Simulation study of a novel target oriented SPECT design using a variable pinhole collimator
Purpose In the past decade, demands for organ specific (target oriented) single‐photon emission computed tomography (SPECT) is increasing, and several groups have conducted studies on developing clinical dedicated SPECT with pinhole collimator to improve the spatial resolution. However, acceptance a...
Gespeichert in:
Veröffentlicht in: | Medical physics (Lancaster) 2017-02, Vol.44 (2), p.470-478 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 478 |
---|---|
container_issue | 2 |
container_start_page | 470 |
container_title | Medical physics (Lancaster) |
container_volume | 44 |
creator | Bae, Seungbin Chun, Jaehee Cha, Hyemi Yeom, Jung Yeol Lee, Kisung Lee, Hakjae |
description | Purpose
In the past decade, demands for organ specific (target oriented) single‐photon emission computed tomography (SPECT) is increasing, and several groups have conducted studies on developing clinical dedicated SPECT with pinhole collimator to improve the spatial resolution. However, acceptance angle of the collimator cannot be adjusted to fit the different ROIs of target objects because the shape of pinhole could not be changed, and the magnifying factor cannot be maximized as the collimator‐to‐detector distance is fixed. Furthermore, those dedicated pinhole SPECTs are typically made for a single purpose and therefore possess a drawback in that it cannot be utilized for any other purpose. In this study, we propose a novel SPECT system using variable pinhole collimator (VP SPECT) whose parameters are flexible.
Methods
The proposed variable pinhole collimator is modeled on conventional pinhole by piling several tungsten layers of different apertures. Depending on the combination of the holes in each layer, a variety of hole diameters and acceptance angles of the pinhole can be made. In addition, VP SPECT system allows attaching the collimator to the object as close as possible to maximize the sensitivity and adjust the distance of the pinhole from the scintillation detector to optimize the system resolution for each rotation angle, automatically. For quantitative measurement, we compared the sensitivity and spatial resolution of VP SPECT with those of conventional pinhole SPECT. To determine the possibility of the clinical and preclinical use of proposed system, a digital mouse whole‐body (MOBY) phantom is used for simulating the live mouse model.
Results
The result of simulation using ultra‐micro hot spot phantom shows that the sensitivity of the proposed VP SPECT system is about 297% of that of the conventional system. While hot rods of diameter 0.6 mm can be distinguished in the image with the proposed VP SPECT system, 1.2‐mm hot rods are barely discernible in the conventional pinhole SPECT image. According to the result of MOBY phantom simulation, heart walls separated by 3 mm were not distinguished in conventional pinhole SPECT images, but were clearly discerned in VP SPECT images.
Conclusions
In this study, we designed a novel pinhole collimator for SPECT and presented preliminary results of target oriented imaging with a simulation study. Currently, we are pursuing strategies to realize the proposed system, with the goal to apply the technology in |
doi_str_mv | 10.1002/mp.12075 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_1854107205</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>1854107205</sourcerecordid><originalsourceid>FETCH-LOGICAL-c3215-5a9fde6be1395ddbb9d377e0a0a7138aa6b6412f0aedbe0162809eb7ba59d3163</originalsourceid><addsrcrecordid>eNp1kF1LwzAYRoMobk7BXyC59KbzTdq066WM-QETB5t3QkmWtzOSNjVpJ_v3VufHlVfPzeHwcAg5ZzBmAPyqasaMQyYOyJAnWRwlHPJDMgTIk4gnIAbkJIRXAEhjAcdkwCcQ8xySIXlemqqzsjWupqHt9I66kkpauy1a2kq_wZY6b7BuUdPlYjZdUY3BbGraBVNvenQrvZHKIm1M_eL6XTtrTSVb50_JUSltwLPvHZGnm9lqehfNH2_vp9fzaB1zJiIh81JjqpDFudBaqVzHWYYgQWYsnkiZqjRhvASJWiGwtL-fo8qUFD3J0nhELvfexru3DkNbVCas0VpZo-tCwSYiYZBxEH_o2rsQPJZF4_uzflcwKD5bFlVTfLXs0Ytva6cq1L_gT7weiPbAu7G4-1dUPCz2wg8oAH0f</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1854107205</pqid></control><display><type>article</type><title>Simulation study of a novel target oriented SPECT design using a variable pinhole collimator</title><source>MEDLINE</source><source>Wiley Online Library Journals Frontfile Complete</source><source>Alma/SFX Local Collection</source><creator>Bae, Seungbin ; Chun, Jaehee ; Cha, Hyemi ; Yeom, Jung Yeol ; Lee, Kisung ; Lee, Hakjae</creator><creatorcontrib>Bae, Seungbin ; Chun, Jaehee ; Cha, Hyemi ; Yeom, Jung Yeol ; Lee, Kisung ; Lee, Hakjae</creatorcontrib><description>Purpose
In the past decade, demands for organ specific (target oriented) single‐photon emission computed tomography (SPECT) is increasing, and several groups have conducted studies on developing clinical dedicated SPECT with pinhole collimator to improve the spatial resolution. However, acceptance angle of the collimator cannot be adjusted to fit the different ROIs of target objects because the shape of pinhole could not be changed, and the magnifying factor cannot be maximized as the collimator‐to‐detector distance is fixed. Furthermore, those dedicated pinhole SPECTs are typically made for a single purpose and therefore possess a drawback in that it cannot be utilized for any other purpose. In this study, we propose a novel SPECT system using variable pinhole collimator (VP SPECT) whose parameters are flexible.
Methods
The proposed variable pinhole collimator is modeled on conventional pinhole by piling several tungsten layers of different apertures. Depending on the combination of the holes in each layer, a variety of hole diameters and acceptance angles of the pinhole can be made. In addition, VP SPECT system allows attaching the collimator to the object as close as possible to maximize the sensitivity and adjust the distance of the pinhole from the scintillation detector to optimize the system resolution for each rotation angle, automatically. For quantitative measurement, we compared the sensitivity and spatial resolution of VP SPECT with those of conventional pinhole SPECT. To determine the possibility of the clinical and preclinical use of proposed system, a digital mouse whole‐body (MOBY) phantom is used for simulating the live mouse model.
Results
The result of simulation using ultra‐micro hot spot phantom shows that the sensitivity of the proposed VP SPECT system is about 297% of that of the conventional system. While hot rods of diameter 0.6 mm can be distinguished in the image with the proposed VP SPECT system, 1.2‐mm hot rods are barely discernible in the conventional pinhole SPECT image. According to the result of MOBY phantom simulation, heart walls separated by 3 mm were not distinguished in conventional pinhole SPECT images, but were clearly discerned in VP SPECT images.
Conclusions
In this study, we designed a novel pinhole collimator for SPECT and presented preliminary results of target oriented imaging with a simulation study. Currently, we are pursuing strategies to realize the proposed system, with the goal to apply the technology into a high‐sensitivity and high‐resolution preclinical SPECT. Should VP SPECT be applied to the clinical setting, we anticipate a high‐sensitivity, high‐resolution system for applications such as heart dedicated SPECT or related fields.</description><identifier>ISSN: 0094-2405</identifier><identifier>EISSN: 2473-4209</identifier><identifier>DOI: 10.1002/mp.12075</identifier><identifier>PMID: 28032904</identifier><language>eng</language><publisher>United States</publisher><subject>Animals ; Computer Simulation ; Equipment Design ; Feasibility Studies ; Heart - diagnostic imaging ; Mice ; Models, Anatomic ; Phantoms, Imaging ; pinhole collimator ; simulation ; single‐photon emission computed tomography ; Tomography, Emission-Computed, Single-Photon - instrumentation ; Tomography, Emission-Computed, Single-Photon - methods ; Tungsten</subject><ispartof>Medical physics (Lancaster), 2017-02, Vol.44 (2), p.470-478</ispartof><rights>2016 American Association of Physicists in Medicine</rights><rights>2016 American Association of Physicists in Medicine.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c3215-5a9fde6be1395ddbb9d377e0a0a7138aa6b6412f0aedbe0162809eb7ba59d3163</citedby><cites>FETCH-LOGICAL-c3215-5a9fde6be1395ddbb9d377e0a0a7138aa6b6412f0aedbe0162809eb7ba59d3163</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://onlinelibrary.wiley.com/doi/pdf/10.1002%2Fmp.12075$$EPDF$$P50$$Gwiley$$H</linktopdf><linktohtml>$$Uhttps://onlinelibrary.wiley.com/doi/full/10.1002%2Fmp.12075$$EHTML$$P50$$Gwiley$$H</linktohtml><link.rule.ids>314,776,780,1411,27901,27902,45550,45551</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/28032904$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Bae, Seungbin</creatorcontrib><creatorcontrib>Chun, Jaehee</creatorcontrib><creatorcontrib>Cha, Hyemi</creatorcontrib><creatorcontrib>Yeom, Jung Yeol</creatorcontrib><creatorcontrib>Lee, Kisung</creatorcontrib><creatorcontrib>Lee, Hakjae</creatorcontrib><title>Simulation study of a novel target oriented SPECT design using a variable pinhole collimator</title><title>Medical physics (Lancaster)</title><addtitle>Med Phys</addtitle><description>Purpose
In the past decade, demands for organ specific (target oriented) single‐photon emission computed tomography (SPECT) is increasing, and several groups have conducted studies on developing clinical dedicated SPECT with pinhole collimator to improve the spatial resolution. However, acceptance angle of the collimator cannot be adjusted to fit the different ROIs of target objects because the shape of pinhole could not be changed, and the magnifying factor cannot be maximized as the collimator‐to‐detector distance is fixed. Furthermore, those dedicated pinhole SPECTs are typically made for a single purpose and therefore possess a drawback in that it cannot be utilized for any other purpose. In this study, we propose a novel SPECT system using variable pinhole collimator (VP SPECT) whose parameters are flexible.
Methods
The proposed variable pinhole collimator is modeled on conventional pinhole by piling several tungsten layers of different apertures. Depending on the combination of the holes in each layer, a variety of hole diameters and acceptance angles of the pinhole can be made. In addition, VP SPECT system allows attaching the collimator to the object as close as possible to maximize the sensitivity and adjust the distance of the pinhole from the scintillation detector to optimize the system resolution for each rotation angle, automatically. For quantitative measurement, we compared the sensitivity and spatial resolution of VP SPECT with those of conventional pinhole SPECT. To determine the possibility of the clinical and preclinical use of proposed system, a digital mouse whole‐body (MOBY) phantom is used for simulating the live mouse model.
Results
The result of simulation using ultra‐micro hot spot phantom shows that the sensitivity of the proposed VP SPECT system is about 297% of that of the conventional system. While hot rods of diameter 0.6 mm can be distinguished in the image with the proposed VP SPECT system, 1.2‐mm hot rods are barely discernible in the conventional pinhole SPECT image. According to the result of MOBY phantom simulation, heart walls separated by 3 mm were not distinguished in conventional pinhole SPECT images, but were clearly discerned in VP SPECT images.
Conclusions
In this study, we designed a novel pinhole collimator for SPECT and presented preliminary results of target oriented imaging with a simulation study. Currently, we are pursuing strategies to realize the proposed system, with the goal to apply the technology into a high‐sensitivity and high‐resolution preclinical SPECT. Should VP SPECT be applied to the clinical setting, we anticipate a high‐sensitivity, high‐resolution system for applications such as heart dedicated SPECT or related fields.</description><subject>Animals</subject><subject>Computer Simulation</subject><subject>Equipment Design</subject><subject>Feasibility Studies</subject><subject>Heart - diagnostic imaging</subject><subject>Mice</subject><subject>Models, Anatomic</subject><subject>Phantoms, Imaging</subject><subject>pinhole collimator</subject><subject>simulation</subject><subject>single‐photon emission computed tomography</subject><subject>Tomography, Emission-Computed, Single-Photon - instrumentation</subject><subject>Tomography, Emission-Computed, Single-Photon - methods</subject><subject>Tungsten</subject><issn>0094-2405</issn><issn>2473-4209</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2017</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp1kF1LwzAYRoMobk7BXyC59KbzTdq066WM-QETB5t3QkmWtzOSNjVpJ_v3VufHlVfPzeHwcAg5ZzBmAPyqasaMQyYOyJAnWRwlHPJDMgTIk4gnIAbkJIRXAEhjAcdkwCcQ8xySIXlemqqzsjWupqHt9I66kkpauy1a2kq_wZY6b7BuUdPlYjZdUY3BbGraBVNvenQrvZHKIm1M_eL6XTtrTSVb50_JUSltwLPvHZGnm9lqehfNH2_vp9fzaB1zJiIh81JjqpDFudBaqVzHWYYgQWYsnkiZqjRhvASJWiGwtL-fo8qUFD3J0nhELvfexru3DkNbVCas0VpZo-tCwSYiYZBxEH_o2rsQPJZF4_uzflcwKD5bFlVTfLXs0Ytva6cq1L_gT7weiPbAu7G4-1dUPCz2wg8oAH0f</recordid><startdate>201702</startdate><enddate>201702</enddate><creator>Bae, Seungbin</creator><creator>Chun, Jaehee</creator><creator>Cha, Hyemi</creator><creator>Yeom, Jung Yeol</creator><creator>Lee, Kisung</creator><creator>Lee, Hakjae</creator><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>201702</creationdate><title>Simulation study of a novel target oriented SPECT design using a variable pinhole collimator</title><author>Bae, Seungbin ; Chun, Jaehee ; Cha, Hyemi ; Yeom, Jung Yeol ; Lee, Kisung ; Lee, Hakjae</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c3215-5a9fde6be1395ddbb9d377e0a0a7138aa6b6412f0aedbe0162809eb7ba59d3163</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2017</creationdate><topic>Animals</topic><topic>Computer Simulation</topic><topic>Equipment Design</topic><topic>Feasibility Studies</topic><topic>Heart - diagnostic imaging</topic><topic>Mice</topic><topic>Models, Anatomic</topic><topic>Phantoms, Imaging</topic><topic>pinhole collimator</topic><topic>simulation</topic><topic>single‐photon emission computed tomography</topic><topic>Tomography, Emission-Computed, Single-Photon - instrumentation</topic><topic>Tomography, Emission-Computed, Single-Photon - methods</topic><topic>Tungsten</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Bae, Seungbin</creatorcontrib><creatorcontrib>Chun, Jaehee</creatorcontrib><creatorcontrib>Cha, Hyemi</creatorcontrib><creatorcontrib>Yeom, Jung Yeol</creatorcontrib><creatorcontrib>Lee, Kisung</creatorcontrib><creatorcontrib>Lee, Hakjae</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Medical physics (Lancaster)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Bae, Seungbin</au><au>Chun, Jaehee</au><au>Cha, Hyemi</au><au>Yeom, Jung Yeol</au><au>Lee, Kisung</au><au>Lee, Hakjae</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Simulation study of a novel target oriented SPECT design using a variable pinhole collimator</atitle><jtitle>Medical physics (Lancaster)</jtitle><addtitle>Med Phys</addtitle><date>2017-02</date><risdate>2017</risdate><volume>44</volume><issue>2</issue><spage>470</spage><epage>478</epage><pages>470-478</pages><issn>0094-2405</issn><eissn>2473-4209</eissn><abstract>Purpose
In the past decade, demands for organ specific (target oriented) single‐photon emission computed tomography (SPECT) is increasing, and several groups have conducted studies on developing clinical dedicated SPECT with pinhole collimator to improve the spatial resolution. However, acceptance angle of the collimator cannot be adjusted to fit the different ROIs of target objects because the shape of pinhole could not be changed, and the magnifying factor cannot be maximized as the collimator‐to‐detector distance is fixed. Furthermore, those dedicated pinhole SPECTs are typically made for a single purpose and therefore possess a drawback in that it cannot be utilized for any other purpose. In this study, we propose a novel SPECT system using variable pinhole collimator (VP SPECT) whose parameters are flexible.
Methods
The proposed variable pinhole collimator is modeled on conventional pinhole by piling several tungsten layers of different apertures. Depending on the combination of the holes in each layer, a variety of hole diameters and acceptance angles of the pinhole can be made. In addition, VP SPECT system allows attaching the collimator to the object as close as possible to maximize the sensitivity and adjust the distance of the pinhole from the scintillation detector to optimize the system resolution for each rotation angle, automatically. For quantitative measurement, we compared the sensitivity and spatial resolution of VP SPECT with those of conventional pinhole SPECT. To determine the possibility of the clinical and preclinical use of proposed system, a digital mouse whole‐body (MOBY) phantom is used for simulating the live mouse model.
Results
The result of simulation using ultra‐micro hot spot phantom shows that the sensitivity of the proposed VP SPECT system is about 297% of that of the conventional system. While hot rods of diameter 0.6 mm can be distinguished in the image with the proposed VP SPECT system, 1.2‐mm hot rods are barely discernible in the conventional pinhole SPECT image. According to the result of MOBY phantom simulation, heart walls separated by 3 mm were not distinguished in conventional pinhole SPECT images, but were clearly discerned in VP SPECT images.
Conclusions
In this study, we designed a novel pinhole collimator for SPECT and presented preliminary results of target oriented imaging with a simulation study. Currently, we are pursuing strategies to realize the proposed system, with the goal to apply the technology into a high‐sensitivity and high‐resolution preclinical SPECT. Should VP SPECT be applied to the clinical setting, we anticipate a high‐sensitivity, high‐resolution system for applications such as heart dedicated SPECT or related fields.</abstract><cop>United States</cop><pmid>28032904</pmid><doi>10.1002/mp.12075</doi><tpages>9</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0094-2405 |
ispartof | Medical physics (Lancaster), 2017-02, Vol.44 (2), p.470-478 |
issn | 0094-2405 2473-4209 |
language | eng |
recordid | cdi_proquest_miscellaneous_1854107205 |
source | MEDLINE; Wiley Online Library Journals Frontfile Complete; Alma/SFX Local Collection |
subjects | Animals Computer Simulation Equipment Design Feasibility Studies Heart - diagnostic imaging Mice Models, Anatomic Phantoms, Imaging pinhole collimator simulation single‐photon emission computed tomography Tomography, Emission-Computed, Single-Photon - instrumentation Tomography, Emission-Computed, Single-Photon - methods Tungsten |
title | Simulation study of a novel target oriented SPECT design using a variable pinhole collimator |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-16T09%3A38%3A02IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Simulation%20study%20of%20a%20novel%20target%20oriented%20SPECT%20design%20using%20a%20variable%20pinhole%20collimator&rft.jtitle=Medical%20physics%20(Lancaster)&rft.au=Bae,%20Seungbin&rft.date=2017-02&rft.volume=44&rft.issue=2&rft.spage=470&rft.epage=478&rft.pages=470-478&rft.issn=0094-2405&rft.eissn=2473-4209&rft_id=info:doi/10.1002/mp.12075&rft_dat=%3Cproquest_cross%3E1854107205%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=1854107205&rft_id=info:pmid/28032904&rfr_iscdi=true |