Analysis of calibration results of radiation survey meters used for area monitoring

Analyses of the results of calibration of survey meters carried out at the Secondary Standards Dosimetry Laboratory (SSDL) in Ghana over a period of 4 y (2008-2011) are reported. The calibration factors (CFs) of the set of survey meters indicated that ∼91.04 % were within the acceptable limit of ± 2...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Radiation protection dosimetry 2013-10, Vol.156 (4), p.506-513, Article 506
Hauptverfasser: Adjei, D, Darko, E O, Annkah, J K, Amoako, J K, Ofori, K, Emi-Reynolds, G, Obeng, M K, Akomaning-Adofo, E, Owusu-Manteaw, P
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 513
container_issue 4
container_start_page 506
container_title Radiation protection dosimetry
container_volume 156
creator Adjei, D
Darko, E O
Annkah, J K
Amoako, J K
Ofori, K
Emi-Reynolds, G
Obeng, M K
Akomaning-Adofo, E
Owusu-Manteaw, P
description Analyses of the results of calibration of survey meters carried out at the Secondary Standards Dosimetry Laboratory (SSDL) in Ghana over a period of 4 y (2008-2011) are reported. The calibration factors (CFs) of the set of survey meters indicated that ∼91.04 % were within the acceptable limit of ± 20.0 %. A higher percentage of the survey meters indicated CFs in the range of 0.95-1.15 except a few of them which indicated values 1.15. The degree of uncertainty in the measurements ranged from 0.03 to 17 % with the majority of them ranging from 0.03 to 6.0 % and a few of them >6.0 %. The results show that most of the survey meters calibrated were within the requirements of the regulations and may provide data for future development of calibration techniques in the country.
doi_str_mv 10.1093/rpd/nct087
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_1500784875</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>1437580532</sourcerecordid><originalsourceid>FETCH-LOGICAL-c320t-dfeb49f78fa945ba833f578cc761178843975b07f7e9aa1510904bdd7dc3a9033</originalsourceid><addsrcrecordid>eNqFkEtLAzEURoMotlY3_gCZpQhjbyZJkyxL8QUFF-p6yOQhkXnUJCP03zt2qgsRXF34OBy4B6FzDNcYJJmHjZm3OoHgB2iKOS1yQmFxiKaAKc0FLWCCTmJ8Ayi4ZPQYTQrCFhxLMUVPy1bV2-hj1rlMq9pXQSXftVmwsa_Tbg7K-HGMffiw26yxyYaY9dGazHUhU8GqrOlan7rg29dTdORUHe3Z_s7Qy-3N8-o-Xz_ePayW61yTAlJunK2odFw4JSmrlCDEMS605guMuRCUSM4q4I5bqRRmw69AK2O40URJIGSGLkfvJnTvvY2pbHzUtq5Va7s-lpgBcEEFZ_-jlHAmgJFiQK9GVIcuxmBduQm-UWFbYii_epdD73LsPcAXe29fNdb8oN-BBwB-2bRPu5gpKF__5fwEl8-Nzg</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1437580532</pqid></control><display><type>article</type><title>Analysis of calibration results of radiation survey meters used for area monitoring</title><source>MEDLINE</source><source>Oxford University Press Journals All Titles (1996-Current)</source><source>Alma/SFX Local Collection</source><creator>Adjei, D ; Darko, E O ; Annkah, J K ; Amoako, J K ; Ofori, K ; Emi-Reynolds, G ; Obeng, M K ; Akomaning-Adofo, E ; Owusu-Manteaw, P</creator><creatorcontrib>Adjei, D ; Darko, E O ; Annkah, J K ; Amoako, J K ; Ofori, K ; Emi-Reynolds, G ; Obeng, M K ; Akomaning-Adofo, E ; Owusu-Manteaw, P</creatorcontrib><description>Analyses of the results of calibration of survey meters carried out at the Secondary Standards Dosimetry Laboratory (SSDL) in Ghana over a period of 4 y (2008-2011) are reported. The calibration factors (CFs) of the set of survey meters indicated that ∼91.04 % were within the acceptable limit of ± 20.0 %. A higher percentage of the survey meters indicated CFs in the range of 0.95-1.15 except a few of them which indicated values &lt;0.55. Some of the survey meters also recorded CFs &gt;1.15. The degree of uncertainty in the measurements ranged from 0.03 to 17 % with the majority of them ranging from 0.03 to 6.0 % and a few of them &gt;6.0 %. The results show that most of the survey meters calibrated were within the requirements of the regulations and may provide data for future development of calibration techniques in the country.</description><identifier>ISSN: 0144-8420</identifier><identifier>EISSN: 1742-3406</identifier><identifier>DOI: 10.1093/rpd/nct087</identifier><identifier>PMID: 23567198</identifier><language>eng</language><publisher>England</publisher><subject>Algorithms ; Calibration ; Cesium Radioisotopes - analysis ; Ghana ; Humans ; Lasers ; Radiation Monitoring - instrumentation ; Radiation Monitoring - methods ; Radiation Protection - methods ; Radiometry - methods ; Reproducibility of Results</subject><ispartof>Radiation protection dosimetry, 2013-10, Vol.156 (4), p.506-513, Article 506</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c320t-dfeb49f78fa945ba833f578cc761178843975b07f7e9aa1510904bdd7dc3a9033</citedby><cites>FETCH-LOGICAL-c320t-dfeb49f78fa945ba833f578cc761178843975b07f7e9aa1510904bdd7dc3a9033</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27923,27924</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/23567198$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Adjei, D</creatorcontrib><creatorcontrib>Darko, E O</creatorcontrib><creatorcontrib>Annkah, J K</creatorcontrib><creatorcontrib>Amoako, J K</creatorcontrib><creatorcontrib>Ofori, K</creatorcontrib><creatorcontrib>Emi-Reynolds, G</creatorcontrib><creatorcontrib>Obeng, M K</creatorcontrib><creatorcontrib>Akomaning-Adofo, E</creatorcontrib><creatorcontrib>Owusu-Manteaw, P</creatorcontrib><title>Analysis of calibration results of radiation survey meters used for area monitoring</title><title>Radiation protection dosimetry</title><addtitle>Radiat Prot Dosimetry</addtitle><description>Analyses of the results of calibration of survey meters carried out at the Secondary Standards Dosimetry Laboratory (SSDL) in Ghana over a period of 4 y (2008-2011) are reported. The calibration factors (CFs) of the set of survey meters indicated that ∼91.04 % were within the acceptable limit of ± 20.0 %. A higher percentage of the survey meters indicated CFs in the range of 0.95-1.15 except a few of them which indicated values &lt;0.55. Some of the survey meters also recorded CFs &gt;1.15. The degree of uncertainty in the measurements ranged from 0.03 to 17 % with the majority of them ranging from 0.03 to 6.0 % and a few of them &gt;6.0 %. The results show that most of the survey meters calibrated were within the requirements of the regulations and may provide data for future development of calibration techniques in the country.</description><subject>Algorithms</subject><subject>Calibration</subject><subject>Cesium Radioisotopes - analysis</subject><subject>Ghana</subject><subject>Humans</subject><subject>Lasers</subject><subject>Radiation Monitoring - instrumentation</subject><subject>Radiation Monitoring - methods</subject><subject>Radiation Protection - methods</subject><subject>Radiometry - methods</subject><subject>Reproducibility of Results</subject><issn>0144-8420</issn><issn>1742-3406</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2013</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNqFkEtLAzEURoMotlY3_gCZpQhjbyZJkyxL8QUFF-p6yOQhkXnUJCP03zt2qgsRXF34OBy4B6FzDNcYJJmHjZm3OoHgB2iKOS1yQmFxiKaAKc0FLWCCTmJ8Ayi4ZPQYTQrCFhxLMUVPy1bV2-hj1rlMq9pXQSXftVmwsa_Tbg7K-HGMffiw26yxyYaY9dGazHUhU8GqrOlan7rg29dTdORUHe3Z_s7Qy-3N8-o-Xz_ePayW61yTAlJunK2odFw4JSmrlCDEMS605guMuRCUSM4q4I5bqRRmw69AK2O40URJIGSGLkfvJnTvvY2pbHzUtq5Va7s-lpgBcEEFZ_-jlHAmgJFiQK9GVIcuxmBduQm-UWFbYii_epdD73LsPcAXe29fNdb8oN-BBwB-2bRPu5gpKF__5fwEl8-Nzg</recordid><startdate>20131001</startdate><enddate>20131001</enddate><creator>Adjei, D</creator><creator>Darko, E O</creator><creator>Annkah, J K</creator><creator>Amoako, J K</creator><creator>Ofori, K</creator><creator>Emi-Reynolds, G</creator><creator>Obeng, M K</creator><creator>Akomaning-Adofo, E</creator><creator>Owusu-Manteaw, P</creator><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>7ST</scope><scope>C1K</scope><scope>SOI</scope></search><sort><creationdate>20131001</creationdate><title>Analysis of calibration results of radiation survey meters used for area monitoring</title><author>Adjei, D ; Darko, E O ; Annkah, J K ; Amoako, J K ; Ofori, K ; Emi-Reynolds, G ; Obeng, M K ; Akomaning-Adofo, E ; Owusu-Manteaw, P</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c320t-dfeb49f78fa945ba833f578cc761178843975b07f7e9aa1510904bdd7dc3a9033</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2013</creationdate><topic>Algorithms</topic><topic>Calibration</topic><topic>Cesium Radioisotopes - analysis</topic><topic>Ghana</topic><topic>Humans</topic><topic>Lasers</topic><topic>Radiation Monitoring - instrumentation</topic><topic>Radiation Monitoring - methods</topic><topic>Radiation Protection - methods</topic><topic>Radiometry - methods</topic><topic>Reproducibility of Results</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Adjei, D</creatorcontrib><creatorcontrib>Darko, E O</creatorcontrib><creatorcontrib>Annkah, J K</creatorcontrib><creatorcontrib>Amoako, J K</creatorcontrib><creatorcontrib>Ofori, K</creatorcontrib><creatorcontrib>Emi-Reynolds, G</creatorcontrib><creatorcontrib>Obeng, M K</creatorcontrib><creatorcontrib>Akomaning-Adofo, E</creatorcontrib><creatorcontrib>Owusu-Manteaw, P</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>Environment Abstracts</collection><collection>Environmental Sciences and Pollution Management</collection><collection>Environment Abstracts</collection><jtitle>Radiation protection dosimetry</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Adjei, D</au><au>Darko, E O</au><au>Annkah, J K</au><au>Amoako, J K</au><au>Ofori, K</au><au>Emi-Reynolds, G</au><au>Obeng, M K</au><au>Akomaning-Adofo, E</au><au>Owusu-Manteaw, P</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Analysis of calibration results of radiation survey meters used for area monitoring</atitle><jtitle>Radiation protection dosimetry</jtitle><addtitle>Radiat Prot Dosimetry</addtitle><date>2013-10-01</date><risdate>2013</risdate><volume>156</volume><issue>4</issue><spage>506</spage><epage>513</epage><pages>506-513</pages><artnum>506</artnum><issn>0144-8420</issn><eissn>1742-3406</eissn><abstract>Analyses of the results of calibration of survey meters carried out at the Secondary Standards Dosimetry Laboratory (SSDL) in Ghana over a period of 4 y (2008-2011) are reported. The calibration factors (CFs) of the set of survey meters indicated that ∼91.04 % were within the acceptable limit of ± 20.0 %. A higher percentage of the survey meters indicated CFs in the range of 0.95-1.15 except a few of them which indicated values &lt;0.55. Some of the survey meters also recorded CFs &gt;1.15. The degree of uncertainty in the measurements ranged from 0.03 to 17 % with the majority of them ranging from 0.03 to 6.0 % and a few of them &gt;6.0 %. The results show that most of the survey meters calibrated were within the requirements of the regulations and may provide data for future development of calibration techniques in the country.</abstract><cop>England</cop><pmid>23567198</pmid><doi>10.1093/rpd/nct087</doi><tpages>8</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0144-8420
ispartof Radiation protection dosimetry, 2013-10, Vol.156 (4), p.506-513, Article 506
issn 0144-8420
1742-3406
language eng
recordid cdi_proquest_miscellaneous_1500784875
source MEDLINE; Oxford University Press Journals All Titles (1996-Current); Alma/SFX Local Collection
subjects Algorithms
Calibration
Cesium Radioisotopes - analysis
Ghana
Humans
Lasers
Radiation Monitoring - instrumentation
Radiation Monitoring - methods
Radiation Protection - methods
Radiometry - methods
Reproducibility of Results
title Analysis of calibration results of radiation survey meters used for area monitoring
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-12T03%3A30%3A33IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Analysis%20of%20calibration%20results%20of%20radiation%20survey%20meters%20used%20for%20area%20monitoring&rft.jtitle=Radiation%20protection%20dosimetry&rft.au=Adjei,%20D&rft.date=2013-10-01&rft.volume=156&rft.issue=4&rft.spage=506&rft.epage=513&rft.pages=506-513&rft.artnum=506&rft.issn=0144-8420&rft.eissn=1742-3406&rft_id=info:doi/10.1093/rpd/nct087&rft_dat=%3Cproquest_cross%3E1437580532%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=1437580532&rft_id=info:pmid/23567198&rfr_iscdi=true