Collection of holes in thick TlBr detectors at low temperature

A 3.5×3.5×4.6mm3 thick TlBr detector with pixellated Au/Cr anodes made by Radiation Monitoring Devices Inc. was studied. The detector has a planar cathode and nine anode pixels surrounded by a guard ring. The pixel pitch is 1.0mm. Digital pulse waveforms of preamplifier outputs were recorded using a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 2012-10, Vol.689, p.7-11
Hauptverfasser: Dönmez, Burçin, He, Zhong, Kim, Hadong, Cirignano, Leonard J., Shah, Kanai S.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A 3.5×3.5×4.6mm3 thick TlBr detector with pixellated Au/Cr anodes made by Radiation Monitoring Devices Inc. was studied. The detector has a planar cathode and nine anode pixels surrounded by a guard ring. The pixel pitch is 1.0mm. Digital pulse waveforms of preamplifier outputs were recorded using a multi-channel GaGe PCI digitizer board. Several experiments were carried out at −20°C, with the detector under bias for over a month. An energy resolution of 1.7% FWHM at 662keV was measured without any correction at −2400V bias. Holes generated at all depths can be collected by the cathode at −2400V bias which made depth correction using the cathode-to-anode ratio technique difficult since both charge carriers contribute to the signal. An energy resolution of 5.1% FWHM at 662keV was obtained from the best pixel electrode without depth correction at +1000V bias. In this positive bias case, the pixel electrode was actually collecting holes. A hole mobility-lifetime of 0.95×10−4cm2/V has been estimated from measurement data.
ISSN:0168-9002
1872-9576
DOI:10.1016/j.nima.2012.06.020