Key comparison BIPM.RI(I)-K2 of the air-kerma standards of the ENEA-INMRI, Italy and the BIPM in low-energy x-rays
A key comparison has been made between the air-kerma standards of the ENEA-INMRI, Italy and the BIPM in the low-energy x-ray range. The results show the standards to be in agreement at the level of the standard uncertainty for the comparison of 1.9 parts in ... No significant trend with radiation qu...
Gespeichert in:
Veröffentlicht in: | Metrologia 2011-01, Vol.48, p.1 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | |
---|---|
container_issue | |
container_start_page | 1 |
container_title | Metrologia |
container_volume | 48 |
creator | Burns, D T Kessler, C Roger, P Toni, M P Pinto, M Bovi, M Cappadozzi, G Silvestri, C |
description | A key comparison has been made between the air-kerma standards of the ENEA-INMRI, Italy and the BIPM in the low-energy x-ray range. The results show the standards to be in agreement at the level of the standard uncertainty for the comparison of 1.9 parts in ... No significant trend with radiation quality is observed. The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database. (ProQuest: ... denotes formulae/symbols omitted.) |
format | Article |
fullrecord | <record><control><sourceid>proquest</sourceid><recordid>TN_cdi_proquest_journals_911196371</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2539154981</sourcerecordid><originalsourceid>FETCH-proquest_journals_9111963713</originalsourceid><addsrcrecordid>eNqNzM0OwUAUhuGJkKifezixIjHSabWjS6RiIhURe5kwKDXDmQq9ez9hb_UtnjdfiTgsHDDKAx6UieO6XkiZH_WrpGbt0XUZ9wLuEJypAjbmfJGYWqNhJBZJbynaokNnHpgd5AcFMkV6UniWYHOptxK39kfxPB5SMU-Wogsil1kBr-Aj7ydINWTmTpVWuC_gQVEWtkEqO5lZ1fxunbQm8Wo8pRc015uy-fpobqhftI4YY1Hoc-b_FT0Be3hH_g</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>911196371</pqid></control><display><type>article</type><title>Key comparison BIPM.RI(I)-K2 of the air-kerma standards of the ENEA-INMRI, Italy and the BIPM in low-energy x-rays</title><source>Institute of Physics Journals</source><creator>Burns, D T ; Kessler, C ; Roger, P ; Toni, M P ; Pinto, M ; Bovi, M ; Cappadozzi, G ; Silvestri, C</creator><creatorcontrib>Burns, D T ; Kessler, C ; Roger, P ; Toni, M P ; Pinto, M ; Bovi, M ; Cappadozzi, G ; Silvestri, C</creatorcontrib><description>A key comparison has been made between the air-kerma standards of the ENEA-INMRI, Italy and the BIPM in the low-energy x-ray range. The results show the standards to be in agreement at the level of the standard uncertainty for the comparison of 1.9 parts in ... No significant trend with radiation quality is observed. The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database. (ProQuest: ... denotes formulae/symbols omitted.)</description><identifier>ISSN: 0026-1394</identifier><identifier>EISSN: 1681-7575</identifier><language>eng</language><publisher>Bristol: IOP Publishing</publisher><subject>Comparative analysis ; Radiation ; Uncertainty ; X-rays</subject><ispartof>Metrologia, 2011-01, Vol.48, p.1</ispartof><rights>Copyright Bureau International des Poids et Mesures 2011</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784</link.rule.ids></links><search><creatorcontrib>Burns, D T</creatorcontrib><creatorcontrib>Kessler, C</creatorcontrib><creatorcontrib>Roger, P</creatorcontrib><creatorcontrib>Toni, M P</creatorcontrib><creatorcontrib>Pinto, M</creatorcontrib><creatorcontrib>Bovi, M</creatorcontrib><creatorcontrib>Cappadozzi, G</creatorcontrib><creatorcontrib>Silvestri, C</creatorcontrib><title>Key comparison BIPM.RI(I)-K2 of the air-kerma standards of the ENEA-INMRI, Italy and the BIPM in low-energy x-rays</title><title>Metrologia</title><description>A key comparison has been made between the air-kerma standards of the ENEA-INMRI, Italy and the BIPM in the low-energy x-ray range. The results show the standards to be in agreement at the level of the standard uncertainty for the comparison of 1.9 parts in ... No significant trend with radiation quality is observed. The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database. (ProQuest: ... denotes formulae/symbols omitted.)</description><subject>Comparative analysis</subject><subject>Radiation</subject><subject>Uncertainty</subject><subject>X-rays</subject><issn>0026-1394</issn><issn>1681-7575</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2011</creationdate><recordtype>article</recordtype><recordid>eNqNzM0OwUAUhuGJkKifezixIjHSabWjS6RiIhURe5kwKDXDmQq9ez9hb_UtnjdfiTgsHDDKAx6UieO6XkiZH_WrpGbt0XUZ9wLuEJypAjbmfJGYWqNhJBZJbynaokNnHpgd5AcFMkV6UniWYHOptxK39kfxPB5SMU-Wogsil1kBr-Aj7ydINWTmTpVWuC_gQVEWtkEqO5lZ1fxunbQm8Wo8pRc015uy-fpobqhftI4YY1Hoc-b_FT0Be3hH_g</recordid><startdate>20110101</startdate><enddate>20110101</enddate><creator>Burns, D T</creator><creator>Kessler, C</creator><creator>Roger, P</creator><creator>Toni, M P</creator><creator>Pinto, M</creator><creator>Bovi, M</creator><creator>Cappadozzi, G</creator><creator>Silvestri, C</creator><general>IOP Publishing</general><scope>7U5</scope><scope>8FD</scope><scope>L7M</scope></search><sort><creationdate>20110101</creationdate><title>Key comparison BIPM.RI(I)-K2 of the air-kerma standards of the ENEA-INMRI, Italy and the BIPM in low-energy x-rays</title><author>Burns, D T ; Kessler, C ; Roger, P ; Toni, M P ; Pinto, M ; Bovi, M ; Cappadozzi, G ; Silvestri, C</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-proquest_journals_9111963713</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2011</creationdate><topic>Comparative analysis</topic><topic>Radiation</topic><topic>Uncertainty</topic><topic>X-rays</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Burns, D T</creatorcontrib><creatorcontrib>Kessler, C</creatorcontrib><creatorcontrib>Roger, P</creatorcontrib><creatorcontrib>Toni, M P</creatorcontrib><creatorcontrib>Pinto, M</creatorcontrib><creatorcontrib>Bovi, M</creatorcontrib><creatorcontrib>Cappadozzi, G</creatorcontrib><creatorcontrib>Silvestri, C</creatorcontrib><collection>Solid State and Superconductivity Abstracts</collection><collection>Technology Research Database</collection><collection>Advanced Technologies Database with Aerospace</collection><jtitle>Metrologia</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Burns, D T</au><au>Kessler, C</au><au>Roger, P</au><au>Toni, M P</au><au>Pinto, M</au><au>Bovi, M</au><au>Cappadozzi, G</au><au>Silvestri, C</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Key comparison BIPM.RI(I)-K2 of the air-kerma standards of the ENEA-INMRI, Italy and the BIPM in low-energy x-rays</atitle><jtitle>Metrologia</jtitle><date>2011-01-01</date><risdate>2011</risdate><volume>48</volume><spage>1</spage><pages>1-</pages><issn>0026-1394</issn><eissn>1681-7575</eissn><abstract>A key comparison has been made between the air-kerma standards of the ENEA-INMRI, Italy and the BIPM in the low-energy x-ray range. The results show the standards to be in agreement at the level of the standard uncertainty for the comparison of 1.9 parts in ... No significant trend with radiation quality is observed. The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database. (ProQuest: ... denotes formulae/symbols omitted.)</abstract><cop>Bristol</cop><pub>IOP Publishing</pub></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0026-1394 |
ispartof | Metrologia, 2011-01, Vol.48, p.1 |
issn | 0026-1394 1681-7575 |
language | eng |
recordid | cdi_proquest_journals_911196371 |
source | Institute of Physics Journals |
subjects | Comparative analysis Radiation Uncertainty X-rays |
title | Key comparison BIPM.RI(I)-K2 of the air-kerma standards of the ENEA-INMRI, Italy and the BIPM in low-energy x-rays |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-25T04%3A02%3A11IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Key%20comparison%20BIPM.RI(I)-K2%20of%20the%20air-kerma%20standards%20of%20the%20ENEA-INMRI,%20Italy%20and%20the%20BIPM%20in%20low-energy%20x-rays&rft.jtitle=Metrologia&rft.au=Burns,%20D%20T&rft.date=2011-01-01&rft.volume=48&rft.spage=1&rft.pages=1-&rft.issn=0026-1394&rft.eissn=1681-7575&rft_id=info:doi/&rft_dat=%3Cproquest%3E2539154981%3C/proquest%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=911196371&rft_id=info:pmid/&rfr_iscdi=true |