Severe Cobalt Intoxication Due to Prosthesis Wear in Repeated Total Hip Arthroplasty

Abstract Our study describes a 55-year-old man with a total hip prosthesis (ceramic femoral head and polyethylene [PE] inlay). After a fall, the ceramic head broke into several pieces, and a subsequent revision surgery with metal femoral head and PE inlay was performed. Three months later, the so fa...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of arthroplasty 2009-08, Vol.24 (5), p.825.e15-825.e20
Hauptverfasser: Oldenburg, Marcus, MD, Wegner, Ralf, MD, Baur, Xaver, MD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract Our study describes a 55-year-old man with a total hip prosthesis (ceramic femoral head and polyethylene [PE] inlay). After a fall, the ceramic head broke into several pieces, and a subsequent revision surgery with metal femoral head and PE inlay was performed. Three months later, the so far healthy patient complained of multiorganic symptoms. The subsequent clinical examinations revealed hypothyroidism, peripheral neuropathy, and cardiomyopathy for the first time. In a second revision surgery, a massive deterioration of the metal femoral head by overlooked particles of the broken ceramic head was found. The cobalt concentration in blood exceeded a never before observed level. After replacement of the scraped femoral head, the cobalt concentration decreased considerably, and most of the symptoms disappeared.
ISSN:0883-5403
1532-8406
DOI:10.1016/j.arth.2008.07.017