Technical Note: An efficient daily QA procedure for proton pencil beam scanning

Purpose The aim of this work was to develop an efficient daily quality assurance (QA) program with strict tolerance levels for pencil beam scanning (PBS) proton radiotherapy featuring simultaneous dosimetric testing on a single, nonuniform field. Methods A nonuniform field measuring beam output, pro...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical physics (Lancaster) 2018-03, Vol.45 (3), p.1040-1049
Hauptverfasser: Younkin, James E., Shen, Jiajian, Bues, Martin, Robertson, Daniel G., Mundy, Daniel W., Clouser, Edward, Liu, Wei, Ding, Xiaoning, Stoker, Joshua B.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 1049
container_issue 3
container_start_page 1040
container_title Medical physics (Lancaster)
container_volume 45
creator Younkin, James E.
Shen, Jiajian
Bues, Martin
Robertson, Daniel G.
Mundy, Daniel W.
Clouser, Edward
Liu, Wei
Ding, Xiaoning
Stoker, Joshua B.
description Purpose The aim of this work was to develop an efficient daily quality assurance (QA) program with strict tolerance levels for pencil beam scanning (PBS) proton radiotherapy featuring simultaneous dosimetric testing on a single, nonuniform field. Methods A nonuniform field measuring beam output, proton range, and spot position was designed for delivery onto a Sun Nuclear Daily‐QA 3 device. A custom acrylic block permitted simultaneous measurement of low‐ and high‐energy proton ranges in addition to beam output. Sensitivities to output, range, and spot position were evaluated to quantitate the device's response. Reproducibility tests were used to identify and control sources of measurement error as well as to assess the QA procedure's robustness. This procedure was implemented in each of our four treatment rooms independently; 4–6 months of daily QA measurements were collected. Results The 1% output, 0.5 mm range, and 1.5 mm spot position tolerances derived from preliminary tests were tighter overall than tolerances found in the literature and equivalent to the limits used for proton system commissioning. The simplicity and automation of the procedure reduced the time required for daily QA to 10 min per treatment room, and competition for beam between multiple treatment rooms was minimized. Conclusions An efficient daily PBS QA procedure can be performed using a single, nonuniform field on a nondedicated QA device. A thorough quantitation of the device's response and careful control of measurement uncertainties allowed daily tolerances to match commissioning standards.
doi_str_mv 10.1002/mp.12787
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_1993990576</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>1993990576</sourcerecordid><originalsourceid>FETCH-LOGICAL-c3847-3eb4e42e04e522d9ec8e5f44e67950feb300cf60e705589abea504bc14562ccb3</originalsourceid><addsrcrecordid>eNp1kF1LwzAUhoMobk7BXyC59KbzNE3axbshfsF0CvM6pOmJRtq0Ni2yf2_npl55dTjw8LznvIScxjCNAdhF1Uxjls2yPTJmPEsizkDukzGA5BHjIEbkKIR3AEgTAYdkxGQiOefZmCxXaN68M7qkj3WHl3TuKVrrjEPf0UK7ck2f57Rpa4NF3yK1dbvZutrTBr1xJc1RVzQY7b3zr8fkwOoy4MluTsjLzfXq6i5aLG_vr-aLyCQznkUJ5hw5Q-AoGCskmhkKyzmmmRRgMU8AjE0BMxBiJvWQIYDnJuYiZcbkyYScb73DLR89hk5VLhgsS-2x7oOK5fCiBJGlf6hp6xBatKppXaXbtYpBbepTVaO-6xvQs521zyssfsGfvgYg2gKfrsT1vyL18LQVfgH_TXfL</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1993990576</pqid></control><display><type>article</type><title>Technical Note: An efficient daily QA procedure for proton pencil beam scanning</title><source>Access via Wiley Online Library</source><source>Alma/SFX Local Collection</source><creator>Younkin, James E. ; Shen, Jiajian ; Bues, Martin ; Robertson, Daniel G. ; Mundy, Daniel W. ; Clouser, Edward ; Liu, Wei ; Ding, Xiaoning ; Stoker, Joshua B.</creator><creatorcontrib>Younkin, James E. ; Shen, Jiajian ; Bues, Martin ; Robertson, Daniel G. ; Mundy, Daniel W. ; Clouser, Edward ; Liu, Wei ; Ding, Xiaoning ; Stoker, Joshua B.</creatorcontrib><description>Purpose The aim of this work was to develop an efficient daily quality assurance (QA) program with strict tolerance levels for pencil beam scanning (PBS) proton radiotherapy featuring simultaneous dosimetric testing on a single, nonuniform field. Methods A nonuniform field measuring beam output, proton range, and spot position was designed for delivery onto a Sun Nuclear Daily‐QA 3 device. A custom acrylic block permitted simultaneous measurement of low‐ and high‐energy proton ranges in addition to beam output. Sensitivities to output, range, and spot position were evaluated to quantitate the device's response. Reproducibility tests were used to identify and control sources of measurement error as well as to assess the QA procedure's robustness. This procedure was implemented in each of our four treatment rooms independently; 4–6 months of daily QA measurements were collected. Results The 1% output, 0.5 mm range, and 1.5 mm spot position tolerances derived from preliminary tests were tighter overall than tolerances found in the literature and equivalent to the limits used for proton system commissioning. The simplicity and automation of the procedure reduced the time required for daily QA to 10 min per treatment room, and competition for beam between multiple treatment rooms was minimized. Conclusions An efficient daily PBS QA procedure can be performed using a single, nonuniform field on a nondedicated QA device. A thorough quantitation of the device's response and careful control of measurement uncertainties allowed daily tolerances to match commissioning standards.</description><identifier>ISSN: 0094-2405</identifier><identifier>EISSN: 2473-4209</identifier><identifier>DOI: 10.1002/mp.12787</identifier><identifier>PMID: 29394447</identifier><language>eng</language><publisher>United States</publisher><subject>daily QA ; PBS ; proton therapy ; quality assurance ; single field</subject><ispartof>Medical physics (Lancaster), 2018-03, Vol.45 (3), p.1040-1049</ispartof><rights>2018 American Association of Physicists in Medicine</rights><rights>2018 American Association of Physicists in Medicine.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c3847-3eb4e42e04e522d9ec8e5f44e67950feb300cf60e705589abea504bc14562ccb3</citedby><cites>FETCH-LOGICAL-c3847-3eb4e42e04e522d9ec8e5f44e67950feb300cf60e705589abea504bc14562ccb3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://onlinelibrary.wiley.com/doi/pdf/10.1002%2Fmp.12787$$EPDF$$P50$$Gwiley$$H</linktopdf><linktohtml>$$Uhttps://onlinelibrary.wiley.com/doi/full/10.1002%2Fmp.12787$$EHTML$$P50$$Gwiley$$H</linktohtml><link.rule.ids>314,780,784,1417,27924,27925,45574,45575</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/29394447$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Younkin, James E.</creatorcontrib><creatorcontrib>Shen, Jiajian</creatorcontrib><creatorcontrib>Bues, Martin</creatorcontrib><creatorcontrib>Robertson, Daniel G.</creatorcontrib><creatorcontrib>Mundy, Daniel W.</creatorcontrib><creatorcontrib>Clouser, Edward</creatorcontrib><creatorcontrib>Liu, Wei</creatorcontrib><creatorcontrib>Ding, Xiaoning</creatorcontrib><creatorcontrib>Stoker, Joshua B.</creatorcontrib><title>Technical Note: An efficient daily QA procedure for proton pencil beam scanning</title><title>Medical physics (Lancaster)</title><addtitle>Med Phys</addtitle><description>Purpose The aim of this work was to develop an efficient daily quality assurance (QA) program with strict tolerance levels for pencil beam scanning (PBS) proton radiotherapy featuring simultaneous dosimetric testing on a single, nonuniform field. Methods A nonuniform field measuring beam output, proton range, and spot position was designed for delivery onto a Sun Nuclear Daily‐QA 3 device. A custom acrylic block permitted simultaneous measurement of low‐ and high‐energy proton ranges in addition to beam output. Sensitivities to output, range, and spot position were evaluated to quantitate the device's response. Reproducibility tests were used to identify and control sources of measurement error as well as to assess the QA procedure's robustness. This procedure was implemented in each of our four treatment rooms independently; 4–6 months of daily QA measurements were collected. Results The 1% output, 0.5 mm range, and 1.5 mm spot position tolerances derived from preliminary tests were tighter overall than tolerances found in the literature and equivalent to the limits used for proton system commissioning. The simplicity and automation of the procedure reduced the time required for daily QA to 10 min per treatment room, and competition for beam between multiple treatment rooms was minimized. Conclusions An efficient daily PBS QA procedure can be performed using a single, nonuniform field on a nondedicated QA device. A thorough quantitation of the device's response and careful control of measurement uncertainties allowed daily tolerances to match commissioning standards.</description><subject>daily QA</subject><subject>PBS</subject><subject>proton therapy</subject><subject>quality assurance</subject><subject>single field</subject><issn>0094-2405</issn><issn>2473-4209</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2018</creationdate><recordtype>article</recordtype><recordid>eNp1kF1LwzAUhoMobk7BXyC59KbzNE3axbshfsF0CvM6pOmJRtq0Ni2yf2_npl55dTjw8LznvIScxjCNAdhF1Uxjls2yPTJmPEsizkDukzGA5BHjIEbkKIR3AEgTAYdkxGQiOefZmCxXaN68M7qkj3WHl3TuKVrrjEPf0UK7ck2f57Rpa4NF3yK1dbvZutrTBr1xJc1RVzQY7b3zr8fkwOoy4MluTsjLzfXq6i5aLG_vr-aLyCQznkUJ5hw5Q-AoGCskmhkKyzmmmRRgMU8AjE0BMxBiJvWQIYDnJuYiZcbkyYScb73DLR89hk5VLhgsS-2x7oOK5fCiBJGlf6hp6xBatKppXaXbtYpBbepTVaO-6xvQs521zyssfsGfvgYg2gKfrsT1vyL18LQVfgH_TXfL</recordid><startdate>201803</startdate><enddate>201803</enddate><creator>Younkin, James E.</creator><creator>Shen, Jiajian</creator><creator>Bues, Martin</creator><creator>Robertson, Daniel G.</creator><creator>Mundy, Daniel W.</creator><creator>Clouser, Edward</creator><creator>Liu, Wei</creator><creator>Ding, Xiaoning</creator><creator>Stoker, Joshua B.</creator><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>201803</creationdate><title>Technical Note: An efficient daily QA procedure for proton pencil beam scanning</title><author>Younkin, James E. ; Shen, Jiajian ; Bues, Martin ; Robertson, Daniel G. ; Mundy, Daniel W. ; Clouser, Edward ; Liu, Wei ; Ding, Xiaoning ; Stoker, Joshua B.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c3847-3eb4e42e04e522d9ec8e5f44e67950feb300cf60e705589abea504bc14562ccb3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2018</creationdate><topic>daily QA</topic><topic>PBS</topic><topic>proton therapy</topic><topic>quality assurance</topic><topic>single field</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Younkin, James E.</creatorcontrib><creatorcontrib>Shen, Jiajian</creatorcontrib><creatorcontrib>Bues, Martin</creatorcontrib><creatorcontrib>Robertson, Daniel G.</creatorcontrib><creatorcontrib>Mundy, Daniel W.</creatorcontrib><creatorcontrib>Clouser, Edward</creatorcontrib><creatorcontrib>Liu, Wei</creatorcontrib><creatorcontrib>Ding, Xiaoning</creatorcontrib><creatorcontrib>Stoker, Joshua B.</creatorcontrib><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Medical physics (Lancaster)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Younkin, James E.</au><au>Shen, Jiajian</au><au>Bues, Martin</au><au>Robertson, Daniel G.</au><au>Mundy, Daniel W.</au><au>Clouser, Edward</au><au>Liu, Wei</au><au>Ding, Xiaoning</au><au>Stoker, Joshua B.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Technical Note: An efficient daily QA procedure for proton pencil beam scanning</atitle><jtitle>Medical physics (Lancaster)</jtitle><addtitle>Med Phys</addtitle><date>2018-03</date><risdate>2018</risdate><volume>45</volume><issue>3</issue><spage>1040</spage><epage>1049</epage><pages>1040-1049</pages><issn>0094-2405</issn><eissn>2473-4209</eissn><abstract>Purpose The aim of this work was to develop an efficient daily quality assurance (QA) program with strict tolerance levels for pencil beam scanning (PBS) proton radiotherapy featuring simultaneous dosimetric testing on a single, nonuniform field. Methods A nonuniform field measuring beam output, proton range, and spot position was designed for delivery onto a Sun Nuclear Daily‐QA 3 device. A custom acrylic block permitted simultaneous measurement of low‐ and high‐energy proton ranges in addition to beam output. Sensitivities to output, range, and spot position were evaluated to quantitate the device's response. Reproducibility tests were used to identify and control sources of measurement error as well as to assess the QA procedure's robustness. This procedure was implemented in each of our four treatment rooms independently; 4–6 months of daily QA measurements were collected. Results The 1% output, 0.5 mm range, and 1.5 mm spot position tolerances derived from preliminary tests were tighter overall than tolerances found in the literature and equivalent to the limits used for proton system commissioning. The simplicity and automation of the procedure reduced the time required for daily QA to 10 min per treatment room, and competition for beam between multiple treatment rooms was minimized. Conclusions An efficient daily PBS QA procedure can be performed using a single, nonuniform field on a nondedicated QA device. A thorough quantitation of the device's response and careful control of measurement uncertainties allowed daily tolerances to match commissioning standards.</abstract><cop>United States</cop><pmid>29394447</pmid><doi>10.1002/mp.12787</doi><tpages>10</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0094-2405
ispartof Medical physics (Lancaster), 2018-03, Vol.45 (3), p.1040-1049
issn 0094-2405
2473-4209
language eng
recordid cdi_proquest_miscellaneous_1993990576
source Access via Wiley Online Library; Alma/SFX Local Collection
subjects daily QA
PBS
proton therapy
quality assurance
single field
title Technical Note: An efficient daily QA procedure for proton pencil beam scanning
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-25T04%3A14%3A16IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Technical%20Note:%20An%20efficient%20daily%20QA%20procedure%20for%20proton%20pencil%20beam%20scanning&rft.jtitle=Medical%20physics%20(Lancaster)&rft.au=Younkin,%20James%20E.&rft.date=2018-03&rft.volume=45&rft.issue=3&rft.spage=1040&rft.epage=1049&rft.pages=1040-1049&rft.issn=0094-2405&rft.eissn=2473-4209&rft_id=info:doi/10.1002/mp.12787&rft_dat=%3Cproquest_cross%3E1993990576%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=1993990576&rft_id=info:pmid/29394447&rfr_iscdi=true