Attenuation length measurements of a liquid scintillator with LabVIEW and reliability evaluation of the device
An attenuation length measurement device was constructed using an oscilloscope and LabVIEW for signal acquisition and processing. The performance of the device has been tested in a variety of ways. The test results show that the set-up has a good stability and high precision (sigma/mean reached 0.4...
Gespeichert in:
Veröffentlicht in: | Chinese physics C 2013-07, Vol.37 (7), p.47-51 |
---|---|
1. Verfasser: | |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 51 |
---|---|
container_issue | 7 |
container_start_page | 47 |
container_title | Chinese physics C |
container_volume | 37 |
creator | 高龙 俞伯祥 丁雅韵 周莉 温良剑 谢宇广 王志刚 蔡啸 孙希磊 方建 薛镇 章爱武 吕绮雯 孙丽君 葛永帅 刘颖彪 牛顺利 胡涛 曹俊 吕军光 |
description | An attenuation length measurement device was constructed using an oscilloscope and LabVIEW for signal acquisition and processing. The performance of the device has been tested in a variety of ways. The test results show that the set-up has a good stability and high precision (sigma/mean reached 0.4 percent). Besides, the accuracy of the measurement system will decrease by about 17 percent if a filter is used. The attenuation length of a gadolinium-loaded liquid scintillator (Gd-LS) was measured as 15.1±0.35 m where Gd-LS was heavily used in the Daya Bay Neutrino Experiment. In addition, one method based on the Beer-Lambert law was proposed to investigate the reliability of the measurement device, the R-square reached 0.9995. Moreover, three purification methods for Linear Alkyl Benzene (LAB) production were compared in the experiment. |
doi_str_mv | 10.1088/1674-1137/37/7/076001 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_1439781011</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><cqvip_id>46376067</cqvip_id><sourcerecordid>1439781011</sourcerecordid><originalsourceid>FETCH-LOGICAL-c312t-1c768ac716e745ef86fa51d3e96dc9df9b635d35f9f8ec9f24812464884b8b13</originalsourceid><addsrcrecordid>eNo9kFFLwzAUhYMoOKc_QYhvvtQlTZqkj2NMHQx8GfoY0vRmi3Tp1qST_Xs7NgYH7sv3HS4HoWdK3ihRakKF5BmlTE6GyAmRghB6g0YkL3jGSJHfotGVuUcPMf4SIvigjlCYpgShN8m3ATcQ1mmDt2Bi38EWQoq4ddjgxu97X-NofUi-aUxqO_znB3Rpqu_F_AebUOMOGm8q3_h0xHAwzaV0KEgbwDUcvIVHdOdME-Hpcsdo9T5fzT6z5dfHYjZdZpbRPGXUSqGMlVSA5AU4JZwpaM2gFLUta1dWghU1K1zpFNjS5VzRnAuuFK9URdkYvZ5rd1277yEmvfXRwvB5gLaPmnJWSkUJPaHFGbVdG2MHTu86vzXdUVOiT_Pq03T6NJ0eIvV53sF7uXibNqz3PqyvIhdsYIRk_5VgemA</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1439781011</pqid></control><display><type>article</type><title>Attenuation length measurements of a liquid scintillator with LabVIEW and reliability evaluation of the device</title><source>IOP Publishing Journals</source><source>Institute of Physics (IOP) Journals - HEAL-Link</source><source>Alma/SFX Local Collection</source><creator>高龙 俞伯祥 丁雅韵 周莉 温良剑 谢宇广 王志刚 蔡啸 孙希磊 方建 薛镇 章爱武 吕绮雯 孙丽君 葛永帅 刘颖彪 牛顺利 胡涛 曹俊 吕军光</creator><creatorcontrib>高龙 俞伯祥 丁雅韵 周莉 温良剑 谢宇广 王志刚 蔡啸 孙希磊 方建 薛镇 章爱武 吕绮雯 孙丽君 葛永帅 刘颖彪 牛顺利 胡涛 曹俊 吕军光</creatorcontrib><description>An attenuation length measurement device was constructed using an oscilloscope and LabVIEW for signal acquisition and processing. The performance of the device has been tested in a variety of ways. The test results show that the set-up has a good stability and high precision (sigma/mean reached 0.4 percent). Besides, the accuracy of the measurement system will decrease by about 17 percent if a filter is used. The attenuation length of a gadolinium-loaded liquid scintillator (Gd-LS) was measured as 15.1±0.35 m where Gd-LS was heavily used in the Daya Bay Neutrino Experiment. In addition, one method based on the Beer-Lambert law was proposed to investigate the reliability of the measurement device, the R-square reached 0.9995. Moreover, three purification methods for Linear Alkyl Benzene (LAB) production were compared in the experiment.</description><identifier>ISSN: 1674-1137</identifier><identifier>EISSN: 0254-3052</identifier><identifier>DOI: 10.1088/1674-1137/37/7/076001</identifier><language>eng</language><subject>Attenuation ; Devices ; LabVIEW ; Law ; Liquids ; Marketing ; Oscilloscopes ; Purification ; Stability ; 中微子实验 ; 可靠性评估 ; 液体闪烁体 ; 直链烷基苯 ; 衰减 ; 设备 ; 长度测量装置</subject><ispartof>Chinese physics C, 2013-07, Vol.37 (7), p.47-51</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c312t-1c768ac716e745ef86fa51d3e96dc9df9b635d35f9f8ec9f24812464884b8b13</citedby><cites>FETCH-LOGICAL-c312t-1c768ac716e745ef86fa51d3e96dc9df9b635d35f9f8ec9f24812464884b8b13</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Uhttp://image.cqvip.com/vip1000/qk/92043A/92043A.jpg</thumbnail><link.rule.ids>315,781,785,27926,27927</link.rule.ids></links><search><creatorcontrib>高龙 俞伯祥 丁雅韵 周莉 温良剑 谢宇广 王志刚 蔡啸 孙希磊 方建 薛镇 章爱武 吕绮雯 孙丽君 葛永帅 刘颖彪 牛顺利 胡涛 曹俊 吕军光</creatorcontrib><title>Attenuation length measurements of a liquid scintillator with LabVIEW and reliability evaluation of the device</title><title>Chinese physics C</title><addtitle>Chinese Physica C</addtitle><description>An attenuation length measurement device was constructed using an oscilloscope and LabVIEW for signal acquisition and processing. The performance of the device has been tested in a variety of ways. The test results show that the set-up has a good stability and high precision (sigma/mean reached 0.4 percent). Besides, the accuracy of the measurement system will decrease by about 17 percent if a filter is used. The attenuation length of a gadolinium-loaded liquid scintillator (Gd-LS) was measured as 15.1±0.35 m where Gd-LS was heavily used in the Daya Bay Neutrino Experiment. In addition, one method based on the Beer-Lambert law was proposed to investigate the reliability of the measurement device, the R-square reached 0.9995. Moreover, three purification methods for Linear Alkyl Benzene (LAB) production were compared in the experiment.</description><subject>Attenuation</subject><subject>Devices</subject><subject>LabVIEW</subject><subject>Law</subject><subject>Liquids</subject><subject>Marketing</subject><subject>Oscilloscopes</subject><subject>Purification</subject><subject>Stability</subject><subject>中微子实验</subject><subject>可靠性评估</subject><subject>液体闪烁体</subject><subject>直链烷基苯</subject><subject>衰减</subject><subject>设备</subject><subject>长度测量装置</subject><issn>1674-1137</issn><issn>0254-3052</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2013</creationdate><recordtype>article</recordtype><recordid>eNo9kFFLwzAUhYMoOKc_QYhvvtQlTZqkj2NMHQx8GfoY0vRmi3Tp1qST_Xs7NgYH7sv3HS4HoWdK3ihRakKF5BmlTE6GyAmRghB6g0YkL3jGSJHfotGVuUcPMf4SIvigjlCYpgShN8m3ATcQ1mmDt2Bi38EWQoq4ddjgxu97X-NofUi-aUxqO_znB3Rpqu_F_AebUOMOGm8q3_h0xHAwzaV0KEgbwDUcvIVHdOdME-Hpcsdo9T5fzT6z5dfHYjZdZpbRPGXUSqGMlVSA5AU4JZwpaM2gFLUta1dWghU1K1zpFNjS5VzRnAuuFK9URdkYvZ5rd1277yEmvfXRwvB5gLaPmnJWSkUJPaHFGbVdG2MHTu86vzXdUVOiT_Pq03T6NJ0eIvV53sF7uXibNqz3PqyvIhdsYIRk_5VgemA</recordid><startdate>20130701</startdate><enddate>20130701</enddate><creator>高龙 俞伯祥 丁雅韵 周莉 温良剑 谢宇广 王志刚 蔡啸 孙希磊 方建 薛镇 章爱武 吕绮雯 孙丽君 葛永帅 刘颖彪 牛顺利 胡涛 曹俊 吕军光</creator><scope>2RA</scope><scope>92L</scope><scope>CQIGP</scope><scope>~WA</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7TB</scope><scope>7U5</scope><scope>8FD</scope><scope>FR3</scope><scope>H8D</scope><scope>L7M</scope></search><sort><creationdate>20130701</creationdate><title>Attenuation length measurements of a liquid scintillator with LabVIEW and reliability evaluation of the device</title><author>高龙 俞伯祥 丁雅韵 周莉 温良剑 谢宇广 王志刚 蔡啸 孙希磊 方建 薛镇 章爱武 吕绮雯 孙丽君 葛永帅 刘颖彪 牛顺利 胡涛 曹俊 吕军光</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c312t-1c768ac716e745ef86fa51d3e96dc9df9b635d35f9f8ec9f24812464884b8b13</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2013</creationdate><topic>Attenuation</topic><topic>Devices</topic><topic>LabVIEW</topic><topic>Law</topic><topic>Liquids</topic><topic>Marketing</topic><topic>Oscilloscopes</topic><topic>Purification</topic><topic>Stability</topic><topic>中微子实验</topic><topic>可靠性评估</topic><topic>液体闪烁体</topic><topic>直链烷基苯</topic><topic>衰减</topic><topic>设备</topic><topic>长度测量装置</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>高龙 俞伯祥 丁雅韵 周莉 温良剑 谢宇广 王志刚 蔡啸 孙希磊 方建 薛镇 章爱武 吕绮雯 孙丽君 葛永帅 刘颖彪 牛顺利 胡涛 曹俊 吕军光</creatorcontrib><collection>中文科技期刊数据库</collection><collection>中文科技期刊数据库-CALIS站点</collection><collection>中文科技期刊数据库-7.0平台</collection><collection>中文科技期刊数据库- 镜像站点</collection><collection>CrossRef</collection><collection>Mechanical & Transportation Engineering Abstracts</collection><collection>Solid State and Superconductivity Abstracts</collection><collection>Technology Research Database</collection><collection>Engineering Research Database</collection><collection>Aerospace Database</collection><collection>Advanced Technologies Database with Aerospace</collection><jtitle>Chinese physics C</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>高龙 俞伯祥 丁雅韵 周莉 温良剑 谢宇广 王志刚 蔡啸 孙希磊 方建 薛镇 章爱武 吕绮雯 孙丽君 葛永帅 刘颖彪 牛顺利 胡涛 曹俊 吕军光</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Attenuation length measurements of a liquid scintillator with LabVIEW and reliability evaluation of the device</atitle><jtitle>Chinese physics C</jtitle><addtitle>Chinese Physica C</addtitle><date>2013-07-01</date><risdate>2013</risdate><volume>37</volume><issue>7</issue><spage>47</spage><epage>51</epage><pages>47-51</pages><issn>1674-1137</issn><eissn>0254-3052</eissn><abstract>An attenuation length measurement device was constructed using an oscilloscope and LabVIEW for signal acquisition and processing. The performance of the device has been tested in a variety of ways. The test results show that the set-up has a good stability and high precision (sigma/mean reached 0.4 percent). Besides, the accuracy of the measurement system will decrease by about 17 percent if a filter is used. The attenuation length of a gadolinium-loaded liquid scintillator (Gd-LS) was measured as 15.1±0.35 m where Gd-LS was heavily used in the Daya Bay Neutrino Experiment. In addition, one method based on the Beer-Lambert law was proposed to investigate the reliability of the measurement device, the R-square reached 0.9995. Moreover, three purification methods for Linear Alkyl Benzene (LAB) production were compared in the experiment.</abstract><doi>10.1088/1674-1137/37/7/076001</doi><tpages>5</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1674-1137 |
ispartof | Chinese physics C, 2013-07, Vol.37 (7), p.47-51 |
issn | 1674-1137 0254-3052 |
language | eng |
recordid | cdi_proquest_miscellaneous_1439781011 |
source | IOP Publishing Journals; Institute of Physics (IOP) Journals - HEAL-Link; Alma/SFX Local Collection |
subjects | Attenuation Devices LabVIEW Law Liquids Marketing Oscilloscopes Purification Stability 中微子实验 可靠性评估 液体闪烁体 直链烷基苯 衰减 设备 长度测量装置 |
title | Attenuation length measurements of a liquid scintillator with LabVIEW and reliability evaluation of the device |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-18T06%3A22%3A02IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Attenuation%20length%20measurements%20of%20a%20liquid%20scintillator%20with%20LabVIEW%20and%20reliability%20evaluation%20of%20the%20device&rft.jtitle=Chinese%20physics%20C&rft.au=%E9%AB%98%E9%BE%99%20%E4%BF%9E%E4%BC%AF%E7%A5%A5%20%E4%B8%81%E9%9B%85%E9%9F%B5%20%E5%91%A8%E8%8E%89%20%E6%B8%A9%E8%89%AF%E5%89%91%20%E8%B0%A2%E5%AE%87%E5%B9%BF%20%E7%8E%8B%E5%BF%97%E5%88%9A%20%E8%94%A1%E5%95%B8%20%E5%AD%99%E5%B8%8C%E7%A3%8A%20%E6%96%B9%E5%BB%BA%20%E8%96%9B%E9%95%87%20%E7%AB%A0%E7%88%B1%E6%AD%A6%20%E5%90%95%E7%BB%AE%E9%9B%AF%20%E5%AD%99%E4%B8%BD%E5%90%9B%20%E8%91%9B%E6%B0%B8%E5%B8%85%20%E5%88%98%E9%A2%96%E5%BD%AA%20%E7%89%9B%E9%A1%BA%E5%88%A9%20%E8%83%A1%E6%B6%9B%20%E6%9B%B9%E4%BF%8A%20%E5%90%95%E5%86%9B%E5%85%89&rft.date=2013-07-01&rft.volume=37&rft.issue=7&rft.spage=47&rft.epage=51&rft.pages=47-51&rft.issn=1674-1137&rft.eissn=0254-3052&rft_id=info:doi/10.1088/1674-1137/37/7/076001&rft_dat=%3Cproquest_cross%3E1439781011%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=1439781011&rft_id=info:pmid/&rft_cqvip_id=46376067&rfr_iscdi=true |