Stereotactic dose planning system used in Leksell Gamma Knife model-B: EGS4 Monte Carlo versus GafChromic films MD-55
The accuracy of the dose planning system (Leksell GammaPlan) used in Gamma Knife (model-B) radiosurgery was verified using both the GafChromic films MD-55 (improved) and the EGS4 Monte Carlo technique. The Monte Carlo technique was employed to calculate the dose distribution along the x-, y- and z-a...
Gespeichert in:
Veröffentlicht in: | Applied radiation and isotopes 2000-09, Vol.53 (3), p.427-430 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 430 |
---|---|
container_issue | 3 |
container_start_page | 427 |
container_title | Applied radiation and isotopes |
container_volume | 53 |
creator | Cheung, Y.C. Yu, K.N. Ho, R.T.K. Yu, C.P. |
description | The accuracy of the dose planning system (Leksell GammaPlan) used in Gamma Knife (model-B) radiosurgery was verified using both the GafChromic films MD-55 (improved) and the EGS4 Monte Carlo technique. The Monte Carlo technique was employed to calculate the dose distribution along the
x-,
y- and
z-axes when a single shot with opening of all 201 sources was delivered at the centre of a simulated water phantom with a diameter of 160 mm. Collimator helmets with different size of the Gamma Knife unit were verified. Good consistency (typical discrepancy less than 2%) was obtained between the results of Monte Carlo and GammaPlan. Small discrepancies, however, were obtained by GafChromic films. Discrepancies, as great as 10% when using the 4 mm collimator helmet, at the low percentage isodose curve along the
z-axis of the measurement results were probably due to the small energy dependency of the GafChromic films. Significant discrepancies were not observed along
x- and
y-axes because such small discrepancies were easily over-washed by other gamma beams coming from the
x- and
y-directions. Similar results showing the discrepancies between the GafChromic films and GammaPlan were obtained when using the 8, 14 and 18 mm collimator helmets. However, the discrepancies along the
z-axis became smaller as scattering effect increased when using larger collimator sizes. We suggest that the Monte Carlo technique should also be applied in stereotactic dose planning system verification as it is an ideal and reliable computational technique. |
doi_str_mv | 10.1016/S0969-8043(99)00285-7 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_13594457</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S0969804399002857</els_id><sourcerecordid>489142</sourcerecordid><originalsourceid>FETCH-LOGICAL-c458t-30c19aedad9e4bd76166520a00506a0af3fa13b8236da2e02c0eabc570480b833</originalsourceid><addsrcrecordid>eNqFkEFv1DAQhS1ERZfCTwD5hOCQMk7ixOZSwVK2FVtxWDhbjj0BQxwvnqRS_z3ZboW4cZrLN-_pfYy9EHAuQDRvd6AbXSioq9davwEolSzaR2wlVFsWWgE8Zqu_yCl7SvQTAGqlyyfsVIBuS1GrFZt3E2ZMk3VTcNwnQr4f7DiG8TunO5ow8pnQ8zDyLf4iHAa-sTFa_nkMPfKYPA7Fh3f8crOr-U0aJ-Rrm4fEbzHTTAvcr3_kFJfwPgyR-M3HQspn7KS3A-Hzh3vGvn26_Lq-KrZfNtfr99vC1VJNRQVOaIveeo1159tGNI0swQJIaCzYvuqtqDpVVo23JULpAG3nZLvshE5V1Rl7dczd5_R7RppMDOSWEXbENJMRldR1LdsFlEfQ5USUsTf7HKLNd0aAOfg2977NQabR2tz7Noe_lw8FcxfR__N1FLwAF0cAl5m3AbMhF3B06ENGNxmfwn8q_gBzII8N</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>13594457</pqid></control><display><type>article</type><title>Stereotactic dose planning system used in Leksell Gamma Knife model-B: EGS4 Monte Carlo versus GafChromic films MD-55</title><source>MEDLINE</source><source>ScienceDirect Journals (5 years ago - present)</source><creator>Cheung, Y.C. ; Yu, K.N. ; Ho, R.T.K. ; Yu, C.P.</creator><creatorcontrib>Cheung, Y.C. ; Yu, K.N. ; Ho, R.T.K. ; Yu, C.P.</creatorcontrib><description>The accuracy of the dose planning system (Leksell GammaPlan) used in Gamma Knife (model-B) radiosurgery was verified using both the GafChromic films MD-55 (improved) and the EGS4 Monte Carlo technique. The Monte Carlo technique was employed to calculate the dose distribution along the
x-,
y- and
z-axes when a single shot with opening of all 201 sources was delivered at the centre of a simulated water phantom with a diameter of 160 mm. Collimator helmets with different size of the Gamma Knife unit were verified. Good consistency (typical discrepancy less than 2%) was obtained between the results of Monte Carlo and GammaPlan. Small discrepancies, however, were obtained by GafChromic films. Discrepancies, as great as 10% when using the 4 mm collimator helmet, at the low percentage isodose curve along the
z-axis of the measurement results were probably due to the small energy dependency of the GafChromic films. Significant discrepancies were not observed along
x- and
y-axes because such small discrepancies were easily over-washed by other gamma beams coming from the
x- and
y-directions. Similar results showing the discrepancies between the GafChromic films and GammaPlan were obtained when using the 8, 14 and 18 mm collimator helmets. However, the discrepancies along the
z-axis became smaller as scattering effect increased when using larger collimator sizes. We suggest that the Monte Carlo technique should also be applied in stereotactic dose planning system verification as it is an ideal and reliable computational technique.</description><identifier>ISSN: 0969-8043</identifier><identifier>EISSN: 1872-9800</identifier><identifier>DOI: 10.1016/S0969-8043(99)00285-7</identifier><identifier>PMID: 10972148</identifier><language>eng</language><publisher>England: Elsevier Ltd</publisher><subject>Dosimetry ; EGS4 ; Film Dosimetry - methods ; GafChromic film ; Gamma Knife ; GammaPlan ; Medical applications ; Monte Carlo ; Monte Carlo Method ; Monte Carlo methods ; Optical collimators ; Phantoms, Imaging ; Radiosurgery - instrumentation ; Radiosurgery - methods ; Radiotherapy Planning, Computer-Assisted - methods ; Radiotherapy, Computer-Assisted - methods ; Surgery ; Water</subject><ispartof>Applied radiation and isotopes, 2000-09, Vol.53 (3), p.427-430</ispartof><rights>2000 Elsevier Science Ltd</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c458t-30c19aedad9e4bd76166520a00506a0af3fa13b8236da2e02c0eabc570480b833</citedby><cites>FETCH-LOGICAL-c458t-30c19aedad9e4bd76166520a00506a0af3fa13b8236da2e02c0eabc570480b833</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://dx.doi.org/10.1016/S0969-8043(99)00285-7$$EHTML$$P50$$Gelsevier$$H</linktohtml><link.rule.ids>314,780,784,3541,27915,27916,45986</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/10972148$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Cheung, Y.C.</creatorcontrib><creatorcontrib>Yu, K.N.</creatorcontrib><creatorcontrib>Ho, R.T.K.</creatorcontrib><creatorcontrib>Yu, C.P.</creatorcontrib><title>Stereotactic dose planning system used in Leksell Gamma Knife model-B: EGS4 Monte Carlo versus GafChromic films MD-55</title><title>Applied radiation and isotopes</title><addtitle>Appl Radiat Isot</addtitle><description>The accuracy of the dose planning system (Leksell GammaPlan) used in Gamma Knife (model-B) radiosurgery was verified using both the GafChromic films MD-55 (improved) and the EGS4 Monte Carlo technique. The Monte Carlo technique was employed to calculate the dose distribution along the
x-,
y- and
z-axes when a single shot with opening of all 201 sources was delivered at the centre of a simulated water phantom with a diameter of 160 mm. Collimator helmets with different size of the Gamma Knife unit were verified. Good consistency (typical discrepancy less than 2%) was obtained between the results of Monte Carlo and GammaPlan. Small discrepancies, however, were obtained by GafChromic films. Discrepancies, as great as 10% when using the 4 mm collimator helmet, at the low percentage isodose curve along the
z-axis of the measurement results were probably due to the small energy dependency of the GafChromic films. Significant discrepancies were not observed along
x- and
y-axes because such small discrepancies were easily over-washed by other gamma beams coming from the
x- and
y-directions. Similar results showing the discrepancies between the GafChromic films and GammaPlan were obtained when using the 8, 14 and 18 mm collimator helmets. However, the discrepancies along the
z-axis became smaller as scattering effect increased when using larger collimator sizes. We suggest that the Monte Carlo technique should also be applied in stereotactic dose planning system verification as it is an ideal and reliable computational technique.</description><subject>Dosimetry</subject><subject>EGS4</subject><subject>Film Dosimetry - methods</subject><subject>GafChromic film</subject><subject>Gamma Knife</subject><subject>GammaPlan</subject><subject>Medical applications</subject><subject>Monte Carlo</subject><subject>Monte Carlo Method</subject><subject>Monte Carlo methods</subject><subject>Optical collimators</subject><subject>Phantoms, Imaging</subject><subject>Radiosurgery - instrumentation</subject><subject>Radiosurgery - methods</subject><subject>Radiotherapy Planning, Computer-Assisted - methods</subject><subject>Radiotherapy, Computer-Assisted - methods</subject><subject>Surgery</subject><subject>Water</subject><issn>0969-8043</issn><issn>1872-9800</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2000</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNqFkEFv1DAQhS1ERZfCTwD5hOCQMk7ixOZSwVK2FVtxWDhbjj0BQxwvnqRS_z3ZboW4cZrLN-_pfYy9EHAuQDRvd6AbXSioq9davwEolSzaR2wlVFsWWgE8Zqu_yCl7SvQTAGqlyyfsVIBuS1GrFZt3E2ZMk3VTcNwnQr4f7DiG8TunO5ow8pnQ8zDyLf4iHAa-sTFa_nkMPfKYPA7Fh3f8crOr-U0aJ-Rrm4fEbzHTTAvcr3_kFJfwPgyR-M3HQspn7KS3A-Hzh3vGvn26_Lq-KrZfNtfr99vC1VJNRQVOaIveeo1159tGNI0swQJIaCzYvuqtqDpVVo23JULpAG3nZLvshE5V1Rl7dczd5_R7RppMDOSWEXbENJMRldR1LdsFlEfQ5USUsTf7HKLNd0aAOfg2977NQabR2tz7Noe_lw8FcxfR__N1FLwAF0cAl5m3AbMhF3B06ENGNxmfwn8q_gBzII8N</recordid><startdate>20000901</startdate><enddate>20000901</enddate><creator>Cheung, Y.C.</creator><creator>Yu, K.N.</creator><creator>Ho, R.T.K.</creator><creator>Yu, C.P.</creator><general>Elsevier Ltd</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope></search><sort><creationdate>20000901</creationdate><title>Stereotactic dose planning system used in Leksell Gamma Knife model-B: EGS4 Monte Carlo versus GafChromic films MD-55</title><author>Cheung, Y.C. ; Yu, K.N. ; Ho, R.T.K. ; Yu, C.P.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c458t-30c19aedad9e4bd76166520a00506a0af3fa13b8236da2e02c0eabc570480b833</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2000</creationdate><topic>Dosimetry</topic><topic>EGS4</topic><topic>Film Dosimetry - methods</topic><topic>GafChromic film</topic><topic>Gamma Knife</topic><topic>GammaPlan</topic><topic>Medical applications</topic><topic>Monte Carlo</topic><topic>Monte Carlo Method</topic><topic>Monte Carlo methods</topic><topic>Optical collimators</topic><topic>Phantoms, Imaging</topic><topic>Radiosurgery - instrumentation</topic><topic>Radiosurgery - methods</topic><topic>Radiotherapy Planning, Computer-Assisted - methods</topic><topic>Radiotherapy, Computer-Assisted - methods</topic><topic>Surgery</topic><topic>Water</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Cheung, Y.C.</creatorcontrib><creatorcontrib>Yu, K.N.</creatorcontrib><creatorcontrib>Ho, R.T.K.</creatorcontrib><creatorcontrib>Yu, C.P.</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><jtitle>Applied radiation and isotopes</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Cheung, Y.C.</au><au>Yu, K.N.</au><au>Ho, R.T.K.</au><au>Yu, C.P.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Stereotactic dose planning system used in Leksell Gamma Knife model-B: EGS4 Monte Carlo versus GafChromic films MD-55</atitle><jtitle>Applied radiation and isotopes</jtitle><addtitle>Appl Radiat Isot</addtitle><date>2000-09-01</date><risdate>2000</risdate><volume>53</volume><issue>3</issue><spage>427</spage><epage>430</epage><pages>427-430</pages><issn>0969-8043</issn><eissn>1872-9800</eissn><abstract>The accuracy of the dose planning system (Leksell GammaPlan) used in Gamma Knife (model-B) radiosurgery was verified using both the GafChromic films MD-55 (improved) and the EGS4 Monte Carlo technique. The Monte Carlo technique was employed to calculate the dose distribution along the
x-,
y- and
z-axes when a single shot with opening of all 201 sources was delivered at the centre of a simulated water phantom with a diameter of 160 mm. Collimator helmets with different size of the Gamma Knife unit were verified. Good consistency (typical discrepancy less than 2%) was obtained between the results of Monte Carlo and GammaPlan. Small discrepancies, however, were obtained by GafChromic films. Discrepancies, as great as 10% when using the 4 mm collimator helmet, at the low percentage isodose curve along the
z-axis of the measurement results were probably due to the small energy dependency of the GafChromic films. Significant discrepancies were not observed along
x- and
y-axes because such small discrepancies were easily over-washed by other gamma beams coming from the
x- and
y-directions. Similar results showing the discrepancies between the GafChromic films and GammaPlan were obtained when using the 8, 14 and 18 mm collimator helmets. However, the discrepancies along the
z-axis became smaller as scattering effect increased when using larger collimator sizes. We suggest that the Monte Carlo technique should also be applied in stereotactic dose planning system verification as it is an ideal and reliable computational technique.</abstract><cop>England</cop><pub>Elsevier Ltd</pub><pmid>10972148</pmid><doi>10.1016/S0969-8043(99)00285-7</doi><tpages>4</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0969-8043 |
ispartof | Applied radiation and isotopes, 2000-09, Vol.53 (3), p.427-430 |
issn | 0969-8043 1872-9800 |
language | eng |
recordid | cdi_proquest_miscellaneous_13594457 |
source | MEDLINE; ScienceDirect Journals (5 years ago - present) |
subjects | Dosimetry EGS4 Film Dosimetry - methods GafChromic film Gamma Knife GammaPlan Medical applications Monte Carlo Monte Carlo Method Monte Carlo methods Optical collimators Phantoms, Imaging Radiosurgery - instrumentation Radiosurgery - methods Radiotherapy Planning, Computer-Assisted - methods Radiotherapy, Computer-Assisted - methods Surgery Water |
title | Stereotactic dose planning system used in Leksell Gamma Knife model-B: EGS4 Monte Carlo versus GafChromic films MD-55 |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-15T01%3A19%3A23IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Stereotactic%20dose%20planning%20system%20used%20in%20Leksell%20Gamma%20Knife%20model-B:%20EGS4%20Monte%20Carlo%20versus%20GafChromic%20films%20MD-55&rft.jtitle=Applied%20radiation%20and%20isotopes&rft.au=Cheung,%20Y.C.&rft.date=2000-09-01&rft.volume=53&rft.issue=3&rft.spage=427&rft.epage=430&rft.pages=427-430&rft.issn=0969-8043&rft.eissn=1872-9800&rft_id=info:doi/10.1016/S0969-8043(99)00285-7&rft_dat=%3Cproquest_cross%3E489142%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=13594457&rft_id=info:pmid/10972148&rft_els_id=S0969804399002857&rfr_iscdi=true |