Noninvasive validation of central and peripheral augmentation index estimated by a novel wrist-worn tonometer

BACKGROUND:The validation of new blood pressure devices with easier use and more rapid data collection may facilitate the incorporation of these measures into clinical practice. We analyze the reliability and validity of Pulse Wave Analysis as reported by a novel wrist-worn device, easy and quick to...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of hypertension 2018-11, Vol.36 (11), p.2204-2214
Hauptverfasser: García-Ortiz, Luis, Recio-Rodríguez, José I, Agudo-Conde, Cristina, Maderuelo-Fernandez, Jose A, Patino-Alonso, Maria C, de Cabo-Laso, Ángela, Rodriguez-Martín, Carmela, Gonzalez-Sanchez, Jesus, Rodriguez-Sanchez, Emiliano, Gómez-Marcos, Manuel A
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 2214
container_issue 11
container_start_page 2204
container_title Journal of hypertension
container_volume 36
creator García-Ortiz, Luis
Recio-Rodríguez, José I
Agudo-Conde, Cristina
Maderuelo-Fernandez, Jose A
Patino-Alonso, Maria C
de Cabo-Laso, Ángela
Rodriguez-Martín, Carmela
Gonzalez-Sanchez, Jesus
Rodriguez-Sanchez, Emiliano
Gómez-Marcos, Manuel A
description BACKGROUND:The validation of new blood pressure devices with easier use and more rapid data collection may facilitate the incorporation of these measures into clinical practice. We analyze the reliability and validity of Pulse Wave Analysis as reported by a novel wrist-worn device, easy and quick to use, comparing central and peripheral augmentation index (AIx) with the same measures from the SphygmoCor. Additionally, we analyzed clinical relevance through the association of the Pulse Wave Analysis, as reported by a novel wrist-worn device, with other cardiovascular parameters. METHODS:Cross-sectional study including 254 participants. Mean age was 51.9 ± 13.4 years and 53% were women. Measurementsperipheral AIx (PAIx) and central AIx (CAIx) by the wrist-worn device and SphygmoCor (MM3); carotid−femoral (cf) pulse wave velocity (PWV) by SphygmoCor (MM3); cardioankle-vascular index (CAVI), ankle−brachial index (ABI) and brachial−ankle (ba) PWV by the Vasera device and carotid−intima media thickness (IMT) by ultrasonography. RESULTS:Intra-observer intra-class correlation coefficient (ICC) for PAIx was 0.886 (95% CI 0.803−0.934) and for CAIx 0.943 (95% CI 0.901−0.968) with Bland Altman limits of agreement −0.75 (−23.8 to 21.8) and 0.08 (−15.7 to 15.9), respectively. Inter-observer ICC for PAIx was 0.952 (95% CI 0.915−0.972) and CAIx 0.893 (95% CI 0.811−0.939) with limits of agreement −0.45 (−13.7 to 12.8) and 0.43 (−17.7 to 18.5), respectively. Comparing the wrist-worn device with SphygmoCor, the ICC was 0.849 (95% CI 0.798−0.887) for PAIx, and 0.783 (95% CI 0.711−0.838) for CAIx. In Bland−Altman, limits of agreement for PAIx 1.03 (−20.67 to 22.73), and for CAIx −2.14 (−24.79 to 20.50). PAIx and CAIx, from the wrist-worn device, correlated with age, CAVI, ABI, baPWV, cfPWV, IMT, glomerular filtration and cardiovascular risk. CONCLUSION:AIx measurements by wrist-worn device shows a good intra-observer and inter-observer reliability, inter-device noninvasive reliability and validity when compared with SphygmoCor, and clinical relevance by association with measures of vascular structure and function, end-organ damage and cardiovascular risk.Trial registrationClinicaltrials.gov NCT02623894; https://clinicaltrials.gov/ct2/show/NCT02623894.
doi_str_mv 10.1097/HJH.0000000000001806
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2047282980</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2047282980</sourcerecordid><originalsourceid>FETCH-LOGICAL-c3566-dde1ee295376fdadec4be7beeac9cf1b4aaf2d8c96ea85eb7fd2d31607c363383</originalsourceid><addsrcrecordid>eNp9kMtOxCAUhonR6Hh5A2NYuqlSaCldGqOOxuhG1w2FUwelMEI747y96KgxLmRDCN-5_B9Chzk5yUldnU5vpifk18kF4RtokhcVy8qyFptoQihnGWcl3UG7MT4nSNQV20Y7tBYFZ1RMUH_nnXELGc0C8EJao-VgvMO-wwrcEKTF0mk8h2DmM_h8jk99-lljxml4wxAH08sBNG5XWGLnF2DxMpg4ZEsfHB688z0MEPbRVidthIOvew89Xl48nE-z2_ur6_Oz20yxkvNMa8gBaF2yindaalBFC1ULIFWturwtpOyoFqrmIEUJbdVpqlnOSaUYZ0ywPXS87jsP_nVM6zW9iQqslQ78GBtKioqKZIEktFijKvgYA3TNPKQwYdXkpPkQ3STRzV_Rqezoa8LY9qB_ir7NJkCsgaW3KXl8seMSQjMDaYfZ_73fAULnjb0</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2047282980</pqid></control><display><type>article</type><title>Noninvasive validation of central and peripheral augmentation index estimated by a novel wrist-worn tonometer</title><source>MEDLINE</source><source>Journals@Ovid Complete</source><creator>García-Ortiz, Luis ; Recio-Rodríguez, José I ; Agudo-Conde, Cristina ; Maderuelo-Fernandez, Jose A ; Patino-Alonso, Maria C ; de Cabo-Laso, Ángela ; Rodriguez-Martín, Carmela ; Gonzalez-Sanchez, Jesus ; Rodriguez-Sanchez, Emiliano ; Gómez-Marcos, Manuel A</creator><creatorcontrib>García-Ortiz, Luis ; Recio-Rodríguez, José I ; Agudo-Conde, Cristina ; Maderuelo-Fernandez, Jose A ; Patino-Alonso, Maria C ; de Cabo-Laso, Ángela ; Rodriguez-Martín, Carmela ; Gonzalez-Sanchez, Jesus ; Rodriguez-Sanchez, Emiliano ; Gómez-Marcos, Manuel A ; on behalf the EVA Investigators group</creatorcontrib><description>BACKGROUND:The validation of new blood pressure devices with easier use and more rapid data collection may facilitate the incorporation of these measures into clinical practice. We analyze the reliability and validity of Pulse Wave Analysis as reported by a novel wrist-worn device, easy and quick to use, comparing central and peripheral augmentation index (AIx) with the same measures from the SphygmoCor. Additionally, we analyzed clinical relevance through the association of the Pulse Wave Analysis, as reported by a novel wrist-worn device, with other cardiovascular parameters. METHODS:Cross-sectional study including 254 participants. Mean age was 51.9 ± 13.4 years and 53% were women. Measurementsperipheral AIx (PAIx) and central AIx (CAIx) by the wrist-worn device and SphygmoCor (MM3); carotid−femoral (cf) pulse wave velocity (PWV) by SphygmoCor (MM3); cardioankle-vascular index (CAVI), ankle−brachial index (ABI) and brachial−ankle (ba) PWV by the Vasera device and carotid−intima media thickness (IMT) by ultrasonography. RESULTS:Intra-observer intra-class correlation coefficient (ICC) for PAIx was 0.886 (95% CI 0.803−0.934) and for CAIx 0.943 (95% CI 0.901−0.968) with Bland Altman limits of agreement −0.75 (−23.8 to 21.8) and 0.08 (−15.7 to 15.9), respectively. Inter-observer ICC for PAIx was 0.952 (95% CI 0.915−0.972) and CAIx 0.893 (95% CI 0.811−0.939) with limits of agreement −0.45 (−13.7 to 12.8) and 0.43 (−17.7 to 18.5), respectively. Comparing the wrist-worn device with SphygmoCor, the ICC was 0.849 (95% CI 0.798−0.887) for PAIx, and 0.783 (95% CI 0.711−0.838) for CAIx. In Bland−Altman, limits of agreement for PAIx 1.03 (−20.67 to 22.73), and for CAIx −2.14 (−24.79 to 20.50). PAIx and CAIx, from the wrist-worn device, correlated with age, CAVI, ABI, baPWV, cfPWV, IMT, glomerular filtration and cardiovascular risk. CONCLUSION:AIx measurements by wrist-worn device shows a good intra-observer and inter-observer reliability, inter-device noninvasive reliability and validity when compared with SphygmoCor, and clinical relevance by association with measures of vascular structure and function, end-organ damage and cardiovascular risk.Trial registrationClinicaltrials.gov NCT02623894; https://clinicaltrials.gov/ct2/show/NCT02623894.</description><identifier>ISSN: 0263-6352</identifier><identifier>EISSN: 1473-5598</identifier><identifier>DOI: 10.1097/HJH.0000000000001806</identifier><identifier>PMID: 29846328</identifier><language>eng</language><publisher>England: Copyright Wolters Kluwer Health, Inc. All rights reserved</publisher><subject>Adult ; Aged ; Ankle Brachial Index ; Blood Pressure ; Carotid Intima-Media Thickness ; Cross-Sectional Studies ; Female ; Humans ; Male ; Manometry - instrumentation ; Middle Aged ; Observer Variation ; Pulse Wave Analysis - instrumentation ; Reproducibility of Results ; Vascular Stiffness ; Wearable Electronic Devices ; Wrist</subject><ispartof>Journal of hypertension, 2018-11, Vol.36 (11), p.2204-2214</ispartof><rights>Copyright © 2018 Wolters Kluwer Health, Inc. All rights reserved.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c3566-dde1ee295376fdadec4be7beeac9cf1b4aaf2d8c96ea85eb7fd2d31607c363383</citedby><cites>FETCH-LOGICAL-c3566-dde1ee295376fdadec4be7beeac9cf1b4aaf2d8c96ea85eb7fd2d31607c363383</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27924,27925</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/29846328$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>García-Ortiz, Luis</creatorcontrib><creatorcontrib>Recio-Rodríguez, José I</creatorcontrib><creatorcontrib>Agudo-Conde, Cristina</creatorcontrib><creatorcontrib>Maderuelo-Fernandez, Jose A</creatorcontrib><creatorcontrib>Patino-Alonso, Maria C</creatorcontrib><creatorcontrib>de Cabo-Laso, Ángela</creatorcontrib><creatorcontrib>Rodriguez-Martín, Carmela</creatorcontrib><creatorcontrib>Gonzalez-Sanchez, Jesus</creatorcontrib><creatorcontrib>Rodriguez-Sanchez, Emiliano</creatorcontrib><creatorcontrib>Gómez-Marcos, Manuel A</creatorcontrib><creatorcontrib>on behalf the EVA Investigators group</creatorcontrib><title>Noninvasive validation of central and peripheral augmentation index estimated by a novel wrist-worn tonometer</title><title>Journal of hypertension</title><addtitle>J Hypertens</addtitle><description>BACKGROUND:The validation of new blood pressure devices with easier use and more rapid data collection may facilitate the incorporation of these measures into clinical practice. We analyze the reliability and validity of Pulse Wave Analysis as reported by a novel wrist-worn device, easy and quick to use, comparing central and peripheral augmentation index (AIx) with the same measures from the SphygmoCor. Additionally, we analyzed clinical relevance through the association of the Pulse Wave Analysis, as reported by a novel wrist-worn device, with other cardiovascular parameters. METHODS:Cross-sectional study including 254 participants. Mean age was 51.9 ± 13.4 years and 53% were women. Measurementsperipheral AIx (PAIx) and central AIx (CAIx) by the wrist-worn device and SphygmoCor (MM3); carotid−femoral (cf) pulse wave velocity (PWV) by SphygmoCor (MM3); cardioankle-vascular index (CAVI), ankle−brachial index (ABI) and brachial−ankle (ba) PWV by the Vasera device and carotid−intima media thickness (IMT) by ultrasonography. RESULTS:Intra-observer intra-class correlation coefficient (ICC) for PAIx was 0.886 (95% CI 0.803−0.934) and for CAIx 0.943 (95% CI 0.901−0.968) with Bland Altman limits of agreement −0.75 (−23.8 to 21.8) and 0.08 (−15.7 to 15.9), respectively. Inter-observer ICC for PAIx was 0.952 (95% CI 0.915−0.972) and CAIx 0.893 (95% CI 0.811−0.939) with limits of agreement −0.45 (−13.7 to 12.8) and 0.43 (−17.7 to 18.5), respectively. Comparing the wrist-worn device with SphygmoCor, the ICC was 0.849 (95% CI 0.798−0.887) for PAIx, and 0.783 (95% CI 0.711−0.838) for CAIx. In Bland−Altman, limits of agreement for PAIx 1.03 (−20.67 to 22.73), and for CAIx −2.14 (−24.79 to 20.50). PAIx and CAIx, from the wrist-worn device, correlated with age, CAVI, ABI, baPWV, cfPWV, IMT, glomerular filtration and cardiovascular risk. CONCLUSION:AIx measurements by wrist-worn device shows a good intra-observer and inter-observer reliability, inter-device noninvasive reliability and validity when compared with SphygmoCor, and clinical relevance by association with measures of vascular structure and function, end-organ damage and cardiovascular risk.Trial registrationClinicaltrials.gov NCT02623894; https://clinicaltrials.gov/ct2/show/NCT02623894.</description><subject>Adult</subject><subject>Aged</subject><subject>Ankle Brachial Index</subject><subject>Blood Pressure</subject><subject>Carotid Intima-Media Thickness</subject><subject>Cross-Sectional Studies</subject><subject>Female</subject><subject>Humans</subject><subject>Male</subject><subject>Manometry - instrumentation</subject><subject>Middle Aged</subject><subject>Observer Variation</subject><subject>Pulse Wave Analysis - instrumentation</subject><subject>Reproducibility of Results</subject><subject>Vascular Stiffness</subject><subject>Wearable Electronic Devices</subject><subject>Wrist</subject><issn>0263-6352</issn><issn>1473-5598</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2018</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp9kMtOxCAUhonR6Hh5A2NYuqlSaCldGqOOxuhG1w2FUwelMEI747y96KgxLmRDCN-5_B9Chzk5yUldnU5vpifk18kF4RtokhcVy8qyFptoQihnGWcl3UG7MT4nSNQV20Y7tBYFZ1RMUH_nnXELGc0C8EJao-VgvMO-wwrcEKTF0mk8h2DmM_h8jk99-lljxml4wxAH08sBNG5XWGLnF2DxMpg4ZEsfHB688z0MEPbRVidthIOvew89Xl48nE-z2_ur6_Oz20yxkvNMa8gBaF2yindaalBFC1ULIFWturwtpOyoFqrmIEUJbdVpqlnOSaUYZ0ywPXS87jsP_nVM6zW9iQqslQ78GBtKioqKZIEktFijKvgYA3TNPKQwYdXkpPkQ3STRzV_Rqezoa8LY9qB_ir7NJkCsgaW3KXl8seMSQjMDaYfZ_73fAULnjb0</recordid><startdate>201811</startdate><enddate>201811</enddate><creator>García-Ortiz, Luis</creator><creator>Recio-Rodríguez, José I</creator><creator>Agudo-Conde, Cristina</creator><creator>Maderuelo-Fernandez, Jose A</creator><creator>Patino-Alonso, Maria C</creator><creator>de Cabo-Laso, Ángela</creator><creator>Rodriguez-Martín, Carmela</creator><creator>Gonzalez-Sanchez, Jesus</creator><creator>Rodriguez-Sanchez, Emiliano</creator><creator>Gómez-Marcos, Manuel A</creator><general>Copyright Wolters Kluwer Health, Inc. All rights reserved</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>201811</creationdate><title>Noninvasive validation of central and peripheral augmentation index estimated by a novel wrist-worn tonometer</title><author>García-Ortiz, Luis ; Recio-Rodríguez, José I ; Agudo-Conde, Cristina ; Maderuelo-Fernandez, Jose A ; Patino-Alonso, Maria C ; de Cabo-Laso, Ángela ; Rodriguez-Martín, Carmela ; Gonzalez-Sanchez, Jesus ; Rodriguez-Sanchez, Emiliano ; Gómez-Marcos, Manuel A</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c3566-dde1ee295376fdadec4be7beeac9cf1b4aaf2d8c96ea85eb7fd2d31607c363383</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2018</creationdate><topic>Adult</topic><topic>Aged</topic><topic>Ankle Brachial Index</topic><topic>Blood Pressure</topic><topic>Carotid Intima-Media Thickness</topic><topic>Cross-Sectional Studies</topic><topic>Female</topic><topic>Humans</topic><topic>Male</topic><topic>Manometry - instrumentation</topic><topic>Middle Aged</topic><topic>Observer Variation</topic><topic>Pulse Wave Analysis - instrumentation</topic><topic>Reproducibility of Results</topic><topic>Vascular Stiffness</topic><topic>Wearable Electronic Devices</topic><topic>Wrist</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>García-Ortiz, Luis</creatorcontrib><creatorcontrib>Recio-Rodríguez, José I</creatorcontrib><creatorcontrib>Agudo-Conde, Cristina</creatorcontrib><creatorcontrib>Maderuelo-Fernandez, Jose A</creatorcontrib><creatorcontrib>Patino-Alonso, Maria C</creatorcontrib><creatorcontrib>de Cabo-Laso, Ángela</creatorcontrib><creatorcontrib>Rodriguez-Martín, Carmela</creatorcontrib><creatorcontrib>Gonzalez-Sanchez, Jesus</creatorcontrib><creatorcontrib>Rodriguez-Sanchez, Emiliano</creatorcontrib><creatorcontrib>Gómez-Marcos, Manuel A</creatorcontrib><creatorcontrib>on behalf the EVA Investigators group</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Journal of hypertension</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>García-Ortiz, Luis</au><au>Recio-Rodríguez, José I</au><au>Agudo-Conde, Cristina</au><au>Maderuelo-Fernandez, Jose A</au><au>Patino-Alonso, Maria C</au><au>de Cabo-Laso, Ángela</au><au>Rodriguez-Martín, Carmela</au><au>Gonzalez-Sanchez, Jesus</au><au>Rodriguez-Sanchez, Emiliano</au><au>Gómez-Marcos, Manuel A</au><aucorp>on behalf the EVA Investigators group</aucorp><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Noninvasive validation of central and peripheral augmentation index estimated by a novel wrist-worn tonometer</atitle><jtitle>Journal of hypertension</jtitle><addtitle>J Hypertens</addtitle><date>2018-11</date><risdate>2018</risdate><volume>36</volume><issue>11</issue><spage>2204</spage><epage>2214</epage><pages>2204-2214</pages><issn>0263-6352</issn><eissn>1473-5598</eissn><abstract>BACKGROUND:The validation of new blood pressure devices with easier use and more rapid data collection may facilitate the incorporation of these measures into clinical practice. We analyze the reliability and validity of Pulse Wave Analysis as reported by a novel wrist-worn device, easy and quick to use, comparing central and peripheral augmentation index (AIx) with the same measures from the SphygmoCor. Additionally, we analyzed clinical relevance through the association of the Pulse Wave Analysis, as reported by a novel wrist-worn device, with other cardiovascular parameters. METHODS:Cross-sectional study including 254 participants. Mean age was 51.9 ± 13.4 years and 53% were women. Measurementsperipheral AIx (PAIx) and central AIx (CAIx) by the wrist-worn device and SphygmoCor (MM3); carotid−femoral (cf) pulse wave velocity (PWV) by SphygmoCor (MM3); cardioankle-vascular index (CAVI), ankle−brachial index (ABI) and brachial−ankle (ba) PWV by the Vasera device and carotid−intima media thickness (IMT) by ultrasonography. RESULTS:Intra-observer intra-class correlation coefficient (ICC) for PAIx was 0.886 (95% CI 0.803−0.934) and for CAIx 0.943 (95% CI 0.901−0.968) with Bland Altman limits of agreement −0.75 (−23.8 to 21.8) and 0.08 (−15.7 to 15.9), respectively. Inter-observer ICC for PAIx was 0.952 (95% CI 0.915−0.972) and CAIx 0.893 (95% CI 0.811−0.939) with limits of agreement −0.45 (−13.7 to 12.8) and 0.43 (−17.7 to 18.5), respectively. Comparing the wrist-worn device with SphygmoCor, the ICC was 0.849 (95% CI 0.798−0.887) for PAIx, and 0.783 (95% CI 0.711−0.838) for CAIx. In Bland−Altman, limits of agreement for PAIx 1.03 (−20.67 to 22.73), and for CAIx −2.14 (−24.79 to 20.50). PAIx and CAIx, from the wrist-worn device, correlated with age, CAVI, ABI, baPWV, cfPWV, IMT, glomerular filtration and cardiovascular risk. CONCLUSION:AIx measurements by wrist-worn device shows a good intra-observer and inter-observer reliability, inter-device noninvasive reliability and validity when compared with SphygmoCor, and clinical relevance by association with measures of vascular structure and function, end-organ damage and cardiovascular risk.Trial registrationClinicaltrials.gov NCT02623894; https://clinicaltrials.gov/ct2/show/NCT02623894.</abstract><cop>England</cop><pub>Copyright Wolters Kluwer Health, Inc. All rights reserved</pub><pmid>29846328</pmid><doi>10.1097/HJH.0000000000001806</doi><tpages>11</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0263-6352
ispartof Journal of hypertension, 2018-11, Vol.36 (11), p.2204-2214
issn 0263-6352
1473-5598
language eng
recordid cdi_proquest_miscellaneous_2047282980
source MEDLINE; Journals@Ovid Complete
subjects Adult
Aged
Ankle Brachial Index
Blood Pressure
Carotid Intima-Media Thickness
Cross-Sectional Studies
Female
Humans
Male
Manometry - instrumentation
Middle Aged
Observer Variation
Pulse Wave Analysis - instrumentation
Reproducibility of Results
Vascular Stiffness
Wearable Electronic Devices
Wrist
title Noninvasive validation of central and peripheral augmentation index estimated by a novel wrist-worn tonometer
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-06T10%3A43%3A33IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Noninvasive%20validation%20of%20central%20and%20peripheral%20augmentation%20index%20estimated%20by%20a%20novel%20wrist-worn%20tonometer&rft.jtitle=Journal%20of%20hypertension&rft.au=Garc%C3%ADa-Ortiz,%20Luis&rft.aucorp=on%20behalf%20the%20EVA%20Investigators%20group&rft.date=2018-11&rft.volume=36&rft.issue=11&rft.spage=2204&rft.epage=2214&rft.pages=2204-2214&rft.issn=0263-6352&rft.eissn=1473-5598&rft_id=info:doi/10.1097/HJH.0000000000001806&rft_dat=%3Cproquest_cross%3E2047282980%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2047282980&rft_id=info:pmid/29846328&rfr_iscdi=true