Methodology on quantification of sonication duration for safe application of MR guided focused ultrasound for liver tumour ablation
•Translational research requires evidence before clinical trials. The described methodology enables a systematic automated quality assurance (QA) tool for use in clinics.•The designed protocols measure the sonication duration and, the delay after the emergency stop successfully and repetitively.•The...
Gespeichert in:
Veröffentlicht in: | Computer methods and programs in biomedicine 2017-12, Vol.152, p.125-130 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 130 |
---|---|
container_issue | |
container_start_page | 125 |
container_title | Computer methods and programs in biomedicine |
container_volume | 152 |
creator | Mihcin, Senay Karakitsios, Ioannis Le, Nhan Strehlow, Jan Demedts, Daniel Schwenke, Michael Haase, Sabrina Preusser, Tobias Melzer, Andreas |
description | •Translational research requires evidence before clinical trials. The described methodology enables a systematic automated quality assurance (QA) tool for use in clinics.•The designed protocols measure the sonication duration and, the delay after the emergency stop successfully and repetitively.•The data presented forms a sample data base for quality assurance purposes by providing evidence.•The developed novel software; TRANS-FUSIMO Treatment System (TTS) meets the specifications required for clinical trials.•The overall approach demonstrates a good practice for translational research for biomedical applications.
Magnetic Resonance Guided Focused Ultrasound (MRgFUS) for liver tumour ablation is a challenging task due to motion caused by breathing and occlusion due the ribcage between the transducer and the tumour. To overcome these challenges, a novel system for liver tumour ablation during free breathing has been designed.
The novel TRANS-FUSIMO Treatment System (TTS, EUFP7) interacts with a Magnetic Resonance (MR) scanner and a focused ultrasound transducer to sonicate to a moving target in liver. To meet the requirements of ISO 13485; a quality management system for medical device design, the system needs to be tested for certain process parameters. The duration of sonication and, the delay after the sonication button is activated, are among the parameters that need to be quantified for efficient and safe ablation of tumour tissue. A novel methodology is developed to quantify these process parameters. A computerised scope is programmed in LabVIEW to collect data via hydrophone; where the coordinates of fiber-optic sensor assembly was fed into the TRANS-FUSIMO treatment software via Magnetic Resonance Imaging (MRI) to sonicate to the tip of the sensor, which is synchronised with the clock of the scope, embedded in a degassed water tank via sensor assembly holder. The sonications were executed for 50 W, 100 W, 150 W for 10 s to quantify the actual sonication duration and the delay after the emergency stop by two independent operators for thirty times. The deviation of the system from the predefined specs was calculated. Student's-T test was used to investigate the user dependency.
The duration of sonication and the delay after the sonication were quantified successfully with the developed method. TTS can sonicate with a maximum deviation of 0.16 s (Std 0.32) from the planned duration and with a delay of 14 ms (Std 0.14) for the emergency stop. Student's |
doi_str_mv | 10.1016/j.cmpb.2017.09.006 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_1954077581</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S016926071730158X</els_id><sourcerecordid>1954077581</sourcerecordid><originalsourceid>FETCH-LOGICAL-c400t-46d1c069ee5e642c553108ffe9310b6582bd6afa2252023f3e5145d3b1ba9bfd3</originalsourceid><addsrcrecordid>eNp9kEFr3DAUhEVoSbZp_0APxcde7DzJlryGXkpo0kBCoCRnIUtPWy225UhWIOf88Wqzm5JTTvMefDMwQ8hXChUFKs62lR7nvmJA2wq6CkAckRVdt6xsueAfyCpDXckEtCfkU4xbAGCci2NywjrgDeNsRZ5vcPnrjR_85qnwU_GQ1LQ467RaXH69LaKfXj-Twv6wPhRRWSzUPA9v2Js_xSY5gyYTOsWsaViCij5N5sU0uEcMxZJGn0Kh-uHF-Zl8tGqI-OWgp-T-4tfd-e_y-vby6vzndakbgKVshKEaRIfIUTRMc15TWFuLXdZe8DXrjVBWsVwMWG1r5LThpu5pr7remvqUfN_nzsE_JIyLHF3UOAxqQp-ipB1voG35mmaU7VEdfIwBrZyDG1V4khTkbny5lbvx5W58CZ3M42fTt0N-6kc0_y2va2fgxx7A3PLRYZBRO5w0GhdQL9J4917-P7mGmJk</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1954077581</pqid></control><display><type>article</type><title>Methodology on quantification of sonication duration for safe application of MR guided focused ultrasound for liver tumour ablation</title><source>MEDLINE</source><source>ScienceDirect Journals (5 years ago - present)</source><creator>Mihcin, Senay ; Karakitsios, Ioannis ; Le, Nhan ; Strehlow, Jan ; Demedts, Daniel ; Schwenke, Michael ; Haase, Sabrina ; Preusser, Tobias ; Melzer, Andreas</creator><creatorcontrib>Mihcin, Senay ; Karakitsios, Ioannis ; Le, Nhan ; Strehlow, Jan ; Demedts, Daniel ; Schwenke, Michael ; Haase, Sabrina ; Preusser, Tobias ; Melzer, Andreas</creatorcontrib><description>•Translational research requires evidence before clinical trials. The described methodology enables a systematic automated quality assurance (QA) tool for use in clinics.•The designed protocols measure the sonication duration and, the delay after the emergency stop successfully and repetitively.•The data presented forms a sample data base for quality assurance purposes by providing evidence.•The developed novel software; TRANS-FUSIMO Treatment System (TTS) meets the specifications required for clinical trials.•The overall approach demonstrates a good practice for translational research for biomedical applications.
Magnetic Resonance Guided Focused Ultrasound (MRgFUS) for liver tumour ablation is a challenging task due to motion caused by breathing and occlusion due the ribcage between the transducer and the tumour. To overcome these challenges, a novel system for liver tumour ablation during free breathing has been designed.
The novel TRANS-FUSIMO Treatment System (TTS, EUFP7) interacts with a Magnetic Resonance (MR) scanner and a focused ultrasound transducer to sonicate to a moving target in liver. To meet the requirements of ISO 13485; a quality management system for medical device design, the system needs to be tested for certain process parameters. The duration of sonication and, the delay after the sonication button is activated, are among the parameters that need to be quantified for efficient and safe ablation of tumour tissue. A novel methodology is developed to quantify these process parameters. A computerised scope is programmed in LabVIEW to collect data via hydrophone; where the coordinates of fiber-optic sensor assembly was fed into the TRANS-FUSIMO treatment software via Magnetic Resonance Imaging (MRI) to sonicate to the tip of the sensor, which is synchronised with the clock of the scope, embedded in a degassed water tank via sensor assembly holder. The sonications were executed for 50 W, 100 W, 150 W for 10 s to quantify the actual sonication duration and the delay after the emergency stop by two independent operators for thirty times. The deviation of the system from the predefined specs was calculated. Student's-T test was used to investigate the user dependency.
The duration of sonication and the delay after the sonication were quantified successfully with the developed method. TTS can sonicate with a maximum deviation of 0.16 s (Std 0.32) from the planned duration and with a delay of 14 ms (Std 0.14) for the emergency stop. Student's T tests indicate that the results do not depend on operators (p > .05).
The evidence obtained via this protocol is crucial for translation- of-research into the clinics for safe application of MRgFUS. The developed protocol could be used for system maintenance in compliance with quality systems in clinics for daily quality assurance routines.</description><identifier>ISSN: 0169-2607</identifier><identifier>EISSN: 1872-7565</identifier><identifier>DOI: 10.1016/j.cmpb.2017.09.006</identifier><identifier>PMID: 29054252</identifier><language>eng</language><publisher>Ireland: Elsevier B.V</publisher><subject>Ablation Techniques - methods ; Computer control of laboratory machines and device ; Experiment and measurement technics, medical device legislation ; Humans ; Liver Neoplasms - diagnostic imaging ; Liver Neoplasms - surgery ; Magnetic Resonance Imaging - methods ; MR guided FUS ; Protocol development ; Quality Assurance, Health Care ; Quality Management Systems ; Safety ; Software ; Sonication - standards ; Sonication duration ; Ultrasonography - methods</subject><ispartof>Computer methods and programs in biomedicine, 2017-12, Vol.152, p.125-130</ispartof><rights>2017 Elsevier B.V.</rights><rights>Copyright © 2017 Elsevier B.V. All rights reserved.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c400t-46d1c069ee5e642c553108ffe9310b6582bd6afa2252023f3e5145d3b1ba9bfd3</citedby><cites>FETCH-LOGICAL-c400t-46d1c069ee5e642c553108ffe9310b6582bd6afa2252023f3e5145d3b1ba9bfd3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://dx.doi.org/10.1016/j.cmpb.2017.09.006$$EHTML$$P50$$Gelsevier$$H</linktohtml><link.rule.ids>314,780,784,3548,27922,27923,45993</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/29054252$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Mihcin, Senay</creatorcontrib><creatorcontrib>Karakitsios, Ioannis</creatorcontrib><creatorcontrib>Le, Nhan</creatorcontrib><creatorcontrib>Strehlow, Jan</creatorcontrib><creatorcontrib>Demedts, Daniel</creatorcontrib><creatorcontrib>Schwenke, Michael</creatorcontrib><creatorcontrib>Haase, Sabrina</creatorcontrib><creatorcontrib>Preusser, Tobias</creatorcontrib><creatorcontrib>Melzer, Andreas</creatorcontrib><title>Methodology on quantification of sonication duration for safe application of MR guided focused ultrasound for liver tumour ablation</title><title>Computer methods and programs in biomedicine</title><addtitle>Comput Methods Programs Biomed</addtitle><description>•Translational research requires evidence before clinical trials. The described methodology enables a systematic automated quality assurance (QA) tool for use in clinics.•The designed protocols measure the sonication duration and, the delay after the emergency stop successfully and repetitively.•The data presented forms a sample data base for quality assurance purposes by providing evidence.•The developed novel software; TRANS-FUSIMO Treatment System (TTS) meets the specifications required for clinical trials.•The overall approach demonstrates a good practice for translational research for biomedical applications.
Magnetic Resonance Guided Focused Ultrasound (MRgFUS) for liver tumour ablation is a challenging task due to motion caused by breathing and occlusion due the ribcage between the transducer and the tumour. To overcome these challenges, a novel system for liver tumour ablation during free breathing has been designed.
The novel TRANS-FUSIMO Treatment System (TTS, EUFP7) interacts with a Magnetic Resonance (MR) scanner and a focused ultrasound transducer to sonicate to a moving target in liver. To meet the requirements of ISO 13485; a quality management system for medical device design, the system needs to be tested for certain process parameters. The duration of sonication and, the delay after the sonication button is activated, are among the parameters that need to be quantified for efficient and safe ablation of tumour tissue. A novel methodology is developed to quantify these process parameters. A computerised scope is programmed in LabVIEW to collect data via hydrophone; where the coordinates of fiber-optic sensor assembly was fed into the TRANS-FUSIMO treatment software via Magnetic Resonance Imaging (MRI) to sonicate to the tip of the sensor, which is synchronised with the clock of the scope, embedded in a degassed water tank via sensor assembly holder. The sonications were executed for 50 W, 100 W, 150 W for 10 s to quantify the actual sonication duration and the delay after the emergency stop by two independent operators for thirty times. The deviation of the system from the predefined specs was calculated. Student's-T test was used to investigate the user dependency.
The duration of sonication and the delay after the sonication were quantified successfully with the developed method. TTS can sonicate with a maximum deviation of 0.16 s (Std 0.32) from the planned duration and with a delay of 14 ms (Std 0.14) for the emergency stop. Student's T tests indicate that the results do not depend on operators (p > .05).
The evidence obtained via this protocol is crucial for translation- of-research into the clinics for safe application of MRgFUS. The developed protocol could be used for system maintenance in compliance with quality systems in clinics for daily quality assurance routines.</description><subject>Ablation Techniques - methods</subject><subject>Computer control of laboratory machines and device</subject><subject>Experiment and measurement technics, medical device legislation</subject><subject>Humans</subject><subject>Liver Neoplasms - diagnostic imaging</subject><subject>Liver Neoplasms - surgery</subject><subject>Magnetic Resonance Imaging - methods</subject><subject>MR guided FUS</subject><subject>Protocol development</subject><subject>Quality Assurance, Health Care</subject><subject>Quality Management Systems</subject><subject>Safety</subject><subject>Software</subject><subject>Sonication - standards</subject><subject>Sonication duration</subject><subject>Ultrasonography - methods</subject><issn>0169-2607</issn><issn>1872-7565</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2017</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp9kEFr3DAUhEVoSbZp_0APxcde7DzJlryGXkpo0kBCoCRnIUtPWy225UhWIOf88Wqzm5JTTvMefDMwQ8hXChUFKs62lR7nvmJA2wq6CkAckRVdt6xsueAfyCpDXckEtCfkU4xbAGCci2NywjrgDeNsRZ5vcPnrjR_85qnwU_GQ1LQ467RaXH69LaKfXj-Twv6wPhRRWSzUPA9v2Js_xSY5gyYTOsWsaViCij5N5sU0uEcMxZJGn0Kh-uHF-Zl8tGqI-OWgp-T-4tfd-e_y-vby6vzndakbgKVshKEaRIfIUTRMc15TWFuLXdZe8DXrjVBWsVwMWG1r5LThpu5pr7remvqUfN_nzsE_JIyLHF3UOAxqQp-ipB1voG35mmaU7VEdfIwBrZyDG1V4khTkbny5lbvx5W58CZ3M42fTt0N-6kc0_y2va2fgxx7A3PLRYZBRO5w0GhdQL9J4917-P7mGmJk</recordid><startdate>201712</startdate><enddate>201712</enddate><creator>Mihcin, Senay</creator><creator>Karakitsios, Ioannis</creator><creator>Le, Nhan</creator><creator>Strehlow, Jan</creator><creator>Demedts, Daniel</creator><creator>Schwenke, Michael</creator><creator>Haase, Sabrina</creator><creator>Preusser, Tobias</creator><creator>Melzer, Andreas</creator><general>Elsevier B.V</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>201712</creationdate><title>Methodology on quantification of sonication duration for safe application of MR guided focused ultrasound for liver tumour ablation</title><author>Mihcin, Senay ; Karakitsios, Ioannis ; Le, Nhan ; Strehlow, Jan ; Demedts, Daniel ; Schwenke, Michael ; Haase, Sabrina ; Preusser, Tobias ; Melzer, Andreas</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c400t-46d1c069ee5e642c553108ffe9310b6582bd6afa2252023f3e5145d3b1ba9bfd3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2017</creationdate><topic>Ablation Techniques - methods</topic><topic>Computer control of laboratory machines and device</topic><topic>Experiment and measurement technics, medical device legislation</topic><topic>Humans</topic><topic>Liver Neoplasms - diagnostic imaging</topic><topic>Liver Neoplasms - surgery</topic><topic>Magnetic Resonance Imaging - methods</topic><topic>MR guided FUS</topic><topic>Protocol development</topic><topic>Quality Assurance, Health Care</topic><topic>Quality Management Systems</topic><topic>Safety</topic><topic>Software</topic><topic>Sonication - standards</topic><topic>Sonication duration</topic><topic>Ultrasonography - methods</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Mihcin, Senay</creatorcontrib><creatorcontrib>Karakitsios, Ioannis</creatorcontrib><creatorcontrib>Le, Nhan</creatorcontrib><creatorcontrib>Strehlow, Jan</creatorcontrib><creatorcontrib>Demedts, Daniel</creatorcontrib><creatorcontrib>Schwenke, Michael</creatorcontrib><creatorcontrib>Haase, Sabrina</creatorcontrib><creatorcontrib>Preusser, Tobias</creatorcontrib><creatorcontrib>Melzer, Andreas</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Computer methods and programs in biomedicine</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Mihcin, Senay</au><au>Karakitsios, Ioannis</au><au>Le, Nhan</au><au>Strehlow, Jan</au><au>Demedts, Daniel</au><au>Schwenke, Michael</au><au>Haase, Sabrina</au><au>Preusser, Tobias</au><au>Melzer, Andreas</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Methodology on quantification of sonication duration for safe application of MR guided focused ultrasound for liver tumour ablation</atitle><jtitle>Computer methods and programs in biomedicine</jtitle><addtitle>Comput Methods Programs Biomed</addtitle><date>2017-12</date><risdate>2017</risdate><volume>152</volume><spage>125</spage><epage>130</epage><pages>125-130</pages><issn>0169-2607</issn><eissn>1872-7565</eissn><abstract>•Translational research requires evidence before clinical trials. The described methodology enables a systematic automated quality assurance (QA) tool for use in clinics.•The designed protocols measure the sonication duration and, the delay after the emergency stop successfully and repetitively.•The data presented forms a sample data base for quality assurance purposes by providing evidence.•The developed novel software; TRANS-FUSIMO Treatment System (TTS) meets the specifications required for clinical trials.•The overall approach demonstrates a good practice for translational research for biomedical applications.
Magnetic Resonance Guided Focused Ultrasound (MRgFUS) for liver tumour ablation is a challenging task due to motion caused by breathing and occlusion due the ribcage between the transducer and the tumour. To overcome these challenges, a novel system for liver tumour ablation during free breathing has been designed.
The novel TRANS-FUSIMO Treatment System (TTS, EUFP7) interacts with a Magnetic Resonance (MR) scanner and a focused ultrasound transducer to sonicate to a moving target in liver. To meet the requirements of ISO 13485; a quality management system for medical device design, the system needs to be tested for certain process parameters. The duration of sonication and, the delay after the sonication button is activated, are among the parameters that need to be quantified for efficient and safe ablation of tumour tissue. A novel methodology is developed to quantify these process parameters. A computerised scope is programmed in LabVIEW to collect data via hydrophone; where the coordinates of fiber-optic sensor assembly was fed into the TRANS-FUSIMO treatment software via Magnetic Resonance Imaging (MRI) to sonicate to the tip of the sensor, which is synchronised with the clock of the scope, embedded in a degassed water tank via sensor assembly holder. The sonications were executed for 50 W, 100 W, 150 W for 10 s to quantify the actual sonication duration and the delay after the emergency stop by two independent operators for thirty times. The deviation of the system from the predefined specs was calculated. Student's-T test was used to investigate the user dependency.
The duration of sonication and the delay after the sonication were quantified successfully with the developed method. TTS can sonicate with a maximum deviation of 0.16 s (Std 0.32) from the planned duration and with a delay of 14 ms (Std 0.14) for the emergency stop. Student's T tests indicate that the results do not depend on operators (p > .05).
The evidence obtained via this protocol is crucial for translation- of-research into the clinics for safe application of MRgFUS. The developed protocol could be used for system maintenance in compliance with quality systems in clinics for daily quality assurance routines.</abstract><cop>Ireland</cop><pub>Elsevier B.V</pub><pmid>29054252</pmid><doi>10.1016/j.cmpb.2017.09.006</doi><tpages>6</tpages><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0169-2607 |
ispartof | Computer methods and programs in biomedicine, 2017-12, Vol.152, p.125-130 |
issn | 0169-2607 1872-7565 |
language | eng |
recordid | cdi_proquest_miscellaneous_1954077581 |
source | MEDLINE; ScienceDirect Journals (5 years ago - present) |
subjects | Ablation Techniques - methods Computer control of laboratory machines and device Experiment and measurement technics, medical device legislation Humans Liver Neoplasms - diagnostic imaging Liver Neoplasms - surgery Magnetic Resonance Imaging - methods MR guided FUS Protocol development Quality Assurance, Health Care Quality Management Systems Safety Software Sonication - standards Sonication duration Ultrasonography - methods |
title | Methodology on quantification of sonication duration for safe application of MR guided focused ultrasound for liver tumour ablation |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-14T08%3A19%3A59IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Methodology%20on%20quantification%20of%20sonication%20duration%20for%20safe%20application%20of%20MR%20guided%20focused%20ultrasound%20for%20liver%20tumour%20ablation&rft.jtitle=Computer%20methods%20and%20programs%20in%20biomedicine&rft.au=Mihcin,%20Senay&rft.date=2017-12&rft.volume=152&rft.spage=125&rft.epage=130&rft.pages=125-130&rft.issn=0169-2607&rft.eissn=1872-7565&rft_id=info:doi/10.1016/j.cmpb.2017.09.006&rft_dat=%3Cproquest_cross%3E1954077581%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=1954077581&rft_id=info:pmid/29054252&rft_els_id=S016926071730158X&rfr_iscdi=true |