Non-spectral interferences encountered with a commercially available high resolution ICP-mass spectrometer
Results of a systematic study concerning non-spectral interferences observed with a commercially available high resolution ICP-mass spectrometer are reported and compared to observations made with a quadrupole-based instrument. In general, matrix effects were observed to be to a large extent compara...
Gespeichert in:
Veröffentlicht in: | Analytical and bioanalytical chemistry 1996-06, Vol.355 (3-4), p.397-400 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 400 |
---|---|
container_issue | 3-4 |
container_start_page | 397 |
container_title | Analytical and bioanalytical chemistry |
container_volume | 355 |
creator | VANHAECKE, F RIONDATO, J MOENS, L DAMS, R |
description | Results of a systematic study concerning non-spectral interferences observed with a commercially available high resolution ICP-mass spectrometer are reported and compared to observations made with a quadrupole-based instrument. In general, matrix effects were observed to be to a large extent comparable for both instruments used. In all cases, the matrix-induced signal suppression or enhancement was seen to depend in a regular way on the mass number of the nuclides monitored. In most cases, the ionization potential of the nuclides has little or no influence on the extent of suppression or enhancement. For As, Se and Te, the introduction of 2.5% ethanol, 0.5 mol/l H(2)SO(4), or to a lesser extent 0.5 mol/l H(3)PO(4), leads to an exceptional increase in the signal intensity for both instruments. Registration of signal behaviour plots (signal intensity as a function of the nebulizer gas flow rate) in different matrices revealed that both the height of the plot and the optimum nebulizer gas flow rate are a function of the matrix composition. Finally, no indication was found that the acceleration of the extracted ions over 8000 V with the high resolution instrument would lead to an alleviation of space charge effects when compared to a quadrupole-based ICP-mass spectrometer. |
doi_str_mv | 10.1007/s0021663550397 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_1859453595</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>1859453595</sourcerecordid><originalsourceid>FETCH-LOGICAL-c322t-280188dd5206e1662192a775ddad83238f40fd55543836cee1d1886b0d8920843</originalsourceid><addsrcrecordid>eNpVkD1v2zAQhomiQe26XTMGHDp0kXP8kqgxMJLWgNFkSGaBJk8xDUp0SKlB_n1k2EjR5Q44PPfg7iXkksGSAVTXGYCzshRKgairT2TOpOAFYwI-kznUoiqgFGJGvua8hyNb8y9kxhRIJbmak_2f2Bf5gHZIJlDfD5haTNhbzHSqcTxO0NFXP-yooTZ2HSbrTQhv1Pw1PphtQLrzzzuaMMcwDj72dL16KDqTMz2ZY4eT5Ru5aE3I-P3cF-Tp7vZx9bvY3P9ar242hRWcDwXXwLR2TnEocXrteLKpKuWccVpwoVsJrVNKSaFFaRGZm_hyC07XHLQUC_Lz5D2k-DJiHprOZ4shmB7jmBumVS2VULWa0OUJtSnmnLBtDsl3Jr01DJpjvs3_-U4LV2f3uO3Q_cPPgU7AjzNgsjWhTaa3Pn9wgnEp61K8A7J2glo</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1859453595</pqid></control><display><type>article</type><title>Non-spectral interferences encountered with a commercially available high resolution ICP-mass spectrometer</title><source>SpringerLink Journals - AutoHoldings</source><creator>VANHAECKE, F ; RIONDATO, J ; MOENS, L ; DAMS, R</creator><creatorcontrib>VANHAECKE, F ; RIONDATO, J ; MOENS, L ; DAMS, R</creatorcontrib><description>Results of a systematic study concerning non-spectral interferences observed with a commercially available high resolution ICP-mass spectrometer are reported and compared to observations made with a quadrupole-based instrument. In general, matrix effects were observed to be to a large extent comparable for both instruments used. In all cases, the matrix-induced signal suppression or enhancement was seen to depend in a regular way on the mass number of the nuclides monitored. In most cases, the ionization potential of the nuclides has little or no influence on the extent of suppression or enhancement. For As, Se and Te, the introduction of 2.5% ethanol, 0.5 mol/l H(2)SO(4), or to a lesser extent 0.5 mol/l H(3)PO(4), leads to an exceptional increase in the signal intensity for both instruments. Registration of signal behaviour plots (signal intensity as a function of the nebulizer gas flow rate) in different matrices revealed that both the height of the plot and the optimum nebulizer gas flow rate are a function of the matrix composition. Finally, no indication was found that the acceleration of the extracted ions over 8000 V with the high resolution instrument would lead to an alleviation of space charge effects when compared to a quadrupole-based ICP-mass spectrometer.</description><identifier>ISSN: 0937-0633</identifier><identifier>ISSN: 1618-2642</identifier><identifier>EISSN: 1432-1130</identifier><identifier>EISSN: 1618-2650</identifier><identifier>DOI: 10.1007/s0021663550397</identifier><identifier>PMID: 15045425</identifier><language>eng</language><publisher>Berlin: Springer</publisher><subject>Analytical chemistry ; Chemistry ; Exact sciences and technology ; Spectrometric and optical methods</subject><ispartof>Analytical and bioanalytical chemistry, 1996-06, Vol.355 (3-4), p.397-400</ispartof><rights>1996 INIST-CNRS</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c322t-280188dd5206e1662192a775ddad83238f40fd55543836cee1d1886b0d8920843</citedby></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>309,310,314,780,784,789,790,23929,23930,25139,27923,27924</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&idt=3124496$$DView record in Pascal Francis$$Hfree_for_read</backlink><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/15045425$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>VANHAECKE, F</creatorcontrib><creatorcontrib>RIONDATO, J</creatorcontrib><creatorcontrib>MOENS, L</creatorcontrib><creatorcontrib>DAMS, R</creatorcontrib><title>Non-spectral interferences encountered with a commercially available high resolution ICP-mass spectrometer</title><title>Analytical and bioanalytical chemistry</title><addtitle>Anal Bioanal Chem</addtitle><description>Results of a systematic study concerning non-spectral interferences observed with a commercially available high resolution ICP-mass spectrometer are reported and compared to observations made with a quadrupole-based instrument. In general, matrix effects were observed to be to a large extent comparable for both instruments used. In all cases, the matrix-induced signal suppression or enhancement was seen to depend in a regular way on the mass number of the nuclides monitored. In most cases, the ionization potential of the nuclides has little or no influence on the extent of suppression or enhancement. For As, Se and Te, the introduction of 2.5% ethanol, 0.5 mol/l H(2)SO(4), or to a lesser extent 0.5 mol/l H(3)PO(4), leads to an exceptional increase in the signal intensity for both instruments. Registration of signal behaviour plots (signal intensity as a function of the nebulizer gas flow rate) in different matrices revealed that both the height of the plot and the optimum nebulizer gas flow rate are a function of the matrix composition. Finally, no indication was found that the acceleration of the extracted ions over 8000 V with the high resolution instrument would lead to an alleviation of space charge effects when compared to a quadrupole-based ICP-mass spectrometer.</description><subject>Analytical chemistry</subject><subject>Chemistry</subject><subject>Exact sciences and technology</subject><subject>Spectrometric and optical methods</subject><issn>0937-0633</issn><issn>1618-2642</issn><issn>1432-1130</issn><issn>1618-2650</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>1996</creationdate><recordtype>article</recordtype><recordid>eNpVkD1v2zAQhomiQe26XTMGHDp0kXP8kqgxMJLWgNFkSGaBJk8xDUp0SKlB_n1k2EjR5Q44PPfg7iXkksGSAVTXGYCzshRKgairT2TOpOAFYwI-kznUoiqgFGJGvua8hyNb8y9kxhRIJbmak_2f2Bf5gHZIJlDfD5haTNhbzHSqcTxO0NFXP-yooTZ2HSbrTQhv1Pw1PphtQLrzzzuaMMcwDj72dL16KDqTMz2ZY4eT5Ru5aE3I-P3cF-Tp7vZx9bvY3P9ar242hRWcDwXXwLR2TnEocXrteLKpKuWccVpwoVsJrVNKSaFFaRGZm_hyC07XHLQUC_Lz5D2k-DJiHprOZ4shmB7jmBumVS2VULWa0OUJtSnmnLBtDsl3Jr01DJpjvs3_-U4LV2f3uO3Q_cPPgU7AjzNgsjWhTaa3Pn9wgnEp61K8A7J2glo</recordid><startdate>19960601</startdate><enddate>19960601</enddate><creator>VANHAECKE, F</creator><creator>RIONDATO, J</creator><creator>MOENS, L</creator><creator>DAMS, R</creator><general>Springer</general><scope>IQODW</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>19960601</creationdate><title>Non-spectral interferences encountered with a commercially available high resolution ICP-mass spectrometer</title><author>VANHAECKE, F ; RIONDATO, J ; MOENS, L ; DAMS, R</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c322t-280188dd5206e1662192a775ddad83238f40fd55543836cee1d1886b0d8920843</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>1996</creationdate><topic>Analytical chemistry</topic><topic>Chemistry</topic><topic>Exact sciences and technology</topic><topic>Spectrometric and optical methods</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>VANHAECKE, F</creatorcontrib><creatorcontrib>RIONDATO, J</creatorcontrib><creatorcontrib>MOENS, L</creatorcontrib><creatorcontrib>DAMS, R</creatorcontrib><collection>Pascal-Francis</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Analytical and bioanalytical chemistry</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>VANHAECKE, F</au><au>RIONDATO, J</au><au>MOENS, L</au><au>DAMS, R</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Non-spectral interferences encountered with a commercially available high resolution ICP-mass spectrometer</atitle><jtitle>Analytical and bioanalytical chemistry</jtitle><addtitle>Anal Bioanal Chem</addtitle><date>1996-06-01</date><risdate>1996</risdate><volume>355</volume><issue>3-4</issue><spage>397</spage><epage>400</epage><pages>397-400</pages><issn>0937-0633</issn><issn>1618-2642</issn><eissn>1432-1130</eissn><eissn>1618-2650</eissn><abstract>Results of a systematic study concerning non-spectral interferences observed with a commercially available high resolution ICP-mass spectrometer are reported and compared to observations made with a quadrupole-based instrument. In general, matrix effects were observed to be to a large extent comparable for both instruments used. In all cases, the matrix-induced signal suppression or enhancement was seen to depend in a regular way on the mass number of the nuclides monitored. In most cases, the ionization potential of the nuclides has little or no influence on the extent of suppression or enhancement. For As, Se and Te, the introduction of 2.5% ethanol, 0.5 mol/l H(2)SO(4), or to a lesser extent 0.5 mol/l H(3)PO(4), leads to an exceptional increase in the signal intensity for both instruments. Registration of signal behaviour plots (signal intensity as a function of the nebulizer gas flow rate) in different matrices revealed that both the height of the plot and the optimum nebulizer gas flow rate are a function of the matrix composition. Finally, no indication was found that the acceleration of the extracted ions over 8000 V with the high resolution instrument would lead to an alleviation of space charge effects when compared to a quadrupole-based ICP-mass spectrometer.</abstract><cop>Berlin</cop><pub>Springer</pub><pmid>15045425</pmid><doi>10.1007/s0021663550397</doi><tpages>4</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0937-0633 |
ispartof | Analytical and bioanalytical chemistry, 1996-06, Vol.355 (3-4), p.397-400 |
issn | 0937-0633 1618-2642 1432-1130 1618-2650 |
language | eng |
recordid | cdi_proquest_miscellaneous_1859453595 |
source | SpringerLink Journals - AutoHoldings |
subjects | Analytical chemistry Chemistry Exact sciences and technology Spectrometric and optical methods |
title | Non-spectral interferences encountered with a commercially available high resolution ICP-mass spectrometer |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-11T11%3A19%3A40IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Non-spectral%20interferences%20encountered%20with%20a%20commercially%20available%20high%20resolution%20ICP-mass%20spectrometer&rft.jtitle=Analytical%20and%20bioanalytical%20chemistry&rft.au=VANHAECKE,%20F&rft.date=1996-06-01&rft.volume=355&rft.issue=3-4&rft.spage=397&rft.epage=400&rft.pages=397-400&rft.issn=0937-0633&rft.eissn=1432-1130&rft_id=info:doi/10.1007/s0021663550397&rft_dat=%3Cproquest_cross%3E1859453595%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=1859453595&rft_id=info:pmid/15045425&rfr_iscdi=true |