Comparison of Suture-Based Anchors and Traditional Bioabsorbable Anchors in Foot and Ankle Surgery

Abstract We compared the pullout strength of a suture-based anchor versus a bioabsorbable anchor in the distal fibula and calcaneus and evaluated the relationship between bone mineral density and peak load to failure. Eight paired cadaveric specimens underwent a modified Broström procedure and Achil...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of foot and ankle surgery 2017-01, Vol.56 (1), p.3-7
Hauptverfasser: Hembree, W. Chad, MD, Tsai, Michael A., BS, Parks, Brent G., MSc, Miller, Stuart D., MD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract We compared the pullout strength of a suture-based anchor versus a bioabsorbable anchor in the distal fibula and calcaneus and evaluated the relationship between bone mineral density and peak load to failure. Eight paired cadaveric specimens underwent a modified Broström procedure and Achilles tendon reattachment. The fibula and calcaneus in the paired specimens received either a suture-based anchor or a bioabsorbable suture anchor. The fibular and calcaneal specimens were loaded to failure, defined as a substantial decrease in the applied load or pullout from the bone. In the fibula, the peak load to failure was significantly greater with the suture-based versus the bioabsorbable anchors (133.3 ± 41.8 N versus 76.8 ± 35.3 N; p  = .002). No significant difference in load with 5 mm of displacement was found between the 2 groups. In the calcaneus, no difference in the peak load to failure was found between the 2 groups, and the peak load to failure with 5 mm of displacement was significantly lower with the suture-based than with the bioabsorbable anchors (52.2 ± 9.8 N versus 75.9 ± 12.4 N; p  = .003). Bone mineral density and peak load to failure were significantly correlated in the fibula with the suture-based anchor. An innovative suture-based anchor had a greater peak load to failure compared with a bioabsorbable anchor in the fibula. In the calcaneus, the load at 5 mm of displacement was significantly lower in the suture-based than in the bioabsorbable group. The correlation findings might indicate the need for a cortical bone shelf with the suture-based anchor. Suture-based anchors could be a viable alternative to bioabsorbable anchors for certain foot and ankle procedures.
ISSN:1067-2516
1542-2224
DOI:10.1053/j.jfas.2016.10.011