Towards a fiber-coupled picowatt cryogenic radiometer
A picowatt cryogenic radiometer (PCR) has been fabricated at the microscale level for electrical substitution optical fiber power measurements. The absorber, electrical heater, and thermometer are all on a micromachined membrane less than 1 mm on a side. Initial measurements with input powers from 5...
Gespeichert in:
Veröffentlicht in: | Optics letters 2012-06, Vol.37 (12), p.2346-2348 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A picowatt cryogenic radiometer (PCR) has been fabricated at the microscale level for electrical substitution optical fiber power measurements. The absorber, electrical heater, and thermometer are all on a micromachined membrane less than 1 mm on a side. Initial measurements with input powers from 50 fW to 20 nW show a response inequivalence between electrical and optical power of 8%. A comparison of the response to electrical and optical input powers between 15 pW to 70 pW yields a repeatability better than ±0.3% (k=2). From our first optical tests, the system has a noise equivalent power of ≈5×10(-15) W/√Hz at 2 Hz, but simple changes to the measurement scheme should yield an NEP 2 orders of magnitude lower. |
---|---|
ISSN: | 0146-9592 1539-4794 |
DOI: | 10.1364/OL.37.002346 |