Stereotactic dose planning system used in Leksell Gamma Knife model-B: EGS4 Monte Carlo versus GafChromic films MD-55

The accuracy of the dose planning system (Leksell GammaPlan) used in Gamma Knife (model-B) radiosurgery was verified using both the GafChromic films MD-55 (improved) and the EGS4 Monte Carlo technique. The Monte Carlo technique was employed to calculate the dose distribution along the x-, y- and z-a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Applied radiation and isotopes 2000-09, Vol.53 (3), p.427-430
Hauptverfasser: Cheung, Y.C., Yu, K.N., Ho, R.T.K., Yu, C.P.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 430
container_issue 3
container_start_page 427
container_title Applied radiation and isotopes
container_volume 53
creator Cheung, Y.C.
Yu, K.N.
Ho, R.T.K.
Yu, C.P.
description The accuracy of the dose planning system (Leksell GammaPlan) used in Gamma Knife (model-B) radiosurgery was verified using both the GafChromic films MD-55 (improved) and the EGS4 Monte Carlo technique. The Monte Carlo technique was employed to calculate the dose distribution along the x-, y- and z-axes when a single shot with opening of all 201 sources was delivered at the centre of a simulated water phantom with a diameter of 160 mm. Collimator helmets with different size of the Gamma Knife unit were verified. Good consistency (typical discrepancy less than 2%) was obtained between the results of Monte Carlo and GammaPlan. Small discrepancies, however, were obtained by GafChromic films. Discrepancies, as great as 10% when using the 4 mm collimator helmet, at the low percentage isodose curve along the z-axis of the measurement results were probably due to the small energy dependency of the GafChromic films. Significant discrepancies were not observed along x- and y-axes because such small discrepancies were easily over-washed by other gamma beams coming from the x- and y-directions. Similar results showing the discrepancies between the GafChromic films and GammaPlan were obtained when using the 8, 14 and 18 mm collimator helmets. However, the discrepancies along the z-axis became smaller as scattering effect increased when using larger collimator sizes. We suggest that the Monte Carlo technique should also be applied in stereotactic dose planning system verification as it is an ideal and reliable computational technique.
doi_str_mv 10.1016/S0969-8043(99)00285-7
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_13594457</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S0969804399002857</els_id><sourcerecordid>489142</sourcerecordid><originalsourceid>FETCH-LOGICAL-c458t-30c19aedad9e4bd76166520a00506a0af3fa13b8236da2e02c0eabc570480b833</originalsourceid><addsrcrecordid>eNqFkEFv1DAQhS1ERZfCTwD5hOCQMk7ixOZSwVK2FVtxWDhbjj0BQxwvnqRS_z3ZboW4cZrLN-_pfYy9EHAuQDRvd6AbXSioq9davwEolSzaR2wlVFsWWgE8Zqu_yCl7SvQTAGqlyyfsVIBuS1GrFZt3E2ZMk3VTcNwnQr4f7DiG8TunO5ow8pnQ8zDyLf4iHAa-sTFa_nkMPfKYPA7Fh3f8crOr-U0aJ-Rrm4fEbzHTTAvcr3_kFJfwPgyR-M3HQspn7KS3A-Hzh3vGvn26_Lq-KrZfNtfr99vC1VJNRQVOaIveeo1159tGNI0swQJIaCzYvuqtqDpVVo23JULpAG3nZLvshE5V1Rl7dczd5_R7RppMDOSWEXbENJMRldR1LdsFlEfQ5USUsTf7HKLNd0aAOfg2977NQabR2tz7Noe_lw8FcxfR__N1FLwAF0cAl5m3AbMhF3B06ENGNxmfwn8q_gBzII8N</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>13594457</pqid></control><display><type>article</type><title>Stereotactic dose planning system used in Leksell Gamma Knife model-B: EGS4 Monte Carlo versus GafChromic films MD-55</title><source>MEDLINE</source><source>ScienceDirect Journals (5 years ago - present)</source><creator>Cheung, Y.C. ; Yu, K.N. ; Ho, R.T.K. ; Yu, C.P.</creator><creatorcontrib>Cheung, Y.C. ; Yu, K.N. ; Ho, R.T.K. ; Yu, C.P.</creatorcontrib><description>The accuracy of the dose planning system (Leksell GammaPlan) used in Gamma Knife (model-B) radiosurgery was verified using both the GafChromic films MD-55 (improved) and the EGS4 Monte Carlo technique. The Monte Carlo technique was employed to calculate the dose distribution along the x-, y- and z-axes when a single shot with opening of all 201 sources was delivered at the centre of a simulated water phantom with a diameter of 160 mm. Collimator helmets with different size of the Gamma Knife unit were verified. Good consistency (typical discrepancy less than 2%) was obtained between the results of Monte Carlo and GammaPlan. Small discrepancies, however, were obtained by GafChromic films. Discrepancies, as great as 10% when using the 4 mm collimator helmet, at the low percentage isodose curve along the z-axis of the measurement results were probably due to the small energy dependency of the GafChromic films. Significant discrepancies were not observed along x- and y-axes because such small discrepancies were easily over-washed by other gamma beams coming from the x- and y-directions. Similar results showing the discrepancies between the GafChromic films and GammaPlan were obtained when using the 8, 14 and 18 mm collimator helmets. However, the discrepancies along the z-axis became smaller as scattering effect increased when using larger collimator sizes. We suggest that the Monte Carlo technique should also be applied in stereotactic dose planning system verification as it is an ideal and reliable computational technique.</description><identifier>ISSN: 0969-8043</identifier><identifier>EISSN: 1872-9800</identifier><identifier>DOI: 10.1016/S0969-8043(99)00285-7</identifier><identifier>PMID: 10972148</identifier><language>eng</language><publisher>England: Elsevier Ltd</publisher><subject>Dosimetry ; EGS4 ; Film Dosimetry - methods ; GafChromic film ; Gamma Knife ; GammaPlan ; Medical applications ; Monte Carlo ; Monte Carlo Method ; Monte Carlo methods ; Optical collimators ; Phantoms, Imaging ; Radiosurgery - instrumentation ; Radiosurgery - methods ; Radiotherapy Planning, Computer-Assisted - methods ; Radiotherapy, Computer-Assisted - methods ; Surgery ; Water</subject><ispartof>Applied radiation and isotopes, 2000-09, Vol.53 (3), p.427-430</ispartof><rights>2000 Elsevier Science Ltd</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c458t-30c19aedad9e4bd76166520a00506a0af3fa13b8236da2e02c0eabc570480b833</citedby><cites>FETCH-LOGICAL-c458t-30c19aedad9e4bd76166520a00506a0af3fa13b8236da2e02c0eabc570480b833</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://dx.doi.org/10.1016/S0969-8043(99)00285-7$$EHTML$$P50$$Gelsevier$$H</linktohtml><link.rule.ids>314,780,784,3541,27915,27916,45986</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/10972148$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Cheung, Y.C.</creatorcontrib><creatorcontrib>Yu, K.N.</creatorcontrib><creatorcontrib>Ho, R.T.K.</creatorcontrib><creatorcontrib>Yu, C.P.</creatorcontrib><title>Stereotactic dose planning system used in Leksell Gamma Knife model-B: EGS4 Monte Carlo versus GafChromic films MD-55</title><title>Applied radiation and isotopes</title><addtitle>Appl Radiat Isot</addtitle><description>The accuracy of the dose planning system (Leksell GammaPlan) used in Gamma Knife (model-B) radiosurgery was verified using both the GafChromic films MD-55 (improved) and the EGS4 Monte Carlo technique. The Monte Carlo technique was employed to calculate the dose distribution along the x-, y- and z-axes when a single shot with opening of all 201 sources was delivered at the centre of a simulated water phantom with a diameter of 160 mm. Collimator helmets with different size of the Gamma Knife unit were verified. Good consistency (typical discrepancy less than 2%) was obtained between the results of Monte Carlo and GammaPlan. Small discrepancies, however, were obtained by GafChromic films. Discrepancies, as great as 10% when using the 4 mm collimator helmet, at the low percentage isodose curve along the z-axis of the measurement results were probably due to the small energy dependency of the GafChromic films. Significant discrepancies were not observed along x- and y-axes because such small discrepancies were easily over-washed by other gamma beams coming from the x- and y-directions. Similar results showing the discrepancies between the GafChromic films and GammaPlan were obtained when using the 8, 14 and 18 mm collimator helmets. However, the discrepancies along the z-axis became smaller as scattering effect increased when using larger collimator sizes. We suggest that the Monte Carlo technique should also be applied in stereotactic dose planning system verification as it is an ideal and reliable computational technique.</description><subject>Dosimetry</subject><subject>EGS4</subject><subject>Film Dosimetry - methods</subject><subject>GafChromic film</subject><subject>Gamma Knife</subject><subject>GammaPlan</subject><subject>Medical applications</subject><subject>Monte Carlo</subject><subject>Monte Carlo Method</subject><subject>Monte Carlo methods</subject><subject>Optical collimators</subject><subject>Phantoms, Imaging</subject><subject>Radiosurgery - instrumentation</subject><subject>Radiosurgery - methods</subject><subject>Radiotherapy Planning, Computer-Assisted - methods</subject><subject>Radiotherapy, Computer-Assisted - methods</subject><subject>Surgery</subject><subject>Water</subject><issn>0969-8043</issn><issn>1872-9800</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2000</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNqFkEFv1DAQhS1ERZfCTwD5hOCQMk7ixOZSwVK2FVtxWDhbjj0BQxwvnqRS_z3ZboW4cZrLN-_pfYy9EHAuQDRvd6AbXSioq9davwEolSzaR2wlVFsWWgE8Zqu_yCl7SvQTAGqlyyfsVIBuS1GrFZt3E2ZMk3VTcNwnQr4f7DiG8TunO5ow8pnQ8zDyLf4iHAa-sTFa_nkMPfKYPA7Fh3f8crOr-U0aJ-Rrm4fEbzHTTAvcr3_kFJfwPgyR-M3HQspn7KS3A-Hzh3vGvn26_Lq-KrZfNtfr99vC1VJNRQVOaIveeo1159tGNI0swQJIaCzYvuqtqDpVVo23JULpAG3nZLvshE5V1Rl7dczd5_R7RppMDOSWEXbENJMRldR1LdsFlEfQ5USUsTf7HKLNd0aAOfg2977NQabR2tz7Noe_lw8FcxfR__N1FLwAF0cAl5m3AbMhF3B06ENGNxmfwn8q_gBzII8N</recordid><startdate>20000901</startdate><enddate>20000901</enddate><creator>Cheung, Y.C.</creator><creator>Yu, K.N.</creator><creator>Ho, R.T.K.</creator><creator>Yu, C.P.</creator><general>Elsevier Ltd</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope></search><sort><creationdate>20000901</creationdate><title>Stereotactic dose planning system used in Leksell Gamma Knife model-B: EGS4 Monte Carlo versus GafChromic films MD-55</title><author>Cheung, Y.C. ; Yu, K.N. ; Ho, R.T.K. ; Yu, C.P.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c458t-30c19aedad9e4bd76166520a00506a0af3fa13b8236da2e02c0eabc570480b833</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2000</creationdate><topic>Dosimetry</topic><topic>EGS4</topic><topic>Film Dosimetry - methods</topic><topic>GafChromic film</topic><topic>Gamma Knife</topic><topic>GammaPlan</topic><topic>Medical applications</topic><topic>Monte Carlo</topic><topic>Monte Carlo Method</topic><topic>Monte Carlo methods</topic><topic>Optical collimators</topic><topic>Phantoms, Imaging</topic><topic>Radiosurgery - instrumentation</topic><topic>Radiosurgery - methods</topic><topic>Radiotherapy Planning, Computer-Assisted - methods</topic><topic>Radiotherapy, Computer-Assisted - methods</topic><topic>Surgery</topic><topic>Water</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Cheung, Y.C.</creatorcontrib><creatorcontrib>Yu, K.N.</creatorcontrib><creatorcontrib>Ho, R.T.K.</creatorcontrib><creatorcontrib>Yu, C.P.</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><jtitle>Applied radiation and isotopes</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Cheung, Y.C.</au><au>Yu, K.N.</au><au>Ho, R.T.K.</au><au>Yu, C.P.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Stereotactic dose planning system used in Leksell Gamma Knife model-B: EGS4 Monte Carlo versus GafChromic films MD-55</atitle><jtitle>Applied radiation and isotopes</jtitle><addtitle>Appl Radiat Isot</addtitle><date>2000-09-01</date><risdate>2000</risdate><volume>53</volume><issue>3</issue><spage>427</spage><epage>430</epage><pages>427-430</pages><issn>0969-8043</issn><eissn>1872-9800</eissn><abstract>The accuracy of the dose planning system (Leksell GammaPlan) used in Gamma Knife (model-B) radiosurgery was verified using both the GafChromic films MD-55 (improved) and the EGS4 Monte Carlo technique. The Monte Carlo technique was employed to calculate the dose distribution along the x-, y- and z-axes when a single shot with opening of all 201 sources was delivered at the centre of a simulated water phantom with a diameter of 160 mm. Collimator helmets with different size of the Gamma Knife unit were verified. Good consistency (typical discrepancy less than 2%) was obtained between the results of Monte Carlo and GammaPlan. Small discrepancies, however, were obtained by GafChromic films. Discrepancies, as great as 10% when using the 4 mm collimator helmet, at the low percentage isodose curve along the z-axis of the measurement results were probably due to the small energy dependency of the GafChromic films. Significant discrepancies were not observed along x- and y-axes because such small discrepancies were easily over-washed by other gamma beams coming from the x- and y-directions. Similar results showing the discrepancies between the GafChromic films and GammaPlan were obtained when using the 8, 14 and 18 mm collimator helmets. However, the discrepancies along the z-axis became smaller as scattering effect increased when using larger collimator sizes. We suggest that the Monte Carlo technique should also be applied in stereotactic dose planning system verification as it is an ideal and reliable computational technique.</abstract><cop>England</cop><pub>Elsevier Ltd</pub><pmid>10972148</pmid><doi>10.1016/S0969-8043(99)00285-7</doi><tpages>4</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0969-8043
ispartof Applied radiation and isotopes, 2000-09, Vol.53 (3), p.427-430
issn 0969-8043
1872-9800
language eng
recordid cdi_proquest_miscellaneous_13594457
source MEDLINE; ScienceDirect Journals (5 years ago - present)
subjects Dosimetry
EGS4
Film Dosimetry - methods
GafChromic film
Gamma Knife
GammaPlan
Medical applications
Monte Carlo
Monte Carlo Method
Monte Carlo methods
Optical collimators
Phantoms, Imaging
Radiosurgery - instrumentation
Radiosurgery - methods
Radiotherapy Planning, Computer-Assisted - methods
Radiotherapy, Computer-Assisted - methods
Surgery
Water
title Stereotactic dose planning system used in Leksell Gamma Knife model-B: EGS4 Monte Carlo versus GafChromic films MD-55
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-15T01%3A19%3A23IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Stereotactic%20dose%20planning%20system%20used%20in%20Leksell%20Gamma%20Knife%20model-B:%20EGS4%20Monte%20Carlo%20versus%20GafChromic%20films%20MD-55&rft.jtitle=Applied%20radiation%20and%20isotopes&rft.au=Cheung,%20Y.C.&rft.date=2000-09-01&rft.volume=53&rft.issue=3&rft.spage=427&rft.epage=430&rft.pages=427-430&rft.issn=0969-8043&rft.eissn=1872-9800&rft_id=info:doi/10.1016/S0969-8043(99)00285-7&rft_dat=%3Cproquest_cross%3E489142%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=13594457&rft_id=info:pmid/10972148&rft_els_id=S0969804399002857&rfr_iscdi=true