Transient under‐sensing of the ventricular lead during abdominal ultrasound as cause of ventricular fibrillation

Pacemaker‐induced arrhythmias represent a very rare complication. Algorithm‐induced ventricular tachycardias have been described but this report is the first to describe a ventricular fibrillation caused by transient undersensing of the ventricular lead during an abdominal ultrasound.

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 2018-08, Vol.41 (8), p.1034-1035
Hauptverfasser: Frontera, Antonio, Klotz, Nicolas, Martin, Ruairidh, Haïssaguerre, Michel, Ritter, Philippe, Bordachar, Pierre
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Pacemaker‐induced arrhythmias represent a very rare complication. Algorithm‐induced ventricular tachycardias have been described but this report is the first to describe a ventricular fibrillation caused by transient undersensing of the ventricular lead during an abdominal ultrasound.
ISSN:0147-8389
1540-8159
DOI:10.1111/pace.13362