An All Solid State Potentiometric Sensor for Monohydrogen Phosphate Ions
An all solid state potentiometric sensor using anthracene thiourea derivative as ionophore was developed. It exhibited a near‐Nernstian slope of 30.8±1.0 mV/decade of activity for HPO42− ions in the concentration range of 1.0×10−7–1.0×10−3 M at pH 7.4. It displayed excellent selectivity for monohydr...
Gespeichert in:
Veröffentlicht in: | Electroanalysis (New York, N.Y.) N.Y.), 2013-08, Vol.25 (8), p.1864-1870 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An all solid state potentiometric sensor using anthracene thiourea derivative as ionophore was developed. It exhibited a near‐Nernstian slope of 30.8±1.0 mV/decade of activity for HPO42− ions in the concentration range of 1.0×10−7–1.0×10−3 M at pH 7.4. It displayed excellent selectivity for monohydrogen phosphate over other anions and the selectivity sequence was determined as HPO42−>SO42−>Cl−>NO3−>OAc−>I−>ClO4−. The developed sensor was evaluated for the analysis of monohydrogen phosphate ions in a standard reference material (SRM 1548) as well as in the potentiometric titration of phosphate ions with a barium chloride solution. |
---|---|
ISSN: | 1040-0397 1521-4109 |
DOI: | 10.1002/elan.201300142 |