Shear Wave Elastography Assessment of Carotid Plaque Stiffness: In Vitro Reproducibility Study
Abstract This study assessed inter- and intra-observer reproducibility of shear wave elastography (SWE) measurements in vessel phantoms simulating soft and hard carotid plaque under steady and pulsatile flow conditions. Supersonic SWE was used to acquire cine-loop data and quantify Young’s modulus i...
Gespeichert in:
Veröffentlicht in: | Ultrasound in medicine & biology 2014, Vol.40 (1), p.200-209 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract This study assessed inter- and intra-observer reproducibility of shear wave elastography (SWE) measurements in vessel phantoms simulating soft and hard carotid plaque under steady and pulsatile flow conditions. Supersonic SWE was used to acquire cine-loop data and quantify Young’s modulus in cryogel vessel phantoms. Data were acquired by two observers, each performing three repeat measurements. Mean Young’s modulus was quantified within 2-mm regions of interest averaged across five frames and, depending on vessel model and observer, ranged from 28 to 240 kPa. The mean inter-frame coefficient of variation (CV) was 0.13 (range: 0.07–0.18) for observer 1 and 0.14 (range: 0.12–0.16) for observer 2, with mean intra-class correlation coefficients (ICCs) of 0.84 and 0.83, respectively. The mean inter-operator CV was 0.13 (range: 0.08–0.20), with a mean ICC of 0.76 (range: 0.69-0.82). Our findings indicate that SWE can quantify Young’s modulus of carotid plaque phantoms with good reproducibility, even in the presence of pulsatile flow. |
---|---|
ISSN: | 0301-5629 1879-291X |
DOI: | 10.1016/j.ultrasmedbio.2013.09.014 |