Improved energy compensation for time-of-flight mass spectrometry

A model for improved energy compensation in time-of-flight (TOF) mass spectrometry has been developed and tested. This model includes effects of both the acceleration and drift region on mass resolution for surface desorption TOF mass spectrometers that employ ion mirrors to improve mass resolution....

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of the American Society for Mass Spectrometry 1994-08, Vol.5 (8), p.779-787
Hauptverfasser: Short, R.T., Todd, P.J.
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A model for improved energy compensation in time-of-flight (TOF) mass spectrometry has been developed and tested. This model includes effects of both the acceleration and drift region on mass resolution for surface desorption TOF mass spectrometers that employ ion mirrors to improve mass resolution. Appropriate placement of an additional stage onto the conventional one- and two-stage mirrors provides compensation for flight time spreads, caused by initial ion kinetic energy distributions, in both regions. Experimental results that validate the model calculations are presented for a modified commercial two-stage ion mirror. For example, m/Δ m for Na + was improved from ∼ 100 to ∼ 200 using only a 200-eV drift energy and a 58-cm drift path.