Evaluating Lens Dose Reduction in Pediatric Neuroradiology Examinations Using Automated Kilovoltage Selection Software

The purpose of this study is to evaluate the potential of an automated kilo-voltage selection software for the reduction of lens dose in pediatric CT scans. Two metal oxide semiconductor field effect transistor (MOSFET) detectors measured the lens dose in two anthropomorphic 1- and 5-year-old phanto...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:American journal of roentgenology (1976) 2018-09, Vol.211 (3), p.635-640
Hauptverfasser: Raudabaugh, Justin, Smith, Aaron K, Moore, Bria, Ramirez-Giraldo, Juan Carlos, Januzis, Natalie, Yoshizumi, Terry
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 640
container_issue 3
container_start_page 635
container_title American journal of roentgenology (1976)
container_volume 211
creator Raudabaugh, Justin
Smith, Aaron K
Moore, Bria
Ramirez-Giraldo, Juan Carlos
Januzis, Natalie
Yoshizumi, Terry
description The purpose of this study is to evaluate the potential of an automated kilo-voltage selection software for the reduction of lens dose in pediatric CT scans. Two metal oxide semiconductor field effect transistor (MOSFET) detectors measured the lens dose in two anthropomorphic 1- and 5-year-old phantoms. These phantoms were scanned using a clinical pediatric brain protocol at 120 kVp as a control with the MDCT scanner. Scans were then repeated using automated kilovoltage software. The automated kilovoltage was set to operate at tube potentials of 120, 110, and 100 kVp. Dose savings were compared with the average lens dose of both eyes between automated kilovoltage and the control setting. Image quality was studied by contrast-to-noise ratios (CNRs) for each setting. The mean (± SD) lens dose from the routine brain scan without automated kilovoltage was 0.92 ± 0.03 cGy and 0.81 ± 0.03 cGy for the 1- and 5-year-old phantoms, respectively. Use of the automated kilovoltage software at 120 kVp, 110 kVp, and 100 kVp resulted in dose reductions of 9.8%, 17.4%, and 19.6%, respectively, for the 1-year-old phantom and 1.2%, 8.6%, and 17.3%, respectively, for the 5-year-old phantom. The CNR for all automated kilovoltage scans was within 11% of the control scans for the 1-year-old and within 6% for the 5-year-old phantom. Our results show that automated kilovoltage software is effective for reducing the radiation dose to the lens of the eye in pediatric patients. Furthermore, the image quality by CNR remained acceptable within 11% of the baseline for all kilovoltage settings used.
doi_str_mv 10.2214/ajr.17.19089
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2061409188</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2061409188</sourcerecordid><originalsourceid>FETCH-LOGICAL-c357t-5f79e8e373dab91d2e86b22ac9637fece458dcf1a1dbf5b844447838f17beacd3</originalsourceid><addsrcrecordid>eNo9kDtv2zAURomiQey62TIHHDtUDilKIjkaqfuKkRR5AN2EK_LKYCCJCUm59b-PHSf5lm85OMMh5JSzeZ7z4hwewpzLOddM6Q9kysuiygQv-EcyZaLimWLi74R8ivGBMSaVlsdkkmtd6CJnU7JZbqAbIblhTVc4RPrNR6Q3aEeTnB-oG-gftA5ScIZe4Rh8AOt859dbuvwPvRtgz0V6H_eKxZh8DwktvXSd3_guwRrpLXZ40N36Nv2DgJ_JUQtdxJPXn5H778u7i5_Z6vrHr4vFKjOilCkrW6lRoZDCQqO5zVFVTZ6D0ZWQLRosSmVNy4Hbpi0bVewmlVAtlw2CsWJGvhy8j8E_jRhT3btosOtgQD_GOmcVL5jmSu3QrwfUBB9jwLZ-DK6HsK05q_el68Xvm5rL-qX0Dj97NY9Nj_YdfksrngHG-3yI</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2061409188</pqid></control><display><type>article</type><title>Evaluating Lens Dose Reduction in Pediatric Neuroradiology Examinations Using Automated Kilovoltage Selection Software</title><source>American Roentgen Ray Society</source><source>Alma/SFX Local Collection</source><creator>Raudabaugh, Justin ; Smith, Aaron K ; Moore, Bria ; Ramirez-Giraldo, Juan Carlos ; Januzis, Natalie ; Yoshizumi, Terry</creator><creatorcontrib>Raudabaugh, Justin ; Smith, Aaron K ; Moore, Bria ; Ramirez-Giraldo, Juan Carlos ; Januzis, Natalie ; Yoshizumi, Terry</creatorcontrib><description>The purpose of this study is to evaluate the potential of an automated kilo-voltage selection software for the reduction of lens dose in pediatric CT scans. Two metal oxide semiconductor field effect transistor (MOSFET) detectors measured the lens dose in two anthropomorphic 1- and 5-year-old phantoms. These phantoms were scanned using a clinical pediatric brain protocol at 120 kVp as a control with the MDCT scanner. Scans were then repeated using automated kilovoltage software. The automated kilovoltage was set to operate at tube potentials of 120, 110, and 100 kVp. Dose savings were compared with the average lens dose of both eyes between automated kilovoltage and the control setting. Image quality was studied by contrast-to-noise ratios (CNRs) for each setting. The mean (± SD) lens dose from the routine brain scan without automated kilovoltage was 0.92 ± 0.03 cGy and 0.81 ± 0.03 cGy for the 1- and 5-year-old phantoms, respectively. Use of the automated kilovoltage software at 120 kVp, 110 kVp, and 100 kVp resulted in dose reductions of 9.8%, 17.4%, and 19.6%, respectively, for the 1-year-old phantom and 1.2%, 8.6%, and 17.3%, respectively, for the 5-year-old phantom. The CNR for all automated kilovoltage scans was within 11% of the control scans for the 1-year-old and within 6% for the 5-year-old phantom. Our results show that automated kilovoltage software is effective for reducing the radiation dose to the lens of the eye in pediatric patients. Furthermore, the image quality by CNR remained acceptable within 11% of the baseline for all kilovoltage settings used.</description><identifier>ISSN: 0361-803X</identifier><identifier>EISSN: 1546-3141</identifier><identifier>DOI: 10.2214/ajr.17.19089</identifier><identifier>PMID: 29949420</identifier><language>eng</language><publisher>United States</publisher><ispartof>American journal of roentgenology (1976), 2018-09, Vol.211 (3), p.635-640</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c357t-5f79e8e373dab91d2e86b22ac9637fece458dcf1a1dbf5b844447838f17beacd3</citedby><cites>FETCH-LOGICAL-c357t-5f79e8e373dab91d2e86b22ac9637fece458dcf1a1dbf5b844447838f17beacd3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,4120,27924,27925</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/29949420$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Raudabaugh, Justin</creatorcontrib><creatorcontrib>Smith, Aaron K</creatorcontrib><creatorcontrib>Moore, Bria</creatorcontrib><creatorcontrib>Ramirez-Giraldo, Juan Carlos</creatorcontrib><creatorcontrib>Januzis, Natalie</creatorcontrib><creatorcontrib>Yoshizumi, Terry</creatorcontrib><title>Evaluating Lens Dose Reduction in Pediatric Neuroradiology Examinations Using Automated Kilovoltage Selection Software</title><title>American journal of roentgenology (1976)</title><addtitle>AJR Am J Roentgenol</addtitle><description>The purpose of this study is to evaluate the potential of an automated kilo-voltage selection software for the reduction of lens dose in pediatric CT scans. Two metal oxide semiconductor field effect transistor (MOSFET) detectors measured the lens dose in two anthropomorphic 1- and 5-year-old phantoms. These phantoms were scanned using a clinical pediatric brain protocol at 120 kVp as a control with the MDCT scanner. Scans were then repeated using automated kilovoltage software. The automated kilovoltage was set to operate at tube potentials of 120, 110, and 100 kVp. Dose savings were compared with the average lens dose of both eyes between automated kilovoltage and the control setting. Image quality was studied by contrast-to-noise ratios (CNRs) for each setting. The mean (± SD) lens dose from the routine brain scan without automated kilovoltage was 0.92 ± 0.03 cGy and 0.81 ± 0.03 cGy for the 1- and 5-year-old phantoms, respectively. Use of the automated kilovoltage software at 120 kVp, 110 kVp, and 100 kVp resulted in dose reductions of 9.8%, 17.4%, and 19.6%, respectively, for the 1-year-old phantom and 1.2%, 8.6%, and 17.3%, respectively, for the 5-year-old phantom. The CNR for all automated kilovoltage scans was within 11% of the control scans for the 1-year-old and within 6% for the 5-year-old phantom. Our results show that automated kilovoltage software is effective for reducing the radiation dose to the lens of the eye in pediatric patients. Furthermore, the image quality by CNR remained acceptable within 11% of the baseline for all kilovoltage settings used.</description><issn>0361-803X</issn><issn>1546-3141</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2018</creationdate><recordtype>article</recordtype><recordid>eNo9kDtv2zAURomiQey62TIHHDtUDilKIjkaqfuKkRR5AN2EK_LKYCCJCUm59b-PHSf5lm85OMMh5JSzeZ7z4hwewpzLOddM6Q9kysuiygQv-EcyZaLimWLi74R8ivGBMSaVlsdkkmtd6CJnU7JZbqAbIblhTVc4RPrNR6Q3aEeTnB-oG-gftA5ScIZe4Rh8AOt859dbuvwPvRtgz0V6H_eKxZh8DwktvXSd3_guwRrpLXZ40N36Nv2DgJ_JUQtdxJPXn5H778u7i5_Z6vrHr4vFKjOilCkrW6lRoZDCQqO5zVFVTZ6D0ZWQLRosSmVNy4Hbpi0bVewmlVAtlw2CsWJGvhy8j8E_jRhT3btosOtgQD_GOmcVL5jmSu3QrwfUBB9jwLZ-DK6HsK05q_el68Xvm5rL-qX0Dj97NY9Nj_YdfksrngHG-3yI</recordid><startdate>201809</startdate><enddate>201809</enddate><creator>Raudabaugh, Justin</creator><creator>Smith, Aaron K</creator><creator>Moore, Bria</creator><creator>Ramirez-Giraldo, Juan Carlos</creator><creator>Januzis, Natalie</creator><creator>Yoshizumi, Terry</creator><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>201809</creationdate><title>Evaluating Lens Dose Reduction in Pediatric Neuroradiology Examinations Using Automated Kilovoltage Selection Software</title><author>Raudabaugh, Justin ; Smith, Aaron K ; Moore, Bria ; Ramirez-Giraldo, Juan Carlos ; Januzis, Natalie ; Yoshizumi, Terry</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c357t-5f79e8e373dab91d2e86b22ac9637fece458dcf1a1dbf5b844447838f17beacd3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2018</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Raudabaugh, Justin</creatorcontrib><creatorcontrib>Smith, Aaron K</creatorcontrib><creatorcontrib>Moore, Bria</creatorcontrib><creatorcontrib>Ramirez-Giraldo, Juan Carlos</creatorcontrib><creatorcontrib>Januzis, Natalie</creatorcontrib><creatorcontrib>Yoshizumi, Terry</creatorcontrib><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>American journal of roentgenology (1976)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Raudabaugh, Justin</au><au>Smith, Aaron K</au><au>Moore, Bria</au><au>Ramirez-Giraldo, Juan Carlos</au><au>Januzis, Natalie</au><au>Yoshizumi, Terry</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Evaluating Lens Dose Reduction in Pediatric Neuroradiology Examinations Using Automated Kilovoltage Selection Software</atitle><jtitle>American journal of roentgenology (1976)</jtitle><addtitle>AJR Am J Roentgenol</addtitle><date>2018-09</date><risdate>2018</risdate><volume>211</volume><issue>3</issue><spage>635</spage><epage>640</epage><pages>635-640</pages><issn>0361-803X</issn><eissn>1546-3141</eissn><abstract>The purpose of this study is to evaluate the potential of an automated kilo-voltage selection software for the reduction of lens dose in pediatric CT scans. Two metal oxide semiconductor field effect transistor (MOSFET) detectors measured the lens dose in two anthropomorphic 1- and 5-year-old phantoms. These phantoms were scanned using a clinical pediatric brain protocol at 120 kVp as a control with the MDCT scanner. Scans were then repeated using automated kilovoltage software. The automated kilovoltage was set to operate at tube potentials of 120, 110, and 100 kVp. Dose savings were compared with the average lens dose of both eyes between automated kilovoltage and the control setting. Image quality was studied by contrast-to-noise ratios (CNRs) for each setting. The mean (± SD) lens dose from the routine brain scan without automated kilovoltage was 0.92 ± 0.03 cGy and 0.81 ± 0.03 cGy for the 1- and 5-year-old phantoms, respectively. Use of the automated kilovoltage software at 120 kVp, 110 kVp, and 100 kVp resulted in dose reductions of 9.8%, 17.4%, and 19.6%, respectively, for the 1-year-old phantom and 1.2%, 8.6%, and 17.3%, respectively, for the 5-year-old phantom. The CNR for all automated kilovoltage scans was within 11% of the control scans for the 1-year-old and within 6% for the 5-year-old phantom. Our results show that automated kilovoltage software is effective for reducing the radiation dose to the lens of the eye in pediatric patients. Furthermore, the image quality by CNR remained acceptable within 11% of the baseline for all kilovoltage settings used.</abstract><cop>United States</cop><pmid>29949420</pmid><doi>10.2214/ajr.17.19089</doi><tpages>6</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0361-803X
ispartof American journal of roentgenology (1976), 2018-09, Vol.211 (3), p.635-640
issn 0361-803X
1546-3141
language eng
recordid cdi_proquest_miscellaneous_2061409188
source American Roentgen Ray Society; Alma/SFX Local Collection
title Evaluating Lens Dose Reduction in Pediatric Neuroradiology Examinations Using Automated Kilovoltage Selection Software
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-04T07%3A14%3A16IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Evaluating%20Lens%20Dose%20Reduction%20in%20Pediatric%20Neuroradiology%20Examinations%20Using%20Automated%20Kilovoltage%20Selection%20Software&rft.jtitle=American%20journal%20of%20roentgenology%20(1976)&rft.au=Raudabaugh,%20Justin&rft.date=2018-09&rft.volume=211&rft.issue=3&rft.spage=635&rft.epage=640&rft.pages=635-640&rft.issn=0361-803X&rft.eissn=1546-3141&rft_id=info:doi/10.2214/ajr.17.19089&rft_dat=%3Cproquest_cross%3E2061409188%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2061409188&rft_id=info:pmid/29949420&rfr_iscdi=true