Cardiac Tamponade caused by Broken Sternal Wire after Pectus Excavatum Repair: A Case Report

Complications of pectus excavatum surgery include pneumothorax, pleuritis, hemothorax, pericardial effusion, displacement of bar, pericarditis and cardiac injury, etc. This is the case of a 15-year-old boy with cardiac tamponade caused by pericarditis who had taken the operation for a pectus excavat...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Annals of Thoracic and Cardiovascular Surgery 2013/02/20, Vol.19(1), pp.52-54
Hauptverfasser: Lee, Soo Hyun, Cho, Bum Sang, Kim, Sung Jin, Lee, Seung Young, Kang, Min Ho, Han, Gi Seok, Cha, Sang-Hoon
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Complications of pectus excavatum surgery include pneumothorax, pleuritis, hemothorax, pericardial effusion, displacement of bar, pericarditis and cardiac injury, etc. This is the case of a 15-year-old boy with cardiac tamponade caused by pericarditis who had taken the operation for a pectus excavatum repair one year previously. The cause was a sternal wire which was used for attachment of the bar to sternum that had fractured and migrated through the pericardium causing a pericardial injury and a pericarditis.
ISSN:1341-1098
2186-1005
DOI:10.5761/atcs.cr.11.01871