Performance characteristics of the Beckman Coulter UniCel DxI 800 TSH (3rd IS) assay

Beckman Coulter recently reformulated their commercial TSH assay with primary calibration to the World Health Organization 3rd TSH international standard. An extensive evaluation of the performance characteristics for this assay was completed. Intra-day and inter-day precision was evaluated using 3...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinica chimica acta 2018-03, Vol.478, p.90-100
Hauptverfasser: Winston-McPherson, Gabrielle N., Samraj, Annie N., Poster, Kristina, Yamaguchi, Diane, Dickerson, Jane A., Drees, Julia C., Holmes, Daniel T., Greene, Dina N.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 100
container_issue
container_start_page 90
container_title Clinica chimica acta
container_volume 478
creator Winston-McPherson, Gabrielle N.
Samraj, Annie N.
Poster, Kristina
Yamaguchi, Diane
Dickerson, Jane A.
Drees, Julia C.
Holmes, Daniel T.
Greene, Dina N.
description Beckman Coulter recently reformulated their commercial TSH assay with primary calibration to the World Health Organization 3rd TSH international standard. An extensive evaluation of the performance characteristics for this assay was completed. Intra-day and inter-day precision was evaluated using 3 concentrations of commercial quality control material. Linearity, reportable range, stability, sensitivity and susceptibility to common inferences were determined using pooled patient specimens. Inter-assay variability was assessed across 5 different platforms (n=47 patient specimens). Intra-day and inter-day CVs were
doi_str_mv 10.1016/j.cca.2017.12.029
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_1980102720</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S0009898117305296</els_id><sourcerecordid>1980102720</sourcerecordid><originalsourceid>FETCH-LOGICAL-c419t-2aa1ea5fcd061ffa0b517a40597270d0fb62cfcc64f77bffd41c4103d769e6183</originalsourceid><addsrcrecordid>eNp9kMtOwzAQRS0EgvL4ADbIS1gkzDhpHIsVlFclJJDari3XGQuXtAE7RfD3GBVYsrJGPvdq5jB2jJAjYHW-yK01uQCUOYochNpiA6xlkRWlEttsAAAqq1WNe2w_xkUaS6hwl-0JJeQQy2LApk8UXBeWZmWJ22cTjO0p-Nh7G3nneP9M_IrsSwL4qFu36ZPPVn5ELb_-GPMagE8n9_y0CA0fT864idF8HrIdZ9pIRz_vAZvd3kxH99nD4914dPmQ2RJVnwljkMzQ2SZt5ZyB-RClKWGopJDQgJtXwjprq9JJOXeuKTEFoWhkpajCujhgp5ve19C9rSn2eumjpbY1K-rWUaOqAUFIAQnFDWpDF2Mgp1-DX5rwqRH0t0y90Emm_papUegkM2VOfurX8yU1f4lfewm42ACUjnz3FHS0npLJxgeyvW46_0_9FyDVgmM</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1980102720</pqid></control><display><type>article</type><title>Performance characteristics of the Beckman Coulter UniCel DxI 800 TSH (3rd IS) assay</title><source>MEDLINE</source><source>Elsevier ScienceDirect Journals</source><creator>Winston-McPherson, Gabrielle N. ; Samraj, Annie N. ; Poster, Kristina ; Yamaguchi, Diane ; Dickerson, Jane A. ; Drees, Julia C. ; Holmes, Daniel T. ; Greene, Dina N.</creator><creatorcontrib>Winston-McPherson, Gabrielle N. ; Samraj, Annie N. ; Poster, Kristina ; Yamaguchi, Diane ; Dickerson, Jane A. ; Drees, Julia C. ; Holmes, Daniel T. ; Greene, Dina N.</creatorcontrib><description>Beckman Coulter recently reformulated their commercial TSH assay with primary calibration to the World Health Organization 3rd TSH international standard. An extensive evaluation of the performance characteristics for this assay was completed. Intra-day and inter-day precision was evaluated using 3 concentrations of commercial quality control material. Linearity, reportable range, stability, sensitivity and susceptibility to common inferences were determined using pooled patient specimens. Inter-assay variability was assessed across 5 different platforms (n=47 patient specimens). Intra-day and inter-day CVs were &lt;10% at all concentrations evaluated. The LOQ, LOD and LOB were 0.0047μIU/ml (10% CV), 0.0012μIU/ml and 0.0005μIU/ml, respectively. Variable bias was observed for the TSH3 assay when evaluated against the previous generation assay and other platforms, but overall TSH3 gave comparable results. The TSH3 assay for UniCel DxI 800, is precise, highly sensitive and comparable to the previous generation assay. The assay is acceptable for clinical testing. •Beckman Coulter reformulated their commercial TSH assay (TSH3).•A comprehensive evaluation of TSH3 was completed.•TSH3 produced comparable results when evaluated against 5 other TSH assays.•TSH demonstrated acceptable intra-day and inter-day (&lt;10%) precision.•TSH3 performs with improved precision and sensitivity.</description><identifier>ISSN: 0009-8981</identifier><identifier>EISSN: 1873-3492</identifier><identifier>DOI: 10.1016/j.cca.2017.12.029</identifier><identifier>PMID: 29275143</identifier><language>eng</language><publisher>Netherlands: Elsevier B.V</publisher><subject>3rd IS ; Access ; Beckman Coulter ; Bias ; Calibration ; Clinical Laboratory Techniques - instrumentation ; Clinical Laboratory Techniques - standards ; DxI ; Harmonization ; Humans ; Immunoassay ; Quality Control ; Reproducibility of Results ; Thyrotropin - analysis ; TSH ; Validation</subject><ispartof>Clinica chimica acta, 2018-03, Vol.478, p.90-100</ispartof><rights>2018 Elsevier B.V.</rights><rights>Copyright © 2018 Elsevier B.V. All rights reserved.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c419t-2aa1ea5fcd061ffa0b517a40597270d0fb62cfcc64f77bffd41c4103d769e6183</citedby><cites>FETCH-LOGICAL-c419t-2aa1ea5fcd061ffa0b517a40597270d0fb62cfcc64f77bffd41c4103d769e6183</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://www.sciencedirect.com/science/article/pii/S0009898117305296$$EHTML$$P50$$Gelsevier$$H</linktohtml><link.rule.ids>314,776,780,3537,27901,27902,65306</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/29275143$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Winston-McPherson, Gabrielle N.</creatorcontrib><creatorcontrib>Samraj, Annie N.</creatorcontrib><creatorcontrib>Poster, Kristina</creatorcontrib><creatorcontrib>Yamaguchi, Diane</creatorcontrib><creatorcontrib>Dickerson, Jane A.</creatorcontrib><creatorcontrib>Drees, Julia C.</creatorcontrib><creatorcontrib>Holmes, Daniel T.</creatorcontrib><creatorcontrib>Greene, Dina N.</creatorcontrib><title>Performance characteristics of the Beckman Coulter UniCel DxI 800 TSH (3rd IS) assay</title><title>Clinica chimica acta</title><addtitle>Clin Chim Acta</addtitle><description>Beckman Coulter recently reformulated their commercial TSH assay with primary calibration to the World Health Organization 3rd TSH international standard. An extensive evaluation of the performance characteristics for this assay was completed. Intra-day and inter-day precision was evaluated using 3 concentrations of commercial quality control material. Linearity, reportable range, stability, sensitivity and susceptibility to common inferences were determined using pooled patient specimens. Inter-assay variability was assessed across 5 different platforms (n=47 patient specimens). Intra-day and inter-day CVs were &lt;10% at all concentrations evaluated. The LOQ, LOD and LOB were 0.0047μIU/ml (10% CV), 0.0012μIU/ml and 0.0005μIU/ml, respectively. Variable bias was observed for the TSH3 assay when evaluated against the previous generation assay and other platforms, but overall TSH3 gave comparable results. The TSH3 assay for UniCel DxI 800, is precise, highly sensitive and comparable to the previous generation assay. The assay is acceptable for clinical testing. •Beckman Coulter reformulated their commercial TSH assay (TSH3).•A comprehensive evaluation of TSH3 was completed.•TSH3 produced comparable results when evaluated against 5 other TSH assays.•TSH demonstrated acceptable intra-day and inter-day (&lt;10%) precision.•TSH3 performs with improved precision and sensitivity.</description><subject>3rd IS</subject><subject>Access</subject><subject>Beckman Coulter</subject><subject>Bias</subject><subject>Calibration</subject><subject>Clinical Laboratory Techniques - instrumentation</subject><subject>Clinical Laboratory Techniques - standards</subject><subject>DxI</subject><subject>Harmonization</subject><subject>Humans</subject><subject>Immunoassay</subject><subject>Quality Control</subject><subject>Reproducibility of Results</subject><subject>Thyrotropin - analysis</subject><subject>TSH</subject><subject>Validation</subject><issn>0009-8981</issn><issn>1873-3492</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2018</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp9kMtOwzAQRS0EgvL4ADbIS1gkzDhpHIsVlFclJJDari3XGQuXtAE7RfD3GBVYsrJGPvdq5jB2jJAjYHW-yK01uQCUOYochNpiA6xlkRWlEttsAAAqq1WNe2w_xkUaS6hwl-0JJeQQy2LApk8UXBeWZmWJ22cTjO0p-Nh7G3nneP9M_IrsSwL4qFu36ZPPVn5ELb_-GPMagE8n9_y0CA0fT864idF8HrIdZ9pIRz_vAZvd3kxH99nD4914dPmQ2RJVnwljkMzQ2SZt5ZyB-RClKWGopJDQgJtXwjprq9JJOXeuKTEFoWhkpajCujhgp5ve19C9rSn2eumjpbY1K-rWUaOqAUFIAQnFDWpDF2Mgp1-DX5rwqRH0t0y90Emm_papUegkM2VOfurX8yU1f4lfewm42ACUjnz3FHS0npLJxgeyvW46_0_9FyDVgmM</recordid><startdate>201803</startdate><enddate>201803</enddate><creator>Winston-McPherson, Gabrielle N.</creator><creator>Samraj, Annie N.</creator><creator>Poster, Kristina</creator><creator>Yamaguchi, Diane</creator><creator>Dickerson, Jane A.</creator><creator>Drees, Julia C.</creator><creator>Holmes, Daniel T.</creator><creator>Greene, Dina N.</creator><general>Elsevier B.V</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>201803</creationdate><title>Performance characteristics of the Beckman Coulter UniCel DxI 800 TSH (3rd IS) assay</title><author>Winston-McPherson, Gabrielle N. ; Samraj, Annie N. ; Poster, Kristina ; Yamaguchi, Diane ; Dickerson, Jane A. ; Drees, Julia C. ; Holmes, Daniel T. ; Greene, Dina N.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c419t-2aa1ea5fcd061ffa0b517a40597270d0fb62cfcc64f77bffd41c4103d769e6183</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2018</creationdate><topic>3rd IS</topic><topic>Access</topic><topic>Beckman Coulter</topic><topic>Bias</topic><topic>Calibration</topic><topic>Clinical Laboratory Techniques - instrumentation</topic><topic>Clinical Laboratory Techniques - standards</topic><topic>DxI</topic><topic>Harmonization</topic><topic>Humans</topic><topic>Immunoassay</topic><topic>Quality Control</topic><topic>Reproducibility of Results</topic><topic>Thyrotropin - analysis</topic><topic>TSH</topic><topic>Validation</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Winston-McPherson, Gabrielle N.</creatorcontrib><creatorcontrib>Samraj, Annie N.</creatorcontrib><creatorcontrib>Poster, Kristina</creatorcontrib><creatorcontrib>Yamaguchi, Diane</creatorcontrib><creatorcontrib>Dickerson, Jane A.</creatorcontrib><creatorcontrib>Drees, Julia C.</creatorcontrib><creatorcontrib>Holmes, Daniel T.</creatorcontrib><creatorcontrib>Greene, Dina N.</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Clinica chimica acta</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Winston-McPherson, Gabrielle N.</au><au>Samraj, Annie N.</au><au>Poster, Kristina</au><au>Yamaguchi, Diane</au><au>Dickerson, Jane A.</au><au>Drees, Julia C.</au><au>Holmes, Daniel T.</au><au>Greene, Dina N.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Performance characteristics of the Beckman Coulter UniCel DxI 800 TSH (3rd IS) assay</atitle><jtitle>Clinica chimica acta</jtitle><addtitle>Clin Chim Acta</addtitle><date>2018-03</date><risdate>2018</risdate><volume>478</volume><spage>90</spage><epage>100</epage><pages>90-100</pages><issn>0009-8981</issn><eissn>1873-3492</eissn><abstract>Beckman Coulter recently reformulated their commercial TSH assay with primary calibration to the World Health Organization 3rd TSH international standard. An extensive evaluation of the performance characteristics for this assay was completed. Intra-day and inter-day precision was evaluated using 3 concentrations of commercial quality control material. Linearity, reportable range, stability, sensitivity and susceptibility to common inferences were determined using pooled patient specimens. Inter-assay variability was assessed across 5 different platforms (n=47 patient specimens). Intra-day and inter-day CVs were &lt;10% at all concentrations evaluated. The LOQ, LOD and LOB were 0.0047μIU/ml (10% CV), 0.0012μIU/ml and 0.0005μIU/ml, respectively. Variable bias was observed for the TSH3 assay when evaluated against the previous generation assay and other platforms, but overall TSH3 gave comparable results. The TSH3 assay for UniCel DxI 800, is precise, highly sensitive and comparable to the previous generation assay. The assay is acceptable for clinical testing. •Beckman Coulter reformulated their commercial TSH assay (TSH3).•A comprehensive evaluation of TSH3 was completed.•TSH3 produced comparable results when evaluated against 5 other TSH assays.•TSH demonstrated acceptable intra-day and inter-day (&lt;10%) precision.•TSH3 performs with improved precision and sensitivity.</abstract><cop>Netherlands</cop><pub>Elsevier B.V</pub><pmid>29275143</pmid><doi>10.1016/j.cca.2017.12.029</doi><tpages>11</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0009-8981
ispartof Clinica chimica acta, 2018-03, Vol.478, p.90-100
issn 0009-8981
1873-3492
language eng
recordid cdi_proquest_miscellaneous_1980102720
source MEDLINE; Elsevier ScienceDirect Journals
subjects 3rd IS
Access
Beckman Coulter
Bias
Calibration
Clinical Laboratory Techniques - instrumentation
Clinical Laboratory Techniques - standards
DxI
Harmonization
Humans
Immunoassay
Quality Control
Reproducibility of Results
Thyrotropin - analysis
TSH
Validation
title Performance characteristics of the Beckman Coulter UniCel DxI 800 TSH (3rd IS) assay
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-10T06%3A39%3A02IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Performance%20characteristics%20of%20the%20Beckman%20Coulter%20UniCel%20DxI%20800%20TSH%20(3rd%20IS)%20assay&rft.jtitle=Clinica%20chimica%20acta&rft.au=Winston-McPherson,%20Gabrielle%20N.&rft.date=2018-03&rft.volume=478&rft.spage=90&rft.epage=100&rft.pages=90-100&rft.issn=0009-8981&rft.eissn=1873-3492&rft_id=info:doi/10.1016/j.cca.2017.12.029&rft_dat=%3Cproquest_cross%3E1980102720%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=1980102720&rft_id=info:pmid/29275143&rft_els_id=S0009898117305296&rfr_iscdi=true