Carotid artery bursting pressure and seal time after multiple uses of a vessel sealing device

Objective To determine any loss of function of a single‐use bipolar vessel sealing device after repeated sterilization and use on canine carotid arteries. Study design Ex vivo; randomized block design (dog). Sample population Carotid artery segments (n = 90) from 15 canine cadavers. Methods Six 10‐m...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Veterinary surgery 2017-05, Vol.46 (4), p.501-506
Hauptverfasser: Blake, Jocelin S., Trumpatori, Brian J., Mathews, Kyle G., Griffith, Emily H.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 506
container_issue 4
container_start_page 501
container_title Veterinary surgery
container_volume 46
creator Blake, Jocelin S.
Trumpatori, Brian J.
Mathews, Kyle G.
Griffith, Emily H.
description Objective To determine any loss of function of a single‐use bipolar vessel sealing device after repeated sterilization and use on canine carotid arteries. Study design Ex vivo; randomized block design (dog). Sample population Carotid artery segments (n = 90) from 15 canine cadavers. Methods Six 10‐mm LigaSure Atlas 20 cm devices (Covidien Inc, Minneapolis, Minnesota) were used to seal canine carotid artery segments in one cadaver, then resterilized using ethylene oxide and reused in a second cadaver, and so on, for a total of 15 uses/15 cadavers per LigaSure. Sealing times and bursting pressures of each carotid artery segment after each use were recorded. Charring and sticking at each seal location was qualitatively assessed. Results Mean (SD) sealing time for all segments was 5.3 (1.4) seconds, ranging from 2.7 to 9.5 seconds. There was no significant change in sealing time for successive uses (P = .117). The mean (SD) bursting pressure for all segments was 1041.3 (316.7) mm Hg, ranging from 355 to 1555 mm Hg. There was no change in bursting pressure across uses (P = .57). Moderate sticking and charring occurred at each time point but was not subjectively different throughout the study. Conclusion The LigaSure Atlas vessel sealing device can be used and resterilized with ethylene oxide up to 15 times with no change in sealing time or bursting pressure of the seal.
doi_str_mv 10.1111/vsu.12648
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_1881262687</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>1881262687</sourcerecordid><originalsourceid>FETCH-LOGICAL-c3538-deffd9bc5867e80b48d1e944bc7275578aa12ac27a280936eae9c7190995ceac3</originalsourceid><addsrcrecordid>eNp10MtKAzEUBuAgiq3VhS8gATe6mDaXuSRLKd6g4EIrbmTIZM7IlLmZTCp9e9NOdSF4NuHAl5_Dj9A5JVPqZ7a2bkpZHIoDNKYRZ4GMyNshGhMa04CHUo7QibUrQogMQ36MRkxwHoahGKP3uTJtX-ZYmR7MBmfO2L5sPnBnwFpnAKsmxxZUhfuy9lvhGa5d1ZddBdhZsLgtsMJrz6Haye33HNalhlN0VKjKwtn-naDl3e3L_CFYPN0_zm8WgeYRF0EORZHLTEciTkCQLBQ5BX9qphOWRFEilKJMaZYoJojkMSiQOqGSSBlpUJpP0NWQ25n204Ht07q0GqpKNdA6m1IhfD8sFomnl3_oqnWm8dd55XuLCOfMq-tBadNaa6BIO1PWymxSStJt56nvPN117u3FPtFlNeS_8qdkD2YD-Cor2PyflL4-L4fIb_kni7A</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1895050332</pqid></control><display><type>article</type><title>Carotid artery bursting pressure and seal time after multiple uses of a vessel sealing device</title><source>MEDLINE</source><source>Access via Wiley Online Library</source><creator>Blake, Jocelin S. ; Trumpatori, Brian J. ; Mathews, Kyle G. ; Griffith, Emily H.</creator><creatorcontrib>Blake, Jocelin S. ; Trumpatori, Brian J. ; Mathews, Kyle G. ; Griffith, Emily H.</creatorcontrib><description>Objective To determine any loss of function of a single‐use bipolar vessel sealing device after repeated sterilization and use on canine carotid arteries. Study design Ex vivo; randomized block design (dog). Sample population Carotid artery segments (n = 90) from 15 canine cadavers. Methods Six 10‐mm LigaSure Atlas 20 cm devices (Covidien Inc, Minneapolis, Minnesota) were used to seal canine carotid artery segments in one cadaver, then resterilized using ethylene oxide and reused in a second cadaver, and so on, for a total of 15 uses/15 cadavers per LigaSure. Sealing times and bursting pressures of each carotid artery segment after each use were recorded. Charring and sticking at each seal location was qualitatively assessed. Results Mean (SD) sealing time for all segments was 5.3 (1.4) seconds, ranging from 2.7 to 9.5 seconds. There was no significant change in sealing time for successive uses (P = .117). The mean (SD) bursting pressure for all segments was 1041.3 (316.7) mm Hg, ranging from 355 to 1555 mm Hg. There was no change in bursting pressure across uses (P = .57). Moderate sticking and charring occurred at each time point but was not subjectively different throughout the study. Conclusion The LigaSure Atlas vessel sealing device can be used and resterilized with ethylene oxide up to 15 times with no change in sealing time or bursting pressure of the seal.</description><identifier>ISSN: 0161-3499</identifier><identifier>EISSN: 1532-950X</identifier><identifier>DOI: 10.1111/vsu.12648</identifier><identifier>PMID: 28334448</identifier><language>eng</language><publisher>United States: Blackwell Publishing Ltd</publisher><subject>Animals ; Arteries ; Biomechanical Phenomena ; Bursting ; Bursting pressure ; Cadaver ; Cadavers ; Carotid arteries ; Carotid Arteries - physiology ; Carotid artery ; Charring ; Dogs ; Dust ; Electrosurgery - veterinary ; Ethylene ; Ethylene oxide ; Mercury ; Pressure ; Sealing ; Segments ; Sterilization ; Surgery ; Surgical Instruments - veterinary ; Vascular Surgical Procedures - instrumentation</subject><ispartof>Veterinary surgery, 2017-05, Vol.46 (4), p.501-506</ispartof><rights>2017 The American College of Veterinary Surgeons</rights><rights>2017 The American College of Veterinary Surgeons.</rights><rights>Copyright 2017 by The American College of Veterinary Surgeons</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c3538-deffd9bc5867e80b48d1e944bc7275578aa12ac27a280936eae9c7190995ceac3</citedby><cites>FETCH-LOGICAL-c3538-deffd9bc5867e80b48d1e944bc7275578aa12ac27a280936eae9c7190995ceac3</cites><orcidid>0000-0003-0899-1243</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://onlinelibrary.wiley.com/doi/pdf/10.1111%2Fvsu.12648$$EPDF$$P50$$Gwiley$$H</linktopdf><linktohtml>$$Uhttps://onlinelibrary.wiley.com/doi/full/10.1111%2Fvsu.12648$$EHTML$$P50$$Gwiley$$H</linktohtml><link.rule.ids>315,781,785,1418,27929,27930,45579,45580</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/28334448$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Blake, Jocelin S.</creatorcontrib><creatorcontrib>Trumpatori, Brian J.</creatorcontrib><creatorcontrib>Mathews, Kyle G.</creatorcontrib><creatorcontrib>Griffith, Emily H.</creatorcontrib><title>Carotid artery bursting pressure and seal time after multiple uses of a vessel sealing device</title><title>Veterinary surgery</title><addtitle>Vet Surg</addtitle><description>Objective To determine any loss of function of a single‐use bipolar vessel sealing device after repeated sterilization and use on canine carotid arteries. Study design Ex vivo; randomized block design (dog). Sample population Carotid artery segments (n = 90) from 15 canine cadavers. Methods Six 10‐mm LigaSure Atlas 20 cm devices (Covidien Inc, Minneapolis, Minnesota) were used to seal canine carotid artery segments in one cadaver, then resterilized using ethylene oxide and reused in a second cadaver, and so on, for a total of 15 uses/15 cadavers per LigaSure. Sealing times and bursting pressures of each carotid artery segment after each use were recorded. Charring and sticking at each seal location was qualitatively assessed. Results Mean (SD) sealing time for all segments was 5.3 (1.4) seconds, ranging from 2.7 to 9.5 seconds. There was no significant change in sealing time for successive uses (P = .117). The mean (SD) bursting pressure for all segments was 1041.3 (316.7) mm Hg, ranging from 355 to 1555 mm Hg. There was no change in bursting pressure across uses (P = .57). Moderate sticking and charring occurred at each time point but was not subjectively different throughout the study. Conclusion The LigaSure Atlas vessel sealing device can be used and resterilized with ethylene oxide up to 15 times with no change in sealing time or bursting pressure of the seal.</description><subject>Animals</subject><subject>Arteries</subject><subject>Biomechanical Phenomena</subject><subject>Bursting</subject><subject>Bursting pressure</subject><subject>Cadaver</subject><subject>Cadavers</subject><subject>Carotid arteries</subject><subject>Carotid Arteries - physiology</subject><subject>Carotid artery</subject><subject>Charring</subject><subject>Dogs</subject><subject>Dust</subject><subject>Electrosurgery - veterinary</subject><subject>Ethylene</subject><subject>Ethylene oxide</subject><subject>Mercury</subject><subject>Pressure</subject><subject>Sealing</subject><subject>Segments</subject><subject>Sterilization</subject><subject>Surgery</subject><subject>Surgical Instruments - veterinary</subject><subject>Vascular Surgical Procedures - instrumentation</subject><issn>0161-3499</issn><issn>1532-950X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2017</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp10MtKAzEUBuAgiq3VhS8gATe6mDaXuSRLKd6g4EIrbmTIZM7IlLmZTCp9e9NOdSF4NuHAl5_Dj9A5JVPqZ7a2bkpZHIoDNKYRZ4GMyNshGhMa04CHUo7QibUrQogMQ36MRkxwHoahGKP3uTJtX-ZYmR7MBmfO2L5sPnBnwFpnAKsmxxZUhfuy9lvhGa5d1ZddBdhZsLgtsMJrz6Haye33HNalhlN0VKjKwtn-naDl3e3L_CFYPN0_zm8WgeYRF0EORZHLTEciTkCQLBQ5BX9qphOWRFEilKJMaZYoJojkMSiQOqGSSBlpUJpP0NWQ25n204Ht07q0GqpKNdA6m1IhfD8sFomnl3_oqnWm8dd55XuLCOfMq-tBadNaa6BIO1PWymxSStJt56nvPN117u3FPtFlNeS_8qdkD2YD-Cor2PyflL4-L4fIb_kni7A</recordid><startdate>201705</startdate><enddate>201705</enddate><creator>Blake, Jocelin S.</creator><creator>Trumpatori, Brian J.</creator><creator>Mathews, Kyle G.</creator><creator>Griffith, Emily H.</creator><general>Blackwell Publishing Ltd</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>8FD</scope><scope>FR3</scope><scope>M7Z</scope><scope>P64</scope><scope>7X8</scope><orcidid>https://orcid.org/0000-0003-0899-1243</orcidid></search><sort><creationdate>201705</creationdate><title>Carotid artery bursting pressure and seal time after multiple uses of a vessel sealing device</title><author>Blake, Jocelin S. ; Trumpatori, Brian J. ; Mathews, Kyle G. ; Griffith, Emily H.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c3538-deffd9bc5867e80b48d1e944bc7275578aa12ac27a280936eae9c7190995ceac3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2017</creationdate><topic>Animals</topic><topic>Arteries</topic><topic>Biomechanical Phenomena</topic><topic>Bursting</topic><topic>Bursting pressure</topic><topic>Cadaver</topic><topic>Cadavers</topic><topic>Carotid arteries</topic><topic>Carotid Arteries - physiology</topic><topic>Carotid artery</topic><topic>Charring</topic><topic>Dogs</topic><topic>Dust</topic><topic>Electrosurgery - veterinary</topic><topic>Ethylene</topic><topic>Ethylene oxide</topic><topic>Mercury</topic><topic>Pressure</topic><topic>Sealing</topic><topic>Segments</topic><topic>Sterilization</topic><topic>Surgery</topic><topic>Surgical Instruments - veterinary</topic><topic>Vascular Surgical Procedures - instrumentation</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Blake, Jocelin S.</creatorcontrib><creatorcontrib>Trumpatori, Brian J.</creatorcontrib><creatorcontrib>Mathews, Kyle G.</creatorcontrib><creatorcontrib>Griffith, Emily H.</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>Technology Research Database</collection><collection>Engineering Research Database</collection><collection>Biochemistry Abstracts 1</collection><collection>Biotechnology and BioEngineering Abstracts</collection><collection>MEDLINE - Academic</collection><jtitle>Veterinary surgery</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Blake, Jocelin S.</au><au>Trumpatori, Brian J.</au><au>Mathews, Kyle G.</au><au>Griffith, Emily H.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Carotid artery bursting pressure and seal time after multiple uses of a vessel sealing device</atitle><jtitle>Veterinary surgery</jtitle><addtitle>Vet Surg</addtitle><date>2017-05</date><risdate>2017</risdate><volume>46</volume><issue>4</issue><spage>501</spage><epage>506</epage><pages>501-506</pages><issn>0161-3499</issn><eissn>1532-950X</eissn><abstract>Objective To determine any loss of function of a single‐use bipolar vessel sealing device after repeated sterilization and use on canine carotid arteries. Study design Ex vivo; randomized block design (dog). Sample population Carotid artery segments (n = 90) from 15 canine cadavers. Methods Six 10‐mm LigaSure Atlas 20 cm devices (Covidien Inc, Minneapolis, Minnesota) were used to seal canine carotid artery segments in one cadaver, then resterilized using ethylene oxide and reused in a second cadaver, and so on, for a total of 15 uses/15 cadavers per LigaSure. Sealing times and bursting pressures of each carotid artery segment after each use were recorded. Charring and sticking at each seal location was qualitatively assessed. Results Mean (SD) sealing time for all segments was 5.3 (1.4) seconds, ranging from 2.7 to 9.5 seconds. There was no significant change in sealing time for successive uses (P = .117). The mean (SD) bursting pressure for all segments was 1041.3 (316.7) mm Hg, ranging from 355 to 1555 mm Hg. There was no change in bursting pressure across uses (P = .57). Moderate sticking and charring occurred at each time point but was not subjectively different throughout the study. Conclusion The LigaSure Atlas vessel sealing device can be used and resterilized with ethylene oxide up to 15 times with no change in sealing time or bursting pressure of the seal.</abstract><cop>United States</cop><pub>Blackwell Publishing Ltd</pub><pmid>28334448</pmid><doi>10.1111/vsu.12648</doi><tpages>6</tpages><orcidid>https://orcid.org/0000-0003-0899-1243</orcidid></addata></record>
fulltext fulltext
identifier ISSN: 0161-3499
ispartof Veterinary surgery, 2017-05, Vol.46 (4), p.501-506
issn 0161-3499
1532-950X
language eng
recordid cdi_proquest_miscellaneous_1881262687
source MEDLINE; Access via Wiley Online Library
subjects Animals
Arteries
Biomechanical Phenomena
Bursting
Bursting pressure
Cadaver
Cadavers
Carotid arteries
Carotid Arteries - physiology
Carotid artery
Charring
Dogs
Dust
Electrosurgery - veterinary
Ethylene
Ethylene oxide
Mercury
Pressure
Sealing
Segments
Sterilization
Surgery
Surgical Instruments - veterinary
Vascular Surgical Procedures - instrumentation
title Carotid artery bursting pressure and seal time after multiple uses of a vessel sealing device
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-16T12%3A10%3A15IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Carotid%20artery%20bursting%20pressure%20and%20seal%20time%20after%20multiple%20uses%20of%20a%20vessel%20sealing%20device&rft.jtitle=Veterinary%20surgery&rft.au=Blake,%20Jocelin%20S.&rft.date=2017-05&rft.volume=46&rft.issue=4&rft.spage=501&rft.epage=506&rft.pages=501-506&rft.issn=0161-3499&rft.eissn=1532-950X&rft_id=info:doi/10.1111/vsu.12648&rft_dat=%3Cproquest_cross%3E1881262687%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=1895050332&rft_id=info:pmid/28334448&rfr_iscdi=true