Accuracy and user performance evaluation of a blood glucose monitoring system
The accuracy of the Contour(®) Plus (Bayer HealthCare LLC, Diabetes Care, Whippany, NJ) blood glucose monitoring system (BGMS) was evaluated in two separate studies. In the laboratory study, fingerstick samples from 100 subjects were tested in duplicate using three test strip lots and assessed per I...
Gespeichert in:
Veröffentlicht in: | Diabetes technology & therapeutics 2015-03, Vol.17 (3), p.152-158 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The accuracy of the Contour(®) Plus (Bayer HealthCare LLC, Diabetes Care, Whippany, NJ) blood glucose monitoring system (BGMS) was evaluated in two separate studies.
In the laboratory study, fingerstick samples from 100 subjects were tested in duplicate using three test strip lots and assessed per International Organization for Standardization (ISO) 15197:2003, Section 7 (≥95% of results within ±15 mg/dL or ±20% of the reference result for samples with glucose concentrations |
---|---|