Inspection of freeform intraocular lens topography by phase measuring deflectometric methods
Manufacturing spherical, aspheric, and freeform surfaces requires testing throughout the development and production process. State-of-the-art topography measurement is limited in applicability for intraocular lenses (IOLs), and there is no dedicated commercial surface measurement system available fo...
Gespeichert in:
Veröffentlicht in: | Applied optics (2004) 2013-06, Vol.52 (18), p.4279-4286 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Manufacturing spherical, aspheric, and freeform surfaces requires testing throughout the development and production process. State-of-the-art topography measurement is limited in applicability for intraocular lenses (IOLs), and there is no dedicated commercial surface measurement system available for freeform IOLs. The purpose of this work was to validate a deflectometric setup for surface measurement, detection of defects, and shape fidelity analysis for the development and production of IOLs. The setup is based on a phase measuring deflectometer with a field-of-view of 80 mm×80 mm and a mean repetition accuracy of 1.6·10(-3) D. The technique is suitable for detection of global and local surface errors, extracted from geometry and topography analysis. For validation according to DIN ISO 5725:2002, spherical IOLs with radii of curvature of 10 and 20 mm, a commercial aspheric IOL, and single-sided freeform IOL samples were used. |
---|---|
ISSN: | 1559-128X 2155-3165 1539-4522 |
DOI: | 10.1364/AO.52.004279 |