Feasibility of an intracranial EEG-fMRI protocol at 3 T: Risk assessment and image quality

Integrating intracranial EEG (iEEG) with functional MRI (iEEG-fMRI) may help elucidate mechanisms underlying the generation of seizures. However, the introduction of iEEG electrodes in the MR environment has inherent risk and data quality implications that require consideration prior to clinical use...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:NeuroImage (Orlando, Fla.) Fla.), 2012-11, Vol.63 (3), p.1237-1248
Hauptverfasser: Boucousis, Shannon, Beers, Craig, Cunningham, Cameron, Gaxiola-Valdez, Ismael, Pittman, Daniel, Goodyear, Bradley, Federico, Paolo
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 1248
container_issue 3
container_start_page 1237
container_title NeuroImage (Orlando, Fla.)
container_volume 63
creator Boucousis, Shannon
Beers, Craig
Cunningham, Cameron
Gaxiola-Valdez, Ismael
Pittman, Daniel
Goodyear, Bradley
Federico, Paolo
description Integrating intracranial EEG (iEEG) with functional MRI (iEEG-fMRI) may help elucidate mechanisms underlying the generation of seizures. However, the introduction of iEEG electrodes in the MR environment has inherent risk and data quality implications that require consideration prior to clinical use. Previous studies of subdural and depth electrodes have confirmed low risk under specific circumstances at 1.5 T and 3 T. However, no studies have assessed risk and image quality related to the feasibility of a full iEEGafMRI protocol. To this end, commercially available platinum subdural grid/strip electrodes (4 x 5 grid or 1 x 8 strip) and 4 or 6-contact depth electrodes were secured to the surface of a custom-made phantom mimicking the conductivity of the human brain. Electrode displacement, temperature increase of electrodes and surrounding phantom material, and voltage fluctuations in electrode contacts were measured in a GE Discovery MR750 3 T MR scanner during a variety of imaging sequences, typical of an iEEG-fMRI protocol. An electrode grid was also used to quantify the spatial extent of susceptibility artifact. The spatial extent of susceptibility artifact in the presence of an electrode was also assessed for typical imaging parameters that maximize BOLD sensitivity at 3 T (TR = 1500 ms; TE = 30 ms; slice thickness = 4 mm; matrix = 64 x 64; field-of-view = 24 cm). Under standard conditions, all electrodes exhibited no measurable displacement and no clinically significant temperature increase (< 1 degree C) during scans employed in a typical iEEG-fMRI experiment, including 60 min of continuous fMRI. However, high SAR sequences, such as fast spin-echo (FSE), produced significant heating in almost all scenarios (> 2.0 degree C) that in some cases exceeded 10 degree C. Induced voltages in the frequency range that could elicit neuronal stimulation (< 10 kHz) were well below the threshold of 100 mV. fMRI signal intensity was significantly reduced within 20 mm of the electrodes for the imaging parameters used in this study. Thus, for the conditions tested, a full iEEG-fMRI protocol poses a low risk at 3 T; however, fMRI sensitivity may be reduced immediately adjacent to the electrodes. In addition, high SAR sequences must be avoided.
doi_str_mv 10.1016/j.neuroimage.2012.08.008
format Article
fullrecord <record><control><sourceid>proquest</sourceid><recordid>TN_cdi_proquest_miscellaneous_1505350754</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>1505350754</sourcerecordid><originalsourceid>FETCH-LOGICAL-p118t-bc9755248d42f789a7f9219813548038e80c08c094c922f09c5e5c350cff47c93</originalsourceid><addsrcrecordid>eNotj7tOAzEURF2ARAj8wy1pdrnXD9amQ9EmRApCikJDEzmOjRz2kay9BX9PeFRTzTkzjAFhSUgP94ey8-PQx9Z--JIj8RJ1iagv2IRQiUITmSt2ndIBEQ1JPWHvc29T3MUm5i_oA9gOYpcH6wbbRdtAXS-K8LJewnHoc-_6BmwGAZtHWMf0CTYln1Lru3xu7uHXDKfR_uBu2GWwTfK3_zllb_N6M3suVq-L5expVRyJdC52zlRKcan3kodKG1sFw8loEkpqFNprdKgdGukM5wGNU145odCFICtnxJTd_XHPE0-jT3nbxuR809jO92Pakjp_V1gpKb4BwX5WUw</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1505350754</pqid></control><display><type>article</type><title>Feasibility of an intracranial EEG-fMRI protocol at 3 T: Risk assessment and image quality</title><source>Access via ScienceDirect (Elsevier)</source><source>ProQuest Central UK/Ireland</source><creator>Boucousis, Shannon ; Beers, Craig ; Cunningham, Cameron ; Gaxiola-Valdez, Ismael ; Pittman, Daniel ; Goodyear, Bradley ; Federico, Paolo</creator><creatorcontrib>Boucousis, Shannon ; Beers, Craig ; Cunningham, Cameron ; Gaxiola-Valdez, Ismael ; Pittman, Daniel ; Goodyear, Bradley ; Federico, Paolo</creatorcontrib><description>Integrating intracranial EEG (iEEG) with functional MRI (iEEG-fMRI) may help elucidate mechanisms underlying the generation of seizures. However, the introduction of iEEG electrodes in the MR environment has inherent risk and data quality implications that require consideration prior to clinical use. Previous studies of subdural and depth electrodes have confirmed low risk under specific circumstances at 1.5 T and 3 T. However, no studies have assessed risk and image quality related to the feasibility of a full iEEGafMRI protocol. To this end, commercially available platinum subdural grid/strip electrodes (4 x 5 grid or 1 x 8 strip) and 4 or 6-contact depth electrodes were secured to the surface of a custom-made phantom mimicking the conductivity of the human brain. Electrode displacement, temperature increase of electrodes and surrounding phantom material, and voltage fluctuations in electrode contacts were measured in a GE Discovery MR750 3 T MR scanner during a variety of imaging sequences, typical of an iEEG-fMRI protocol. An electrode grid was also used to quantify the spatial extent of susceptibility artifact. The spatial extent of susceptibility artifact in the presence of an electrode was also assessed for typical imaging parameters that maximize BOLD sensitivity at 3 T (TR = 1500 ms; TE = 30 ms; slice thickness = 4 mm; matrix = 64 x 64; field-of-view = 24 cm). Under standard conditions, all electrodes exhibited no measurable displacement and no clinically significant temperature increase (&lt; 1 degree C) during scans employed in a typical iEEG-fMRI experiment, including 60 min of continuous fMRI. However, high SAR sequences, such as fast spin-echo (FSE), produced significant heating in almost all scenarios (&gt; 2.0 degree C) that in some cases exceeded 10 degree C. Induced voltages in the frequency range that could elicit neuronal stimulation (&lt; 10 kHz) were well below the threshold of 100 mV. fMRI signal intensity was significantly reduced within 20 mm of the electrodes for the imaging parameters used in this study. Thus, for the conditions tested, a full iEEG-fMRI protocol poses a low risk at 3 T; however, fMRI sensitivity may be reduced immediately adjacent to the electrodes. In addition, high SAR sequences must be avoided.</description><identifier>ISSN: 1053-8119</identifier><identifier>DOI: 10.1016/j.neuroimage.2012.08.008</identifier><language>eng</language><ispartof>NeuroImage (Orlando, Fla.), 2012-11, Vol.63 (3), p.1237-1248</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>315,782,786,27931,27932,64394</link.rule.ids></links><search><creatorcontrib>Boucousis, Shannon</creatorcontrib><creatorcontrib>Beers, Craig</creatorcontrib><creatorcontrib>Cunningham, Cameron</creatorcontrib><creatorcontrib>Gaxiola-Valdez, Ismael</creatorcontrib><creatorcontrib>Pittman, Daniel</creatorcontrib><creatorcontrib>Goodyear, Bradley</creatorcontrib><creatorcontrib>Federico, Paolo</creatorcontrib><title>Feasibility of an intracranial EEG-fMRI protocol at 3 T: Risk assessment and image quality</title><title>NeuroImage (Orlando, Fla.)</title><description>Integrating intracranial EEG (iEEG) with functional MRI (iEEG-fMRI) may help elucidate mechanisms underlying the generation of seizures. However, the introduction of iEEG electrodes in the MR environment has inherent risk and data quality implications that require consideration prior to clinical use. Previous studies of subdural and depth electrodes have confirmed low risk under specific circumstances at 1.5 T and 3 T. However, no studies have assessed risk and image quality related to the feasibility of a full iEEGafMRI protocol. To this end, commercially available platinum subdural grid/strip electrodes (4 x 5 grid or 1 x 8 strip) and 4 or 6-contact depth electrodes were secured to the surface of a custom-made phantom mimicking the conductivity of the human brain. Electrode displacement, temperature increase of electrodes and surrounding phantom material, and voltage fluctuations in electrode contacts were measured in a GE Discovery MR750 3 T MR scanner during a variety of imaging sequences, typical of an iEEG-fMRI protocol. An electrode grid was also used to quantify the spatial extent of susceptibility artifact. The spatial extent of susceptibility artifact in the presence of an electrode was also assessed for typical imaging parameters that maximize BOLD sensitivity at 3 T (TR = 1500 ms; TE = 30 ms; slice thickness = 4 mm; matrix = 64 x 64; field-of-view = 24 cm). Under standard conditions, all electrodes exhibited no measurable displacement and no clinically significant temperature increase (&lt; 1 degree C) during scans employed in a typical iEEG-fMRI experiment, including 60 min of continuous fMRI. However, high SAR sequences, such as fast spin-echo (FSE), produced significant heating in almost all scenarios (&gt; 2.0 degree C) that in some cases exceeded 10 degree C. Induced voltages in the frequency range that could elicit neuronal stimulation (&lt; 10 kHz) were well below the threshold of 100 mV. fMRI signal intensity was significantly reduced within 20 mm of the electrodes for the imaging parameters used in this study. Thus, for the conditions tested, a full iEEG-fMRI protocol poses a low risk at 3 T; however, fMRI sensitivity may be reduced immediately adjacent to the electrodes. In addition, high SAR sequences must be avoided.</description><issn>1053-8119</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2012</creationdate><recordtype>article</recordtype><recordid>eNotj7tOAzEURF2ARAj8wy1pdrnXD9amQ9EmRApCikJDEzmOjRz2kay9BX9PeFRTzTkzjAFhSUgP94ey8-PQx9Z--JIj8RJ1iagv2IRQiUITmSt2ndIBEQ1JPWHvc29T3MUm5i_oA9gOYpcH6wbbRdtAXS-K8LJewnHoc-_6BmwGAZtHWMf0CTYln1Lru3xu7uHXDKfR_uBu2GWwTfK3_zllb_N6M3suVq-L5expVRyJdC52zlRKcan3kodKG1sFw8loEkpqFNprdKgdGukM5wGNU145odCFICtnxJTd_XHPE0-jT3nbxuR809jO92Pakjp_V1gpKb4BwX5WUw</recordid><startdate>20121115</startdate><enddate>20121115</enddate><creator>Boucousis, Shannon</creator><creator>Beers, Craig</creator><creator>Cunningham, Cameron</creator><creator>Gaxiola-Valdez, Ismael</creator><creator>Pittman, Daniel</creator><creator>Goodyear, Bradley</creator><creator>Federico, Paolo</creator><scope>7QO</scope><scope>7TK</scope><scope>7U1</scope><scope>7U2</scope><scope>8FD</scope><scope>C1K</scope><scope>FR3</scope><scope>P64</scope></search><sort><creationdate>20121115</creationdate><title>Feasibility of an intracranial EEG-fMRI protocol at 3 T: Risk assessment and image quality</title><author>Boucousis, Shannon ; Beers, Craig ; Cunningham, Cameron ; Gaxiola-Valdez, Ismael ; Pittman, Daniel ; Goodyear, Bradley ; Federico, Paolo</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-p118t-bc9755248d42f789a7f9219813548038e80c08c094c922f09c5e5c350cff47c93</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2012</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Boucousis, Shannon</creatorcontrib><creatorcontrib>Beers, Craig</creatorcontrib><creatorcontrib>Cunningham, Cameron</creatorcontrib><creatorcontrib>Gaxiola-Valdez, Ismael</creatorcontrib><creatorcontrib>Pittman, Daniel</creatorcontrib><creatorcontrib>Goodyear, Bradley</creatorcontrib><creatorcontrib>Federico, Paolo</creatorcontrib><collection>Biotechnology Research Abstracts</collection><collection>Neurosciences Abstracts</collection><collection>Risk Abstracts</collection><collection>Safety Science and Risk</collection><collection>Technology Research Database</collection><collection>Environmental Sciences and Pollution Management</collection><collection>Engineering Research Database</collection><collection>Biotechnology and BioEngineering Abstracts</collection><jtitle>NeuroImage (Orlando, Fla.)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Boucousis, Shannon</au><au>Beers, Craig</au><au>Cunningham, Cameron</au><au>Gaxiola-Valdez, Ismael</au><au>Pittman, Daniel</au><au>Goodyear, Bradley</au><au>Federico, Paolo</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Feasibility of an intracranial EEG-fMRI protocol at 3 T: Risk assessment and image quality</atitle><jtitle>NeuroImage (Orlando, Fla.)</jtitle><date>2012-11-15</date><risdate>2012</risdate><volume>63</volume><issue>3</issue><spage>1237</spage><epage>1248</epage><pages>1237-1248</pages><issn>1053-8119</issn><abstract>Integrating intracranial EEG (iEEG) with functional MRI (iEEG-fMRI) may help elucidate mechanisms underlying the generation of seizures. However, the introduction of iEEG electrodes in the MR environment has inherent risk and data quality implications that require consideration prior to clinical use. Previous studies of subdural and depth electrodes have confirmed low risk under specific circumstances at 1.5 T and 3 T. However, no studies have assessed risk and image quality related to the feasibility of a full iEEGafMRI protocol. To this end, commercially available platinum subdural grid/strip electrodes (4 x 5 grid or 1 x 8 strip) and 4 or 6-contact depth electrodes were secured to the surface of a custom-made phantom mimicking the conductivity of the human brain. Electrode displacement, temperature increase of electrodes and surrounding phantom material, and voltage fluctuations in electrode contacts were measured in a GE Discovery MR750 3 T MR scanner during a variety of imaging sequences, typical of an iEEG-fMRI protocol. An electrode grid was also used to quantify the spatial extent of susceptibility artifact. The spatial extent of susceptibility artifact in the presence of an electrode was also assessed for typical imaging parameters that maximize BOLD sensitivity at 3 T (TR = 1500 ms; TE = 30 ms; slice thickness = 4 mm; matrix = 64 x 64; field-of-view = 24 cm). Under standard conditions, all electrodes exhibited no measurable displacement and no clinically significant temperature increase (&lt; 1 degree C) during scans employed in a typical iEEG-fMRI experiment, including 60 min of continuous fMRI. However, high SAR sequences, such as fast spin-echo (FSE), produced significant heating in almost all scenarios (&gt; 2.0 degree C) that in some cases exceeded 10 degree C. Induced voltages in the frequency range that could elicit neuronal stimulation (&lt; 10 kHz) were well below the threshold of 100 mV. fMRI signal intensity was significantly reduced within 20 mm of the electrodes for the imaging parameters used in this study. Thus, for the conditions tested, a full iEEG-fMRI protocol poses a low risk at 3 T; however, fMRI sensitivity may be reduced immediately adjacent to the electrodes. In addition, high SAR sequences must be avoided.</abstract><doi>10.1016/j.neuroimage.2012.08.008</doi><tpages>12</tpages></addata></record>
fulltext fulltext
identifier ISSN: 1053-8119
ispartof NeuroImage (Orlando, Fla.), 2012-11, Vol.63 (3), p.1237-1248
issn 1053-8119
language eng
recordid cdi_proquest_miscellaneous_1505350754
source Access via ScienceDirect (Elsevier); ProQuest Central UK/Ireland
title Feasibility of an intracranial EEG-fMRI protocol at 3 T: Risk assessment and image quality
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-04T09%3A21%3A30IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Feasibility%20of%20an%20intracranial%20EEG-fMRI%20protocol%20at%203%20T:%20Risk%20assessment%20and%20image%20quality&rft.jtitle=NeuroImage%20(Orlando,%20Fla.)&rft.au=Boucousis,%20Shannon&rft.date=2012-11-15&rft.volume=63&rft.issue=3&rft.spage=1237&rft.epage=1248&rft.pages=1237-1248&rft.issn=1053-8119&rft_id=info:doi/10.1016/j.neuroimage.2012.08.008&rft_dat=%3Cproquest%3E1505350754%3C/proquest%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=1505350754&rft_id=info:pmid/&rfr_iscdi=true