A solid state detector for intraoperative imaging

A beta detecting probe for intraoperative tumour imaging and localisation has been developed. The probe is based on a charge coupled device which detects the particles directly in its sensitive volume. The active area of the device is encapsulated within a carbon-fibre loaded plastic housing 75 /spl...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on nuclear science 1999-08, Vol.46 (4), p.1172-1176
Hauptverfasser: Visvikis, D., MacDonald, J.H.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A beta detecting probe for intraoperative tumour imaging and localisation has been developed. The probe is based on a charge coupled device which detects the particles directly in its sensitive volume. The active area of the device is encapsulated within a carbon-fibre loaded plastic housing 75 /spl mu/m. Images are displayed in real time on a computer screen with a readout frequency of 1 MHz. A study of the effect of dark current on image formation at body temperature has demonstrated a reduction in the signal of /spl sim/55%. The advantage of this device for intraoperative tumour imaging is its minimal sensitivity to the background gamma radiation. This was confirmed using a /sup 18/F source separated from the CCD with a 2 cm perspex block to absorb the positrons. The linearity of the probe was also tested over an activity range of 1-100 /spl mu/Ci. Although the intrinsic resolution of the CCD was measured to be equal to 66 /spl mu/m, the image resolution of the device was
ISSN:0018-9499
1558-1578