Towards malaria microscopy at the point-of-contact: an assessment of the diagnostic performance of the Newton Nm1 microscope in Uganda
Malaria microscopy in sub-Saharan Africa is often restricted by access to light microscopes. To address this gap, a novel portable inverted monocular microscope, the Newton Nm1, was designed and is now commercially available. Its diagnostic performance was assessed in a blinded-slide trial at ×1000...
Gespeichert in:
Veröffentlicht in: | Parasitology 2014-12, Vol.141 (14), p.1819-1825 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Malaria microscopy in sub-Saharan Africa is often restricted by access to light microscopes. To address this gap, a novel portable inverted monocular microscope, the Newton Nm1, was designed and is now commercially available. Its diagnostic performance was assessed in a blinded-slide trial at ×1000 (oil) of Giemsa-stained thick blood films against a conventional microscope as undertaken by four Ugandan Ministry of Health technicians. With the Newton Nm1, diagnostic performance was: sensitivity 93·5% (95% confidence interval (CI) 78·6–99·2%), specificity 100·0% (95% CI 82·4–100·0%), positive predictive value 100·0% (95% CI 88·1–100·0%) and negative predictive value 90·5% (95% CI 69·6–98·8%). Discordance was due to a systematic error underestimating parasitaemia by ~45%; when counting Plasmodium parasites against 200 white blood cells, blood films with low parasitaemia (i.e. |
---|---|
ISSN: | 0031-1820 1469-8161 |
DOI: | 10.1017/S0031182014000833 |