared to MSCT performed on a 128-slice CT scanner thanks to its higher spatial resolution, with comparable equivalent doses but different effective doses to the organs.</description><subject>Anatomy</subject><subject>Cadaver</subject><subject>Cone-beam computed tomography</subject><subject>Ear</subject><subject>Humans</subject><subject>Multidetector computed tomography</subject><subject>Radiation Dosage</subject><subject>Temporal bone</subject><subject>Temporal Bone - anatomy & histology</subject><subject>Tomography Scanners, X-Ray Computed</subject><subject>Tomography, X-Ray Computed - instrumentation</subject><subject>Tomography, X-Ray Computed - methods</subject><issn>0720-048X</issn><issn>1872-7727</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2018</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp9kEFr3DAQhUVoSbZpf0Gh6NiL3ZFkra1CD2FJmkCglz30JmRpFGRsayPZgfz7aLNpjj0NDG_mvfcR8pVBzYBtfww1Dsm4mgPramhrYOKMbFjX8qptefuBbKDlUEHT_b0gn3IeAEA2ip-TCwFMAgi2IcPNaJbqYGYc6W5PnzDlNVPGuyqPweJxF2a64HSIyYy0jzPSMJmHMD_8pFc5Y84TzguN_nWN9HE1Y1ieqZkdLemCWUKcqYsZP5OP3owZv7zNS7K_ud7vbqv7P7_vdlf3lRVSLZVFuy3Bney54LLhPVOcqbZRHWeMeQuN9B6MlL7jSnbQ-B6cQu9AeQZOXJLvp7eHFB9XzIueQrY4jqVjXLPmoKQUfCtEkYqT1KaYc0KvD6m0SM-agT4y1oN-ZayPjDW0ujAuV9_eDNZ-Qvd-8w9qEfw6CbC0fAqYdLYBZ4suJLSLdjH81-AFFg-M1w</recordid><startdate>201809</startdate><enddate>201809</enddate><creator>Piergallini, Lorenzo</creator><creator>Scola, Elisa</creator><creator>Tuscano, Bruno</creator><creator>Brambilla, Roberto</creator><creator>Campoleoni, Mauro</creator><creator>Raimondi, Gabriella</creator><creator>Lombardi, Luciano</creator><creator>Di Berardino, Federica</creator><creator>Zanetti, Diego</creator><creator>Sina, Clara</creator><creator>Triulzi, Fabio</creator><creator>Conte, Giorgio</creator><general>Elsevier B.V</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><orcidid>https://orcid.org/0000-0002-8116-4108</orcidid><orcidid>https://orcid.org/0000-0001-7514-5624</orcidid><orcidid>https://orcid.org/0000-0001-9056-4122</orcidid><orcidid>https://orcid.org/0000-0002-1562-7389</orcidid></search><sort><creationdate>201809</creationdate><title>Flat-panel CT versus 128-slice CT in temporal bone imaging: Assessment of image quality and radiation dose</title><author>Piergallini, Lorenzo ; Scola, Elisa ; Tuscano, Bruno ; Brambilla, Roberto ; Campoleoni, Mauro ; Raimondi, Gabriella ; Lombardi, Luciano ; Di Berardino, Federica ; Zanetti, Diego ; Sina, Clara ; Triulzi, Fabio ; Conte, Giorgio</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c359t-cec6072d5b232542b1921974982111fc045ff0a55f8295804fb0d9efd09f10d3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2018</creationdate><topic>Anatomy</topic><topic>Cadaver</topic><topic>Cone-beam computed tomography</topic><topic>Ear</topic><topic>Humans</topic><topic>Multidetector computed tomography</topic><topic>Radiation Dosage</topic><topic>Temporal bone</topic><topic>Temporal Bone - anatomy & histology</topic><topic>Tomography Scanners, X-Ray Computed</topic><topic>Tomography, X-Ray Computed - instrumentation</topic><topic>Tomography, X-Ray Computed - methods</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Piergallini, Lorenzo</creatorcontrib><creatorcontrib>Scola, Elisa</creatorcontrib><creatorcontrib>Tuscano, Bruno</creatorcontrib><creatorcontrib>Brambilla, Roberto</creatorcontrib><creatorcontrib>Campoleoni, Mauro</creatorcontrib><creatorcontrib>Raimondi, Gabriella</creatorcontrib><creatorcontrib>Lombardi, Luciano</creatorcontrib><creatorcontrib>Di Berardino, Federica</creatorcontrib><creatorcontrib>Zanetti, Diego</creatorcontrib><creatorcontrib>Sina, Clara</creatorcontrib><creatorcontrib>Triulzi, Fabio</creatorcontrib><creatorcontrib>Conte, Giorgio</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>European journal of radiology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Piergallini, Lorenzo</au><au>Scola, Elisa</au><au>Tuscano, Bruno</au><au>Brambilla, Roberto</au><au>Campoleoni, Mauro</au><au>Raimondi, Gabriella</au><au>Lombardi, Luciano</au><au>Di Berardino, Federica</au><au>Zanetti, Diego</au><au>Sina, Clara</au><au>Triulzi, Fabio</au><au>Conte, Giorgio</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Flat-panel CT versus 128-slice CT in temporal bone imaging: Assessment of image quality and radiation dose</atitle><jtitle>European journal of radiology</jtitle><addtitle>Eur J Radiol</addtitle><date>2018-09</date><risdate>2018</risdate><volume>106</volume><spage>106</spage><epage>113</epage><pages>106-113</pages><issn>0720-048X</issn><eissn>1872-7727</eissn><abstract>We compared the image quality and radiation dose of flat-panel CT (FPCT) and multi-slice CT (MSCT) performed respectively with an angiographic unit and a 128-slice CT scanner. We investigated whether the higher spatial resolution of FPCT translated into higher image quality and we sought to eliminate inter-subject variability by scanning temporal bone specimens with both techniques.
Fifteen temporal bone specimens were imaged with FPCT and MSCT. Two neuroradiologists experienced in otoradiology evaluated 30 anatomical structures with a 0–2 score; 18 structures important from a clinical perspective were assigned a twofold value in calculation of the overall score. The radiation dose was calculated through the use of an anthropomorphic phantom.
The image quality was significantly higher for FPCT than MSCT for 10 of the 30 anatomical structures; the overall score was also significantly higher for FPCT (p = 0.001).
The equivalent dose of the two techniques was very similar, but with different effective doses to the organs.
FPCT performed on an angiographic unit provides higher image quality in temporal bone assessment compared to MSCT performed on a 128-slice CT scanner thanks to its higher spatial resolution, with comparable equivalent doses but different effective doses to the organs.</abstract><cop>Ireland</cop><pub>Elsevier B.V</pub><pmid>30150031</pmid><doi>10.1016/j.ejrad.2018.07.013</doi><tpages>8</tpages><orcidid>https://orcid.org/0000-0002-8116-4108</orcidid><orcidid>https://orcid.org/0000-0001-7514-5624</orcidid><orcidid>https://orcid.org/0000-0001-9056-4122</orcidid><orcidid>https://orcid.org/0000-0002-1562-7389</orcidid></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0720-048X |
ispartof | European journal of radiology, 2018-09, Vol.106, p.106-113 |
issn | 0720-048X 1872-7727 |
language | eng |
recordid | cdi_proquest_miscellaneous_2095532633 |
source | MEDLINE; Elsevier ScienceDirect Journals Complete |
subjects | Anatomy Cadaver Cone-beam computed tomography Ear Humans Multidetector computed tomography Radiation Dosage Temporal bone Temporal Bone - anatomy & histology Tomography Scanners, X-Ray Computed Tomography, X-Ray Computed - instrumentation Tomography, X-Ray Computed - methods |
title | Flat-panel CT versus 128-slice CT in temporal bone imaging: Assessment of image quality and radiation dose |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-24T01%3A00%3A47IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Flat-panel%20CT%20versus%20128-slice%20CT%20in%20temporal%20bone%20imaging:%20Assessment%20of%20image%20quality%20and%20radiation%20dose&rft.jtitle=European%20journal%20of%20radiology&rft.au=Piergallini,%20Lorenzo&rft.date=2018-09&rft.volume=106&rft.spage=106&rft.epage=113&rft.pages=106-113&rft.issn=0720-048X&rft.eissn=1872-7727&rft_id=info:doi/10.1016/j.ejrad.2018.07.013&rft_dat=%3Cproquest_cross%3E2095532633%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2095532633&rft_id=info:pmid/30150031&rft_els_id=S0720048X18302523&rfr_iscdi=true |