Nanometer differential mobility analyzer (Nano-DMA): Design and numerical modeling
The development of a nanometer differential mobility analyzer (Nano-DMA) optimized for the particle size range from 3 to 50 nm based on cylindrical configuration was described. The design was facilitated by numerical model which incorporates the flow field, electric field, and aerosol transport calc...
Gespeichert in:
Veröffentlicht in: | Journal of aerosol science 1996-01, Vol.27, p.no. Sul 1-no. Sul 1 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The development of a nanometer differential mobility analyzer (Nano-DMA) optimized for the particle size range from 3 to 50 nm based on cylindrical configuration was described. The design was facilitated by numerical model which incorporates the flow field, electric field, and aerosol transport calculations inside DMA. The nano-DMA was designed to achieve high resolution measurement of nanometer aerosol down to 3 nm. Transfer function of the nano-DMA was obtained using the numerical model. Result shows a significant improvement in performance of nano-DMA. |
---|---|