Transcatheter closure of patent ductus arteriosus using the amplatzer duct occluder in infants under one year of age

Percutaneous closure of patent ductus arteriosus (PDA) is a well-established technique. We evaluated the usefulness of the Amplatzer duct occluder for the percutaneous closure of patent ductus arteriosus in 29 children under 1 year of age. The patients' mean age was 8.9±2.8 months and their mea...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Revista española de cardiologia 2009-08, Vol.62 (8), p.867-874
Hauptverfasser: Parra-Bravo, Rafael, Cruz-Ramírez, Arturo, Rebolledo-Pineda, Víctor, Robles-Cervantes, Jorge, Chávez-Fernández, Alejandro, Beirana-Palencia, Luisa, Jiménez-Montufar, Laura, de Jesús Estrada-Loza, María, Estrada-Flores, José, Báez-Zamudio, Nelson, Escobar-Ponce, Mariano
Format: Artikel
Sprache:eng ; spa
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 874
container_issue 8
container_start_page 867
container_title Revista española de cardiologia
container_volume 62
creator Parra-Bravo, Rafael
Cruz-Ramírez, Arturo
Rebolledo-Pineda, Víctor
Robles-Cervantes, Jorge
Chávez-Fernández, Alejandro
Beirana-Palencia, Luisa
Jiménez-Montufar, Laura
de Jesús Estrada-Loza, María
Estrada-Flores, José
Báez-Zamudio, Nelson
Escobar-Ponce, Mariano
description Percutaneous closure of patent ductus arteriosus (PDA) is a well-established technique. We evaluated the usefulness of the Amplatzer duct occluder for the percutaneous closure of patent ductus arteriosus in 29 children under 1 year of age. The patients' mean age was 8.9±2.8 months and their mean weight was 6.4±1.5kg. In addition, 24.1% of patients were aged 6 months or less and 17.2% weighed 5kg or less. All completed follow-up (0.5-36 months). The minimum PDA diameter was 3.16±1.24 mm. The device was implanted successfully in 26 patients (89.6%). Failures were due to either the device migrating to the descending aorta, persistent moderate leakage, or to difficulty in the advancement of the device. No deaths were associated with the procedure. Three major complications occurred in two patients (10.3%). Aortography showed initial total occlusion in 65.5% of patients. At 3-month follow-up, total occlusion was observed in 96.1% of patients. Ultimately, closure was achieved successfully in 25 of the 29 patients (86.2%). During follow-up, four patients exhibited mild left pulmonary artery stenosis and one exhibited mild stenosis of the descending aorta. In children under 1 year of age, percutaneous closure of patent ductus arteriosus using an Amplatzer occluder was a safe and effective procedure. It is possible that improvements in the design of the occluder could decrease the complication rate.
doi_str_mv 10.1016/S0300-8932(09)72069-X
format Article
fullrecord <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_proquest_miscellaneous_1413162734</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>1413162734</sourcerecordid><originalsourceid>FETCH-LOGICAL-p141t-462c6691339e67f545f073072299c344813053534665032b0d26ae8e844ae47a3</originalsourceid><addsrcrecordid>eNo1UEtLAzEQDoLYWv0JSo71sJrXJpujFF9Q8GAFb0uana0ru9k1j0P99aZYYWCY-R7DNwhdUXJLCZV3b4QTUlSasyXRN4oRqYuPEzSnpdIFY4LN0HkIX4SwkgtxhmaME0G4quYobrxxwZr4CRE8tv0Ykgc8tngyEVzETbIxBWx8hrsMBpxC53Y4C7AZpt7En6w7sPBobZ-aPHUuV2tczGR3WIwO8B6MP_iaHVyg09b0AS6PfYHeHx82q-di_fr0srpfFxMVNBZCMiulppxrkKotRdkSxYliTGubg1SUk5LnSFKWhLMtaZg0UEElhAGhDF-g5Z_v5MfvBCHWQxcs9L1xMKZQ5yucSqa4yNTrIzVtB2jqyXeD8fv6_1P8Fwxiaco</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1413162734</pqid></control><display><type>article</type><title>Transcatheter closure of patent ductus arteriosus using the amplatzer duct occluder in infants under one year of age</title><source>Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals</source><source>ScienceDirect Journals (5 years ago - present)</source><creator>Parra-Bravo, Rafael ; Cruz-Ramírez, Arturo ; Rebolledo-Pineda, Víctor ; Robles-Cervantes, Jorge ; Chávez-Fernández, Alejandro ; Beirana-Palencia, Luisa ; Jiménez-Montufar, Laura ; de Jesús Estrada-Loza, María ; Estrada-Flores, José ; Báez-Zamudio, Nelson ; Escobar-Ponce, Mariano</creator><creatorcontrib>Parra-Bravo, Rafael ; Cruz-Ramírez, Arturo ; Rebolledo-Pineda, Víctor ; Robles-Cervantes, Jorge ; Chávez-Fernández, Alejandro ; Beirana-Palencia, Luisa ; Jiménez-Montufar, Laura ; de Jesús Estrada-Loza, María ; Estrada-Flores, José ; Báez-Zamudio, Nelson ; Escobar-Ponce, Mariano</creatorcontrib><description>Percutaneous closure of patent ductus arteriosus (PDA) is a well-established technique. We evaluated the usefulness of the Amplatzer duct occluder for the percutaneous closure of patent ductus arteriosus in 29 children under 1 year of age. The patients' mean age was 8.9±2.8 months and their mean weight was 6.4±1.5kg. In addition, 24.1% of patients were aged 6 months or less and 17.2% weighed 5kg or less. All completed follow-up (0.5-36 months). The minimum PDA diameter was 3.16±1.24 mm. The device was implanted successfully in 26 patients (89.6%). Failures were due to either the device migrating to the descending aorta, persistent moderate leakage, or to difficulty in the advancement of the device. No deaths were associated with the procedure. Three major complications occurred in two patients (10.3%). Aortography showed initial total occlusion in 65.5% of patients. At 3-month follow-up, total occlusion was observed in 96.1% of patients. Ultimately, closure was achieved successfully in 25 of the 29 patients (86.2%). During follow-up, four patients exhibited mild left pulmonary artery stenosis and one exhibited mild stenosis of the descending aorta. In children under 1 year of age, percutaneous closure of patent ductus arteriosus using an Amplatzer occluder was a safe and effective procedure. It is possible that improvements in the design of the occluder could decrease the complication rate.</description><identifier>EISSN: 1579-2242</identifier><identifier>DOI: 10.1016/S0300-8932(09)72069-X</identifier><identifier>PMID: 23040378</identifier><language>eng ; spa</language><publisher>Spain</publisher><ispartof>Revista española de cardiologia, 2009-08, Vol.62 (8), p.867-874</ispartof><rights>Copyright © 2009 Sociedad Española de Cardiología. Published by Elsevier Espana. All rights reserved.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27922,27923</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/23040378$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Parra-Bravo, Rafael</creatorcontrib><creatorcontrib>Cruz-Ramírez, Arturo</creatorcontrib><creatorcontrib>Rebolledo-Pineda, Víctor</creatorcontrib><creatorcontrib>Robles-Cervantes, Jorge</creatorcontrib><creatorcontrib>Chávez-Fernández, Alejandro</creatorcontrib><creatorcontrib>Beirana-Palencia, Luisa</creatorcontrib><creatorcontrib>Jiménez-Montufar, Laura</creatorcontrib><creatorcontrib>de Jesús Estrada-Loza, María</creatorcontrib><creatorcontrib>Estrada-Flores, José</creatorcontrib><creatorcontrib>Báez-Zamudio, Nelson</creatorcontrib><creatorcontrib>Escobar-Ponce, Mariano</creatorcontrib><title>Transcatheter closure of patent ductus arteriosus using the amplatzer duct occluder in infants under one year of age</title><title>Revista española de cardiologia</title><addtitle>Rev Esp Cardiol</addtitle><description>Percutaneous closure of patent ductus arteriosus (PDA) is a well-established technique. We evaluated the usefulness of the Amplatzer duct occluder for the percutaneous closure of patent ductus arteriosus in 29 children under 1 year of age. The patients' mean age was 8.9±2.8 months and their mean weight was 6.4±1.5kg. In addition, 24.1% of patients were aged 6 months or less and 17.2% weighed 5kg or less. All completed follow-up (0.5-36 months). The minimum PDA diameter was 3.16±1.24 mm. The device was implanted successfully in 26 patients (89.6%). Failures were due to either the device migrating to the descending aorta, persistent moderate leakage, or to difficulty in the advancement of the device. No deaths were associated with the procedure. Three major complications occurred in two patients (10.3%). Aortography showed initial total occlusion in 65.5% of patients. At 3-month follow-up, total occlusion was observed in 96.1% of patients. Ultimately, closure was achieved successfully in 25 of the 29 patients (86.2%). During follow-up, four patients exhibited mild left pulmonary artery stenosis and one exhibited mild stenosis of the descending aorta. In children under 1 year of age, percutaneous closure of patent ductus arteriosus using an Amplatzer occluder was a safe and effective procedure. It is possible that improvements in the design of the occluder could decrease the complication rate.</description><issn>1579-2242</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2009</creationdate><recordtype>article</recordtype><recordid>eNo1UEtLAzEQDoLYWv0JSo71sJrXJpujFF9Q8GAFb0uana0ru9k1j0P99aZYYWCY-R7DNwhdUXJLCZV3b4QTUlSasyXRN4oRqYuPEzSnpdIFY4LN0HkIX4SwkgtxhmaME0G4quYobrxxwZr4CRE8tv0Ykgc8tngyEVzETbIxBWx8hrsMBpxC53Y4C7AZpt7En6w7sPBobZ-aPHUuV2tczGR3WIwO8B6MP_iaHVyg09b0AS6PfYHeHx82q-di_fr0srpfFxMVNBZCMiulppxrkKotRdkSxYliTGubg1SUk5LnSFKWhLMtaZg0UEElhAGhDF-g5Z_v5MfvBCHWQxcs9L1xMKZQ5yucSqa4yNTrIzVtB2jqyXeD8fv6_1P8Fwxiaco</recordid><startdate>200908</startdate><enddate>200908</enddate><creator>Parra-Bravo, Rafael</creator><creator>Cruz-Ramírez, Arturo</creator><creator>Rebolledo-Pineda, Víctor</creator><creator>Robles-Cervantes, Jorge</creator><creator>Chávez-Fernández, Alejandro</creator><creator>Beirana-Palencia, Luisa</creator><creator>Jiménez-Montufar, Laura</creator><creator>de Jesús Estrada-Loza, María</creator><creator>Estrada-Flores, José</creator><creator>Báez-Zamudio, Nelson</creator><creator>Escobar-Ponce, Mariano</creator><scope>NPM</scope><scope>7X8</scope></search><sort><creationdate>200908</creationdate><title>Transcatheter closure of patent ductus arteriosus using the amplatzer duct occluder in infants under one year of age</title><author>Parra-Bravo, Rafael ; Cruz-Ramírez, Arturo ; Rebolledo-Pineda, Víctor ; Robles-Cervantes, Jorge ; Chávez-Fernández, Alejandro ; Beirana-Palencia, Luisa ; Jiménez-Montufar, Laura ; de Jesús Estrada-Loza, María ; Estrada-Flores, José ; Báez-Zamudio, Nelson ; Escobar-Ponce, Mariano</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-p141t-462c6691339e67f545f073072299c344813053534665032b0d26ae8e844ae47a3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng ; spa</language><creationdate>2009</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Parra-Bravo, Rafael</creatorcontrib><creatorcontrib>Cruz-Ramírez, Arturo</creatorcontrib><creatorcontrib>Rebolledo-Pineda, Víctor</creatorcontrib><creatorcontrib>Robles-Cervantes, Jorge</creatorcontrib><creatorcontrib>Chávez-Fernández, Alejandro</creatorcontrib><creatorcontrib>Beirana-Palencia, Luisa</creatorcontrib><creatorcontrib>Jiménez-Montufar, Laura</creatorcontrib><creatorcontrib>de Jesús Estrada-Loza, María</creatorcontrib><creatorcontrib>Estrada-Flores, José</creatorcontrib><creatorcontrib>Báez-Zamudio, Nelson</creatorcontrib><creatorcontrib>Escobar-Ponce, Mariano</creatorcontrib><collection>PubMed</collection><collection>MEDLINE - Academic</collection><jtitle>Revista española de cardiologia</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Parra-Bravo, Rafael</au><au>Cruz-Ramírez, Arturo</au><au>Rebolledo-Pineda, Víctor</au><au>Robles-Cervantes, Jorge</au><au>Chávez-Fernández, Alejandro</au><au>Beirana-Palencia, Luisa</au><au>Jiménez-Montufar, Laura</au><au>de Jesús Estrada-Loza, María</au><au>Estrada-Flores, José</au><au>Báez-Zamudio, Nelson</au><au>Escobar-Ponce, Mariano</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Transcatheter closure of patent ductus arteriosus using the amplatzer duct occluder in infants under one year of age</atitle><jtitle>Revista española de cardiologia</jtitle><addtitle>Rev Esp Cardiol</addtitle><date>2009-08</date><risdate>2009</risdate><volume>62</volume><issue>8</issue><spage>867</spage><epage>874</epage><pages>867-874</pages><eissn>1579-2242</eissn><abstract>Percutaneous closure of patent ductus arteriosus (PDA) is a well-established technique. We evaluated the usefulness of the Amplatzer duct occluder for the percutaneous closure of patent ductus arteriosus in 29 children under 1 year of age. The patients' mean age was 8.9±2.8 months and their mean weight was 6.4±1.5kg. In addition, 24.1% of patients were aged 6 months or less and 17.2% weighed 5kg or less. All completed follow-up (0.5-36 months). The minimum PDA diameter was 3.16±1.24 mm. The device was implanted successfully in 26 patients (89.6%). Failures were due to either the device migrating to the descending aorta, persistent moderate leakage, or to difficulty in the advancement of the device. No deaths were associated with the procedure. Three major complications occurred in two patients (10.3%). Aortography showed initial total occlusion in 65.5% of patients. At 3-month follow-up, total occlusion was observed in 96.1% of patients. Ultimately, closure was achieved successfully in 25 of the 29 patients (86.2%). During follow-up, four patients exhibited mild left pulmonary artery stenosis and one exhibited mild stenosis of the descending aorta. In children under 1 year of age, percutaneous closure of patent ductus arteriosus using an Amplatzer occluder was a safe and effective procedure. It is possible that improvements in the design of the occluder could decrease the complication rate.</abstract><cop>Spain</cop><pmid>23040378</pmid><doi>10.1016/S0300-8932(09)72069-X</doi><tpages>8</tpages></addata></record>
fulltext fulltext
identifier EISSN: 1579-2242
ispartof Revista española de cardiologia, 2009-08, Vol.62 (8), p.867-874
issn 1579-2242
language eng ; spa
recordid cdi_proquest_miscellaneous_1413162734
source Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals; ScienceDirect Journals (5 years ago - present)
title Transcatheter closure of patent ductus arteriosus using the amplatzer duct occluder in infants under one year of age
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-14T03%3A28%3A52IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Transcatheter%20closure%20of%20patent%20ductus%20arteriosus%20using%20the%20amplatzer%20duct%20occluder%20in%20infants%20under%20one%20year%20of%20age&rft.jtitle=Revista%20espa%C3%B1ola%20de%20cardiologia&rft.au=Parra-Bravo,%20Rafael&rft.date=2009-08&rft.volume=62&rft.issue=8&rft.spage=867&rft.epage=874&rft.pages=867-874&rft.eissn=1579-2242&rft_id=info:doi/10.1016/S0300-8932(09)72069-X&rft_dat=%3Cproquest_pubme%3E1413162734%3C/proquest_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=1413162734&rft_id=info:pmid/23040378&rfr_iscdi=true