Computed tomographic venography for varicose veins of the lower extremities: prospective comparison of 80-kVp and conventional 120-kVp protocols
To prospectively investigate the feasibility of an 80-kilovolt (peak) (kVp) protocol in computed tomographic venography for varicose veins of the lower extremities by comparison with conventional 120-kVp protocol. Attenuation values and signal-to-noise ratio of iodine contrast medium (CM) were deter...
Gespeichert in:
Veröffentlicht in: | Journal of computer assisted tomography 2012-09, Vol.36 (5), p.583-590 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 590 |
---|---|
container_issue | 5 |
container_start_page | 583 |
container_title | Journal of computer assisted tomography |
container_volume | 36 |
creator | Cho, Eun-Suk Kim, Joo Hee Kim, Sungjun Yu, Jeong-Sik Chung, Jae-Joon Yoon, Choon-Sik Lee, Hyeon-Kyeong Lee, Kyung Hee |
description | To prospectively investigate the feasibility of an 80-kilovolt (peak) (kVp) protocol in computed tomographic venography for varicose veins of the lower extremities by comparison with conventional 120-kVp protocol.
Attenuation values and signal-to-noise ratio of iodine contrast medium (CM) were determined in a water phantom for 2 tube voltages (80 kVp and 120 kVp). Among 100 patients, 50 patients were scanned with 120 kVp and 150 effective milliampere second (mAs(eff)), and the other 50 patients were scanned with 80 kVp and 390 mAs(eff) after the administration of 1.7-mL/kg CM (370 mg of iodine per milliliter). The 2 groups were compared for venous attenuation, contrast-to-noise ratio, and subjective degree of venous enhancement, image noise, and overall diagnostic image quality.
In the phantom, the attenuation value and signal-to-noise ratio value for iodine CM at 80 kVp were 63.8% and 33.0% higher, respectively, than those obtained at 120 kVp. The mean attenuation of the measured veins of the lower extremities was 148.3 Hounsfield units (HU) for the 80-kVp protocol and 94.8 HU for the 120-kVp protocol. Contrast-to-noise ratio was also significantly higher with the 80-kVp protocol. The overall diagnostic image quality of the 3-dimensional volume-rendered images was good with both protocols. The subjective score for venous enhancement was higher at the 80-kVp protocol. The mean volume computed tomography dose index of the 80-kVp (5.6 mGy) protocol was 23.3% lower than that of the 120-kVp (7.3 mGy) protocol.
The use of the 80-kVp protocol improved overall venous attenuation, especially in perforating vein, and provided similarly high diagnostic image quality with a lower radiation dose when compared to the conventional 120-kVp protocol. |
doi_str_mv | 10.1097/RCT.0b013e3182621ea9 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_1113221223</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>1113221223</sourcerecordid><originalsourceid>FETCH-LOGICAL-c289t-16e9a18a4ff790f96c0fc4e5285c4fe8c8eef36ce5713103ce5f586dd943dddb3</originalsourceid><addsrcrecordid>eNqFkc9O3DAQxq2qqCy0b1BVPvYS8NiJY_dWrfgnISEh4Bp5nXFxm8TB9i7wFjwyRgsceuE0o5nv9400HyHfgR0A0-3h5fLqgK0YCBSguOSARn8iC2gErwTUzWeyYEKKSrXQ7JK9lP4yBq0Q9Reyy7nWXAJbkKdlGOd1xp7mMIY_0cy33tINTtv-kboQ6cZEb0PCMvdTosHRfIt0CPcYKT7kiKPPHtMvOseQZrTZb5DaYly4FKYXQLHq381MzdSXxVT8sw-TGSjw7aKQOdgwpK9kx5kh4bfXuk-uj4-ulqfV-cXJ2fL3eWW50rkCidqAMrVzrWZOS8ucrbHhqrG1Q2UVohPSYtOCACZK4xol-17Xou_7ldgnP7e-5fLdGlPuRp8sDoOZMKxTBwCCc-BcfCxlUnMmQbVFWm-ltnwiRXTdHP1o4mMRdS-xdSW27v_YCvbj9cJ6NWL_Dr3lJJ4Bvv6XYg</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1069206187</pqid></control><display><type>article</type><title>Computed tomographic venography for varicose veins of the lower extremities: prospective comparison of 80-kVp and conventional 120-kVp protocols</title><source>MEDLINE</source><source>Journals@Ovid Complete</source><creator>Cho, Eun-Suk ; Kim, Joo Hee ; Kim, Sungjun ; Yu, Jeong-Sik ; Chung, Jae-Joon ; Yoon, Choon-Sik ; Lee, Hyeon-Kyeong ; Lee, Kyung Hee</creator><creatorcontrib>Cho, Eun-Suk ; Kim, Joo Hee ; Kim, Sungjun ; Yu, Jeong-Sik ; Chung, Jae-Joon ; Yoon, Choon-Sik ; Lee, Hyeon-Kyeong ; Lee, Kyung Hee</creatorcontrib><description>To prospectively investigate the feasibility of an 80-kilovolt (peak) (kVp) protocol in computed tomographic venography for varicose veins of the lower extremities by comparison with conventional 120-kVp protocol.
Attenuation values and signal-to-noise ratio of iodine contrast medium (CM) were determined in a water phantom for 2 tube voltages (80 kVp and 120 kVp). Among 100 patients, 50 patients were scanned with 120 kVp and 150 effective milliampere second (mAs(eff)), and the other 50 patients were scanned with 80 kVp and 390 mAs(eff) after the administration of 1.7-mL/kg CM (370 mg of iodine per milliliter). The 2 groups were compared for venous attenuation, contrast-to-noise ratio, and subjective degree of venous enhancement, image noise, and overall diagnostic image quality.
In the phantom, the attenuation value and signal-to-noise ratio value for iodine CM at 80 kVp were 63.8% and 33.0% higher, respectively, than those obtained at 120 kVp. The mean attenuation of the measured veins of the lower extremities was 148.3 Hounsfield units (HU) for the 80-kVp protocol and 94.8 HU for the 120-kVp protocol. Contrast-to-noise ratio was also significantly higher with the 80-kVp protocol. The overall diagnostic image quality of the 3-dimensional volume-rendered images was good with both protocols. The subjective score for venous enhancement was higher at the 80-kVp protocol. The mean volume computed tomography dose index of the 80-kVp (5.6 mGy) protocol was 23.3% lower than that of the 120-kVp (7.3 mGy) protocol.
The use of the 80-kVp protocol improved overall venous attenuation, especially in perforating vein, and provided similarly high diagnostic image quality with a lower radiation dose when compared to the conventional 120-kVp protocol.</description><identifier>ISSN: 0363-8715</identifier><identifier>EISSN: 1532-3145</identifier><identifier>DOI: 10.1097/RCT.0b013e3182621ea9</identifier><identifier>PMID: 22992610</identifier><language>eng</language><publisher>United States</publisher><subject>Adult ; Computed tomography ; Contrast Media ; Feasibility Studies ; Humans ; Iohexol - analogs & derivatives ; Lower Extremity - blood supply ; Lower Extremity - diagnostic imaging ; Male ; Middle Aged ; Peripheral Vascular Diseases - diagnostic imaging ; Phantoms, Imaging ; Phlebography - methods ; Prospective Studies ; Radiographic Image Interpretation, Computer-Assisted ; Signal-To-Noise Ratio ; Tomography, X-Ray Computed - methods</subject><ispartof>Journal of computer assisted tomography, 2012-09, Vol.36 (5), p.583-590</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><cites>FETCH-LOGICAL-c289t-16e9a18a4ff790f96c0fc4e5285c4fe8c8eef36ce5713103ce5f586dd943dddb3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,776,780,27903,27904</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/22992610$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Cho, Eun-Suk</creatorcontrib><creatorcontrib>Kim, Joo Hee</creatorcontrib><creatorcontrib>Kim, Sungjun</creatorcontrib><creatorcontrib>Yu, Jeong-Sik</creatorcontrib><creatorcontrib>Chung, Jae-Joon</creatorcontrib><creatorcontrib>Yoon, Choon-Sik</creatorcontrib><creatorcontrib>Lee, Hyeon-Kyeong</creatorcontrib><creatorcontrib>Lee, Kyung Hee</creatorcontrib><title>Computed tomographic venography for varicose veins of the lower extremities: prospective comparison of 80-kVp and conventional 120-kVp protocols</title><title>Journal of computer assisted tomography</title><addtitle>J Comput Assist Tomogr</addtitle><description>To prospectively investigate the feasibility of an 80-kilovolt (peak) (kVp) protocol in computed tomographic venography for varicose veins of the lower extremities by comparison with conventional 120-kVp protocol.
Attenuation values and signal-to-noise ratio of iodine contrast medium (CM) were determined in a water phantom for 2 tube voltages (80 kVp and 120 kVp). Among 100 patients, 50 patients were scanned with 120 kVp and 150 effective milliampere second (mAs(eff)), and the other 50 patients were scanned with 80 kVp and 390 mAs(eff) after the administration of 1.7-mL/kg CM (370 mg of iodine per milliliter). The 2 groups were compared for venous attenuation, contrast-to-noise ratio, and subjective degree of venous enhancement, image noise, and overall diagnostic image quality.
In the phantom, the attenuation value and signal-to-noise ratio value for iodine CM at 80 kVp were 63.8% and 33.0% higher, respectively, than those obtained at 120 kVp. The mean attenuation of the measured veins of the lower extremities was 148.3 Hounsfield units (HU) for the 80-kVp protocol and 94.8 HU for the 120-kVp protocol. Contrast-to-noise ratio was also significantly higher with the 80-kVp protocol. The overall diagnostic image quality of the 3-dimensional volume-rendered images was good with both protocols. The subjective score for venous enhancement was higher at the 80-kVp protocol. The mean volume computed tomography dose index of the 80-kVp (5.6 mGy) protocol was 23.3% lower than that of the 120-kVp (7.3 mGy) protocol.
The use of the 80-kVp protocol improved overall venous attenuation, especially in perforating vein, and provided similarly high diagnostic image quality with a lower radiation dose when compared to the conventional 120-kVp protocol.</description><subject>Adult</subject><subject>Computed tomography</subject><subject>Contrast Media</subject><subject>Feasibility Studies</subject><subject>Humans</subject><subject>Iohexol - analogs & derivatives</subject><subject>Lower Extremity - blood supply</subject><subject>Lower Extremity - diagnostic imaging</subject><subject>Male</subject><subject>Middle Aged</subject><subject>Peripheral Vascular Diseases - diagnostic imaging</subject><subject>Phantoms, Imaging</subject><subject>Phlebography - methods</subject><subject>Prospective Studies</subject><subject>Radiographic Image Interpretation, Computer-Assisted</subject><subject>Signal-To-Noise Ratio</subject><subject>Tomography, X-Ray Computed - methods</subject><issn>0363-8715</issn><issn>1532-3145</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2012</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNqFkc9O3DAQxq2qqCy0b1BVPvYS8NiJY_dWrfgnISEh4Bp5nXFxm8TB9i7wFjwyRgsceuE0o5nv9400HyHfgR0A0-3h5fLqgK0YCBSguOSARn8iC2gErwTUzWeyYEKKSrXQ7JK9lP4yBq0Q9Reyy7nWXAJbkKdlGOd1xp7mMIY_0cy33tINTtv-kboQ6cZEb0PCMvdTosHRfIt0CPcYKT7kiKPPHtMvOseQZrTZb5DaYly4FKYXQLHq381MzdSXxVT8sw-TGSjw7aKQOdgwpK9kx5kh4bfXuk-uj4-ulqfV-cXJ2fL3eWW50rkCidqAMrVzrWZOS8ucrbHhqrG1Q2UVohPSYtOCACZK4xol-17Xou_7ldgnP7e-5fLdGlPuRp8sDoOZMKxTBwCCc-BcfCxlUnMmQbVFWm-ltnwiRXTdHP1o4mMRdS-xdSW27v_YCvbj9cJ6NWL_Dr3lJJ4Bvv6XYg</recordid><startdate>20120901</startdate><enddate>20120901</enddate><creator>Cho, Eun-Suk</creator><creator>Kim, Joo Hee</creator><creator>Kim, Sungjun</creator><creator>Yu, Jeong-Sik</creator><creator>Chung, Jae-Joon</creator><creator>Yoon, Choon-Sik</creator><creator>Lee, Hyeon-Kyeong</creator><creator>Lee, Kyung Hee</creator><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>7QO</scope><scope>8FD</scope><scope>FR3</scope><scope>P64</scope></search><sort><creationdate>20120901</creationdate><title>Computed tomographic venography for varicose veins of the lower extremities: prospective comparison of 80-kVp and conventional 120-kVp protocols</title><author>Cho, Eun-Suk ; Kim, Joo Hee ; Kim, Sungjun ; Yu, Jeong-Sik ; Chung, Jae-Joon ; Yoon, Choon-Sik ; Lee, Hyeon-Kyeong ; Lee, Kyung Hee</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c289t-16e9a18a4ff790f96c0fc4e5285c4fe8c8eef36ce5713103ce5f586dd943dddb3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2012</creationdate><topic>Adult</topic><topic>Computed tomography</topic><topic>Contrast Media</topic><topic>Feasibility Studies</topic><topic>Humans</topic><topic>Iohexol - analogs & derivatives</topic><topic>Lower Extremity - blood supply</topic><topic>Lower Extremity - diagnostic imaging</topic><topic>Male</topic><topic>Middle Aged</topic><topic>Peripheral Vascular Diseases - diagnostic imaging</topic><topic>Phantoms, Imaging</topic><topic>Phlebography - methods</topic><topic>Prospective Studies</topic><topic>Radiographic Image Interpretation, Computer-Assisted</topic><topic>Signal-To-Noise Ratio</topic><topic>Tomography, X-Ray Computed - methods</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Cho, Eun-Suk</creatorcontrib><creatorcontrib>Kim, Joo Hee</creatorcontrib><creatorcontrib>Kim, Sungjun</creatorcontrib><creatorcontrib>Yu, Jeong-Sik</creatorcontrib><creatorcontrib>Chung, Jae-Joon</creatorcontrib><creatorcontrib>Yoon, Choon-Sik</creatorcontrib><creatorcontrib>Lee, Hyeon-Kyeong</creatorcontrib><creatorcontrib>Lee, Kyung Hee</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>Biotechnology Research Abstracts</collection><collection>Technology Research Database</collection><collection>Engineering Research Database</collection><collection>Biotechnology and BioEngineering Abstracts</collection><jtitle>Journal of computer assisted tomography</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Cho, Eun-Suk</au><au>Kim, Joo Hee</au><au>Kim, Sungjun</au><au>Yu, Jeong-Sik</au><au>Chung, Jae-Joon</au><au>Yoon, Choon-Sik</au><au>Lee, Hyeon-Kyeong</au><au>Lee, Kyung Hee</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Computed tomographic venography for varicose veins of the lower extremities: prospective comparison of 80-kVp and conventional 120-kVp protocols</atitle><jtitle>Journal of computer assisted tomography</jtitle><addtitle>J Comput Assist Tomogr</addtitle><date>2012-09-01</date><risdate>2012</risdate><volume>36</volume><issue>5</issue><spage>583</spage><epage>590</epage><pages>583-590</pages><issn>0363-8715</issn><eissn>1532-3145</eissn><abstract>To prospectively investigate the feasibility of an 80-kilovolt (peak) (kVp) protocol in computed tomographic venography for varicose veins of the lower extremities by comparison with conventional 120-kVp protocol.
Attenuation values and signal-to-noise ratio of iodine contrast medium (CM) were determined in a water phantom for 2 tube voltages (80 kVp and 120 kVp). Among 100 patients, 50 patients were scanned with 120 kVp and 150 effective milliampere second (mAs(eff)), and the other 50 patients were scanned with 80 kVp and 390 mAs(eff) after the administration of 1.7-mL/kg CM (370 mg of iodine per milliliter). The 2 groups were compared for venous attenuation, contrast-to-noise ratio, and subjective degree of venous enhancement, image noise, and overall diagnostic image quality.
In the phantom, the attenuation value and signal-to-noise ratio value for iodine CM at 80 kVp were 63.8% and 33.0% higher, respectively, than those obtained at 120 kVp. The mean attenuation of the measured veins of the lower extremities was 148.3 Hounsfield units (HU) for the 80-kVp protocol and 94.8 HU for the 120-kVp protocol. Contrast-to-noise ratio was also significantly higher with the 80-kVp protocol. The overall diagnostic image quality of the 3-dimensional volume-rendered images was good with both protocols. The subjective score for venous enhancement was higher at the 80-kVp protocol. The mean volume computed tomography dose index of the 80-kVp (5.6 mGy) protocol was 23.3% lower than that of the 120-kVp (7.3 mGy) protocol.
The use of the 80-kVp protocol improved overall venous attenuation, especially in perforating vein, and provided similarly high diagnostic image quality with a lower radiation dose when compared to the conventional 120-kVp protocol.</abstract><cop>United States</cop><pmid>22992610</pmid><doi>10.1097/RCT.0b013e3182621ea9</doi><tpages>8</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0363-8715 |
ispartof | Journal of computer assisted tomography, 2012-09, Vol.36 (5), p.583-590 |
issn | 0363-8715 1532-3145 |
language | eng |
recordid | cdi_proquest_miscellaneous_1113221223 |
source | MEDLINE; Journals@Ovid Complete |
subjects | Adult Computed tomography Contrast Media Feasibility Studies Humans Iohexol - analogs & derivatives Lower Extremity - blood supply Lower Extremity - diagnostic imaging Male Middle Aged Peripheral Vascular Diseases - diagnostic imaging Phantoms, Imaging Phlebography - methods Prospective Studies Radiographic Image Interpretation, Computer-Assisted Signal-To-Noise Ratio Tomography, X-Ray Computed - methods |
title | Computed tomographic venography for varicose veins of the lower extremities: prospective comparison of 80-kVp and conventional 120-kVp protocols |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-27T21%3A54%3A43IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Computed%20tomographic%20venography%20for%20varicose%20veins%20of%20the%20lower%20extremities:%20prospective%20comparison%20of%2080-kVp%20and%20conventional%20120-kVp%20protocols&rft.jtitle=Journal%20of%20computer%20assisted%20tomography&rft.au=Cho,%20Eun-Suk&rft.date=2012-09-01&rft.volume=36&rft.issue=5&rft.spage=583&rft.epage=590&rft.pages=583-590&rft.issn=0363-8715&rft.eissn=1532-3145&rft_id=info:doi/10.1097/RCT.0b013e3182621ea9&rft_dat=%3Cproquest_cross%3E1113221223%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=1069206187&rft_id=info:pmid/22992610&rfr_iscdi=true |