Amperometric enzyme electrode for free cholesterol determination prepared with cholesterol oxidase immobilized in poly(vinylferrocenium) film

The steady state amperometric measurements of free cholesterol were performed by an enzyme electrode which was developed through electrostatic immobilization of the cholesterol oxidase in poly(vinylferrocenium) perchlorate, PVF +ClO 4 −, film that was coated on a Pt electrode. Amperometric measureme...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Enzyme and microbial technology 2007-01, Vol.40 (2), p.262-265
Hauptverfasser: ÖZER, B. Cem, ÖZYÖRÜK, Haluk, CELEBI, Serdar S, YILDIZ, Atilla
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The steady state amperometric measurements of free cholesterol were performed by an enzyme electrode which was developed through electrostatic immobilization of the cholesterol oxidase in poly(vinylferrocenium) perchlorate, PVF +ClO 4 −, film that was coated on a Pt electrode. Amperometric measurements were performed at an applied potential of 0.7 V versus SCE. The effects of enzyme concentration, polymer film thickness, working temperature, dissolved oxygen gas and pH of solution on the response electrode were elucidated. The activation energy and the apparent Michaelis–Menten constant were calculated for enzyme electrode system. The linear working range, the sensitivity and the stability of the biosensor were also determined. The mechanism for the operation of the enzyme electrode was proposed.
ISSN:0141-0229
1879-0909
DOI:10.1016/j.enzmictec.2006.04.007