Assessment of novel digital and smartphone goniometers for measurement of canine stifle joint angles
OBJECTIVE To evaluate accuracy and reliability of 3 novel goniometers for measurement of canine stifle joint angles and compare the results with those obtained with a universal goniometer (UG). SAMPLE 8 pelvic limbs from 4 canine cadavers. PROCEDURES Each limb was secured to a wooden platform at 3 a...
Gespeichert in:
Veröffentlicht in: | American journal of veterinary research 2016-07, Vol.77 (7), p.749-755 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | OBJECTIVE To evaluate accuracy and reliability of 3 novel goniometers for measurement of canine stifle joint angles and compare the results with those obtained with a universal goniometer (UG). SAMPLE 8 pelvic limbs from 4 canine cadavers. PROCEDURES Each limb was secured to a wooden platform at 3 arbitrarily selected fixed stifle joint angles. Goniometry was performed with 2 smartphone-based applications (novel goniometers A and B), a digital goniometer (novel goniometer C), and a UG; 3 evaluators performed measurements in triplicate for each angle with each device. Results were compared with stifle joint angle measurements on radiographs (used as a gold standard). Accuracy was determined by calculation of bias and total error, coefficients of variation were calculated to estimate reliability, and strength of linear association between radiographic and goniometer measurements was assessed by calculation of correlation coefficients. RESULTS Mean coefficient of variation was lowest for the UG (4.88%), followed by novel goniometers B (7.37%), A (7.57%), and C (12.71%). Correlation with radiographic measurements was highest for the UG (r = 0.97), followed by novel goniometers B (0.93), A (0.90), and C (0.78). Constant bias was present for all devices except novel goniometer B. The UG and novel goniometer A had positive constant bias; novel goniometer C had negative constant bias. Total error at 50° and 100° angles was > 5% for all devices. CONCLUSIONS AND CLINICAL RELEVANCE None of the devices accurately represented radiographically measured stifle joint angles. Additional veterinary studies are indicated prior to the use of novel goniometers in dogs. |
---|---|
ISSN: | 0002-9645 1943-5681 |
DOI: | 10.2460/ajvr.77.7.749 |