Asymmetric Motion Distribution Between Components of a Mobile-Core Lumbar Disc Prosthesis: An Explanation of Unequal Wear Distribution in Explanted CHARITÉ Polyethylene Cores

BACKGROUND:The biconvex mobile core of the CHARITÉ lumbar disc prosthesis forms two joints (spherical bearings) with the metal end plates. We quantified the intra-prosthesis motion to test the hypothesis that the total prosthesis motion would not be equally distributed between the two bearings of im...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of bone and joint surgery. American volume 2012-05, Vol.94 (9), p.846-854
Hauptverfasser: Patwardhan, Avinash G, Havey, Robert M, Wharton, Nicholas D, Tsitsopoulos, Parmenion P, Newman, Patrick, Carandang, Gerard, Voronov, Leonard I
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 854
container_issue 9
container_start_page 846
container_title Journal of bone and joint surgery. American volume
container_volume 94
creator Patwardhan, Avinash G
Havey, Robert M
Wharton, Nicholas D
Tsitsopoulos, Parmenion P
Newman, Patrick
Carandang, Gerard
Voronov, Leonard I
description BACKGROUND:The biconvex mobile core of the CHARITÉ lumbar disc prosthesis forms two joints (spherical bearings) with the metal end plates. We quantified the intra-prosthesis motion to test the hypothesis that the total prosthesis motion would not be equally distributed between the two bearings of implanted CHARITÉ discs, which might explain the unequal wear distribution reported in explanted cores. METHODS:The hypothesis was tested by studying the flexion-extension motion responses of (1) twenty-six monosegmental CHARITÉ III discs implanted in nineteen human cadaveric lumbar spines, and (2) twenty-one CHARITÉ III discs (fifteen monosegmental, six bisegmental) implanted in eighteen patients in other published clinical studies. Intra-prosthesis motions were quantified with use of a radiographic image analysis technique. RESULTS:Eighty-eight percent of the CHARITÉ discs implanted in cadaveric specimens exhibited larger motion at the superior bearing, with 54% demonstrating more than twice as much motion at the superior bearing as at the inferior bearing. The ratio of motion at the superior bearing to motion at the inferior bearing averaged 2.68 ± 1.84, which was significantly larger than 1.0 (p < 0.001). Ninety percent of prostheses implanted in patients showed larger motion at the superior bearing. The motion ratio averaged 2.39 ± 2.47 for monosegmental cases and 2.55 ± 2.66 for all cases; both ratios were significantly larger than 1.0 (p < 0.05). CONCLUSIONS:We found preferentially larger motion at the superior bearing of the CHARITÉ discs implanted in human cadaveric lumbar spines and in patients, regardless of the implanted level. CLINICAL RELEVANCE:The findings regarding unequal motion distribution between the dual bearings of mobile-core disc prostheses are relevant to improving in vitro wear testing protocols to better replicate in vivo wear-producing conditions.
doi_str_mv 10.2106/JBJS.J.00638
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_1011179755</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>1011179755</sourcerecordid><originalsourceid>FETCH-LOGICAL-c3231-875db3d559972294fecf0fa78d1883dcc489b13ef882a50e5061fd9f9570de223</originalsourceid><addsrcrecordid>eNpFkc1u1DAUhS0EokNhxxp5g8SCDNf2OD_spqHQjgZRQSuWkZNcawJOPLUdTecR-jQ8BC-G56d0ZV3ru-fco0PIawZTziD9sDhb_JgupgCpyJ-QCZNCJkzk6VMyAeAsKYSUJ-SF978AYDaD7Dk54VxKnmZyQv7M_bbvMbiuoV9t6OxAP3U-jvW4H84wbBAHWtp-bQccgqdWUxXZujOYlNYhXY59rdxur6FXzvqwQt_5j3Q-0PO7tVGD2kvFvZsBb0dl6E888I8-3QMbsKXlxfz75fXfe3plzRbDamtwQLrz8i_JM62Mx1fH95TcfD6_Li-S5bcvl-V8mTSCC5bkmWxr0UpZFBnnxUxjo0GrLG9Znou2aWZ5UTOBOs-5koASUqbbQhcygxY5F6fk3UF37eztiD5UfYyHJl6IdvQVA8ZYVmRSRvT9AW1idu9QV2vX9cptI1TtKqp2FVWLal9RxN8clce6x_Y__NBJBN4eAeUbZbRTQ9P5Ry5mArH3nR24jTUBnf9txg26aoXKhFU0i22nXCQcGAcJHJL4w5j4BwSZq1c</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1011179755</pqid></control><display><type>article</type><title>Asymmetric Motion Distribution Between Components of a Mobile-Core Lumbar Disc Prosthesis: An Explanation of Unequal Wear Distribution in Explanted CHARITÉ Polyethylene Cores</title><source>MEDLINE</source><source>Alma/SFX Local Collection</source><source>Journals@Ovid Complete</source><creator>Patwardhan, Avinash G ; Havey, Robert M ; Wharton, Nicholas D ; Tsitsopoulos, Parmenion P ; Newman, Patrick ; Carandang, Gerard ; Voronov, Leonard I</creator><creatorcontrib>Patwardhan, Avinash G ; Havey, Robert M ; Wharton, Nicholas D ; Tsitsopoulos, Parmenion P ; Newman, Patrick ; Carandang, Gerard ; Voronov, Leonard I</creatorcontrib><description>BACKGROUND:The biconvex mobile core of the CHARITÉ lumbar disc prosthesis forms two joints (spherical bearings) with the metal end plates. We quantified the intra-prosthesis motion to test the hypothesis that the total prosthesis motion would not be equally distributed between the two bearings of implanted CHARITÉ discs, which might explain the unequal wear distribution reported in explanted cores. METHODS:The hypothesis was tested by studying the flexion-extension motion responses of (1) twenty-six monosegmental CHARITÉ III discs implanted in nineteen human cadaveric lumbar spines, and (2) twenty-one CHARITÉ III discs (fifteen monosegmental, six bisegmental) implanted in eighteen patients in other published clinical studies. Intra-prosthesis motions were quantified with use of a radiographic image analysis technique. RESULTS:Eighty-eight percent of the CHARITÉ discs implanted in cadaveric specimens exhibited larger motion at the superior bearing, with 54% demonstrating more than twice as much motion at the superior bearing as at the inferior bearing. The ratio of motion at the superior bearing to motion at the inferior bearing averaged 2.68 ± 1.84, which was significantly larger than 1.0 (p &lt; 0.001). Ninety percent of prostheses implanted in patients showed larger motion at the superior bearing. The motion ratio averaged 2.39 ± 2.47 for monosegmental cases and 2.55 ± 2.66 for all cases; both ratios were significantly larger than 1.0 (p &lt; 0.05). CONCLUSIONS:We found preferentially larger motion at the superior bearing of the CHARITÉ discs implanted in human cadaveric lumbar spines and in patients, regardless of the implanted level. CLINICAL RELEVANCE:The findings regarding unequal motion distribution between the dual bearings of mobile-core disc prostheses are relevant to improving in vitro wear testing protocols to better replicate in vivo wear-producing conditions.</description><identifier>ISSN: 0021-9355</identifier><identifier>EISSN: 1535-1386</identifier><identifier>DOI: 10.2106/JBJS.J.00638</identifier><identifier>PMID: 22552675</identifier><identifier>CODEN: JBJSA3</identifier><language>eng</language><publisher>Boston, MA: Copyright by The Journal of Bone and Joint Surgery, Incorporated</publisher><subject>Adult ; Biological and medical sciences ; Cadaver ; Diseases of the osteoarticular system ; Equipment Failure Analysis ; Female ; Humans ; Intervertebral Disc - surgery ; Joint Prosthesis ; Lumbar Vertebrae - surgery ; Male ; Medical sciences ; Middle Aged ; Motion ; Orthopedic surgery ; Polyethylene ; Range of Motion, Articular - physiology ; Surgery (general aspects). Transplantations, organ and tissue grafts. Graft diseases ; Weight-Bearing - physiology</subject><ispartof>Journal of bone and joint surgery. American volume, 2012-05, Vol.94 (9), p.846-854</ispartof><rights>Copyright 2012 by The Journal of Bone and Joint Surgery, Incorporated</rights><rights>2015 INIST-CNRS</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><cites>FETCH-LOGICAL-c3231-875db3d559972294fecf0fa78d1883dcc489b13ef882a50e5061fd9f9570de223</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,776,780,27901,27902</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&amp;idt=25990355$$DView record in Pascal Francis$$Hfree_for_read</backlink><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/22552675$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Patwardhan, Avinash G</creatorcontrib><creatorcontrib>Havey, Robert M</creatorcontrib><creatorcontrib>Wharton, Nicholas D</creatorcontrib><creatorcontrib>Tsitsopoulos, Parmenion P</creatorcontrib><creatorcontrib>Newman, Patrick</creatorcontrib><creatorcontrib>Carandang, Gerard</creatorcontrib><creatorcontrib>Voronov, Leonard I</creatorcontrib><title>Asymmetric Motion Distribution Between Components of a Mobile-Core Lumbar Disc Prosthesis: An Explanation of Unequal Wear Distribution in Explanted CHARITÉ Polyethylene Cores</title><title>Journal of bone and joint surgery. American volume</title><addtitle>J Bone Joint Surg Am</addtitle><description>BACKGROUND:The biconvex mobile core of the CHARITÉ lumbar disc prosthesis forms two joints (spherical bearings) with the metal end plates. We quantified the intra-prosthesis motion to test the hypothesis that the total prosthesis motion would not be equally distributed between the two bearings of implanted CHARITÉ discs, which might explain the unequal wear distribution reported in explanted cores. METHODS:The hypothesis was tested by studying the flexion-extension motion responses of (1) twenty-six monosegmental CHARITÉ III discs implanted in nineteen human cadaveric lumbar spines, and (2) twenty-one CHARITÉ III discs (fifteen monosegmental, six bisegmental) implanted in eighteen patients in other published clinical studies. Intra-prosthesis motions were quantified with use of a radiographic image analysis technique. RESULTS:Eighty-eight percent of the CHARITÉ discs implanted in cadaveric specimens exhibited larger motion at the superior bearing, with 54% demonstrating more than twice as much motion at the superior bearing as at the inferior bearing. The ratio of motion at the superior bearing to motion at the inferior bearing averaged 2.68 ± 1.84, which was significantly larger than 1.0 (p &lt; 0.001). Ninety percent of prostheses implanted in patients showed larger motion at the superior bearing. The motion ratio averaged 2.39 ± 2.47 for monosegmental cases and 2.55 ± 2.66 for all cases; both ratios were significantly larger than 1.0 (p &lt; 0.05). CONCLUSIONS:We found preferentially larger motion at the superior bearing of the CHARITÉ discs implanted in human cadaveric lumbar spines and in patients, regardless of the implanted level. CLINICAL RELEVANCE:The findings regarding unequal motion distribution between the dual bearings of mobile-core disc prostheses are relevant to improving in vitro wear testing protocols to better replicate in vivo wear-producing conditions.</description><subject>Adult</subject><subject>Biological and medical sciences</subject><subject>Cadaver</subject><subject>Diseases of the osteoarticular system</subject><subject>Equipment Failure Analysis</subject><subject>Female</subject><subject>Humans</subject><subject>Intervertebral Disc - surgery</subject><subject>Joint Prosthesis</subject><subject>Lumbar Vertebrae - surgery</subject><subject>Male</subject><subject>Medical sciences</subject><subject>Middle Aged</subject><subject>Motion</subject><subject>Orthopedic surgery</subject><subject>Polyethylene</subject><subject>Range of Motion, Articular - physiology</subject><subject>Surgery (general aspects). Transplantations, organ and tissue grafts. Graft diseases</subject><subject>Weight-Bearing - physiology</subject><issn>0021-9355</issn><issn>1535-1386</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2012</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNpFkc1u1DAUhS0EokNhxxp5g8SCDNf2OD_spqHQjgZRQSuWkZNcawJOPLUdTecR-jQ8BC-G56d0ZV3ru-fco0PIawZTziD9sDhb_JgupgCpyJ-QCZNCJkzk6VMyAeAsKYSUJ-SF978AYDaD7Dk54VxKnmZyQv7M_bbvMbiuoV9t6OxAP3U-jvW4H84wbBAHWtp-bQccgqdWUxXZujOYlNYhXY59rdxur6FXzvqwQt_5j3Q-0PO7tVGD2kvFvZsBb0dl6E888I8-3QMbsKXlxfz75fXfe3plzRbDamtwQLrz8i_JM62Mx1fH95TcfD6_Li-S5bcvl-V8mTSCC5bkmWxr0UpZFBnnxUxjo0GrLG9Znou2aWZ5UTOBOs-5koASUqbbQhcygxY5F6fk3UF37eztiD5UfYyHJl6IdvQVA8ZYVmRSRvT9AW1idu9QV2vX9cptI1TtKqp2FVWLal9RxN8clce6x_Y__NBJBN4eAeUbZbRTQ9P5Ry5mArH3nR24jTUBnf9txg26aoXKhFU0i22nXCQcGAcJHJL4w5j4BwSZq1c</recordid><startdate>20120502</startdate><enddate>20120502</enddate><creator>Patwardhan, Avinash G</creator><creator>Havey, Robert M</creator><creator>Wharton, Nicholas D</creator><creator>Tsitsopoulos, Parmenion P</creator><creator>Newman, Patrick</creator><creator>Carandang, Gerard</creator><creator>Voronov, Leonard I</creator><general>Copyright by The Journal of Bone and Joint Surgery, Incorporated</general><general>Journal of Bone and Joint Surgery Incorporated</general><scope>IQODW</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>20120502</creationdate><title>Asymmetric Motion Distribution Between Components of a Mobile-Core Lumbar Disc Prosthesis: An Explanation of Unequal Wear Distribution in Explanted CHARITÉ Polyethylene Cores</title><author>Patwardhan, Avinash G ; Havey, Robert M ; Wharton, Nicholas D ; Tsitsopoulos, Parmenion P ; Newman, Patrick ; Carandang, Gerard ; Voronov, Leonard I</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c3231-875db3d559972294fecf0fa78d1883dcc489b13ef882a50e5061fd9f9570de223</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2012</creationdate><topic>Adult</topic><topic>Biological and medical sciences</topic><topic>Cadaver</topic><topic>Diseases of the osteoarticular system</topic><topic>Equipment Failure Analysis</topic><topic>Female</topic><topic>Humans</topic><topic>Intervertebral Disc - surgery</topic><topic>Joint Prosthesis</topic><topic>Lumbar Vertebrae - surgery</topic><topic>Male</topic><topic>Medical sciences</topic><topic>Middle Aged</topic><topic>Motion</topic><topic>Orthopedic surgery</topic><topic>Polyethylene</topic><topic>Range of Motion, Articular - physiology</topic><topic>Surgery (general aspects). Transplantations, organ and tissue grafts. Graft diseases</topic><topic>Weight-Bearing - physiology</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Patwardhan, Avinash G</creatorcontrib><creatorcontrib>Havey, Robert M</creatorcontrib><creatorcontrib>Wharton, Nicholas D</creatorcontrib><creatorcontrib>Tsitsopoulos, Parmenion P</creatorcontrib><creatorcontrib>Newman, Patrick</creatorcontrib><creatorcontrib>Carandang, Gerard</creatorcontrib><creatorcontrib>Voronov, Leonard I</creatorcontrib><collection>Pascal-Francis</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Journal of bone and joint surgery. American volume</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Patwardhan, Avinash G</au><au>Havey, Robert M</au><au>Wharton, Nicholas D</au><au>Tsitsopoulos, Parmenion P</au><au>Newman, Patrick</au><au>Carandang, Gerard</au><au>Voronov, Leonard I</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Asymmetric Motion Distribution Between Components of a Mobile-Core Lumbar Disc Prosthesis: An Explanation of Unequal Wear Distribution in Explanted CHARITÉ Polyethylene Cores</atitle><jtitle>Journal of bone and joint surgery. American volume</jtitle><addtitle>J Bone Joint Surg Am</addtitle><date>2012-05-02</date><risdate>2012</risdate><volume>94</volume><issue>9</issue><spage>846</spage><epage>854</epage><pages>846-854</pages><issn>0021-9355</issn><eissn>1535-1386</eissn><coden>JBJSA3</coden><abstract>BACKGROUND:The biconvex mobile core of the CHARITÉ lumbar disc prosthesis forms two joints (spherical bearings) with the metal end plates. We quantified the intra-prosthesis motion to test the hypothesis that the total prosthesis motion would not be equally distributed between the two bearings of implanted CHARITÉ discs, which might explain the unequal wear distribution reported in explanted cores. METHODS:The hypothesis was tested by studying the flexion-extension motion responses of (1) twenty-six monosegmental CHARITÉ III discs implanted in nineteen human cadaveric lumbar spines, and (2) twenty-one CHARITÉ III discs (fifteen monosegmental, six bisegmental) implanted in eighteen patients in other published clinical studies. Intra-prosthesis motions were quantified with use of a radiographic image analysis technique. RESULTS:Eighty-eight percent of the CHARITÉ discs implanted in cadaveric specimens exhibited larger motion at the superior bearing, with 54% demonstrating more than twice as much motion at the superior bearing as at the inferior bearing. The ratio of motion at the superior bearing to motion at the inferior bearing averaged 2.68 ± 1.84, which was significantly larger than 1.0 (p &lt; 0.001). Ninety percent of prostheses implanted in patients showed larger motion at the superior bearing. The motion ratio averaged 2.39 ± 2.47 for monosegmental cases and 2.55 ± 2.66 for all cases; both ratios were significantly larger than 1.0 (p &lt; 0.05). CONCLUSIONS:We found preferentially larger motion at the superior bearing of the CHARITÉ discs implanted in human cadaveric lumbar spines and in patients, regardless of the implanted level. CLINICAL RELEVANCE:The findings regarding unequal motion distribution between the dual bearings of mobile-core disc prostheses are relevant to improving in vitro wear testing protocols to better replicate in vivo wear-producing conditions.</abstract><cop>Boston, MA</cop><pub>Copyright by The Journal of Bone and Joint Surgery, Incorporated</pub><pmid>22552675</pmid><doi>10.2106/JBJS.J.00638</doi><tpages>9</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0021-9355
ispartof Journal of bone and joint surgery. American volume, 2012-05, Vol.94 (9), p.846-854
issn 0021-9355
1535-1386
language eng
recordid cdi_proquest_miscellaneous_1011179755
source MEDLINE; Alma/SFX Local Collection; Journals@Ovid Complete
subjects Adult
Biological and medical sciences
Cadaver
Diseases of the osteoarticular system
Equipment Failure Analysis
Female
Humans
Intervertebral Disc - surgery
Joint Prosthesis
Lumbar Vertebrae - surgery
Male
Medical sciences
Middle Aged
Motion
Orthopedic surgery
Polyethylene
Range of Motion, Articular - physiology
Surgery (general aspects). Transplantations, organ and tissue grafts. Graft diseases
Weight-Bearing - physiology
title Asymmetric Motion Distribution Between Components of a Mobile-Core Lumbar Disc Prosthesis: An Explanation of Unequal Wear Distribution in Explanted CHARITÉ Polyethylene Cores
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-14T09%3A11%3A23IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Asymmetric%20Motion%20Distribution%20Between%20Components%20of%20a%20Mobile-Core%20Lumbar%20Disc%20Prosthesis:%20An%20Explanation%20of%20Unequal%20Wear%20Distribution%20in%20Explanted%20CHARIT%C3%89%20Polyethylene%20Cores&rft.jtitle=Journal%20of%20bone%20and%20joint%20surgery.%20American%20volume&rft.au=Patwardhan,%20Avinash%20G&rft.date=2012-05-02&rft.volume=94&rft.issue=9&rft.spage=846&rft.epage=854&rft.pages=846-854&rft.issn=0021-9355&rft.eissn=1535-1386&rft.coden=JBJSA3&rft_id=info:doi/10.2106/JBJS.J.00638&rft_dat=%3Cproquest_cross%3E1011179755%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=1011179755&rft_id=info:pmid/22552675&rfr_iscdi=true