Verification of high dose rate super(192)Ir source position during brachytherapy treatment
A system for in vivo tracking of 1 Ci super(192)Ir source during brachytherapy treatment has been built using high resistivity silicon pad detectors as image sensors and knife-edge lead pinholes as collimators. The sensors consist of 256 pads arranged in 32 x8 grid with pad size [MathML equation] an...
Gespeichert in:
Veröffentlicht in: | Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 2010-05, Vol.617 (1-3), p.206-208 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A system for in vivo tracking of 1 Ci super(192)Ir source during brachytherapy treatment has been built using high resistivity silicon pad detectors as image sensors and knife-edge lead pinholes as collimators. The sensors consist of 256 pads arranged in 32 x8 grid with pad size [MathML equation] and 1 mm thickness. The sensors have two metal layers, enabling connection of readout electronics (VATAGP3_1 chips) at the edge of the detector. With source self-images obtained from a dual-pinhole system, location of the source can be reconstructed in three dimensions in real time, allowing on-line detection of deviations from planned treatment. The system was tested with 1 Ci super(192)Ir clinical source in air and plexi-glass phantom. The movements of the source could be tracked in a field of view of approximately [MathML equation] with absolute precision of about 5 mm. Positions of the source, relative to the first measured source position, could be mapped with precision of around 3 mm. |
---|---|
ISSN: | 0168-9002 |
DOI: | 10.1016/j.nima.2009.09.122 |