Image-based MRI gradient estimation
In order to reduce geometric distortion phenomena in MR images, every MRI system main magnet undergoes a shimming process. Since this process aims at optimizing magnetic field homogeneity within a so-called uniformity sphere, image quality outside this sphere is neglected. Since the fields vary smoo...
Gespeichert in:
Veröffentlicht in: | Magnetic resonance imaging 2018-06, Vol.49, p.138-144 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In order to reduce geometric distortion phenomena in MR images, every MRI system main magnet undergoes a shimming process. Since this process aims at optimizing magnetic field homogeneity within a so-called uniformity sphere, image quality outside this sphere is neglected.
Since the fields vary smoothly in space, MR signal-to-noise ratio is still non-zero just outside the uniformity region, but correction of MR image distortion fails due to lack of magnetic field knowledge outside it.
We propose a novel algorithm for measuring all the fields involved in the generation of images. Our proposal is based on exploitation of the distortion which can be observed in images of a known phantom. The proposed method will enable measurement of the fields in a region that can be bigger than the uniformity sphere depending on the phantom dimensions. |
---|---|
ISSN: | 0730-725X 1873-5894 |
DOI: | 10.1016/j.mri.2017.12.028 |