Carotid CT-angiography: Low versus standard volume contrast media and low kV protocol for 128-slice MDCT

Abstract Availability and utilization of computed tomography angiography has been increasing recently. We aimed to assess the effectiveness of low amount of contrast media and low kV value in order to reduce possible side effects of contrast media and to provide optimization of kV value in the evalu...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:European journal of radiology 2012-09, Vol.81 (9), p.2144-2147
Hauptverfasser: Kayan, Mustafa, Köroğlu, Mert, Yeşildağ, Ahmet, Ceylan, Ergun, Aktaş, Aykut Recep, Yasar, Selçuk, Aynali, Giray, Parlak, Cem, Munduz, Mehmet, Gürses, Cemil
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 2147
container_issue 9
container_start_page 2144
container_title European journal of radiology
container_volume 81
creator Kayan, Mustafa
Köroğlu, Mert
Yeşildağ, Ahmet
Ceylan, Ergun
Aktaş, Aykut Recep
Yasar, Selçuk
Aynali, Giray
Parlak, Cem
Munduz, Mehmet
Gürses, Cemil
description Abstract Availability and utilization of computed tomography angiography has been increasing recently. We aimed to assess the effectiveness of low amount of contrast media and low kV value in order to reduce possible side effects of contrast media and to provide optimization of kV value in the evaluation of the carotid artery with multi-detector computed tomography angiography. Forty one patients were randomized into two groups. Contrast media was administered at a dose of 1 ml/kg in group A patients and of 0.5 ml/kg in group B patients. kV value of 120 in group A and 100 in group B were chosen. Bolus tracking technique was used. Attenuation values of certain arterial segments were measured, and values over 200 HU were considered as significant. North American Symptomatic Carotid Endartherectomy Trial criteria were utilized in the evaluation of stenosis. Image quality in arterial segments of all cases was found to be sufficient for diagnosis. Arterial attenuation values were found to be higher in group B than group A. When compared separately in all arterial segments, there was no statistically significant difference between the groups. For stenosis, 615 arterial segments were evaluated. Moderate stenosis in eight segments and severe stenosis in three segments were identified in group A. Occlusion in three segments, severe stenosis in three segments, and moderate stenosis in 25 segments were detected in group B. Better image quality can be obtained, and the amount of contrast media can be reduced using low kV technique in carotid artery multi-detector computed tomography angiography examination.
doi_str_mv 10.1016/j.ejrad.2011.05.006
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_1033155681</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>1_s2_0_S0720048X11004839</els_id><sourcerecordid>1033155681</sourcerecordid><originalsourceid>FETCH-LOGICAL-c414t-c486ec25ba6dca9b91e7b178a6765c35ae0600bbc4a7dc0e0cf3a0ecd9197e033</originalsourceid><addsrcrecordid>eNqFkU1v1DAQhi0EokvhFyAhH7kkjJ3ETpBAQuGjSIs4sCBuluPMtk698dZOFu2_x2FbDr304jn4eWfsZwh5ySBnwMSbIcch6D7nwFgOVQ4gHpEVqyXPpOTyMVmB5JBBWf8-I89iHACgKhv-lJxxJnkh6nJFrlod_GR72m4yPV5afxn0_ur4lq79H3rAEOdI46THXoeeHrybd0iNH6eg40R32FtN0yV1ib7-Rfeplzfe0a0PlPE6i84apN8-tpvn5MlWu4gvbus5-fn506a9yNbfv3xtP6wzU7JySmct0PCq06I3uukahrJjstZCisoUlUYQAF1nSi17AwhmW2hA0zeskQhFcU5en_qmt9zMGCe1s9Ggc3pEP0fFEsOqStQsocUJNcHHGHCr9sHudDgmSC2K1aD-KVaLYgWVSopT6tXtgLlLAv5n7pwm4N0JwPTNg8WgorE4miQroJlU7-0DA97fyxtnR2u0u8YjxsHPYUwGFVORK1A_li0vS2ZsKUVT_AVUXKLk</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1033155681</pqid></control><display><type>article</type><title>Carotid CT-angiography: Low versus standard volume contrast media and low kV protocol for 128-slice MDCT</title><source>MEDLINE</source><source>Access via ScienceDirect (Elsevier)</source><creator>Kayan, Mustafa ; Köroğlu, Mert ; Yeşildağ, Ahmet ; Ceylan, Ergun ; Aktaş, Aykut Recep ; Yasar, Selçuk ; Aynali, Giray ; Parlak, Cem ; Munduz, Mehmet ; Gürses, Cemil</creator><creatorcontrib>Kayan, Mustafa ; Köroğlu, Mert ; Yeşildağ, Ahmet ; Ceylan, Ergun ; Aktaş, Aykut Recep ; Yasar, Selçuk ; Aynali, Giray ; Parlak, Cem ; Munduz, Mehmet ; Gürses, Cemil</creatorcontrib><description>Abstract Availability and utilization of computed tomography angiography has been increasing recently. We aimed to assess the effectiveness of low amount of contrast media and low kV value in order to reduce possible side effects of contrast media and to provide optimization of kV value in the evaluation of the carotid artery with multi-detector computed tomography angiography. Forty one patients were randomized into two groups. Contrast media was administered at a dose of 1 ml/kg in group A patients and of 0.5 ml/kg in group B patients. kV value of 120 in group A and 100 in group B were chosen. Bolus tracking technique was used. Attenuation values of certain arterial segments were measured, and values over 200 HU were considered as significant. North American Symptomatic Carotid Endartherectomy Trial criteria were utilized in the evaluation of stenosis. Image quality in arterial segments of all cases was found to be sufficient for diagnosis. Arterial attenuation values were found to be higher in group B than group A. When compared separately in all arterial segments, there was no statistically significant difference between the groups. For stenosis, 615 arterial segments were evaluated. Moderate stenosis in eight segments and severe stenosis in three segments were identified in group A. Occlusion in three segments, severe stenosis in three segments, and moderate stenosis in 25 segments were detected in group B. Better image quality can be obtained, and the amount of contrast media can be reduced using low kV technique in carotid artery multi-detector computed tomography angiography examination.</description><identifier>ISSN: 0720-048X</identifier><identifier>EISSN: 1872-7727</identifier><identifier>DOI: 10.1016/j.ejrad.2011.05.006</identifier><identifier>PMID: 21723684</identifier><language>eng</language><publisher>Ireland: Elsevier Ireland Ltd</publisher><subject>Adult ; Aged ; Aged, 80 and over ; Angiography - methods ; Carotid Arteries - diagnostic imaging ; Carotid artery ; Carotid Artery Diseases - diagnostic imaging ; Contrast Media - administration &amp; dosage ; CT angiography ; Dose-Response Relationship, Drug ; Dose-Response Relationship, Radiation ; Female ; Humans ; Iohexol - administration &amp; dosage ; Iohexol - analogs &amp; derivatives ; Low dose ; Low kV ; Male ; Middle Aged ; Radiation Dosage ; Radiology ; Reproducibility of Results ; Sensitivity and Specificity ; Tomography, X-Ray Computed - methods</subject><ispartof>European journal of radiology, 2012-09, Vol.81 (9), p.2144-2147</ispartof><rights>Elsevier Ireland Ltd</rights><rights>2011 Elsevier Ireland Ltd</rights><rights>Copyright © 2011 Elsevier Ireland Ltd. All rights reserved.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c414t-c486ec25ba6dca9b91e7b178a6765c35ae0600bbc4a7dc0e0cf3a0ecd9197e033</citedby><cites>FETCH-LOGICAL-c414t-c486ec25ba6dca9b91e7b178a6765c35ae0600bbc4a7dc0e0cf3a0ecd9197e033</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://dx.doi.org/10.1016/j.ejrad.2011.05.006$$EHTML$$P50$$Gelsevier$$H</linktohtml><link.rule.ids>314,780,784,3550,27924,27925,45995</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/21723684$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Kayan, Mustafa</creatorcontrib><creatorcontrib>Köroğlu, Mert</creatorcontrib><creatorcontrib>Yeşildağ, Ahmet</creatorcontrib><creatorcontrib>Ceylan, Ergun</creatorcontrib><creatorcontrib>Aktaş, Aykut Recep</creatorcontrib><creatorcontrib>Yasar, Selçuk</creatorcontrib><creatorcontrib>Aynali, Giray</creatorcontrib><creatorcontrib>Parlak, Cem</creatorcontrib><creatorcontrib>Munduz, Mehmet</creatorcontrib><creatorcontrib>Gürses, Cemil</creatorcontrib><title>Carotid CT-angiography: Low versus standard volume contrast media and low kV protocol for 128-slice MDCT</title><title>European journal of radiology</title><addtitle>Eur J Radiol</addtitle><description>Abstract Availability and utilization of computed tomography angiography has been increasing recently. We aimed to assess the effectiveness of low amount of contrast media and low kV value in order to reduce possible side effects of contrast media and to provide optimization of kV value in the evaluation of the carotid artery with multi-detector computed tomography angiography. Forty one patients were randomized into two groups. Contrast media was administered at a dose of 1 ml/kg in group A patients and of 0.5 ml/kg in group B patients. kV value of 120 in group A and 100 in group B were chosen. Bolus tracking technique was used. Attenuation values of certain arterial segments were measured, and values over 200 HU were considered as significant. North American Symptomatic Carotid Endartherectomy Trial criteria were utilized in the evaluation of stenosis. Image quality in arterial segments of all cases was found to be sufficient for diagnosis. Arterial attenuation values were found to be higher in group B than group A. When compared separately in all arterial segments, there was no statistically significant difference between the groups. For stenosis, 615 arterial segments were evaluated. Moderate stenosis in eight segments and severe stenosis in three segments were identified in group A. Occlusion in three segments, severe stenosis in three segments, and moderate stenosis in 25 segments were detected in group B. Better image quality can be obtained, and the amount of contrast media can be reduced using low kV technique in carotid artery multi-detector computed tomography angiography examination.</description><subject>Adult</subject><subject>Aged</subject><subject>Aged, 80 and over</subject><subject>Angiography - methods</subject><subject>Carotid Arteries - diagnostic imaging</subject><subject>Carotid artery</subject><subject>Carotid Artery Diseases - diagnostic imaging</subject><subject>Contrast Media - administration &amp; dosage</subject><subject>CT angiography</subject><subject>Dose-Response Relationship, Drug</subject><subject>Dose-Response Relationship, Radiation</subject><subject>Female</subject><subject>Humans</subject><subject>Iohexol - administration &amp; dosage</subject><subject>Iohexol - analogs &amp; derivatives</subject><subject>Low dose</subject><subject>Low kV</subject><subject>Male</subject><subject>Middle Aged</subject><subject>Radiation Dosage</subject><subject>Radiology</subject><subject>Reproducibility of Results</subject><subject>Sensitivity and Specificity</subject><subject>Tomography, X-Ray Computed - methods</subject><issn>0720-048X</issn><issn>1872-7727</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2012</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNqFkU1v1DAQhi0EokvhFyAhH7kkjJ3ETpBAQuGjSIs4sCBuluPMtk698dZOFu2_x2FbDr304jn4eWfsZwh5ySBnwMSbIcch6D7nwFgOVQ4gHpEVqyXPpOTyMVmB5JBBWf8-I89iHACgKhv-lJxxJnkh6nJFrlod_GR72m4yPV5afxn0_ur4lq79H3rAEOdI46THXoeeHrybd0iNH6eg40R32FtN0yV1ib7-Rfeplzfe0a0PlPE6i84apN8-tpvn5MlWu4gvbus5-fn506a9yNbfv3xtP6wzU7JySmct0PCq06I3uukahrJjstZCisoUlUYQAF1nSi17AwhmW2hA0zeskQhFcU5en_qmt9zMGCe1s9Ggc3pEP0fFEsOqStQsocUJNcHHGHCr9sHudDgmSC2K1aD-KVaLYgWVSopT6tXtgLlLAv5n7pwm4N0JwPTNg8WgorE4miQroJlU7-0DA97fyxtnR2u0u8YjxsHPYUwGFVORK1A_li0vS2ZsKUVT_AVUXKLk</recordid><startdate>20120901</startdate><enddate>20120901</enddate><creator>Kayan, Mustafa</creator><creator>Köroğlu, Mert</creator><creator>Yeşildağ, Ahmet</creator><creator>Ceylan, Ergun</creator><creator>Aktaş, Aykut Recep</creator><creator>Yasar, Selçuk</creator><creator>Aynali, Giray</creator><creator>Parlak, Cem</creator><creator>Munduz, Mehmet</creator><creator>Gürses, Cemil</creator><general>Elsevier Ireland Ltd</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>20120901</creationdate><title>Carotid CT-angiography: Low versus standard volume contrast media and low kV protocol for 128-slice MDCT</title><author>Kayan, Mustafa ; Köroğlu, Mert ; Yeşildağ, Ahmet ; Ceylan, Ergun ; Aktaş, Aykut Recep ; Yasar, Selçuk ; Aynali, Giray ; Parlak, Cem ; Munduz, Mehmet ; Gürses, Cemil</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c414t-c486ec25ba6dca9b91e7b178a6765c35ae0600bbc4a7dc0e0cf3a0ecd9197e033</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2012</creationdate><topic>Adult</topic><topic>Aged</topic><topic>Aged, 80 and over</topic><topic>Angiography - methods</topic><topic>Carotid Arteries - diagnostic imaging</topic><topic>Carotid artery</topic><topic>Carotid Artery Diseases - diagnostic imaging</topic><topic>Contrast Media - administration &amp; dosage</topic><topic>CT angiography</topic><topic>Dose-Response Relationship, Drug</topic><topic>Dose-Response Relationship, Radiation</topic><topic>Female</topic><topic>Humans</topic><topic>Iohexol - administration &amp; dosage</topic><topic>Iohexol - analogs &amp; derivatives</topic><topic>Low dose</topic><topic>Low kV</topic><topic>Male</topic><topic>Middle Aged</topic><topic>Radiation Dosage</topic><topic>Radiology</topic><topic>Reproducibility of Results</topic><topic>Sensitivity and Specificity</topic><topic>Tomography, X-Ray Computed - methods</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Kayan, Mustafa</creatorcontrib><creatorcontrib>Köroğlu, Mert</creatorcontrib><creatorcontrib>Yeşildağ, Ahmet</creatorcontrib><creatorcontrib>Ceylan, Ergun</creatorcontrib><creatorcontrib>Aktaş, Aykut Recep</creatorcontrib><creatorcontrib>Yasar, Selçuk</creatorcontrib><creatorcontrib>Aynali, Giray</creatorcontrib><creatorcontrib>Parlak, Cem</creatorcontrib><creatorcontrib>Munduz, Mehmet</creatorcontrib><creatorcontrib>Gürses, Cemil</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>European journal of radiology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Kayan, Mustafa</au><au>Köroğlu, Mert</au><au>Yeşildağ, Ahmet</au><au>Ceylan, Ergun</au><au>Aktaş, Aykut Recep</au><au>Yasar, Selçuk</au><au>Aynali, Giray</au><au>Parlak, Cem</au><au>Munduz, Mehmet</au><au>Gürses, Cemil</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Carotid CT-angiography: Low versus standard volume contrast media and low kV protocol for 128-slice MDCT</atitle><jtitle>European journal of radiology</jtitle><addtitle>Eur J Radiol</addtitle><date>2012-09-01</date><risdate>2012</risdate><volume>81</volume><issue>9</issue><spage>2144</spage><epage>2147</epage><pages>2144-2147</pages><issn>0720-048X</issn><eissn>1872-7727</eissn><abstract>Abstract Availability and utilization of computed tomography angiography has been increasing recently. We aimed to assess the effectiveness of low amount of contrast media and low kV value in order to reduce possible side effects of contrast media and to provide optimization of kV value in the evaluation of the carotid artery with multi-detector computed tomography angiography. Forty one patients were randomized into two groups. Contrast media was administered at a dose of 1 ml/kg in group A patients and of 0.5 ml/kg in group B patients. kV value of 120 in group A and 100 in group B were chosen. Bolus tracking technique was used. Attenuation values of certain arterial segments were measured, and values over 200 HU were considered as significant. North American Symptomatic Carotid Endartherectomy Trial criteria were utilized in the evaluation of stenosis. Image quality in arterial segments of all cases was found to be sufficient for diagnosis. Arterial attenuation values were found to be higher in group B than group A. When compared separately in all arterial segments, there was no statistically significant difference between the groups. For stenosis, 615 arterial segments were evaluated. Moderate stenosis in eight segments and severe stenosis in three segments were identified in group A. Occlusion in three segments, severe stenosis in three segments, and moderate stenosis in 25 segments were detected in group B. Better image quality can be obtained, and the amount of contrast media can be reduced using low kV technique in carotid artery multi-detector computed tomography angiography examination.</abstract><cop>Ireland</cop><pub>Elsevier Ireland Ltd</pub><pmid>21723684</pmid><doi>10.1016/j.ejrad.2011.05.006</doi><tpages>4</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0720-048X
ispartof European journal of radiology, 2012-09, Vol.81 (9), p.2144-2147
issn 0720-048X
1872-7727
language eng
recordid cdi_proquest_miscellaneous_1033155681
source MEDLINE; Access via ScienceDirect (Elsevier)
subjects Adult
Aged
Aged, 80 and over
Angiography - methods
Carotid Arteries - diagnostic imaging
Carotid artery
Carotid Artery Diseases - diagnostic imaging
Contrast Media - administration & dosage
CT angiography
Dose-Response Relationship, Drug
Dose-Response Relationship, Radiation
Female
Humans
Iohexol - administration & dosage
Iohexol - analogs & derivatives
Low dose
Low kV
Male
Middle Aged
Radiation Dosage
Radiology
Reproducibility of Results
Sensitivity and Specificity
Tomography, X-Ray Computed - methods
title Carotid CT-angiography: Low versus standard volume contrast media and low kV protocol for 128-slice MDCT
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-24T21%3A26%3A29IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Carotid%20CT-angiography:%20Low%20versus%20standard%20volume%20contrast%20media%20and%20low%20kV%20protocol%20for%20128-slice%20MDCT&rft.jtitle=European%20journal%20of%20radiology&rft.au=Kayan,%20Mustafa&rft.date=2012-09-01&rft.volume=81&rft.issue=9&rft.spage=2144&rft.epage=2147&rft.pages=2144-2147&rft.issn=0720-048X&rft.eissn=1872-7727&rft_id=info:doi/10.1016/j.ejrad.2011.05.006&rft_dat=%3Cproquest_cross%3E1033155681%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=1033155681&rft_id=info:pmid/21723684&rft_els_id=1_s2_0_S0720048X11004839&rfr_iscdi=true