Real-Time Measurement of Polyvinylpyrrolidone Eluted from Polysulfone Membrane Dialyzers Based on the Ultraviolet Spectrum

The objective of this study was to develop a new method for measuring polyvinylpyrrolidone (PVP) eluted from polysulfone (PSu) membrane dialyzers. The Müller method is generally used for the measurement of PVP, in which the PVP concentration is determined by measuring the absorbance after a red colo...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:ASAIO journal (1992) 2019-07, Vol.65 (5), p.522-526
Hauptverfasser: Shimamoto, Yoshimasa, Matsushita, Seigo, Yamamoto, Takatoshi, Nakamura, Yuki, Kamada, Aki, Miyata, Masahiro, Umimoto, Koichi
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 526
container_issue 5
container_start_page 522
container_title ASAIO journal (1992)
container_volume 65
creator Shimamoto, Yoshimasa
Matsushita, Seigo
Yamamoto, Takatoshi
Nakamura, Yuki
Kamada, Aki
Miyata, Masahiro
Umimoto, Koichi
description The objective of this study was to develop a new method for measuring polyvinylpyrrolidone (PVP) eluted from polysulfone (PSu) membrane dialyzers. The Müller method is generally used for the measurement of PVP, in which the PVP concentration is determined by measuring the absorbance after a red color is generated by the formation of PVP-iodine complexes when iodine is added to a sample. In contrast, our method does not require any reagents and allows real-time measurement of PVP by the ultraviolet absorption spectrum (UV-s method). In this study, the UV-s method and the Müller method were used to measure PVP eluted from two types of PSu membrane dialyzer (PS-1.6UW sterilized by autoclaving [n = 10] and APS-15SA sterilized by gamma radiation [n = 10]). Polyvinylpyrrolidone concentrations measured by the two methods showed a significant positive correlation (rs = 0.99, p = 0.0006). The PVP concentration (median [25th–75th percentiles]) in PS-1.6UW dialyzer washings obtained by rinsing with physiologic saline was 2.0 (1.18–4.85) mg/L with the Müller method and 3.35 (2.38–4.23) mg/L with the UV-s method, showing no significant difference. However, the PVP concentration in APS-15SA dialyzer washings was 0 (0–0.35) mg/L by the Müller method and 0.95 (0.45–2.58) mg/L by the UV-s method, and there was a significant difference between the two methods. In conclusion, the low concentration of PVP eluted from a PSu dialyzer sterilized by gamma radiation was hardly detected by the Müller method but could be clearly detected by the new UV-s method. These findings suggest that the UV-s method could be used for real-time measurement of PVP eluted from PSu membrane dialyzers.
doi_str_mv 10.1097/MAT.0000000000000829
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2049934096</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2049934096</sourcerecordid><originalsourceid>FETCH-LOGICAL-c4229-6b86076cf1c2e06e31e6911657ae5bb690d8e3b361ee2840b996b64ebb5f8f083</originalsourceid><addsrcrecordid>eNp9kE1PAyEQhonRWL_-gTEcvWyFZZfCsX6btNFoTbxtYDubrrKlAluz_fWiVWM8yIXJ8Mw75EHokJI-JXJwMh5O-uT3EancQDs0ZyKRGXvajDXJRZJKynto1_tnQuIjo9uol0rBGWdkB63uQZlkUjeAx6B866CBecC2wnfWdMt63plF55w19dTOAV-YNsAUV842n4BvTfXRH0OjnYrFea1MtwLn8anykbRzHGaAH01wallbAwE_LKAMrm320ValjIeDr3sPPV5eTM6uk9Ht1c3ZcJSUWZrKhGvByYCXFS1TIBwYBS4p5flAQa41l2QqgGnGKUAqMqKl5JpnoHVeiYoItoeO17kLZ19b8KFoal-CMfG_tvVFSjIpWUYkj2i2RktnvXdQFQtXN8p1BSXFh_UiWi_-Wo9jR18bWt3A9GfoW3MExBp4syZEOS-mfQNXzKL7MPs_-x2wdpB0</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2049934096</pqid></control><display><type>article</type><title>Real-Time Measurement of Polyvinylpyrrolidone Eluted from Polysulfone Membrane Dialyzers Based on the Ultraviolet Spectrum</title><source>MEDLINE</source><source>Journals@Ovid LWW Legacy Archive</source><source>Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals</source><creator>Shimamoto, Yoshimasa ; Matsushita, Seigo ; Yamamoto, Takatoshi ; Nakamura, Yuki ; Kamada, Aki ; Miyata, Masahiro ; Umimoto, Koichi</creator><creatorcontrib>Shimamoto, Yoshimasa ; Matsushita, Seigo ; Yamamoto, Takatoshi ; Nakamura, Yuki ; Kamada, Aki ; Miyata, Masahiro ; Umimoto, Koichi</creatorcontrib><description>The objective of this study was to develop a new method for measuring polyvinylpyrrolidone (PVP) eluted from polysulfone (PSu) membrane dialyzers. The Müller method is generally used for the measurement of PVP, in which the PVP concentration is determined by measuring the absorbance after a red color is generated by the formation of PVP-iodine complexes when iodine is added to a sample. In contrast, our method does not require any reagents and allows real-time measurement of PVP by the ultraviolet absorption spectrum (UV-s method). In this study, the UV-s method and the Müller method were used to measure PVP eluted from two types of PSu membrane dialyzer (PS-1.6UW sterilized by autoclaving [n = 10] and APS-15SA sterilized by gamma radiation [n = 10]). Polyvinylpyrrolidone concentrations measured by the two methods showed a significant positive correlation (rs = 0.99, p = 0.0006). The PVP concentration (median [25th–75th percentiles]) in PS-1.6UW dialyzer washings obtained by rinsing with physiologic saline was 2.0 (1.18–4.85) mg/L with the Müller method and 3.35 (2.38–4.23) mg/L with the UV-s method, showing no significant difference. However, the PVP concentration in APS-15SA dialyzer washings was 0 (0–0.35) mg/L by the Müller method and 0.95 (0.45–2.58) mg/L by the UV-s method, and there was a significant difference between the two methods. In conclusion, the low concentration of PVP eluted from a PSu dialyzer sterilized by gamma radiation was hardly detected by the Müller method but could be clearly detected by the new UV-s method. These findings suggest that the UV-s method could be used for real-time measurement of PVP eluted from PSu membrane dialyzers.</description><identifier>ISSN: 1058-2916</identifier><identifier>EISSN: 1538-943X</identifier><identifier>DOI: 10.1097/MAT.0000000000000829</identifier><identifier>PMID: 29863630</identifier><language>eng</language><publisher>United States: Copyright by the American Society for Artificial Internal Organs</publisher><subject>Humans ; Kidneys, Artificial ; Membranes, Artificial ; Polymers - chemistry ; Povidone - analysis ; Renal Dialysis - methods ; Spectrophotometry, Ultraviolet - methods ; Sulfones - chemistry</subject><ispartof>ASAIO journal (1992), 2019-07, Vol.65 (5), p.522-526</ispartof><rights>Copyright © 2019 by the American Society for Artificial Internal Organs</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c4229-6b86076cf1c2e06e31e6911657ae5bb690d8e3b361ee2840b996b64ebb5f8f083</citedby><cites>FETCH-LOGICAL-c4229-6b86076cf1c2e06e31e6911657ae5bb690d8e3b361ee2840b996b64ebb5f8f083</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,776,780,27903,27904</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/29863630$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Shimamoto, Yoshimasa</creatorcontrib><creatorcontrib>Matsushita, Seigo</creatorcontrib><creatorcontrib>Yamamoto, Takatoshi</creatorcontrib><creatorcontrib>Nakamura, Yuki</creatorcontrib><creatorcontrib>Kamada, Aki</creatorcontrib><creatorcontrib>Miyata, Masahiro</creatorcontrib><creatorcontrib>Umimoto, Koichi</creatorcontrib><title>Real-Time Measurement of Polyvinylpyrrolidone Eluted from Polysulfone Membrane Dialyzers Based on the Ultraviolet Spectrum</title><title>ASAIO journal (1992)</title><addtitle>ASAIO J</addtitle><description>The objective of this study was to develop a new method for measuring polyvinylpyrrolidone (PVP) eluted from polysulfone (PSu) membrane dialyzers. The Müller method is generally used for the measurement of PVP, in which the PVP concentration is determined by measuring the absorbance after a red color is generated by the formation of PVP-iodine complexes when iodine is added to a sample. In contrast, our method does not require any reagents and allows real-time measurement of PVP by the ultraviolet absorption spectrum (UV-s method). In this study, the UV-s method and the Müller method were used to measure PVP eluted from two types of PSu membrane dialyzer (PS-1.6UW sterilized by autoclaving [n = 10] and APS-15SA sterilized by gamma radiation [n = 10]). Polyvinylpyrrolidone concentrations measured by the two methods showed a significant positive correlation (rs = 0.99, p = 0.0006). The PVP concentration (median [25th–75th percentiles]) in PS-1.6UW dialyzer washings obtained by rinsing with physiologic saline was 2.0 (1.18–4.85) mg/L with the Müller method and 3.35 (2.38–4.23) mg/L with the UV-s method, showing no significant difference. However, the PVP concentration in APS-15SA dialyzer washings was 0 (0–0.35) mg/L by the Müller method and 0.95 (0.45–2.58) mg/L by the UV-s method, and there was a significant difference between the two methods. In conclusion, the low concentration of PVP eluted from a PSu dialyzer sterilized by gamma radiation was hardly detected by the Müller method but could be clearly detected by the new UV-s method. These findings suggest that the UV-s method could be used for real-time measurement of PVP eluted from PSu membrane dialyzers.</description><subject>Humans</subject><subject>Kidneys, Artificial</subject><subject>Membranes, Artificial</subject><subject>Polymers - chemistry</subject><subject>Povidone - analysis</subject><subject>Renal Dialysis - methods</subject><subject>Spectrophotometry, Ultraviolet - methods</subject><subject>Sulfones - chemistry</subject><issn>1058-2916</issn><issn>1538-943X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2019</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp9kE1PAyEQhonRWL_-gTEcvWyFZZfCsX6btNFoTbxtYDubrrKlAluz_fWiVWM8yIXJ8Mw75EHokJI-JXJwMh5O-uT3EancQDs0ZyKRGXvajDXJRZJKynto1_tnQuIjo9uol0rBGWdkB63uQZlkUjeAx6B866CBecC2wnfWdMt63plF55w19dTOAV-YNsAUV842n4BvTfXRH0OjnYrFea1MtwLn8anykbRzHGaAH01wallbAwE_LKAMrm320ValjIeDr3sPPV5eTM6uk9Ht1c3ZcJSUWZrKhGvByYCXFS1TIBwYBS4p5flAQa41l2QqgGnGKUAqMqKl5JpnoHVeiYoItoeO17kLZ19b8KFoal-CMfG_tvVFSjIpWUYkj2i2RktnvXdQFQtXN8p1BSXFh_UiWi_-Wo9jR18bWt3A9GfoW3MExBp4syZEOS-mfQNXzKL7MPs_-x2wdpB0</recordid><startdate>201907</startdate><enddate>201907</enddate><creator>Shimamoto, Yoshimasa</creator><creator>Matsushita, Seigo</creator><creator>Yamamoto, Takatoshi</creator><creator>Nakamura, Yuki</creator><creator>Kamada, Aki</creator><creator>Miyata, Masahiro</creator><creator>Umimoto, Koichi</creator><general>Copyright by the American Society for Artificial Internal Organs</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>201907</creationdate><title>Real-Time Measurement of Polyvinylpyrrolidone Eluted from Polysulfone Membrane Dialyzers Based on the Ultraviolet Spectrum</title><author>Shimamoto, Yoshimasa ; Matsushita, Seigo ; Yamamoto, Takatoshi ; Nakamura, Yuki ; Kamada, Aki ; Miyata, Masahiro ; Umimoto, Koichi</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c4229-6b86076cf1c2e06e31e6911657ae5bb690d8e3b361ee2840b996b64ebb5f8f083</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2019</creationdate><topic>Humans</topic><topic>Kidneys, Artificial</topic><topic>Membranes, Artificial</topic><topic>Polymers - chemistry</topic><topic>Povidone - analysis</topic><topic>Renal Dialysis - methods</topic><topic>Spectrophotometry, Ultraviolet - methods</topic><topic>Sulfones - chemistry</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Shimamoto, Yoshimasa</creatorcontrib><creatorcontrib>Matsushita, Seigo</creatorcontrib><creatorcontrib>Yamamoto, Takatoshi</creatorcontrib><creatorcontrib>Nakamura, Yuki</creatorcontrib><creatorcontrib>Kamada, Aki</creatorcontrib><creatorcontrib>Miyata, Masahiro</creatorcontrib><creatorcontrib>Umimoto, Koichi</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>ASAIO journal (1992)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Shimamoto, Yoshimasa</au><au>Matsushita, Seigo</au><au>Yamamoto, Takatoshi</au><au>Nakamura, Yuki</au><au>Kamada, Aki</au><au>Miyata, Masahiro</au><au>Umimoto, Koichi</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Real-Time Measurement of Polyvinylpyrrolidone Eluted from Polysulfone Membrane Dialyzers Based on the Ultraviolet Spectrum</atitle><jtitle>ASAIO journal (1992)</jtitle><addtitle>ASAIO J</addtitle><date>2019-07</date><risdate>2019</risdate><volume>65</volume><issue>5</issue><spage>522</spage><epage>526</epage><pages>522-526</pages><issn>1058-2916</issn><eissn>1538-943X</eissn><abstract>The objective of this study was to develop a new method for measuring polyvinylpyrrolidone (PVP) eluted from polysulfone (PSu) membrane dialyzers. The Müller method is generally used for the measurement of PVP, in which the PVP concentration is determined by measuring the absorbance after a red color is generated by the formation of PVP-iodine complexes when iodine is added to a sample. In contrast, our method does not require any reagents and allows real-time measurement of PVP by the ultraviolet absorption spectrum (UV-s method). In this study, the UV-s method and the Müller method were used to measure PVP eluted from two types of PSu membrane dialyzer (PS-1.6UW sterilized by autoclaving [n = 10] and APS-15SA sterilized by gamma radiation [n = 10]). Polyvinylpyrrolidone concentrations measured by the two methods showed a significant positive correlation (rs = 0.99, p = 0.0006). The PVP concentration (median [25th–75th percentiles]) in PS-1.6UW dialyzer washings obtained by rinsing with physiologic saline was 2.0 (1.18–4.85) mg/L with the Müller method and 3.35 (2.38–4.23) mg/L with the UV-s method, showing no significant difference. However, the PVP concentration in APS-15SA dialyzer washings was 0 (0–0.35) mg/L by the Müller method and 0.95 (0.45–2.58) mg/L by the UV-s method, and there was a significant difference between the two methods. In conclusion, the low concentration of PVP eluted from a PSu dialyzer sterilized by gamma radiation was hardly detected by the Müller method but could be clearly detected by the new UV-s method. These findings suggest that the UV-s method could be used for real-time measurement of PVP eluted from PSu membrane dialyzers.</abstract><cop>United States</cop><pub>Copyright by the American Society for Artificial Internal Organs</pub><pmid>29863630</pmid><doi>10.1097/MAT.0000000000000829</doi><tpages>5</tpages></addata></record>
fulltext fulltext
identifier ISSN: 1058-2916
ispartof ASAIO journal (1992), 2019-07, Vol.65 (5), p.522-526
issn 1058-2916
1538-943X
language eng
recordid cdi_proquest_miscellaneous_2049934096
source MEDLINE; Journals@Ovid LWW Legacy Archive; Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals
subjects Humans
Kidneys, Artificial
Membranes, Artificial
Polymers - chemistry
Povidone - analysis
Renal Dialysis - methods
Spectrophotometry, Ultraviolet - methods
Sulfones - chemistry
title Real-Time Measurement of Polyvinylpyrrolidone Eluted from Polysulfone Membrane Dialyzers Based on the Ultraviolet Spectrum
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-23T03%3A48%3A24IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Real-Time%20Measurement%20of%20Polyvinylpyrrolidone%20Eluted%20from%20Polysulfone%20Membrane%20Dialyzers%20Based%20on%20the%20Ultraviolet%20Spectrum&rft.jtitle=ASAIO%20journal%20(1992)&rft.au=Shimamoto,%20Yoshimasa&rft.date=2019-07&rft.volume=65&rft.issue=5&rft.spage=522&rft.epage=526&rft.pages=522-526&rft.issn=1058-2916&rft.eissn=1538-943X&rft_id=info:doi/10.1097/MAT.0000000000000829&rft_dat=%3Cproquest_cross%3E2049934096%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2049934096&rft_id=info:pmid/29863630&rfr_iscdi=true